Victor 01S Temperature Source Signal Generator User Manual

Model: 01S | Brand: Victor

1. Introduction

The Victor 01S is a high-precision temperature source signal generator designed for calibrating and testing various temperature sensors and electrical signals. It features a basic accuracy of 0.05% and a 5-digit output display, making it suitable for professional and industrial applications. This manual provides detailed instructions for the safe and effective use of your device.

Victor 01S Temperature Source Signal Generator front view with features highlighted
Figure 1: Victor 01S Temperature Source Signal Generator with key features.

2. Safety Information

Please read and understand all safety warnings and operating instructions before using this device. Failure to do so may result in injury or damage to the instrument.

  • Always use the device within its specified voltage and current limits.
  • Do not operate the device in wet environments or near explosive gases.
  • Ensure proper connection of test leads to avoid electrical shock or damage.
  • Replace batteries promptly when the low pressure tips indicator appears.
  • Do not attempt to repair or modify the device yourself. Refer servicing to qualified personnel.

3. Product Overview

3.1. Panel Introduction

Detailed front panel diagram of Victor 01S with numbered labels
Figure 2: Front Panel Layout and Components.
  1. LCD display screen
  2. Turn temperature compensation on/off (RJ-ON button)
  3. Power on/off (Green button)
  4. Switch output function (FUNC button)
  5. Select output range (RANGE button)
  6. Output setting button (CONFIG button)
  7. Analog output resistance low-end terminal
  8. 3W: 3-wire resistance measurement terminal
  9. 4W: 4-wire resistance measurement SENSE terminal
  10. Analog output resistance high-end terminal
  11. Select temperature unit (°C/°F button)
  12. Backlight/flashlight button (Yellow button)
  13. ZERO button
  14. Configuration button (Arrow keys)
  15. V: Voltage output (+) output terminal
  16. COM: Voltage output common (-) (return) terminal

3.2. Key Features

  • HD large screen display for clear readings.
  • Flashlight lighting for use in low-light conditions.
  • Low pressure tips for battery status indication.
  • Automatic shut-down to conserve battery life.
  • Clear function (ZERO button).
  • Thermocouple (TC) function.
  • Thermal resistance (RTD) function.
  • Unit conversion (°C/°F).
  • High-precision measurement with 0.05% accuracy and 5-digit output.
Close-up details of Victor 01S features including screen, buttons, and battery compartment
Figure 3: Detailed views of the device's display, buttons, and battery compartment.

3.3. Package Contents

Contents of the Victor 01S product package
Figure 4: Included Packaging Accessories.
  1. Packing box
  2. Temperature Signal Generator (Victor 01S)
  3. Alligator clips
  4. Industrial Test Leads
  5. Latch key
  6. Thermocouple adapter
  7. User Manual
  8. Product qualification certificate

4. Setup

4.1. Battery Installation

The Victor 01S requires three (3) 1.5V AA batteries for operation. To install or replace batteries:

  1. Locate the battery compartment on the back of the device.
  2. Open the battery cover.
  3. Insert three 1.5V AA batteries, ensuring correct polarity (+/-).
  4. Close the battery cover securely.
Side view of Victor 01S showing battery compartment and kickstand
Figure 5: Battery compartment location.

4.2. Connecting Test Leads

Connect the industrial test leads or alligator clips to the appropriate terminals on the device based on the desired measurement or output function (e.g., V, COM, 3W, 4W, Ω). Ensure connections are firm and secure.

5. Operating Instructions

5.1. Basic Operation

  1. Power On/Off: Press the green power button (3) to turn the device on or off.
  2. Function Selection: Press the 'FUNC' button (4) to cycle through different output functions (e.g., DCV, OMH, TC, RTD).
  3. Range Selection: Use the 'RANGE' button (5) to select the appropriate measurement or output range for the selected function.
  4. Output Setting: Use the 'CONFIG' button (6) and arrow keys (14) to adjust the output value.
  5. Zero Function: Press the 'ZERO' button (13) to clear or zero the current reading/output.
  6. Unit Conversion: Press the '°C/°F' button (11) to switch between Celsius and Fahrenheit temperature units.
  7. Backlight/Flashlight: Press the yellow button (12) to activate the display backlight or the flashlight.

5.2. Thermocouple (TC) Simulation

The Victor 01S can simulate 11 types of thermocouple signals. Connect the thermocouple adapter and test leads to the appropriate terminals.

Diagram showing 11 types of thermocouple signals and their temperature ranges
Figure 6: Supported Thermocouple Types and Ranges.
  • R-type: 0°C ~ 1767°C
  • S-type: 0°C ~ 1767°C
  • K-type: -200°C ~ 1372°C
  • E-type: -200°C ~ 1000°C
  • J-type: -200°C ~ 1200°C
  • T-type: -250°C ~ 400°C
  • N-type: -200°C ~ 1300°C
  • B-type: 600°C ~ 1820°C
  • A-type: 0°C ~ 2500°C
  • C-type: 0°C ~ 2310°C
  • D-type: 0°C ~ 2310°C

5.3. Thermal Resistance (RTD) Simulation

The device can output 5 types of RTD signals. Connect test leads to the 3W or 4W terminals as required.

Diagram showing 5 types of thermal resistance signals and their temperature ranges
Figure 7: Supported Thermal Resistance Types and Ranges.
  • Pt100: -200°C ~ 800°C
  • Cu50: -50°C ~ 150°C
  • Pt200: -200°C ~ 630°C
  • Pt1000: -200°C ~ 630°C
  • Pt500: -200°C ~ 630°C

6. Specifications

The following table details the analog output functions and their respective ranges, resolutions, and accuracies. Specifications are valid for one year after calibration, at 23°C ±5°C, 20-70% RH, accuracy = ± (% set value + reading).

Table of analog output function specifications for Victor 01S
Figure 8: Analog Output Function Specifications.
Analog Output Function Specifications
Output FunctionMeasurement RangeOutput RangeResolutionAccuracy
DCV-10.000mV ~ 110.00mV0 ~ 1.1000V0.01mV0.05% + 0.03mV
DCV1V-0.1000V ~ 1.1000V0.0001V0.05% + 0.2mV
OMH400Ω0 ~ 400.0Ω0.1Ω0.05% + 2Ω
OMH4KΩ0 ~ 4.000KΩ0.05% + 20Ω
TC (R)0°C ~ 1767°C0°C ~ 1767°C1°C0°C ~ 100°C: 1.5°C
100°C ~ 1767°C: 1.2°C
TC (S)0°C ~ 1767°C0°C ~ 1767°C1°C0°C ~ 100°C: 1.5°C
100°C ~ 1767°C: 1.2°C
TC (K)-200°C ~ 1372°C-200°C ~ 1372°C0.1°C-200°C ~ 0°C: 0.9°C
0°C ~ 1000°C: 0.6°C
1000°C ~ 1372°C: 0.7°C
TC (E)-200°C ~ 1000°C-200°C ~ 1000°C0.1°C-200°C ~ 0°C: 0.8°C
0°C ~ 800°C: 0.5°C
800°C ~ 1000°C: 0.6°C
TC (J)-200°C ~ 1200°C-200°C ~ 1200°C0.1°C-200°C ~ 0°C: 0.8°C
0°C ~ 1000°C: 0.5°C
1000°C ~ 1200°C: 0.6°C
TC (T)-250.0°C ~ 400.0°C-250.0°C ~ 400.0°C0.1°C-250.0°C ~ -100°C: 0.6°C
-100°C ~ 400°C: 0.5°C
TC (N)-200°C ~ 1300°C-200°C ~ 1300°C0.1°C-200°C ~ 0°C: 0.8°C
0°C ~ 1000°C: 0.6°C
1000°C ~ 1300°C: 0.7°C
TC (B)600°C ~ 1820°C600°C ~ 1820°C1°C600°C ~ 1500°C: 2.0°C
1500°C ~ 1820°C: 2.4°C
TC (A)0°C ~ 2500°C0°C ~ 2500°C1°C0°C ~ 1600°C: 2.0°C
1600°C ~ 2500°C: 3.0°C
TC (C)0°C ~ 2310°C0°C ~ 2310°C1°C0°C ~ 270°C: 2.5°C
270°C ~ 2310°C: 3.0°C
TC (D)0°C ~ 2310°C0°C ~ 2310°C1°C0°C ~ 270°C: 2.5°C
270°C ~ 2310°C: 3.0°C
RTD (Pt100(385))-200°C ~ 800°C-200°C ~ 800°C0.1°C-200°C ~ 0°C: 0.3°C
0°C ~ 600°C: 0.2°C
600°C ~ 800°C: 0.3°C
RTD (Pt200(385))-200°C ~ 630°C-200°C ~ 630°C0.1°C-200°C ~ 0°C: 0.3°C
0°C ~ 630°C: 0.2°C
RTD (Pt500(385))-200°C ~ 630°C-200°C ~ 630°C0.1°C-200°C ~ 0°C: 0.3°C
0°C ~ 630°C: 0.2°C
RTD (Pt1000(385))-200°C ~ 630°C-200°C ~ 630°C0.1°C-200°C ~ 0°C: 0.3°C
0°C ~ 630°C: 0.2°C
RTD (Cu50)-50°C ~ 150°C-50°C ~ 150°C0.1°C-50°C ~ 150°C: 0.6°C

6.1. General Specifications

  • Measurement Accuracy: 0.05%
  • Battery Type: 3 x 1.5V AA batteries (not included)
  • Maximum Applied Voltage at Output Terminal: Approximately 30Vpk
  • Maximum Applied Current at Output Terminal: Approximately 25mA
  • Load Characteristics: Capacitive load ≥0.01uF (DCV/OHM/TC/RTD)
  • Temperature Coefficient: 0.1 × Basic accuracy/°C (Temperature range <18°C or >28°C)
  • Dimensions (L x W x H): 24 cm x 19 cm x 7 cm (approximate package dimensions)
  • Weight: 0.799 kg (approximate package weight)

7. Maintenance

  • Cleaning: Use a soft, damp cloth to clean the device. Do not use abrasive cleaners or solvents.
  • Battery Replacement: Replace batteries when the low battery indicator appears to ensure accurate readings and prevent data loss.
  • Storage: Store the device in a cool, dry place away from direct sunlight and extreme temperatures. Remove batteries if storing for extended periods.
  • Calibration: For optimal accuracy, periodic calibration by a qualified service center is recommended, especially after one year of use.

8. Troubleshooting

  • Device does not power on: Check battery installation and ensure batteries are not depleted.
  • Inaccurate readings: Verify correct function and range selection. Ensure test leads are properly connected and not damaged. Consider recalibration if issues persist.
  • Display is dim: Activate the backlight using the yellow button. If still dim, replace batteries.
  • No output signal: Check output settings and connections. Ensure the device is in the correct output mode.

9. User Tips

While specific user tips from reviews or Q&A are not available, here are some general recommendations for using the Victor 01S:

  • Always double-check your connections before applying power or generating a signal to prevent damage to the device or the circuit under test.
  • Familiarize yourself with the different thermocouple and RTD types and their respective temperature ranges to ensure you select the correct setting for your application.
  • Utilize the automatic shut-down feature to prolong battery life, especially during prolonged periods of inactivity.
  • Keep the device clean and free from dust and moisture to maintain its performance and longevity.

10. Manufacturing Process Video

Watch this video to see the manufacturing and quality control processes for Victor products, ensuring high standards and reliability.

Video: Victor Product Manufacturing and Testing Process.

11. Warranty and Support

For warranty information, technical support, or service inquiries, please contact your authorized Victor dealer or the manufacturer directly. Keep your purchase receipt and product qualification certificate for reference.

Related Documents - 01S

Preview VICTOR 4105A Digital Grounding Resistance Meter Instruction Manual
Comprehensive instruction manual for the VICTOR 4105A Digital Earth Tester, covering overview, safety precautions, work principle, technical specifications, measurement methods, battery installation, and troubleshooting.
Preview VICTOR Electronic Testing and Measuring Instruments Catalog | Digital Multimeters, Oscilloscopes, and More
Explore the comprehensive range of VICTOR electronic testing and measuring instruments, including digital multimeters, clamp meters, oscilloscopes, function generators, and more. High-quality, reliable equipment for professional use.
Preview Victor 8145C Dual Display Digital Multimeter User Manual and Specifications
Comprehensive user manual for the Victor 8145C Dual Display Digital Multimeter, covering features, operation, safety, specifications, and maintenance. Includes detailed instructions and technical data.
Preview Victor 1226 Calculator Quick Start Guide
A quick start guide for the Victor 1226 calculator, covering setup, paper loading, time and date setting, troubleshooting, and available accessories.
Preview Victor 6013B Smart SMD Component Tester Instruction Manual
Detailed instruction manual for the Victor 6013B Smart SMD component tester, covering its introduction, safety rules, button operations, technical specifications, testing procedures for resistors, capacitors, and diodes, and maintenance guidelines.
Preview Victor Personal Massager User Manual and Joyhub App Guide
Comprehensive user manual for the Victor personal massager, detailing product features, app connectivity with Joyhub, usage instructions, care, warnings, and warranty information. Supports multiple languages.