1. Setup
1.1 What's in the Box?
- 1 x Yieryi 5-in-1 Water Quality Tester (YG-9909)
- 1 x User Manual
- 1 x pH 4.00 Calibration Powder
- 1 x pH 6.86 Calibration Powder
- 1 x pH 9.18 Calibration Powder
- 1 x Screwdriver

Image: Contents of the YG-9909 package, including the tester, calibration powders, screwdriver, and manual.
1.2 Battery Installation
The YG-9909 requires 3 x 1.5V LR44 coin cell batteries (not included).
- Twist and remove the battery cap located at the top of the device.
- Insert three LR44 batteries, ensuring correct polarity (+/-).
- Replace the battery cap and twist to secure it.

Image: Close-up of the battery compartment showing the placement of LR44 batteries.
1.3 Electrode Replacement
The electrode of the YG-9909 is replaceable. Follow these steps:
- Using the provided screwdriver, unscrew the four small screws around the electrode base.
- Carefully remove the screw sleeve.
- Unplug the old electrode from the main unit.
- Plug in the new electrode, ensuring it is seated correctly.
- Reattach the screw sleeve and tighten the four screws.
Video: Product demonstration showing electrode replacement (from 0:14 to 0:28).
2. Operating Instructions
2.1 Basic Operation
- Power On/Off: Press the ON/OFF button.
- Mode Switching: Press the MODE/CAL button to cycle through pH, EC, TDS, and Salinity measurement modes. The current mode will be displayed on the screen.
- Temperature Unit: Press the °C/°F button to switch between Celsius and Fahrenheit.
- Lock Reading (HOLD): Press the HOLD button to freeze the current reading on the display. Press it again to release.
Video: Product demonstration showing mode switching (0:02-0:08) and temperature unit switching (0:09-0:11).
2.2 pH Calibration
The tester is factory calibrated. Recalibration is recommended periodically or if readings appear inaccurate. Always perform pH 6.86 calibration first, then pH 4.01 and pH 9.18.
- Prepare pH 6.86 Solution: Empty one packet of pH 6.86 powder into a clean cup. Add 250ml of purified or distilled water. Stir well and wait 5-10 minutes for the powder to fully dissolve.
- Power On & Select Mode: Turn on the meter. Press the MODE/CAL button until 'pH' is displayed at the top right of the screen.
- Calibrate:
- Immerse the electrode into the pH 6.86 solution and stir gently.
- Wait for the reading to stabilize.
- Long press the MODE/CAL button for 3 seconds until 'CLEA' appears on the screen. Release the button.
- Wait for the reading to stop flashing; calibration is complete when the reading stabilizes.
- Repeat for pH 4.01 and pH 9.18: Use the same steps to prepare and calibrate with pH 4.01 and pH 9.18 solutions.

Image: pH calibration process showing the meter with buffer solutions.
2.3 EC/TDS/Salinity Calibration
The EC, TDS, and Salinity values are pre-calibrated at the factory and typically do not require user calibration. The meter is ready for use out of the box for these parameters.
- Factory TDS Calibration Point: 1382 ppm
- Factory EC Calibration Point: 1413 µS/cm
- Factory Salinity Calibration Point: 35 ppt
3. Maintenance
- Probe Handling: The glass electrode is fragile. Handle it gently and avoid touching the bottom or edges of the container during testing to prevent breakage.
- Cleaning: After each use, rinse the electrode thoroughly with purified or distilled water and gently wipe it dry. Do not let the electrode dry out completely.
- Electrode Lifespan: Electrodes are consumable with a typical shelf life of 6 months. It is recommended to replace them every 6 months for optimal accuracy.
- Battery Replacement: Replace the batteries when the low battery indicator flashes on the display to ensure accurate readings.
4. Troubleshooting
- Value Jumps During Test: It is normal for the value to jump when the test probe encounters air due to resistance. Gently stir the solution to remove air bubbles around the electrode.
- Meter Cannot Be Calibrated: Ensure you are using fresh calibration solutions and following the calibration steps precisely. If the issue persists, contact customer support.
- Product Damaged or Not Working: If you receive a damaged product or it fails to operate normally, please contact the seller or customer support immediately for assistance.
5. Specifications
| Parameter | Range | Resolution | Accuracy |
|---|---|---|---|
| Model | YG-9909 | ||
| pH | 0.00 - 14.00 pH | 0.01 pH | ±0.05 pH |
| TDS | 0 - 10000 ppm, 10.1 - 200.0 ppt | 1 ppm, 0.1 ppt | ±2% F.S. |
| EC | 0 - 10000 µS/cm | 1 µS/cm | ±2% F.S. |
| Salinity | 0 - 20%, 0 - 10000 ppm, 10.1 - 200.0 ppt | 0.01%, 1 ppm, 0.1 ppt | ±2% F.S. |
| Temperature | 0.0°C - 60.0°C (32.0°F - 140°F) | 0.1°C / 0.1°F | ±0.5°C |
| ATC (Automatic Temperature Compensation) | 0.1°C - 60.0°C | ||
| Work Environment | 0 - 60°C (32 - 140°F); RH 100% | ||
| Waterproof Rating | IP66 | ||
| Dimensions | 183 x 42 x 42 mm | ||
| Display | LCD Backlight Display | ||
| Battery | 3 x 1.5V LR44 (not included) | ||
| Automatic Shut-off | 135 seconds | ||
6. User Tips
- Always rinse the electrode with purified water after each use and keep it dry to maintain accuracy and prolong lifespan.
- Follow the calibration instructions carefully, as proper calibration is key to accurate pH readings. The included calibration powders are designed for this device.
- Store the meter in a cool, dry place when not in use.
7. Warranty and Support
- Main Unit Warranty: The main unit comes with a 1-year warranty for non-artificial damage.
- Electrode Warranty: Electrodes are consumable and come with a 6-month warranty against non-human damage.
- Support: For any questions regarding accessories, product issues, or warranty claims, please contact the seller or manufacturer's customer support. They aim to provide a satisfactory solution as soon as possible.