1. Product Overview
The Heltec Lithium Battery Capacity Tester series, including models like HT-BCT50A, HT-DC50ABP, HT-BCT10A30V, HT-CC20ABP, HT-CC40ABP, HT-ED10AC20, and HT-ED50AC8, are designed for precise battery capacity testing. These devices are suitable for various battery types such as 18650, NCM, LiFePO4, and VRLA batteries, covering a wide range of voltages and currents.
The HT-BCT50A model, for example, is specifically designed for battery cells within 0.3-5V and 1-2000Ah. It boasts an accuracy of ±0.001A/V and integrates functions for battery capacity testing, charging, discharging, and setting work steps via an upper computer interface. It accurately collects real-time battery voltage signals, performs data storage, and processes operations. The testers are characterized by high precision, strong timeliness, simple operation, and practical reliability.


2. Product Features
- High-Precision Measurement: Automatically collects battery voltage for analysis and continuously outputs a constant current.
- Software Control: Can be controlled by software via a computer connection for convenient and efficient work.
- Adjustable Current: Current is adjustable, with peak values up to 50A (model dependent).
- Integrated Charging System: Built-in charging system for integrated control of charging and discharging.
- User-Friendly Interface: Simplified operation interface, supporting both Chinese and English languages.
- Comprehensive Safety: Includes protections against over-voltage, reverse battery connection, battery disconnection, high internal machine temperature, and over-current.
- Multi-Channel Operation (Specific Models): Some models feature multiple isolated and independently operating channels, allowing for parallel operation and voltage balancing.
- Data Logging & Analysis: Real-time data collection, storage, and graphical analysis via host computer software.
Video Demonstration:
3. Specifications
3.1 General Specifications (HT-BCT50A)
- Scope of Application: Battery cells within 0.3-5V and 1-2000Ah
- Charging and Discharging Range: 0.3 to 5V adjustable; 0.3 to 50A adjustable
- Work Steps: Charging/Discharging/Standing/Cycling (5 times per unit, 9999 times online)
- Communication: USB, Windows XP or above system, supports Chinese and English
- Auxiliary Function: Recharge or discharge equalization
- Protective Function: Battery over-voltage/reverse battery connection/battery disconnection/high temperature inside the machine/over-voltage and over-current protection
3.2 Calibration and Performance
- Calibration Equipment: Voltage standard source: Fluke 8845A; Current standard source: Guwen PCS-1000I
- Accuracy Range: Voltage ±0.1%, current ±0.1% (Accuracy guarantee period is within one year from the date of purchase)
- Heat Dissipation: Temperature-controlled fan turns on at 40°C, protected at 83°C.
- Power Supply: AC200-220V 50/60HZ (110V can be preset)
- Working Environment: Ventilate at 0-40°C to prevent heat accumulation.
- Size and Weight: Width: 167 mm, height: 165 mm, depth: 240mm, net weight: 2.6Kg
3.3 Model-Specific Specifications
| Model | Charging Range | Discharging Range | Accuracy |
|---|---|---|---|
| HT-BCT50A4C | 0.3-5V/0.5-50A | 0.3-5V/0.5-50A | V±0.1%/A±0.2% |
| HT-DC50ABP | NONE | 5-120V/1-50A | V±0.1%/A±0.1% |
| HT-BCT10A30V | 1-30V/0.5-10A | 1-30V/0.5-10A | V±0.1%/A±0.1% |
| HT-BCT50A | 0.3-5V/0.3-50A | 0.3-5V/0.3-50A | V±0.1%/A±0.1% |
| HT-CC20ABP | 9-99V/0.5-10A | 9-99V/0.5-20A | ±0.1V/±0.1A |
| HT-CC40ABP | 9-99V/0.5-20A | 9-99V/0.5-40A | ±0.1V/±0.1A |
| HT-ED10AC20 | 1-5V/0.5-10A | 1-5V/0.5-10A | ±0.02V/±0.02A |
| HT-ED50AC8 | 1-5V/0.5-50A | 1-5V/0.5-50A | ±0.02V/±0.02A |


4. Setup
4.1 Package Contents
The typical package for the HT-BCT50A includes:
- Tester Unit (e.g., HT-BCT50A)
- Testing Clips (alligator clips)
- Power Cord
- USB Data Wire
- User Manual (this document)
- Testing Fixture (optional accessory)
4.2 Device Connections
Familiarize yourself with the device's interfaces and controls before connecting. The HT-BCT50A features a front panel with a display, control knob, Start/Stop button, and power switch. The side and back panels include the power cord interface, data transmission interface, and heat dissipation vents.
Battery Connection Guidelines:
- Only use the battery clamps and connection wires compatible with the capacity tester. Do not extend or modify them.
- Both small and large alligator clips must be clamped onto the battery tabs. The small alligator clip is for voltage sampling, while the large alligator clip is for over-current.
- Ensure the contact area between the large crocodile clamp and the battery tab is sufficiently large. Avoid clamping screws, nickel sheets, or wires, as this can cause overheating, damage, or abnormal test interruption.
- The small crocodile clip must be clamped at the root of the battery terminals to ensure accurate voltage sampling and prevent inaccurate capacity testing.
4.3 Software Installation and Connection
The battery testing can be conducted through the host computer software. Follow these steps for installation and connection:
- Step 1: Find and open the installation software.
- Step 2: Choose the installation location. It is recommended not to install on the C drive.
- Step 3: After installation, open the software for the first time. Select the network (ensure public network is checked if firewall is enabled).
- Step 4: Click "Connect". In the window that appears, select the device you need to connect to. Once it turns blue, click "Connect Device". A window for "IP Settings" will appear; click "OK" by default.
After successful connection, battery information will be displayed, including current voltage, operating current, test capacity, test time, step sequence number, operating status, test status, temperature, and device model.
5. Operating Instructions
5.1 Basic Device Operation
Always turn on the machine first, then clamp the battery.
- Start Up: Connect the power cord to the back of the machine and press the power switch button on the front.
- Adjust Mode: Rotate the button to adjust the working mode. Press the Settings knob to enter the Settings page.
- Set Parameters: Rotate left and right to adjust parameters. Press to confirm. After correctly setting all parameters, save and exit.
5.2 Detailed Parameter Settings
Charging Mode Settings:
- Charging Cut-off Voltage: 2.7-2.8V for lithium titanate, 4.1-4.2V for 18650/ternary/polymer, 3.6-3.65V for lithium iron phosphate.
- Charging Current: Set to 10-20% of the battery cell capacity. Refer to battery specification sheet for optimal settings.
- Cut-off Current: When charging current is less than this value, it's judged as fully charged. Recommended to set at 0.2-1A.
Discharge Mode Settings:
- Discharge Cut-off Voltage: Lithium titanate 1.6-1.7V, 18650/ternary/polymer 2.75-2.8V, lithium iron phosphate 2.4-2.5V.
- Discharge Current: Set to 10-50% of the cell capacity. Refer to battery specification sheet for optimal settings.
Loop Mode Settings:
After setting charging and discharging parameters, set the following:
- Save Voltage: Cut-off voltage for the last charge in cycle mode, between charging and discharging cut-off voltages.
- Cycle: Define the sequence of double-charge, discharge, and charge steps.
- Standing Time: Cooling time for the battery cell in cycle mode, generally 10 minutes.
Discharge Current Setting:
- For cells <10Ah, discharge rate <1A; for 10-50Ah, <5A; for >50Ah, <20A.
- Discharge cut-off current: If discharge current is less than this value, it is judged as the end of equalization. Default is 0.01A.
5.3 Working Modes
After returning to the main page, rotate the Settings buttons left and right to switch to the desired working mode (Charge, Discharge, Cycle, or Balanced Mode). Press the Start/Stop button to enter the working state, and press it again to pause.
5.4 Data Analysis (Device Display)
Once a test is completed (with an alarm prompt), the result page will automatically pop up. Manually record the test results before proceeding to the next battery. If a cycle is repeated, press the Start/Stop button multiple times to view results for each cycle.
The device displays changes in voltage and current curves during charging and discharging. The yellow curve represents voltage variation over time, and the green curve represents current variation over time.
5.5 Host Computer Software Operation
Main Software Interface:
After installing the software and connecting the instrument, click "Connect" and then "Search". The device's IP address will appear. Click "Connect" to control the instrument. The interface will display device information, including connection status and number of connected devices. Double-clicking a device shows detailed settings.
Step Settings:
Click the "Step Settings" button to enter the interface. Create a new step by naming it, then click "Add Step" to set charging and discharging parameters, cycles, etc., according to the battery type.
Data Storage Conditions:
To save test data, ensure the corresponding option is checked. You can set the following conditions for data storage:
- Time Interval: Minimum time interval for data saving.
- Voltage Interval: Voltage difference condition for data saving.
- Current Interval: Current difference values for data storage.
Avoid setting intervals too small to prevent slow computer operation.
Startup and Control:
Right-click "Settings/Start" to select and start work steps. Ensure work steps are correctly configured. You can fill in battery batch number and remarks for easy memory and query.
Restore/Stop: The restore function allows the device to continue from the last stopped step. The stop function temporarily halts operation, which can be resumed later.
Data Analysis (Software):
After a channel has run or completed, test records are available. Select the analysis function to retrieve the record file and launch the data curve analyzer.
Data Export/Import:
Data can be imported by right-clicking the main menu grid and selecting "Analysis", or by using the "Import Data" button in the analysis software. Data can be exported as an XLS report after importing and viewing.
Multi-Curve Comparison:
The software allows comparison of differences between multiple data record files or a single data file.
6. Maintenance
- Fan Operation: The temperature-controlled fan activates at 40°C and is protected at 83°C. Regularly check and maintain the fan operation to ensure proper cooling.
- Environment: Operate the device in a well-ventilated area between 0-40°C to prevent heat accumulation.
- Cleaning: The interior of the instrument contains precision components. Prevent water or water mist from entering the instrument, as this will cause damage.
7. Troubleshooting
Note: 90% of faults are caused by poor contact between the fixture and battery tab, excessive charging/discharging current, or poor cell quality. Ensure good contact and appropriate settings.
Common Issues and Solutions:
- Why does the machine alarm?
Reasons: Reversed positive/negative poles, over-temperature, over-voltage.
Solution: Check if the fixture is clamped in reverse, if heat dissipation is normal, and if the battery voltage exceeds 5V. - Why is there no result data after the test is completed?
Reason: A work step did not run for more than 2 minutes.
Method: First, perform a single charge/discharge for 5 minutes, then proceed with the normal test process. - Why does the cut-off voltage exceed the set value during charging?
Reason 1: Poor contact between the fixture and the battery tab.
Solution: Increase contact area and ensure good contact.
Reason 2: Charging current seriously exceeds the battery's capacity.
Solution: Set parameters correctly and reasonably per battery specification sheet. - Why is there a serious deviation in the test result data?
Reason 1: Incorrect charging cut-off voltage/current settings.
Solution: Set parameters correctly and reasonably per battery specification sheet.
Reason 2: Incorrect discharge cut-off voltage/current settings.
Solution: Set parameters correctly and reasonably per battery specification sheet. - Why do test results of the same battery vary multiple times?
Reason 1: Inconsistent test environment (e.g., ambient temperature, cut-off voltages/currents, standing time).
Reason 2: Fixture influence (e.g., inconsistent contact area of clips).
Reason 3: Chemical properties of lithium batteries can be affected by subtle external factors. - Why can't the set discharge cut-off voltage be reached?
Reason 1: Discharge current is too large, causing rapid voltage drop.
Solution: Reduce the discharge current.
Reason 2: Poor contact between the fixture and the battery tab.
Solution: Increase contact area and ensure good contact. - Why can't the set charging cut-off voltage be reached?
Reason 1: Saturation current is too large, leading to incomplete charging.
Solution: Reduce the saturation current.
Reason 2: Charging current is too large, causing rapid voltage rise.
Solution: Reduce the charging current.
Reason 3: Poor battery cells (excessive self-discharge).
Solution: Phase out this battery cell.
Reason 4: Poor contact between the fixture and the battery tab.
Solution: Increase contact area to ensure good contact. - Why does the voltage rise/drop a few hours after test/equalization?
Reason 1: Due to physical characteristics, voltage naturally drops after full charge or rises after discharge.
Solution: These are inherent physical properties and cannot be changed.
8. User Tips
- Improving Test Accuracy: The accuracy of capacity testing can be affected by speed. To improve accuracy and mitigate the Pickett effect (decrease in capacity due to faster speed), it is recommended to set a smaller discharge current value.
- Battery Condition: This capacity tester is designed for healthy batteries. It cannot accurately test damaged batteries, including those with increased internal resistance, terminal voltage below discharge termination, or internal short circuits/open circuits.
- Contact Quality: Always ensure excellent contact between the testing clips and battery terminals. Poor contact is a common cause of inaccurate readings and operational issues.
9. Safety Warnings and Important Attention
General Attention:
- For batteries with capacity less than 10AH, the charging current should be ≤ 5A for safety.
- During charging and discharging, personnel must monitor the instrument and battery. Do not use unattended.
- Do not disassemble or modify the capacity tester yourself. Unauthorized modifications will void the warranty and user will be held responsible for any accidents.
- Prevent water or mist from entering the instrument.
- Data will not be recorded if operation lasts less than 2 minutes. Data recording starts after 2 minutes (overwriting previous data).
Importance Reminder:
- Parameter Settings: Always select voltage settings that match the battery type (e.g., Ternary lithium battery 2.8V, lithium iron phosphate 2.5V). Incorrect settings can damage the battery.
- Voltage Limit: Do not test batteries with a voltage exceeding 5V. Do not connect in reverse; otherwise, the warranty will be void.
- Battery Condition: Do not test batteries without a protection board, non-rechargeable batteries, zero voltage batteries, or swollen/damaged batteries.
- Lithium Battery Packs: For lithium battery packs with a port protection board, charging must be done through the charging port and discharging through the discharging port. Cycle mode is strictly prohibited.
- Radiator Surface: The radiator surface can reach high temperatures and is connected to the positive terminal of the battery. Short-circuiting and touching are strictly prohibited.
- Cut-off Settings: Charging cut-off voltage/current and discharging cut-off voltage/current must be set correctly and reasonably.
- Environment: Tests must be conducted in an open and supervised environment, free from high-temperature, flammable, explosive, or valuable items.
- Grounding: Use a well-grounded socket to prevent leakage or induced electricity.
WARNING!
The battery charging and discharging parameters must be set correctly; otherwise, it may cause the battery to explode and catch fire.
The battery under test is less than 30V, and the positive and negative terminals must not be reversed.
10. Warranty and Support
Warranty Regulations:
- Warranty Period: 1 year.
- Warranty Scope: Limited to normal use. Man-made damage, self-disassembly, modification, repair not in accordance with instructions, and damage from external force majeure factors are not covered.
- Accessories: Test fixtures are consumables and do not have a warranty.
- Service Contact: For warranty service, contact your dealer. If unable to reach the dealer, contact Heltec directly via email or phone.
- International Service: There are no after-sales service stations outside China. For international warranty service, the product must be sent to Heltec for repair, with freight costs borne by the customer.
For the complete and original user manual in PDF format, please refer to the official document: Heltec HT-BCT50A User Manual (PDF).