CERTEON CKHW

Contents

Infrared Temperature Sensor CKHW Series User Manual

Model: CKHW | Brand: CERTEON

1. Introduction

The CERTEON CKHW Series Infrared Temperature Sensor is a non-contact temperature measurement device designed for continuous industrial monitoring. It detects infrared radiation emitted from the target surface and converts it into a standard industrial output signal. This sensor is suitable for integration into automation systems via 4-20 mA, voltage, or serial communication outputs. It is primarily intended for non-metal surface temperature measurement in industrial environments.

Infrared Temperature Sensor with display showing temperature range, output signals, and dimensions.
Figure 1: CERTEON CKHW Series Infrared Temperature Sensor with associated display unit.

2. Specifications

The following table details the key specifications of the CKHW Series Infrared Temperature Sensor:

ParameterDetails
BrandCERTEON
ModelCKHW
Material304 Stainless Steel
Temperature Range-30°C to 1300°C (customizable)
Output Signals4-20mA (standard), customizable (RS485, RS232, 0-5V, 0-10V)
Power Supply12-36V DC (standard), customizable 5-36V DC
SizeM18 x 1 threaded body, thread length 40 mm, overall length 113 mm
Cable Length1.5m (standard), customizable
Protection ClassIP65
Optical ResolutionHigher D:S ratio means better accuracy
Measurement PrincipleInfrared radiation sensing for surface temperature calculation
Target RequirementMinimum diameter 10 mm; stationary or slow-moving objects only
Product dimensions diagram showing sensor length of 98mm and diameter of 18.7mm with M18x1 thread.
Figure 2: Product Dimensions Diagram.
Diagram illustrating the principle of infrared temperature sensing, showing a measured object emitting infrared rays detected by the sensor.
Figure 3: Infrared Radiation Sensing Principle.

3. Applications

The CKHW Series Infrared Temperature Sensor is suitable for a wide range of industrial applications, including:

  • Food processing
  • Textiles
  • Plastics
  • Ceramics
  • Machinery such as tea dryers and pancake machines
  • Equipment involving coal or graphite
  • Measurement of surfaces like glass, water, and chemically treated metals
Image showing various applications of the infrared temperature sensor in different industrial settings.
Figure 4: Diverse industrial applications of the sensor.

4. Features

Key features of the CKHW Series Infrared Temperature Sensor include:

  • Advanced infrared optical lens for improved accuracy.
  • Millisecond-level response speed with multiple output options.
  • Non-contact design for hard-to-reach or mobile targets.
  • Compatible with a wide range of accessories for added functionality.
Diagram showing the sensor probe, quick response, and non-contact measurement suitable for various non-metallic media.
Figure 5: Non-contact infrared temperature measurement capability.

5. Setup

5.1 Preparation

  • Ensure the selected temperature range and output signal match your application requirements.
  • Confirm that the power supply voltage complies with the sensor rating (12-36V DC standard, customizable 5-36V DC).
  • Verify that the target surface is non-metallic. If measuring metal surfaces, apply high-temperature resistant paint or tape to the measurement spot to ensure accurate emissivity.
  • Ensure the target diameter exceeds 10 mm.

5.2 Installation

  • Mount the sensor using the M18 threaded connection. Secure firmly to avoid vibration.
  • Maintain a clear optical path between the sensor lens and the target surface.
  • Avoid installation near strong electromagnetic interference sources.
  • Connect wiring according to the selected output configuration. For 4-20 mA two-wire systems, connect in series with the control loop according to industrial standards.
Infrared temperature sensor mounted with a bracket, showing threaded fixing and IP65 protection.
Figure 6: Sensor mounted with bracket, highlighting threaded fixing and IP65 protection.
Diagram showing the sensor probe, threaded fixing, IP65 protection, and stable mounting brackets.
Figure 7: Convenient temperature measuring instrument features.

6. Operating Instructions

6.1 Operation

  • Power on the device after completing wiring checks.
  • Allow stabilization time as required by the system.
  • Align the sensor perpendicular to the target surface for optimal accuracy.
  • Ensure the object is stationary or moving slowly. The device is not suitable for high-speed moving targets.

6.2 Data Reading

  • For 4-20 mA output, interpret the current signal according to the configured temperature range.
  • For voltage or RS485/RS232 output, configure the receiving device or PLC according to communication protocol settings provided by the seller.
  • Use compatible display or control instruments if required.
Diagram showing the internal low-power circuit board of the sensor, emphasizing anti-interference and stable output.
Figure 8: Internal circuit board design for stable output.

7. Maintenance

  • Keep the optical lens clean. Use a soft, dry cloth or industrial lens tissue for cleaning. Do not use corrosive solvents.
  • Store the device in a dry and ventilated environment.
  • Avoid mechanical shock during transportation and storage.
  • If the device will not be used for an extended period, disconnect from power supply and store in original packaging.
  • Do not disassemble the device.

8. Troubleshooting

This section addresses common issues and their potential solutions.

8.1 Common Issues & Solutions

  • Inaccurate Readings on Metal Surfaces: This sensor is designed for non-metal surfaces. For metal measurement, surface treatment (e.g., applying high-temperature resistant paint or tape) is required to ensure proper emissivity.
  • No Reading for Fast-Moving Objects: The device is not suitable for high-speed moving targets. Ensure the target is stationary or moving slowly.
  • No Output Signal: Check wiring connections according to the selected output configuration. Ensure the power supply voltage is within the rated range (12-36V DC).
  • Sensor Not Powering On: Verify the power supply connection and voltage. Ensure correct polarity if applicable.

9. User Tips

  • For optimal performance, always ensure the target object has a minimum diameter of 10mm.
  • If you require a specific temperature range outside the standard options, contact the seller for customizable options between -30°C and 1300°C.
  • Consider purchasing optional accessories like intelligent display controllers, cooling shields, various mounts, laser aimers, or sound-light alarms for enhanced functionality and protection in specific environments. These accessories are sold separately.
  • Always confirm the required temperature range and output signal before placing an order to ensure the sensor meets your specific needs.

10. Warranty and Support

This document is provided for product instruction purposes only and does not constitute legal advice. The seller is responsible for ensuring compliance with destination market regulations.

10.1 Important Warnings (EU Market Reference)

  • This device operates on DC power. Incorrect wiring may cause electrical hazard or equipment damage.
  • Do not use in explosive atmospheres unless certified accordingly.
  • Infrared sensors do not measure internal temperature; surface emissivity affects accuracy.
  • Do not stare into optional laser aiming devices if equipped.
  • Keep out of reach of children.
  • Proper disposal: Follow WEEE directives for electronic waste.
  • Note: The above warning information is for reference only. Final warning language must comply with the official language requirements of the destination EU country.

10.2 Further Information and Support

For detailed item guides and documents, please refer to the official user manual PDF: View User Manual (PDF).

For technical consultations or customized product requirements, please contact CERTEON Automation Store directly:

Contact information for CERTEON Automation Store, including WhatsApp and Email.
Figure 9: CERTEON Automation Store Contact Information.

© 2024 CERTEON. All rights reserved.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.