Hantek 6254BD 6074BD 6104BD 6204BD

Contents

Hantek 4-Channel Digital Oscilloscope User Manual

Models: 6254BD, 6074BD, 6104BD, 6204BD

Brand: Hantek

1. Introduction

This manual provides comprehensive instructions for the Hantek 4-Channel Digital Oscilloscope models 6254BD, 6074BD, 6104BD, and 6204BD. These portable oscilloscopes are designed for production testing, research and design, and troubleshooting of analog circuits. They also feature an integrated arbitrary waveform signal generator.

Please read this manual thoroughly before operating the device to ensure safe and correct usage.

2. General Safety Summary

Review the following safety precautions carefully before operating the device to avoid personal injuries or damage to the device and any connected products. Use the device only as specified in this user's guide.

2.1 To Avoid Fire or Personal Injury

  • Use Proper Power Cord: Use only the power cord specified for this product and certified for the country of use.
  • Connect and Disconnect Properly: Do not connect or disconnect probes or test leads while they are connected to a voltage source. Connect the probe output to the measurement device before connecting the probe to the circuit under test. Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement device.
  • Observe All Terminal Ratings: To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
  • Use Proper Probe: To avoid shock hazard, use a properly rated probe for your measurement.
  • Avoid Circuit or Wire Exposure: Do not touch exposed connections and components when power is on.
  • Do Not Operate With Suspected Failures: If suspected damage occurs with the device, have it inspected by qualified service personnel before further operations.
  • Provide Proper Ventilation: Refer to the installation instructions for proper ventilation of the device.
  • Do not operate in Wet/Damp Conditions.
  • Do not operate in an Explosive Atmosphere.
  • Keep Product Surfaces Clean and Dry.

3. System Requirements

To run the oscilloscope software, your computer configuration should meet the following requirements:

3.1 Minimum System Requirements

  • Operating System: Windows XP/Vista/Win7
  • Processor: Upwards of 1.00G processor
  • Memory: 256MB
  • Disk Space: 500MB free space
  • Screen Resolution: 800 x 600
  • Operating System: Windows XP SP3 System
  • Processor: 2.4G Processor
  • Memory: 1GB
  • Disk Space: 80GB
  • Screen Resolution: 1024 x 768 or 1280 x 1024
  • DPI Setting: Normal Size (96DPI)

4. Setup and Installation

4.1 Software Installation

Caution: You must install the software before using the oscilloscope.

  1. While in Windows, insert the installation CD into the CD-ROM drive.
  2. The installation should start up automatically. If not, navigate to the CD-ROM drive in Windows Explorer and run Setup.exe.
  3. The software installation wizard will start. Click 'Next' to continue.
  4. Choose a destination directory. Click 'Next' to continue.
  5. Review the setup information. Click 'Next' to start copying files.
  6. The installation status dialog will be displayed during file copying.
  7. The system configuration will be updated.
  8. The installation is complete.
Hantek oscilloscope software installation screenshot
Screenshot of the Hantek oscilloscope software installation process.

4.2 Driver Installation

Example for HANTEK 6000B:

  1. Connect the A-Type Plug of the USB cable to your PC's USB port.
  2. Connect the other A-Type Plug of the USB cable to the device's USB port.
  3. When "New hardware is found" appears, right-click and select "Manage".
  4. In Device Manager, locate the "DSO Device" under "Other devices". Right-click it and select "Update Driver Software".
  5. Select "Browse my computer for driver software".
  6. Navigate to the driver path (e.g., C:\Program Files\HantekXXXX\Driver) and click 'Next'.
  7. If prompted by Windows Security, click "Install this driver software anyway".
  8. The driver software will install.
  9. Once successfully installed, the device manager will show "Hantek6000B" under "Universal Serial Bus controllers".
Hantek oscilloscope USB connection diagram
Diagram showing USB cable connection to the oscilloscope and PC.
Hantek oscilloscope rear panel with USB ports
Rear panel of the Hantek oscilloscope showing USBXI and USBIA ports.
Hantek oscilloscope front panel with BNC connectors
Front panel of the Hantek oscilloscope with CH1, CH2, CH3, CH4 BNC input connectors.

5. General Check and Function Verification

After receiving your oscilloscope, perform the following checks:

5.1 Check the Shipping Container for Damage

Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically.

5.2 Check the Accessories

Verify that all accessories listed in the "Accessories" section of this guide are present and undamaged. If any contents are incomplete or damaged, notify the franchiser.

5.3 Check the Instrument

In case there is any mechanical damage or defect, or if the instrument does not operate properly or fails performance tests, please notify the franchiser.

5.4 Probe Compensation

Perform this function to match the characteristics of the probe and the channel input. This should be performed whenever attaching a probe to any input channel for the first time.

  1. Set the attenuation switch on the probe to 10X. Connect the probe to CH1 of the oscilloscope. Ensure the probe hook-tip is firmly inserted onto the probe.
  2. Attach the probe tip to the Probe Compensator and the reference lead to the ground connector.
  3. Select CH1 and then press the "AUTOSET" button in the menu or toolbar.
  4. Check the shape of the displayed waveform.
Probe compensation diagram showing correctly compensated, over compensated, and under compensated waveforms
Waveform examples for correctly compensated, over compensated, and under compensated probes.

WARNING: To avoid electric shock while using the probe, ensure the insulation of the cable is perfect, and do not touch the metallic portions of the probe head while it is connected to a voltage source.

5.5 Self Calibration

The self-calibration routine optimizes the oscilloscope signal path for maximum measurement accuracy. Run this routine if the ambient temperature changes by 5°C or more. For accurate calibration, power on the oscilloscope and wait twenty minutes to ensure it is warmed up. Disconnect any probes or cables from the input connectors. Then, access the "Utility -> Calibration" option and follow the on-screen directions. The self-calibration routine takes several minutes.

6. Operating Basics

After software and driver installation, launch the Hantek oscilloscope software. The user interface will be displayed.

6.1 User Interface Overview

The software interface displays waveforms and provides control settings. Key elements include:

  • Main Menu: All settings are accessible here.
  • Toolbar: Quick access to common functions.
  • Trigger Information: Displays edge trigger slope, source, and level.
  • DDS Panel (for HANTEK 6000BD models): For function/arbitrary waveform generator controls.
  • Horizontal Panel: Adjust Time/Div and format.
  • Vertical Panel: Turn channels ON/OFF, adjust Volt/Div, coupling, and probe attenuation.
  • Trigger Panel: Change trigger mode, sweep, source, and slope.
  • System Time: Displays current time.
  • Edge Trigger Level Marker: Indicates the trigger level on the waveform.
  • Main Time Base Setting: Displays the current time base.
  • CH1/CH2/CH3/CH4 Information: Shows channel coupling, vertical scale factors, and bandwidth limits.
  • Connect Status: Indicates connection status.
  • Reference Point Markers: Shows reference points for displayed waveforms.
  • Horizontal Trigger Position Marker: Indicates horizontal trigger position.
  • Trigger Status: AUTO (acquiring without trigger), Trig'D (trigger detected), STOP (acquisition stopped), RUN (acquiring).
Hantek oscilloscope software user interface screenshot
Overview of the Hantek oscilloscope software user interface.

6.2 Menu System

The main menu provides access to various functions:

  • File: Load or save data, setup, images, print.
  • View: Change user interface elements (Toolbar, Status Bar, Side Bar).
  • Setup: Configure reference, math, trigger, vertical, and horizontal settings.
  • Display: Change waveform display type (Grid, Intensity, Grid Background Color).
  • Cursor: Set cursor measurement type (Source, Type).
  • Measure: Set measurement parameters.
  • Acquire: Control run/stop, zoom, buffer length, acquisition mode, interpolation, autoset.
  • Utility: Access pass/fail, F/C, calibration, factory setup, language.
  • Vehicle: Diagnosis setup, clear reference waveform.
  • Window: Arrange windows (Cascade, Vertical, Horizontal, Arrange Icons).
  • Help: Access help file.

6.3 Vertical System Setup

To configure vertical settings:

  1. Click "Setup -> Vertical" in the main menu.
  2. Select Channel: Choose CH1, CH2, CH3, or CH4.
  3. ON/OFF: Turn the selected channel on or off.
  4. VOLTS/DIV: Set the voltage range per division.
  5. Coupling: Set to DC, AC, or GND. DC blocks AC components, AC blocks DC components, GND disconnects the input.
  6. Probe: Set the probe attenuation factor (e.g., x1, x10) to ensure correct vertical scale reading.
  7. BW Limit: Reject frequency components higher than 20MHz.
  8. Invert: Invert the selected waveform.

6.4 Horizontal System Setup

To configure horizontal settings:

  1. Click "Setup -> Horizontal" in the main menu.
  2. Time/DIV: Set the time base parameters (scale factor for main or window time base).
  3. Format: Set the waveform display format (Y-T, X-Y). Y-T shows voltage vs. time, X-Y shows CH1 vs. CH2.
  4. Horizontal Position: Double-click the channel button to set the trigger point to the horizontal center of the screen. Drag the 'T' marker on screen to change the horizontal position.

6.5 Trigger System Setup

To configure trigger settings:

  1. Click "Setup -> Trigger" in the main menu or the trigger icon in the toolbar.
  2. Trigger Mode: Sets the trigger mode (Edge, Pulse, Video, Alternative).
  3. Trigger Sweep: Selects AUTO (acquires waveform even if no trigger), NORMAL (acquires only when triggered), or SINGLE (acquires once then stops).
  4. Trigger Source: Selects the trigger source (CH1, CH2, CH3, CH4).
  5. Trigger Slope: Selects the edge trigger slope (Positive or Negative).
  6. Pulse Trigger: Detects abnormal signals by setting pulse width conditions (+Less, +Equal, +More, -Less, -Equal, -More).
  7. Video Trigger: Triggers on NTSC, PAL, or SECAM video signals.
  8. ALT System: Allows setting trigger type to Edge or Pulse.
  9. HF Rejection: Eliminates trigger higher-frequency (20MHz above).

7. Measurements

7.1 Simple Measurement

To acquire and display a signal:

  1. Connect the signal to CH1 using a probe.
  2. Click the "AUTOSET" button on the toolbar or in the "Acquire" menu. The oscilloscope will automatically adjust settings for a stable display.
  3. To measure frequency and Vpp:
    • Click "Measure -> Horizontal -> Frequency".
    • Click "Measure -> Vertical -> Peak-to-Peak".
  4. To clear measurements, click "Measure -> Clear Measure".
Hantek oscilloscope software displaying sine wave with measurements
Example of a sine wave displayed on the oscilloscope software with measurements.

7.2 Cursor Measurements

Use cursors for quick time and amplitude measurements.

7.2.1 Measuring Peak Frequency or Time (Vertical Cursors)

  1. Click "Cursor -> Source", select the desired channel (e.g., CH1).
  2. Click "Cursor -> Type", select "Vertical".
  3. Click and drag the left mouse button to position the vertical cursors.
  4. The frequency and time difference will be shown in the status bar.

7.2.2 Measuring Amplitude (Horizontal Cursors)

  1. Click "Cursor -> Source", select the desired channel.
  2. Click "Cursor -> Type", select "Horizontal".
  3. Click and drag the left mouse button to position the horizontal cursors.
  4. The voltage difference will be shown in the status bar.

7.2.3 Tracing Amplitude at a Fixed X-axis Position

  1. Click "Cursor -> Source", select the desired channel.
  2. Click "Cursor -> Type", select "Trace".
  3. Click the cursor at the desired X-axis position on the waveform.

7.3 Automatic Measurements

The oscilloscope provides 20 parametric auto measurements (12 voltage and 8 time measurements).

  1. Click "Measure" in the main menu.
  2. Source: Select the channel for measurement (CH1, CH2, CH3, CH4, or MATH).
  3. Vertical Measurements: Maximum, Minimum, Peak To Peak, Top, Base, Middle, RMS, Amplitude, Mean, Cycle Mean, Preshoot, Overshoot.
  4. Horizontal Measurements: Period, Frequency, Rise Time, Fall Time, +Duty Cycle, -Duty Cycle, +Pulse Width, -Pulse Width.
  5. Clear Measure: Clear all displayed measurements.
  6. Edit Measure: Select up to 8 measurements to display simultaneously.

8. Advanced Functions

8.1 Math Functions

The oscilloscope supports various mathematical operations on waveforms.

  1. Click "MATH" in the Channel menu to set up the MATH channel.
  2. ON/OFF: Turn the MATH channel on or off.
  3. Source A/B: Set the input sources for the math operation.
  4. Operate: Select the operation type:
    • A + B: Add source A and source B.
    • A - B: Subtract source B from source A.
    • A x B: Multiply source A by source B.
    • A / B: Divide source A by source B.
    • FFT: Convert a time-domain signal into its frequency components (spectrum).
  5. Volt/DIV: Set the resolution of the math channel.
  6. Probe: Set the math channel probe attenuation.
  7. Invert: Invert the math channel waveform.
Hantek oscilloscope software displaying multiple waveforms with math functions
Example of multiple waveforms displayed with math functions applied.

8.2 Reference Waveforms

You can save and load reference waveforms for comparison.

  1. Click "REF" in the "Setup" menu.
  2. ON/OFF: Turn the reference channel on or off.
  3. Volt/DIV: Set the resolution of the reference channel.
  4. Load: Load a reference waveform from a ".rfc" file.
  5. Save: Save the current waveform as a ".rfc" file.

8.3 Zoom In/Out and Drag Waveforms

To analyze specific parts of a waveform:

  1. Click "Stop" to halt waveform acquisition.
  2. Use "Zoom Out" or "Zoom In" in the "Acquire" menu, or left/right click on the display screen.
  3. Adjust "Time/Div" in the Horizontal menu or panel to zoom.
  4. To drag the waveform, use the "Drag" function in the "Acquire" menu to modify its position.

8.4 Acquisition Modes

The oscilloscope offers different acquisition modes:

  • Normal: Samples the signal at evenly spaced intervals.
  • Average: Acquires multiple waveforms, averages them, and displays the result to reduce random noise.
  • Peak Detect: Finds the maximum and minimum values in each sampling interval to show the waveform.

8.5 X-Y Operation

The X-Y plot analyzes the correlation between two channels, displaying a Lissajous diagram. This is useful for comparing frequencies, amplitudes, and phases.

  1. Set probe attenuation to "x10" for both probes.
  2. Connect CH1 probe to the input and CH2 probe to the output of the circuit.
  3. Click the "AUTOSET" button.
  4. Adjust vertical scale and offset for similar amplitude signals on both channels.
  5. Select "X-Y" format in the Horizontal window.
  6. Apply the Ellipse Method to observe phase difference (Sine = A/B or C/D).

8.6 Waveform Generator (for HANTEK 6000BD models)

The HANTEK 6000BD models include a function/arbitrary waveform generator.

  1. Select the DDS control panel in the sidebar.
  2. On/Off: Enable or disable waveform output.
  3. Sync: Synchronize the output.
  4. Wave Type: Select from Sine, Square, AM/FM, Ramp, Trapezia, Gause, Arb, Exponent, White.
  5. Parameters: Adjust Frequency (0-75MHz), Amplitude (0-3.5V), Y Offset (-7-7V), Phase (0.0-1.0), and Duty Cycle (0.0-1.0).
Hantek oscilloscope software waveform generator interface
Screenshot of the waveform generator interface.

Video demonstrating the use of the Hantek oscilloscope for automotive EGR valve testing.

8.7 Pass/Fail Test

The Pass/Fail function monitors signals and outputs pass or fail signals by comparing the input signal with a pre-created mask.

  1. Click "Pass/Fail" in the "Utility" menu.
  2. Control: Turn ON/OFF, select Source (CH1, CH2, CH3, CH4), Output (FAIL/PASS), and Stop When Output.
  3. Mask Setting: Set Vertical and Horizontal limit ranges.
  4. Create: Create the Pass/Fail area according to the mask.
  5. Save/Load: Save or load mask setups.
  6. Operate: Click "Start" to begin the test, "Stop" to end.

8.8 Waveform Record and Replay

The record function allows you to capture and replay input waveforms.

  1. Click "Record" in the "Utility" menu.
  2. Path: Specify the directory to save the recorded data.
  3. Storage: Record waveform data as a ".dfr" file. The maximum file length is about 1GB.
  4. Play Back: Click to start playing back recorded waveforms.
  5. Interval(S): Set the recording interval.
  6. Start/Stop: Start or stop recording waveforms.

9. Maintenance

9.1 General Care

Do not store or leave the oscilloscope where it will be exposed to direct sunlight for long periods of time.

9.2 Cleaning

Inspect the device and probes as often as operating conditions require. Ensure the device is disconnected from all power sources before cleaning.

  1. Remove loose dust on the outside of the oscilloscope and probes with a lint-free cloth. Be careful to avoid scratching the clear glass display filter.
  2. Use a soft cloth dampened with water to clean the device.

Caution: To avoid damage to the device or probes, do not expose them to sprays, liquids, or solvents. Do not use any abrasive or chemical cleaning agents on the surface of the device or probes.

10. Troubleshooting

If you encounter issues with your Hantek Digital Oscilloscope, consider the following steps:

  • Software Not Launching: Ensure the software and drivers are correctly installed as per sections 4.1 and 4.2. Check system requirements (section 3).
  • No Waveform Display:
    • Verify the USB connection between the oscilloscope and your PC.
    • Perform a "Function Check" (section 5.4) to ensure the oscilloscope is operating correctly.
    • Check probe connections and compensation (section 5.4).
    • Use the "AUTOSET" function (section 7.1) to automatically adjust settings.
    • Ensure the correct channel is selected and turned ON in the Vertical System settings (section 6.3).
    • Check trigger settings (section 6.5) if the waveform is unstable or not appearing.
  • Inaccurate Measurements:
    • Perform "Probe Compensation" (section 5.4).
    • Run "Self Calibration" (section 5.5), especially if the ambient temperature has changed significantly.
    • Verify probe attenuation settings in the Vertical System (section 6.3).
  • Waveform Generator Issues: Ensure the correct wave type and parameters are set in the Waveform Generator panel (section 8.6).
  • Software Freezing/Crashing: Ensure your computer meets the recommended system requirements (section 3.2). Close other demanding applications.

If problems persist, refer to the full user manual PDF or contact Hantek support.

11. Specifications

CategoryParameterValue
VerticalAnalog Channels4
Bandwidth70MHz/100MHz/200MHz/250MHz (-3dB)
Input Impedance1MΩ; Capacitance: 25 pF±3pF
Input Sensitivity2mV/div to 10V/div
Input CouplingAC/DC/GND
Vertical Resolution8 bits
Memory Depth64K/CH, 32K/2CH, 16K/4CH
Max Input Protection400V (DC+AC Peak)
DC Gain Accuracy±3%
Bandwidth Limit20MHz
HorizontalReal-Time Sampling Rate1GS/s
Time Base Range2ns/div to 1000s/div (1-2-5 sequences)
Time Base Precision±50ppm
TriggerSourceCH1, CH2, CH3, CH4
ModeEdge, Pulse, Video, Alternative
X-Y ModeX-Axis Input: CH1, Y-Axis Input: CH2, Phase Shift: Max. 3 degree
Waveform Signal ProcessOperations+, -, x, ÷, FFT, Invert
Cursors MeasurementCross, Trace, Horizontal, Vertical
Auto MeasurementVpp, Vamp, Vmax, Vmin, Vtop, Vmid, Vbase, Vavg, Vrms, Vcrms, Preshoot, Overshoot, Frequency, Period, Rise Time, Fall Time, Positive Width, Negative Width, Duty Cycle
Arbitrary Waveform Generator (6000BD)Waveform FrequencyDC~25MHz
DAC2K~200MHz adjustable
Frequency Resolution0.10%
Wave Amplitude±3.5V Max.
General FeaturesInterfaceUSB 2.0
Power SourceDual port USB
Volume175mm x 105mm x 25mm
Weight0.9KG

12. Accessories

The standard accessories included with your Hantek Digital Oscilloscope are:

  • Oscilloscope Host: 1 unit
  • USB Cable: 1 unit (for connecting to PC)
  • Passive Probes: 2 units (1.5m, 1:1/10:1 attenuation, rated 100Vrms CAT III in X1, 300Vrms CAT II in X10)
  • Alligator Clip Test Leads: 2 units
  • Signal Source Test Line: 1 unit
  • Software Installation CD: Contains software and user manual.
  • Quick Guide: 1 unit
  • BNC to BNC Cable: 1 unit (for HANTEK 6000BD models)
  • HT25: 1 unit (for HANTEK 6000BE models)
Hantek oscilloscope and included accessories
Hantek oscilloscope with probes, USB cable, and other accessories.

13. User Tips

  • For optimal performance, ensure your PC meets the recommended system requirements, especially for memory and processor speed.
  • Always perform probe compensation and self-calibration for accurate measurements, particularly after significant temperature changes.
  • When using the waveform generator, experiment with different wave types and parameters to understand their effects on the output signal.
  • Utilize the "Average" acquisition mode to reduce noise in repetitive signals for clearer analysis.
  • The X-Y mode is excellent for phase and frequency comparisons between two signals.

14. Support and Further Information

For the latest software, drivers, and detailed product information, please visit the official Hantek website:

For technical assistance or warranty information, please contact your local distributor or Hantek customer support through their official website.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.