MESTEK CM87 Pro

Contents

MESTEK 1000A AC/DC True RMS Auto Range Digital Clamp Meter User Manual

Models: CM87A, CM87B, CM87C, CM87 Pro

Brand: MESTEK

1. Introduction

This manual provides detailed instructions for the safe and effective use of your MESTEK 1000A AC/DC True RMS Auto Range Digital Clamp Meter. This versatile instrument is designed for various electrical measurements, including AC/DC current, AC/DC voltage, resistance, capacitance, frequency, and temperature. It features True RMS, VFD, and INRUSH capabilities, making it suitable for power maintenance, electrical installation, and industrial testing.

MESTEK 1000A AC/DC True RMS Auto Range Digital Clamp Meter with probes and temperature probe
MESTEK 1000A AC/DC True RMS Auto Range Digital Clamp Meter with included accessories.
Unboxing video for the MESTEK Digital Clamp Meter, showing the contents of the package.

2. Safety Information

Always adhere to safety precautions when using electrical testing equipment. Failure to do so may result in electric shock, injury, or damage to the meter or equipment under test.

  • Ensure the meter is in good working condition before use.
  • Do not exceed the maximum input values for any function.
  • Use appropriate personal protective equipment (PPE).
  • Avoid touching exposed conductors or circuit components.
  • Do not operate the meter in wet environments or explosive atmospheres.
  • Replace batteries promptly when the low battery indicator appears.

3. Product Overview and Setup

Familiarize yourself with the components of your clamp meter before operation.

Diagram of MESTEK clamp meter with labeled parts and dimensions
Appearance Introduction: Key components and dimensions of the MESTEK Clamp Meter.

3.1 Parts List

Image showing the contents of the MESTEK clamp meter package
The package typically includes: (1) Clamp shaped meter, (2) K-type thermocouple, (3) User Manual, (4) Test probes, (5) Instrument storage bag, (6) Packaging box.

3.2 Battery Installation

The meter requires 2 x 1.5V AAA batteries. Ensure correct polarity when installing. The low battery indication will appear when the voltage drops below 2.4V.

3.3 Connecting Test Probes

Insert the red test lead into the 'INPUT' jack and the black test lead into the 'COM' jack for voltage, resistance, capacitance, frequency, diode, and continuity measurements.

4. Operating Instructions

The MESTEK clamp meter offers a variety of measurement functions. Rotate the function knob to select the desired mode.

4.1 AC/DC Current Measurement (Clamp)

To measure current, place a single wire of the tested circuit at the center of the clamp head. Ensure the clamp is fully closed. The instrument screen will display the current value.

Images demonstrating AC and DC current measurement with the clamp meter
Top: Measuring AC current in an electrical panel. Bottom: Measuring DC current from a car battery.

4.1.1 DC Current Zeroing

In DC current mode, long press the 'ZERO' key to eliminate any base number caused by magnetic field interference, ensuring more accurate measurement results.

Image showing the clamp meter display before and after DC current zeroing
Before resetting to zero (00.39A) and after resetting to zero (00.00A).

4.2 AC/DC Voltage Measurement

Connect the test probes to the circuit. The meter will automatically identify whether the voltage is AC or DC and display the measured value.

Images showing AC voltage measurement (750V) and DC voltage measurement (1000V)
Top: Measuring AC voltage in an electrical panel. Bottom: Measuring DC voltage from a car battery.

4.2.1 AUTO-VLOZ (Automatic Voltage Recognition)

The instrument automatically identifies AC or DC voltage based on the measured signal.

Images showing automatic recognition of AC and DC voltage
Automatic recognition of AC voltage (left) and DC voltage (right).

4.3 Resistance Measurement

Connect the test probes across the component to measure resistance. The screen will display the resistance value in Ohms.

Image showing resistance measurement using test probes
Resistance measurement in progress.

4.4 Capacitance Measurement

Connect the test probes across the capacitor to measure capacitance. The screen will display the capacitance value.

Image showing capacitance measurement using test probes
Capacitance measurement in progress.

4.5 Frequency Measurement

The screen displays the frequency value of the tested signal source. The instrument synchronously displays the percentage of the measured signal.

Image showing frequency measurement using test probes
Frequency measurement in an electrical panel.

4.6 Contact Temperature Measurement

When measuring objects or liquids, place the K-type thermocouple probe in contact with the object being measured and press the 'SEL' key briefly to convert between °C/°F.

Image showing contact temperature measurement with a K-type thermocouple
Measuring temperature of frozen goods using the K-type thermocouple.

4.7 Diode Test

Connect the test probes across the diode. The meter will display the forward voltage drop.

Image showing diode test using test probes
Diode test in progress.

4.8 Continuity Buzzer

Connect the test probes across the circuit. If the resistance is below a certain threshold, the buzzer will sound, indicating continuity.

Image showing continuity test using test probes
Continuity test in progress.

4.9 Non-Contact Voltage (NCV) Detection

When the NCV sensing area of the instrument is brought close to a circuit, different alarm prompts (audible and visual) will be sounded according to the voltage intensity. The screen will display 'L' for low voltage or 'H' for high voltage with a red backlight.

Images showing NCV detection for low and high voltage
NCV detection indicating low voltage (left) and high voltage (right).

4.10 LIVE Wire Recognition

Use a probe to touch the tested conductor for detection. The red backlight is illuminated, the buzzer sounds an alarm, and the tested wire is identified as a live wire.

Image showing LIVE wire recognition with the clamp meter
LIVE wire recognition in action.

4.11 Data Hold (HOLD)

Short press the 'HOLD' key to lock the current measurement data on the display. The buzzer will sound a warning every 2 seconds to indicate the locked state.

Image showing the HOLD data lock function on the clamp meter
The HOLD function locks the displayed measurement.

4.12 MIN/MAX Measurement

This function allows locking test data in AC/DC voltage and AC/DC current modes to capture minimum and maximum values.

Image showing MIN/MAX measurement function on the clamp meter
MIN/MAX measurement for capturing extreme values.

4.13 VFD (Variable Frequency Drive) Measurement

Enabling VFD in AC voltage and current mode can effectively reduce interference from high-frequency signals on measurement results.

4.14 INRUSH Current Measurement

Measures the initial surge current when a device is powered on (available on CM87C and CM87 Pro models).

4.15 Flashlight

The meter includes a built-in flashlight for illuminating dark work areas.

Image showing the flashlight feature of the clamp meter
The integrated flashlight provides illumination for convenience.

5. Maintenance

5.1 Cleaning

Wipe the meter with a dry, soft cloth. Do not use abrasives or solvents.

5.2 Battery Replacement

When the low battery indicator (less than 2.4V) appears on the display, replace the 2 x 1.5V AAA batteries immediately to ensure accurate readings.

5.3 Storage

If the meter is not used for a long period, remove the batteries to prevent leakage. Store the meter in a cool, dry place, away from direct sunlight and extreme temperatures.

6. Troubleshooting

If the meter does not function correctly, check the following:

  • No display or dim display: Check battery level and replace if necessary. Ensure batteries are installed correctly.
  • Inaccurate readings: Verify the correct function mode is selected. Ensure test leads are properly connected and not damaged. For DC current, perform a zeroing operation.
  • No continuity beep: Check if the circuit is truly continuous and within the resistance threshold for the buzzer.
  • NCV not detecting: Ensure the NCV sensing area is close enough to the AC voltage source.

If problems persist, contact MESTEK customer support.

7. Specifications

The following table outlines the technical parameters for the MESTEK Digital Clamp Meter models:

Technical Parameter table for MESTEK Digital Clamp Meter
Detailed technical specifications for the MESTEK Clamp Meter.
MESTEK Digital Clamp Meter Parameter Comparison
FeatureCM87ACM87BCM87CCM87 Pro
DC current✗✗0-600A0-1000A
AC current0-600A0-600A0-600A0-1000A
K type temperature probe✗✓✓✓
DC voltage0-600V0-600V0-600V0-1000V
AC voltage0-600V0-600V0-600V0-750V
Capacitance0-4mF0-60mF0-60mF0-60mF
Resistance0-40MΩ0-60MΩ0-60MΩ0-60MΩ
Inrush current✗✗0-600A0-1000A
Display counts3999599959995999
Operation rangeAuto rangeAuto rangeAuto rangeAuto range
Frequency measure✓0-10MHZ0-10MHZ0-10MHZ
Overload protection✓✓✓✓
Continuity buzzer✓✓✓✓
Jaw opening42mm42mm42mm42mm
True-RMS✓✓✓✓
Data hold✓✓✓✓
Backlight✓✓✓✓
Auto power off✓✓✓✓
NCV✓✓✓✓
LIVE wire recognition✓✓✓✓
Diode✓✓✓✓
Input impedance1MΩ1MΩ1MΩ1MΩ
Low battery indication< 2.4V< 2.4V< 2.4V< 2.4V
Battery2*1.5V AAA2*1.5V AAA2*1.5V AAA2*1.5V AAA

General Specifications:

  • Dimensions: 218*75*39mm
  • Weight: ~0.48kg (excluding battery)
  • Operating Temperature: 0-40°C
  • Storage Temperature: -10°C ~ 60°C
  • Display Type: Digital Only
  • Maximum Display Value: 5999 (for CM87B, CM87C, CM87 Pro) / 3999 (for CM87A)
  • Over Range Display: OL or -OL
  • Sample Time: About 3 times per second
  • Automatic Shutdown Time: About 15 minutes
  • Backlight Automatic Off Time: About 3 minutes
  • Flashlight Automatic Shutdown Time: About 5 minutes
  • Jaw Size: 30mm
  • Certification: CE, FCC, RoHS

8. Warranty and Support

MESTEK products are designed for reliability and performance. For warranty information, technical support, or service inquiries, please refer to the official MESTEK website or contact your retailer. Keep your purchase receipt as proof of purchase.

9. User Tips

  • Always start with the highest range: When measuring an unknown value, always select the highest range setting first and then decrease it if necessary to avoid overloading the meter.
  • Check test leads: Before each use, visually inspect test leads for any damage or frayed insulation.
  • Isolate power: Whenever possible, de-energize the circuit before making resistance, capacitance, or continuity measurements.
  • Single wire for clamp: For current measurements with the clamp, ensure only one conductor passes through the jaw. Measuring multiple conductors will result in an inaccurate reading (or zero if currents are balanced).
  • Use Data Hold: For measurements in hard-to-reach areas, use the 'HOLD' function to freeze the reading on the display for easier viewing.
  • Mind the environment: Avoid using the meter in extreme temperatures or high humidity, as this can affect accuracy and safety.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.