OWON SPM6103

OWON SPM6103 Programmable Lab Power Supply User Manual

Model: SPM6103

Brand: OWON

1. Introduction

The OWON SPM6103 is a high-performance programmable DC power supply designed for laboratory and industrial applications. It features a wide output range of 0-60V and 0-10A, with a maximum power output of 300W. This device integrates a high-resolution 4½ digit multimeter, offering convenient measurement capabilities for voltage, current, resistance, capacitance, diode, and continuity. Its compact design, multiple protection functions, and SCPI communication support make it a versatile tool for various testing and experimental needs, including battery charging, aging tests, and electronic device repair.

2. Safety Information

WARNING:

  • To avoid electric shock, the power cord protective grounding conductor must be connected to ground.
  • Do not remove covers. Refer servicing to qualified personnel.
  • Always ensure the power supply is turned off and disconnected from the mains before making any connections or adjustments to avoid electric shock or damage to the device.
  • Operate the device in a well-ventilated area to prevent overheating.
Rear view of the OWON SPM6103 with safety warning and labeled components

Image: Rear panel of the SPM6103 showing the power cord socket, fuse, power switch, intelligent temperature control fan, and USB communication port, along with a safety warning regarding grounding and servicing.

3. Package Contents

Please check the package carefully upon receipt to ensure all items are present and undamaged:

  • 1 x OWON SPM6103 Programmable DC Power Supply
  • 1 x Power Cord
  • 1 x USB Cable
  • 1 x Multimeter Lead Set
  • 1 x Ground Wire
  • 1 x Spare Fuse
  • 1 x User Manual (English)

4. Product Overview

4.1 Front Panel Layout

Labeled front panel of the OWON SPM6103 showing all controls and display

Image: Detailed view of the SPM6103 front panel with numbered labels indicating the function of each control and display area.

  1. 2.8-inch LCD: Displays voltage, current, power, settings, and multimeter readings.
  2. Voltage Setting / AC/DC Voltage Switch: Adjusts output voltage and switches between AC/DC voltage measurement modes for the multimeter.
  3. Current Setting / AC/DC Current Switch: Adjusts output current and switches between AC/DC current measurement modes for the multimeter.
  4. Ω/OVP / ÷/OCP Button: Sets Over Voltage Protection (OVP) and Over Current Protection (OCP) values. Also switches between resistance, capacitance, diode, and continuity measurement types for the multimeter.
  5. DISP Button: Switches between different display modes.
  6. Mode Button: Toggles between DC power supply mode and Multimeter mode.
  7. Adjustment Knob: Used for fine-tuning voltage, current, and other settings.
  8. On/Off Button: Independent control for output. Long press to activate automatic output 5 seconds after booting.
  9. Multimeter Voltage and Current Sockets: Input terminals for multimeter leads (red for positive, black for negative).
  10. Channel Output Terminals: Main output terminals for the DC power supply.
  11. Front USB Port: Provides 5V/1A charging output.

4.2 Rear Panel Layout

Rear panel of the OWON SPM6103 with labeled components

Image: Rear panel of the SPM6103 showing the intelligent temperature control fan, power cord socket, fuse, power switch, and USB device communication port.

  • Intelligent Temperature Control Fan: Automatically adjusts speed based on internal temperature to ensure optimal cooling.
  • Power Cord Socket: Connects the device to the main power supply.
  • Fuse: Overcurrent protection for the device.
  • Power Switch: Turns the main power of the unit on or off.
  • USB Device Communication Port: Supports SCPI communication protocol for remote control and data logging.

5. Setup

  1. Unpacking: Carefully remove the power supply and all accessories from the packaging.
  2. Placement: Place the unit on a stable, level surface with adequate ventilation around the intelligent temperature control fan at the rear.
  3. Power Connection: Connect the provided power cord to the power cord socket on the rear panel. Ensure the other end is connected to a properly grounded AC power outlet.
  4. USB Communication (Optional): If remote control via SCPI is desired, connect the USB cable from the device's rear USB port to your computer.
  5. Initial Power On: Flip the power switch on the rear panel to the 'ON' position. The display should light up.

6. Operating Instructions

6.1 DC Power Supply Mode

Display showing CV/CC output, memory settings, and list waveform editing

Image: Screenshots of the SPM6103 display illustrating Constant Voltage (CV) and Constant Current (CC) output modes, memory shortcut parameters, and list waveform editing output.

  1. Selecting Power Supply Mode: Press the 'Mode' button until the display indicates Power Supply mode (e.g., showing voltage and current output).
  2. Setting Voltage and Current: Use the 'Voltage Setting' and 'Current Setting' buttons, then rotate the 'Adjustment Knob' to set the desired output voltage and current limits.
  3. Output On/Off: Press the 'On/Off' button to enable or disable the output. The button will illuminate when the output is active. A long press can set the automatic output function to activate 5 seconds after booting.
  4. Overvoltage Protection (OVP) and Overcurrent Protection (OCP): Press the 'Ω/OVP / ÷/OCP' button to access OVP/OCP settings. Use the adjustment knob to set the maximum voltage and current thresholds. These functions protect your circuit by automatically shutting down the output if the set limits are exceeded.
  5. Constant Voltage (CV) and Constant Current (CC) Output: The device automatically switches between CV and CC modes to protect the circuit. If the actual current exceeds the preset current value, the power supply automatically switches to CC mode.
  6. Memory Functions: The device supports 4 groups of memory shortcut parameters for quick recall of frequently used settings. Use the 'Memory' button (F1) to save and recall these settings.
  7. List Waveform Editing Output: The unit allows for editing up to 10 groups of timing output, enabling complex test sequences.

6.2 Multimeter Mode

OWON SPM6103 in multimeter mode measuring components

Image: The SPM6103 being used as a multimeter to measure components on a circuit board, demonstrating its integrated functionality.

  1. Selecting Multimeter Mode: Press the 'Mode' button until the display shows multimeter measurement options (e.g., V, A, Ω, F).
  2. Connecting Multimeter Leads: Insert the red multimeter lead into the positive (+) multimeter voltage/current socket and the black lead into the negative (-) COM socket.
  3. Selecting Measurement Type: Use the 'Ω/OVP / ÷/OCP' button to cycle through different measurement types: Voltage (DCV/ACV), Current (DCA/ACA), Capacitance, Resistance, Diode, and Continuity.
  4. Performing Measurements: Connect the multimeter leads to the circuit or component you wish to measure. The reading will be displayed on the LCD.

7. Maintenance

  • Cleaning: Use a soft, dry cloth to clean the exterior of the device. Do not use abrasive cleaners or solvents.
  • Ventilation: Ensure the rear fan is free from obstructions to allow for proper airflow and cooling.
  • Fuse Replacement: If the device fails to power on, check the fuse located on the rear panel. Replace it with a fuse of the same type and rating (e.g., 250V, 15A for 200-240V AC input, or as specified on the unit).
  • Storage: When not in use for extended periods, store the power supply in a cool, dry environment, away from direct sunlight and extreme temperatures.

8. Troubleshooting

  • No Power:
    • Check if the power cord is securely connected to both the unit and the power outlet.
    • Ensure the rear power switch is in the 'ON' position.
    • Inspect the fuse on the rear panel. If blown, replace it with a fuse of the correct rating.
  • No Output Voltage/Current:
    • Verify that the 'On/Off' output button on the front panel is pressed and illuminated.
    • Check OVP/OCP settings. If the output is being limited by these protections, adjust them or disable them temporarily for testing.
    • Ensure the unit is in Power Supply mode, not Multimeter mode.
  • Inaccurate Multimeter Readings:
    • Ensure multimeter leads are correctly connected to the appropriate input sockets.
    • Verify that the correct measurement type (Voltage, Current, Resistance, etc.) is selected using the 'Ω/OVP / ÷/OCP' button.
    • Check the condition of the multimeter leads for any damage.
  • Unit Overheating:
    • Ensure the intelligent temperature control fan at the rear is not obstructed and is operating.
    • Reduce the load on the power supply if operating at its maximum limits for extended periods.
    • Ensure adequate ventilation around the unit.

9. Specifications

9.1 Power Supply Parameters

ParameterValue
Voltage (rms)≤3mVrms (0-30V), ≤5mVrms (30-60V)
Current (rms)≤30mAp-p
Output Temperature Coefficient (0℃-40℃) - Voltage100ppm/℃
Output Temperature Coefficient (0℃-40℃) - Current200ppm/℃
Readback Temperature Coefficient - Voltage100ppm/℃
Readback Temperature Coefficient - Current200ppm/℃
Response Time (50%-100% rated load)≤1.0m/s
Data Storage4 groups of data
Working Temperature0℃~40℃
Storage Temperature-20℃~60℃
Relative Humidity≤90%RH; no condensation
Operating Height≤2000 meters
CoolingFan cooling, temperature intelligent speed control
Communication PortUSB communication, compatible with SCPI protocol
USB Charging Port5V/1A charging

9.2 Multimeter Parameters

Measurement TypeRangeAccuracy
Full Scale Reading4½ digits
Auto RangeYes
DC Voltage (DCV)200.00mV, 2.0000V, 20.000V, 200.00V, 1000V±(0.3%+10digit) to ±(0.3%±5digit)
AC Voltage (ACV)200.00mV, 2.0000V, 20.000V, 200.00V, 750V±(0.8%+10digit) to ±(1%±10digit)
DC Current (DCA)200.00mA, 10.000A±(0.8%±10digit) to ±(2.5%±10digit)
AC Current (ACA)200.00mA, 10.000A±(1%±10digit) to ±(2.8%±10digit)
Capacitance20.000nF to 2.0000mF±(3.0%±10digit)
Resistance200.00Ω to 100MΩ±(0.8%±10digit) to ±(5%±10digit)
Diode Test0-2V
Continuity TestYes

9.3 Physical Dimensions

MeasurementValue
Item Size (L x W x H)226 x 82 x 142mm / 8.89 x 3.23 x 5.59in
Item Weight1580g / 3.48lb
Package Weight2057g / 4.53lb
Package Size (L x W x H)34 x 13 x 19.5cm / 13.39 x 5.12 x 7.68in
Dimensions of the OWON SPM6103 power supply

Image: Side view of the SPM6103 with dimensions: 142mm/5.59in (height), 226mm/8.89in (depth), 82mm/3.23in (width).

10. User Tips

  • Gradual Adjustment: When powering sensitive circuits, always start with lower voltage and current settings and gradually increase them to the desired levels to prevent damage.
  • Pre-check Connections: Before enabling the output, always double-check all connections to your circuit to ensure they are correct and secure.
  • Utilize Memory Presets: For repetitive tasks, save your frequently used voltage and current settings into the memory groups for quick recall, saving time and reducing setup errors.
  • Monitor Temperature: While the unit has an intelligent cooling fan, ensure it's not placed in an enclosed space to allow for optimal heat dissipation, especially during high-power operation.

11. Warranty and Support

For warranty information, technical support, or service inquiries, please refer to the official OWON website or contact your authorized dealer. Keep your purchase receipt as proof of purchase for any warranty claims.

Related Documents - SPM6103

Preview OWON Digital Multimeter Quick Start Guide: Safety, Operation, and Connectivity
A comprehensive guide for the OWON digital multimeter, detailing safety instructions, quick start procedures, measurement connections, mobile app integration via Bluetooth, and essential care and cleaning tips.
Preview OWON ODP3031 Linear Programmable DC Power Supply User Manual
Comprehensive user manual for the OWON ODP3031 Linear Programmable DC Power Supply, detailing its features, operation, safety precautions, troubleshooting, and technical specifications.
Preview OWON ODP Series Linear Programmable DC Power Supply User Manual
User manual for the OWON ODP Series Linear Programmable DC Power Supply, covering general safety, characteristics, quick start guide, front panel operation, PC communication, troubleshooting, technical specifications, and appendices.
Preview OWON ODP Series Triple Output DC Power Supply User Manual
Comprehensive user manual for OWON ODP Series Triple Output Linear Programmable DC Power Supplies (ODP3033, ODP3063, ODP6033), covering setup, operation, safety, and technical specifications.
Preview Owon P Series Single Channel DC Power Supply User Manual
Comprehensive user manual for the Owon P Series single channel DC power supply, detailing safety precautions, panel operations, system settings, troubleshooting, and general care.
Preview OWON FDS Series Quick Guide: User Manual and Operation Instructions
Get started with your OWON FDS Series oscilloscope. This quick guide covers essential setup, safety, and operational details for its oscilloscope, function generator, power supply, and multimeter features. Visit www.owon.com.hk/download for full support.