INFICON PSG500 Spare Sensor DN 16 KF Instruction Manual

Model: PSG500 Series

1. Introduction

This manual provides essential information for the proper installation, operation, and maintenance of the INFICON PSG500 Spare Sensor DN 16 KF. The PSG500 series sensors are designed for reliable vacuum measurement in various applications. Please read this manual thoroughly before using the product to ensure safe and efficient operation.

INFICON PSG500 Spare Sensor packaging
Figure 1: INFICON PSG500 Spare Sensor packaging, showing model number 350-920.
Open box with INFICON PSG500 sensor and installation manual
Figure 2: Contents of the PSG500 sensor package, including the sensor and installation manual.

2. Setup and Installation

The INFICON PSG500 Spare Sensor is designed for integration into existing vacuum systems. Proper installation is crucial for accurate readings and system integrity.

2.1 Connection Types

The sensor typically utilizes DN 16 KF vacuum connections. Ensure your system is compatible with this standard. The provided diagram illustrates common connection methods.

Diagram of DN 16 ISO-KF and DN 16 CF-R vacuum connections
Figure 3: Illustration of DN 16 ISO-KF and DN 16 CF-R vacuum connection types for the sensor.

2.2 Installation Steps (General Guidelines)

  1. Safety First: Ensure the vacuum system is depressurized and isolated before beginning installation. Wear appropriate personal protective equipment.
  2. Inspect Components: Carefully unbox the sensor and inspect it for any signs of damage. Verify that all necessary gaskets and clamps for the DN 16 KF connection are present and in good condition.
  3. Prepare Connection Point: Clean the mating surfaces of the vacuum system's port and the sensor's flange to ensure a leak-tight seal.
  4. Mount the Sensor: Position the sensor onto the vacuum port. Use a centering ring and clamp (not included, typically part of the vacuum system) to secure the DN 16 KF connection. Tighten the clamp evenly to prevent leaks.
  5. Electrical Connection: Connect the sensor's electrical cable to the appropriate vacuum gauge controller or power supply, following the wiring instructions provided with your specific controller. Refer to the specifications for correct supply voltage.
  6. Leak Check: After installation, perform a thorough leak check of the vacuum system to ensure the sensor connection is sealed properly.

For detailed, model-specific installation instructions, please refer to the full installation manual that accompanies the product or consult INFICON documentation.

3. Operating Instructions

Once installed and connected, the INFICON PSG500 sensor operates by providing a voltage output proportional to the measured vacuum pressure. The sensor requires a compatible vacuum gauge controller for power and to interpret the output signal.

3.1 Powering On

Ensure the sensor is correctly wired to its controller. Apply the specified supply voltage (+14V to +30V DC) to the controller. The sensor will typically begin measuring immediately upon receiving power.

3.2 Measurement Range

The PSG500 series sensors are designed to measure vacuum pressures within a specific range. Refer to the 'Specifications' section for the exact measurement range of your sensor model.

3.3 Output Signal

The sensor provides a voltage output (V) that corresponds to the vacuum pressure. This signal is interpreted by the connected vacuum gauge controller to display the pressure reading. Consult your controller's manual for details on interpreting the sensor's output.

3.4 Admissible Temperature

Ensure the operating environment temperature remains within the admissible range of +5°C to +60°C to maintain sensor accuracy and longevity.

4. Maintenance

Regular maintenance helps ensure the longevity and accuracy of your INFICON PSG500 sensor. Specific maintenance procedures may vary depending on the application and vacuum environment.

  • Cleaning: The sensor element should generally not be cleaned unless absolutely necessary and only by trained personnel following manufacturer guidelines. Contamination can affect accuracy. The materials exposed to vacuum are glass, Ni, NiFe, and stainless steel (DIN 1.4301 / 1.4305 / 1.4435).
  • Inspection: Periodically inspect the sensor and its connections for any signs of wear, damage, or contamination. Check vacuum seals (gaskets, O-rings) for degradation.
  • Calibration: Over time, sensors may drift. Regular calibration against a known standard is recommended to maintain measurement accuracy. Refer to INFICON's calibration procedures or contact technical support.
  • Storage: If the sensor is to be stored for an extended period, ensure it is kept in a clean, dry environment within the specified storage temperature range (-20°C to +65°C).

For detailed maintenance schedules and procedures, consult the official INFICON documentation for the PSG500 series.

5. Troubleshooting

This section provides general troubleshooting tips. For complex issues, it is recommended to contact technical support.

  • No Reading/Incorrect Reading:
    • Check all electrical connections between the sensor and the controller.
    • Verify the power supply voltage is within the specified range (+14V to +30V DC).
    • Ensure the vacuum system is properly sealed and there are no leaks.
    • Confirm the sensor is compatible with the connected controller.
    • Check if the sensor element is contaminated or damaged.
  • Unstable Readings:
    • Inspect for intermittent electrical connections.
    • Check for vibrations or temperature fluctuations in the operating environment.
    • Verify the vacuum system stability.

If these steps do not resolve the issue, consult the comprehensive troubleshooting guide in the full product manual or contact INFICON technical support.

6. Technical Specifications

The following table outlines the key technical specifications for the INFICON PSG500 series sensors. Values are typical and may vary slightly between specific models (e.g., PSG500 tungsten, PSG502-S nickel, etc.).

Technical specifications table for INFICON PSG500 series sensors
Figure 4: Detailed technical specifications for various PSG500 series sensor models.
ParameterUnitPSG500 tungstenPSG500-S tungsten, setpointsPSG502-S nickel, setpointsPSG510-S tungsten, setpoints, ceramic feedthroughsPSG512-S nickel, setpoints, ceramic feedthroughs
Measurement rangeV+1.9 ... +10.0+1.9 ... +10.0+1.9 ... +10.0+1.9 ... +10.0+1.9 ... +10.0
Accuracy (N2)
1 x 10-3 ... 100 mbar% of reading±15%±15%±15%±15%±15%
5 x 10-4 ... 1 x 10-3 mbar% of reading±50%±50%±50%±50%±50%
100 ... 1000 mbar% of reading±50%±50%±50%±50%±50%
Repeatability (air)
1 x 10-3 ... 100 mbar% of reading2%2%2%2%2%
Admissible temperature
Operation°C+5 ... +60+5 ... +60+5 ... +60+5 ... +60+5 ... +60
Vacuum connection (1)80 / 250 (2)80 / 250 (2)80 / 250 (2)80 / 250 (2)80 / 250 (2)
Storage°C-20 ... +65-20 ... +65-20 ... +65-20 ... +65-20 ... +65
Supply voltage
At gaugeV (dc)+14 ... +30+14 ... +30+14 ... +30+14 ... +30+14 ... +30
RippleV (p-p)≤1≤1≤1≤1≤1
Power consumptionW≤1≤1≤1≤1≤1
Setpoint
Number of setpointsnone1111
Setting rangembar-1 x 10-3 ... 10001 x 10-3 ... 10001 x 10-3 ... 10001 x 10-3 ... 1000
Hysteresis% of reading-±2%±2%±2%±2%
Relay contact: Switching timems-<100<100<100<100
Materials exposed to vacuumglass, Ni, NiFe, DIN 1.4301 / 1.4305 / 1.4435glass, Ni, NiFe, DIN 1.4301 / 1.4305 / 1.4435glass, Ni, NiFe, DIN 1.4301 / 1.4305 / 1.4435glass, Ni, NiFe, DIN 1.4301 / 1.4305 / 1.4435glass, Ni, NiFe, DIN 1.4301 / 1.4305 / 1.4435

Note: (1) Connection type, (2) Max bakeout temperature for sensor head / electronics.

7. User Tips

Based on general best practices for vacuum sensors:

  • Always handle the sensor with care to avoid damaging the sensitive vacuum-exposed components.
  • Ensure the vacuum system is clean and free of particulate matter before installing the sensor to prevent contamination.
  • When replacing a sensor, ensure the new sensor's specifications match the requirements of your vacuum system and controller.
  • If experiencing inconsistent readings, double-check all connections and ensure the sensor is operating within its specified temperature range.

No specific user tips were available from customer reviews or Q&A for this product.

8. Warranty and Support

The seller, Semiconductor Equipment & Parts Hub Store, offers the following warranty and support:

  • Warranty: New items come with a one-year warranty. Used items have a six-month warranty.
  • Technical Support: Technical support is provided.
  • Repair Services: The store specializes in semiconductor parts and Applied Materials components, offering repair services.

For technical inquiries, repair services, or warranty claims, please contact the seller directly via email:

Email: ksjxiao hua@163.com

Seller's contact information and warranty details