Yieryi 5-in-1 Water Quality Tester YG-9909 User Manual

Model: YG-9909

Yieryi 5-in-1 Water Quality Tester YG-9909

The Yieryi YG-9909 is a versatile device for measuring pH, Salinity, TDS, EC, and Temperature in various water applications.

1. Setup

1.1 What's in the Box?

  • 1 x Yieryi 5-in-1 Water Quality Tester (YG-9909)
  • 1 x User Manual
  • 1 x pH 4.00 Calibration Powder
  • 1 x pH 6.86 Calibration Powder
  • 1 x pH 9.18 Calibration Powder
  • 1 x Screwdriver
Package contents of Yieryi YG-9909 tester

Image: Contents of the YG-9909 package, including the tester, calibration powders, screwdriver, and manual.

1.2 Battery Installation

The YG-9909 requires 3 x 1.5V LR44 coin cell batteries (not included).

  1. Twist and remove the battery cap located at the top of the device.
  2. Insert three LR44 batteries, ensuring correct polarity (+/-).
  3. Replace the battery cap and twist to secure it.
Battery compartment of Yieryi YG-9909

Image: Close-up of the battery compartment showing the placement of LR44 batteries.

1.3 Electrode Replacement

The electrode of the YG-9909 is replaceable. Follow these steps:

  1. Using the provided screwdriver, unscrew the four small screws around the electrode base.
  2. Carefully remove the screw sleeve.
  3. Unplug the old electrode from the main unit.
  4. Plug in the new electrode, ensuring it is seated correctly.
  5. Reattach the screw sleeve and tighten the four screws.

Video: Product demonstration showing electrode replacement (from 0:14 to 0:28).

2. Operating Instructions

2.1 Basic Operation

  • Power On/Off: Press the ON/OFF button.
  • Mode Switching: Press the MODE/CAL button to cycle through pH, EC, TDS, and Salinity measurement modes. The current mode will be displayed on the screen.
  • Temperature Unit: Press the °C/°F button to switch between Celsius and Fahrenheit.
  • Lock Reading (HOLD): Press the HOLD button to freeze the current reading on the display. Press it again to release.

Video: Product demonstration showing mode switching (0:02-0:08) and temperature unit switching (0:09-0:11).

2.2 pH Calibration

The tester is factory calibrated. Recalibration is recommended periodically or if readings appear inaccurate. Always perform pH 6.86 calibration first, then pH 4.01 and pH 9.18.

  1. Prepare pH 6.86 Solution: Empty one packet of pH 6.86 powder into a clean cup. Add 250ml of purified or distilled water. Stir well and wait 5-10 minutes for the powder to fully dissolve.
  2. Power On & Select Mode: Turn on the meter. Press the MODE/CAL button until 'pH' is displayed at the top right of the screen.
  3. Calibrate:
    • Immerse the electrode into the pH 6.86 solution and stir gently.
    • Wait for the reading to stabilize.
    • Long press the MODE/CAL button for 3 seconds until 'CLEA' appears on the screen. Release the button.
    • Wait for the reading to stop flashing; calibration is complete when the reading stabilizes.
  4. Repeat for pH 4.01 and pH 9.18: Use the same steps to prepare and calibrate with pH 4.01 and pH 9.18 solutions.
pH Calibration setup with solutions

Image: pH calibration process showing the meter with buffer solutions.

2.3 EC/TDS/Salinity Calibration

The EC, TDS, and Salinity values are pre-calibrated at the factory and typically do not require user calibration. The meter is ready for use out of the box for these parameters.

  • Factory TDS Calibration Point: 1382 ppm
  • Factory EC Calibration Point: 1413 µS/cm
  • Factory Salinity Calibration Point: 35 ppt

3. Maintenance

  • Probe Handling: The glass electrode is fragile. Handle it gently and avoid touching the bottom or edges of the container during testing to prevent breakage.
  • Cleaning: After each use, rinse the electrode thoroughly with purified or distilled water and gently wipe it dry. Do not let the electrode dry out completely.
  • Electrode Lifespan: Electrodes are consumable with a typical shelf life of 6 months. It is recommended to replace them every 6 months for optimal accuracy.
  • Battery Replacement: Replace the batteries when the low battery indicator flashes on the display to ensure accurate readings.

4. Troubleshooting

  • Value Jumps During Test: It is normal for the value to jump when the test probe encounters air due to resistance. Gently stir the solution to remove air bubbles around the electrode.
  • Meter Cannot Be Calibrated: Ensure you are using fresh calibration solutions and following the calibration steps precisely. If the issue persists, contact customer support.
  • Product Damaged or Not Working: If you receive a damaged product or it fails to operate normally, please contact the seller or customer support immediately for assistance.

5. Specifications

ParameterRangeResolutionAccuracy
ModelYG-9909
pH0.00 - 14.00 pH0.01 pH±0.05 pH
TDS0 - 10000 ppm, 10.1 - 200.0 ppt1 ppm, 0.1 ppt±2% F.S.
EC0 - 10000 µS/cm1 µS/cm±2% F.S.
Salinity0 - 20%, 0 - 10000 ppm, 10.1 - 200.0 ppt0.01%, 1 ppm, 0.1 ppt±2% F.S.
Temperature0.0°C - 60.0°C (32.0°F - 140°F)0.1°C / 0.1°F±0.5°C
ATC (Automatic Temperature Compensation)0.1°C - 60.0°C
Work Environment0 - 60°C (32 - 140°F); RH 100%
Waterproof RatingIP66
Dimensions183 x 42 x 42 mm
DisplayLCD Backlight Display
Battery3 x 1.5V LR44 (not included)
Automatic Shut-off135 seconds

6. User Tips

  • Always rinse the electrode with purified water after each use and keep it dry to maintain accuracy and prolong lifespan.
  • Follow the calibration instructions carefully, as proper calibration is key to accurate pH readings. The included calibration powders are designed for this device.
  • Store the meter in a cool, dry place when not in use.

7. Warranty and Support

  • Main Unit Warranty: The main unit comes with a 1-year warranty for non-artificial damage.
  • Electrode Warranty: Electrodes are consumable and come with a 6-month warranty against non-human damage.
  • Support: For any questions regarding accessories, product issues, or warranty claims, please contact the seller or manufacturer's customer support. They aim to provide a satisfactory solution as soon as possible.