1. Introduction

The Mestek CM87Pro Digital Clamp Meter is a professional-grade, automatic range clamp meter designed for accurate and reliable electrical measurements. It features True RMS for precise readings of non-sinusoidal waveforms, Inrush current measurement for motor startup analysis, and a Variable Frequency Drive (VFD) function to reduce interference in AC measurements. This manual provides detailed instructions for the safe and effective use of your CM87Pro clamp meter.

The CM87Pro offers a 6000 counting range and can measure up to 1000A AC/DC current, 1000V DC voltage, and 750V AC voltage, along with resistance, capacitance, frequency, and temperature. It is an indispensable tool for power maintenance, electrical installation, and industrial inspection.

Mestek CM87Pro Digital Clamp Meter with probes, highlighting key features like 1000A AC/DC, True RMS, Inrush, VFD, and VA color screen.

Figure 1: Mestek CM87Pro Digital Clamp Meter, highlighting key features.

2. Package Contents

Upon unpacking, please verify that all items listed below are present and in good condition. If any items are missing or damaged, contact your vendor immediately.

  • Mestek CM87Pro Digital Clamp Meter
  • K-type Thermocouple
  • Test Probes (Red and Black)
  • Instrument Storage Bag
  • User Manual (this document)
  • Packaging Box

Video 1: Unboxing and contents of the Mestek CM87Pro Digital Clamp Meter.

Image showing the contents of the Mestek CM87Pro package: Clamp shaped table (clamp meter), K-type thermocouple, User Manual, Test probe, Instrument storage bag, Packaging box.

Figure 2: Visual representation of the package contents.

3. Product Overview

Familiarize yourself with the various parts and controls of the Mestek CM87Pro Digital Clamp Meter for optimal use.

Diagram of Mestek CM87Pro Digital Clamp Meter with labeled parts: NCV sensing area, Current sensing zone, LED lamp, Flashlight/Data Hold Key, Trigger, Function knob, Function switch key, VFD/INRUSH, ZERO, Max/Min switching, Display, Black probe, Red probe, Dimensions.

Figure 3: Detailed diagram of the clamp meter's components and their functions.

  • NCV Sensing Area: Used for non-contact voltage detection.
  • Current Sensing Zone: The area within the clamp for current measurement.
  • LED Lamp: Provides illumination for dark work areas.
  • Flashlight/Data Hold Key: Activates the flashlight and holds the current display reading.
  • Trigger: Opens and closes the clamp jaw.
  • Function Knob: Selects the desired measurement function.
  • Function Switch Key (SEL): Toggles between different sub-functions within a mode (e.g., AC/DC, °C/°F).
  • VFD/INRUSH Button: Activates Variable Frequency Drive measurement or Inrush current measurement.
  • ZERO Button: Resets DC current readings to zero.
  • Max/Min Switching: Toggles between maximum and minimum recorded values.
  • Display: VA color screen showing measurement readings and indicators.
  • COM Input: Common terminal for test probes (black).
  • INPUT Terminal: Input terminal for test probes (red).

4. Setup

4.1 Battery Installation

The Mestek CM87Pro requires two 1.5V AAA batteries (not included). To install or replace batteries:

  1. Ensure the clamp meter is turned OFF.
  2. Locate the battery compartment on the back of the device.
  3. Use a screwdriver to open the battery compartment cover.
  4. Insert two 1.5V AAA batteries, observing the correct polarity (+/-).
  5. Replace the battery compartment cover and secure it with the screw.

A low battery indication will appear on the display when the battery voltage drops below 2.4V.

4.2 Connecting Test Probes

For voltage, resistance, capacitance, frequency, diode, and continuity measurements, connect the test probes:

  1. Insert the black test lead into the 'COM' input jack.
  2. Insert the red test lead into the 'INPUT' jack.

4.3 Connecting K-type Thermocouple

For temperature measurements, connect the K-type thermocouple:

  1. Ensure the clamp meter is turned OFF.
  2. Insert the K-type thermocouple into the 'COM' and 'INPUT' jacks, observing polarity if indicated.

5. Operating Modes

Turn the function knob to select the desired measurement mode. Use the 'SEL' button to switch between sub-functions within a mode.

Icons representing various functions: AC/DC current, AC/DC voltage, AUTO-V LoZ, Resistance, Capacitance, Diode, On/off buzzing, Frequency, Temperature, NCV, Live, MAX/MIN, Zeroing function/data hold, VFD, Inrush, Flashlight.

Figure 4: Visual list of the clamp meter's functions.

5.1 Current Measurement (AC/DC)

The CM87Pro can measure up to 1000A AC/DC current.

  1. Turn the function knob to the '1000A AC/DC CLAMP METER' position.
  2. Press the 'SEL' button to switch between AC and DC current modes if necessary.
  3. Open the clamp jaw using the trigger.
  4. Place a single wire of the tested circuit at the center of the clamp head. Ensure the clamp is fully closed.
  5. The instrument screen will display the current value.
Mestek CM87Pro measuring AC current on red wires and DC current on a car battery.

Figure 5: Measuring AC and DC current with the clamp meter.

5.2 Voltage Measurement (AC/DC & AUTO-V LoZ)

The meter automatically recognizes AC or DC voltage up to 1000V DC and 750V AC.

  1. Turn the function knob to the 'AUTO-V LoZ' position for automatic recognition or specific AC/DC voltage ranges.
  2. Connect the test probes to the circuit in parallel.
  3. The screen will display the voltage value.
  4. For AC voltage and current mode, the VFD function can be enabled to reduce interference from high-frequency signals.
Mestek CM87Pro measuring AC voltage in an electrical panel and DC voltage from a car battery, highlighting VFD function.

Figure 6: Measuring AC and DC voltage with VFD function.

5.3 Resistance Measurement

  1. Turn the function knob to the 'Ω' (Resistance) position.
  2. Connect the test probes across the component to be measured.
  3. The display will show the resistance value.

5.4 Capacitance Measurement

  1. Turn the function knob to the 'Capacitance' position.
  2. Connect the test probes across the capacitor.
  3. The display will show the capacitance value.

5.5 Frequency Measurement (Hz%)

  1. Turn the function knob to the 'Hz%' (Frequency) position.
  2. Connect the test probes to the signal source.
  3. The screen displays the frequency value and the percentage of the measured signal.
Mestek CM87Pro measuring frequency in an electrical panel.

Figure 7: Measuring frequency.

5.6 Temperature Measurement

Requires the K-type thermocouple.

  1. Turn the function knob to the 'Temperature' position.
  2. Connect the K-type thermocouple to the 'COM' and 'INPUT' jacks.
  3. Place the thermocouple probe in contact with the object being measured.
  4. Briefly press the 'SEL' key to switch between Celsius (°C) and Fahrenheit (°F).
Mestek CM87Pro measuring temperature of frozen fish using a K-type thermocouple probe.

Figure 8: Performing contact temperature measurement.

5.7 Non-Contact Voltage (NCV) Detection

  1. Turn the function knob to the 'NCV' position.
  2. Bring the NCV sensing area of the instrument close to the circuit.
  3. Different alarm prompts will be sounded according to the voltage intensity, and the screen will display 'L' for low voltage or 'H' for high voltage, with a red backlight for high voltage.
Mestek CM87Pro performing NCV detection, showing low voltage and high voltage indications near a wall socket.

Figure 9: Non-contact AC voltage detection (NCV).

5.8 Live Wire Recognition (LIVE)

  1. Turn the function knob to the 'LIVE' position.
  2. Use a test probe to touch the tested conductor for detection.
  3. A red backlight will illuminate, and the buzzer will sound an alarm if the tested wire is a live wire.
Mestek CM87Pro performing live wire recognition near a wall socket, showing a red backlight and alarm.

Figure 10: Live wire recognition in action.

5.9 Diode Test

  1. Turn the function knob to the 'Diode' position.
  2. Connect the test probes across the diode.
  3. The display will show the forward voltage drop.

5.10 Continuity Test (On/Off Buzzing)

  1. Turn the function knob to the 'Continuity' position.
  2. Connect the test probes across the circuit or component.
  3. A buzzer will sound if continuity is detected (resistance is low).

5.11 Inrush Current Measurement

Measures the peak current when a machine starts.

  1. In AC current mode, press and hold the 'INRUSH' button.
  2. Clamp around a single wire of the circuit.
  3. The meter will capture and display the surge current.
  4. Always disconnect the load power supply before measuring.
Mestek CM87Pro measuring surge current during motor startup, with a diagram illustrating peak current.

Figure 11: Surge current (INRUSH) measurement.

5.12 Data Hold (HOLD)

To lock the current display reading:

  1. Short press the 'HOLD' key.
  2. The display will show 'HOLD', and a buzzer will sound every 2 seconds to indicate the locked state.
  3. Press 'HOLD' again to release the data hold function.
Mestek CM87Pro displaying a held measurement value with a 'HOLD' indicator.

Figure 12: Data Hold function.

5.13 DC Current Zero Reset (ZERO)

To eliminate baseline numbers caused by magnetic field interference in DC current mode:

  1. In DC current mode, long press the 'ZERO' key.
  2. The display will reset to 00.00A, providing more accurate measurement results.
Mestek CM87Pro display before and after DC current zero reset, showing a change from 00.39A to 00.00A.

Figure 13: DC current zero reset.

5.14 MAX/MIN Measurement

To record maximum and minimum values:

  1. In AC/DC voltage or AC/DC current mode, press the 'MAX/MIN' button.
  2. The meter will start recording the maximum and minimum values.
  3. Press again to cycle through MAX, MIN, and current readings.
Mestek CM87Pro showing its flashlight, VA color screen, and MIN/MAX measurement function.

Figure 14: Flashlight, VA color screen, and MIN/MAX measurement function.

6. Maintenance

6.1 Cleaning

Wipe the device with a dry, soft cloth. Do not use abrasives or solvents. Keep the clamp jaw free of dust and debris.

6.2 Battery Replacement

Refer to section 4.1 for battery installation/replacement instructions. Always replace batteries promptly when the low battery indicator appears to ensure accurate readings.

6.3 Storage

When not in use for extended periods, remove the batteries to prevent leakage. Store the clamp meter in its protective storage bag in a cool, dry place, away from direct sunlight and extreme temperatures.

7. Troubleshooting

  • No display or faint display: Check battery installation and replace batteries if necessary.
  • Inaccurate readings: Ensure correct mode selection, proper probe connection, and that the clamp jaw is fully closed around a single conductor for current measurements. Perform DC current zero reset if measuring DC current.
  • No NCV detection: Ensure the NCV sensing area is brought close enough to the live conductor.
  • Meter does not turn on: Check batteries. If issues persist, contact support.

8. Specifications

FeatureSpecification
Brand NameMESTEK
Model NumberCM87 Pro
DIY SuppliesElectrical
OriginMainland China
CertificationCE, FCC, RoHS
Display TypeDigital Only
Display Counts5999
Operation RangeAuto range
Full Range Overload Protection Voltage600V
Max Input Voltage (COM to INPUT)DC 1000V / AC 750V
Sample TimeAbout 3 times per second
DC Current0-1000A
AC Current0-1000A
DC Voltage0-1000V
AC Voltage0-750V
Capacitance0-60mF
Resistance0-60MΩ
Frequency Measure0-10MHZ
K-type Temperature ProbeIncluded
Inrush Current0-1000A
Overload ProtectionSupport
Continuity BuzzerYes
Jaw Opening30mm
True-RMSYes
Data HoldYes
BacklightSupport
Auto Power OffAbout 15 minutes
Backlight Auto Off TimeAbout 3 minutes
Flashlight Auto Shutdown TimeAbout 5 minutes
NCVSupport
LIVE Wire RecognitionYes
DiodeSupport
Input Impedance1MΩ
Low Battery IndicationWhen less than 2.4V
Battery2*1.5V AAA
Dimensions218 x 75 x 39mm
Weight~240g (excluding battery)
Operating Temperature0-50 °C
Storage Temperature-10-60 °C
Technical specifications table for the clamp meter, including full range overload protection, max input voltage, display type, max display value, over range display, sample time, auto shutdown time, backlight auto off time, flashlight auto shutdown time, power supply, operation temperature, storage temperature, size, jaw size, and weight.

Figure 15: Technical Parameter Table.

Comparison table of Mestek CM87A, CM87B, CM87C, and CM87Pro models, detailing DC/AC current, K-type temperature probe, DC/AC voltage, Capacitance, Resistance, Inrush current, Display counts, Operation range, Frequency, Overload protection, Continuity buzzer, Jaw opening, True-RMS, Data hold, Backlight, Auto power off, NCV, LIVE wire recognition, Diode, Input impedance, Low battery indication, and Battery.

Figure 16: Digital Clamp Meter Parameter Comparison (CM87Pro is the top version).

9. User Tips

  • For accurate current measurements, always ensure the clamp jaw is fully closed around a single conductor. Measuring multiple conductors simultaneously will result in incorrect readings.
  • When measuring DC current, use the 'ZERO' function to eliminate any residual readings caused by magnetic fields before taking your measurement.
  • The VFD function is particularly useful when working with variable frequency drives or other noisy electrical environments to get more stable AC voltage and current readings.
  • Utilize the flashlight in dimly lit areas for better visibility of test points.

10. Warranty and Support

For warranty information and technical support, please refer to the documentation provided at the point of purchase or contact the manufacturer directly. Keep your purchase receipt as proof of purchase.

Mestek CM87Pro Digital Clamp Meter User Manual - Version 1.0


Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.