Hantek HT824

Contents

Hantek HT824 Multifunction Process Calibrator User Manual

High-precision Five And A Half Signal Source Multimeter for Voltage, Current, and Resistance

1. Introduction

The Hantek HT824 Multifunction Process Calibrator is a high-accuracy, high-stability instrument designed for a wide range of industrial and laboratory applications. It can simulate and measure various electrical signals, including millivolt and volt voltages, milliampere currents, and different types of thermocouple (TC) and Resistance Temperature Detector (RTD) signals. This manual provides essential information for the safe and effective use of your calibrator.

2. Package Contents

Verify that all items listed below are included in your package. If any items are missing or damaged, please contact your supplier.

Hantek HT824 Calibrator and accessories in a carrying case

Figure 2.1: Hantek HT824 Calibrator and included accessories.

  1. Chassis (Main Unit)
  2. Data Line (USB Cable)
  3. Test Leads
  4. Chassis Key
  5. Thermocouple
  6. Instructions Manual
  7. CD (Software/Drivers)
  8. Power Cord

3. Product Overview

The Hantek HT824 features a clear 5 1/2 digit input/output display and an intuitive keypad for operation. Below are images illustrating the device from various angles.

Front view of the Hantek HT824 calibrator
Figure 3.1: Front View
Angled view of the Hantek HT824 calibrator with kickstand
Figure 3.2: Angled View with Kickstand
Rear view of the Hantek HT824 calibrator
Figure 3.3: Rear View

4. Specifications

The following tables detail the technical specifications of the Hantek HT824 Multifunction Process Calibrator.

4.1 DC Voltage Measurement

Parameter Range (Upper Screen)*1 Range (Bottom Screen)*2 Range
Range 0-30.000V 0-24.000V 0-90.000mV
Resolution 0.001V 0.001V 0.001mV
Accuracy 0.1%+5 0.05%+5 0.05%+5

*1 and *2: Input resistance is greater than 1MΩ

4.2 DC Voltage Output

Parameter Range Range
Range 0—20.000V 0—90.000mV
Resolution 0.001V 0.001mV
Accuracy 0.05%+5 0.05%+5

4.3 Millivolt Measurement and Output

Parameter Value
Range -10.000mV—80.000mV
Resolution 0.001mV
Accuracy 0.05%+5

*1 Press TC(16) to select this function. Signal is at thermocouple micro input/output TC port.

4.4 DC Current (Milliampere) Measurement

Parameter Range (Upper Screen)*1 Range (Bottom Screen)*2
Range 0—24.000mA 0—24.000mA
Resolution 0.001mA 0.001mA
Accuracy 0.05%+5 0.05%+5

4.5 DC Current (Milliampere) Output

Parameter Value
Range 0—24.000mA
Resolution 0.001mA
Accuracy 0.05%+5

SIMU (Simulation): When outputting current, external voltage is greater than 12V, less than 28V.

SOUR (Source): When outputting current, signal driving ability is 1000Ω in 20mA.

4.6 Resistance Measurement

Parameter Range 1 Range 2
Ohm Range 0—400.00Ω 400.0—4000.0Ω
Accuracy ±Ω (4 Wire) 0.02%+5 0.02%+5
Accuracy ±Ω (2 & 3 Wire) 0.05%+5 0.05%+5
  • Excitation current: 0.2mA.
  • Maximum input voltage: 30V.
  • 2 Wire: Do not include wire resistance.
  • 3 Wire: Assume to use the matched testing wire, the total resistance cannot be greater than 100Ω.

4.7 Resistance Output

Parameter Range 1 Range 2
Range 5.00Ω—400.00Ω 400.0—1500.0Ω
Excitation current from the measurement instrument 0.15mA-2mA 0.05mA-0.8mA
Accuracy ±Ω 0.05%+5 0.05%+5
Resolution 0.1Ω 0.1Ω

4.8 Temperature - Thermocouple (ITS-90)

Type Range Resolution Measurement and Output Accuracy ±℃
S-50.0℃ to 0℃0.1℃/0.1℉2℃
0℃ to 500.0℃0.1℃/0.1℉1.5℃
500.0℃ to 1760.0℃0.1℃/0.1℉1.3℃
R-50.0℃ to 0℃0.1℃/0.1℉2℃
0℃ to 500.0℃0.1℃/0.1℉1.5℃
500.0℃ to 1760.0℃0.1℃/0.1℉1.3℃
B200℃ to 800℃1℃/1℉2.5℃
800℃ to 1800℃1℃/1℉2.3℃
K-200.0℃ to 1370.0℃0.1℃/0.1℉1.3℃
N-200.0℃ to 1300.0℃0.1℃/0.1℉1.3℃
E-200.0℃ to 1000.0℃0.1℃/0.1℉1℃
J-200.0℃ to 1200.0℃0.1℃/0.1℉1℃
T-200.0℃ to 400.0℃0.1℃/0.1℉1℃

If open cold junction compensation, there should be additional ±0.5℃.

4.9 Temperature - Thermal Resistance

Type Range Accuracy ±℃ (Testing 4 wire ℃) Accuracy ±℃ (Testing 2 & 3 wire ℃) Output ℃
Pt100-385-200.0℃-850.0℃0.8℃1℃0.8℃
Pt100-3926-200.0℃-850.0℃0.8℃1℃0.8℃
Pt100-JIS-200.0℃-850.0℃0.8℃1℃0.8℃
Pt200-385-200.0℃-250.0℃0.7℃0.8℃0.7℃
250.0℃-630.0℃1.3℃2.1℃1.3℃
Pt500-385-200.0℃-500.0℃0.8℃1.1℃0.8℃
500.0℃—630.0℃1℃1.5℃1℃
Pt1000-385-200.0℃-100.0℃0.7℃0.7℃0.8℃
100.0℃-630.0℃0.7℃0.8℃0.8℃
Cu100-50.0℃-150.0℃1℃1.2℃1℃
Cu50-50.0℃-150.0℃1℃1.2℃1℃
  • Resolution: 0.1°C, 0.1°F
  • Allowable excitation current (output): Pt100-385, Pt100-392, Pt100-JIS, Pt200-385: 0.15 to 3.0 mA
  • Pt500-385: 0.05 to 0.80mA; Pt1000-385: 0.05 to 0.40mA
  • 2 Wire: Do not include wire resistance.
  • 3 Wire: Assume to use the matched testing wire, the total resistance cannot be greater than 100Ω.

4.10 Comprehensive Index

Parameter Value
Operation Temperature0℃ to 50℃
Storage Temperature-20℃ to 70℃
Operation HeightAverage elevation 3000 meters below.
Relative Humidity (No condensation)75% (30℃ to 40℃)
45% (40℃ to 50℃)
35% (50℃ to 55℃)
<10℃, no control
StabilityOut of the range 23±5℃, each degree increase ±0.005% of the range.
VibrationThe random vibration, 2g, 5 to 500Hz
SecurityEN 61010-1:1992
Protection LevelPollution grade Ⅱ

5. Initial Setup

  1. Unpacking and Inspection: Carefully remove the calibrator and all accessories from the packaging. Inspect for any signs of damage.
  2. Charging the Battery: The HT824 uses an advanced built-in 7.4V/7800mAh lithium battery. Connect the provided power cord to the calibrator and a suitable power outlet to charge the device before first use.
  3. Software Installation (Optional): Insert the provided CD into your computer to install any necessary software or drivers for PC connectivity and data analysis.
  4. Power On: Press the power button to turn on the calibrator. The display will illuminate.

6. Operation

The HT824 offers a variety of measurement and output functions. The screen displays connection diagrams for probes and plugs to help prevent incorrect operation.

6.1 General Operation Features

  • Display: 5 1/2 digit input/output display for precise readings.
  • Cold Junction Compensation: Supports manual and automatic cold junction compensation for thermocouple measurements.
  • Direct Temperature: The calibrator can directly measure and output temperature values.
  • Signal Output: Features auto triangular wave and step wave signal output.
  • Data Storage: Capable of storing 8 sets of measurement and output data, which can be recalled for review.
  • Memory Function: All setting values and the function from the last operation are automatically saved after power-off.
  • Backlight: The display includes a backlight for visibility in various lighting conditions.
  • Step/Slope Output: The output function includes manual/auto step and slope output, allowing arbitrary setting of 0%, 25%, and 100% values within the signal output range to test transmitter response speed.

6.2 Measurement and Output Modes

The calibrator supports various measurement and output modes. The following images illustrate the software interface for different functions:

Software interface showing Voltage measurement and RTD PT100/385 temperature output
Figure 6.1: Voltage Measurement & RTD Output
Software interface showing Voltage measurement and TC (Thermocouple) output
Figure 6.2: Voltage Measurement & TC Output
Software interface showing Voltage measurement and Resistance output
Figure 6.3: Voltage Measurement & Resistance Output
Software interface showing Voltage measurement and Current (mA) output
Figure 6.4: Voltage Measurement & Current Output
Software interface showing Voltage measurement and Voltage (V) output
Figure 6.5: Voltage Measurement & Voltage Output
Software interface showing RTD PT100/385 temperature measurement and output
Figure 6.6: RTD Temperature Measurement & Output

Refer to the on-screen prompts and the detailed instructions in the full user manual (PDF) for specific steps on selecting modes, connecting probes, and interpreting results.

7. Maintenance

  • Cleaning: Use a soft, dry cloth to clean the exterior of the calibrator. Do not use abrasive cleaners or solvents.
  • Battery Care: To prolong battery life, avoid fully discharging the battery frequently. Charge the device regularly, especially if it will be stored for an extended period.
  • Storage: Store the calibrator in a cool, dry place, away from direct sunlight and extreme temperatures. Use the provided carrying case for protection during storage and transport.
  • Calibration: For continued accuracy, periodic calibration by a qualified service center is recommended.

8. Troubleshooting

  • Device Not Powering On: Ensure the battery is charged. Connect the power adapter and try again. If the issue persists, the battery may need replacement or there could be an internal fault.
  • Inaccurate Readings: Check that the correct measurement mode is selected and that probes are properly connected. Ensure the device is within its specified operating temperature and humidity ranges. For temperature measurements, verify cold junction compensation settings.
  • No Signal Output: Confirm the output mode is correctly configured and the output range is appropriate for the connected device. Check all connections.
  • PC Connection Issues: Ensure the USB data line is securely connected and that the necessary drivers from the provided CD are installed on your computer.

9. User Tips

  • The advanced built-in battery management system and 7.4V/7800mAh lithium battery are designed to provide over 8 hours of continuous operation, making it reliable for extended on-site measurement tasks.
  • Always pay attention to the connection diagrams displayed on the screen when connecting probes and plugs to avoid incorrect operation and potential damage to the device or the circuit being tested.
  • Utilize the data storage feature to save up to 8 sets of measurement and output data for later review or analysis, which can be particularly useful for repetitive tasks or trend monitoring.

10. Warranty and Support

For detailed warranty information, technical support, or service inquiries, please refer to the official Hantek documentation or contact your point of purchase.

A comprehensive user manual in PDF format is available for download or viewing: Hantek HT824 User Manual (PDF).

Certifications for Hantek HT824 including CE, FCC, RoHS, ISO, 3C

Figure 10.1: Product Certifications

11. Product Video

Watch this video for a visual demonstration of the Hantek HT824 Multifunction Process Calibrator.

Video 11.1: Hantek HT824 Product Demonstration

© 2023 Hantek. All rights reserved.