JUNCTEK PSG9080

Contents

JUNTEK PSG9080/PSG9060 Programmable Signal Generator User Manual

High-Precision Dual Channel Function/Arbitrary Waveform Signal Generator

1. Introduction

The JUNTEK PSG9080 and PSG9060 are programmable dual-channel function/arbitrary waveform signal generators designed for high-precision signal generation. These instruments can produce a variety of waveforms, including sine, square, triangle, pulse, and arbitrary waves. The PSG9080 offers a frequency range from 1 nHz to 80 MHz, while the PSG9060 has a frequency range up to 60 MHz. Both models feature modulation, frequency sweep, measurement signal frequency, and programming functions.

The device can simultaneously display the output signal, amplitude, phase, duty cycle, and frequency. With excellent amplitude-frequency characteristics, multi-functionality, and high performance, these generators are suitable for education, research and development, production, testing, and other industrial applications. The instrument features an exquisite design and a 3.2-inch high-resolution color LCD screen for an enhanced visual experience.

JUNTEK PSG9080 Programmable Signal Generator front view with display showing dual sine waves.
Figure 1.1: Front view of the JUNTEK PSG9080 Programmable Signal Generator.

Product Demonstration Video

Video 1.1: A brief demonstration of the PSG9080's features and operation, including language selection, dual-channel output, system settings, modulation, measurement, sweep, voltage control, and programming functions.

2. Safety Requirements

Please review the following safety precautions carefully before operating the instrument to prevent personal injury or damage to the instrument and any connected products. Always follow these instructions for proper use.

  • Use Proper Power Cord: Only use the exclusive power cord designed for the instrument and authorized for use in your local country.
  • Connect the Probe Correctly: If using a probe, do not connect the ground lead to high voltage, as it shares the same potential as the ground.
  • Observe All Terminal Ratings: To avoid fire or shock hazards, observe all ratings and markers on the instrument. Refer to this manual for more information before connecting.
  • Use Proper Over-voltage Protection: Ensure no over-voltage (e.g., from lightning) reaches the product, as this could expose the operator to electric shock.
  • Do Not Operate Without Covers: Do not operate the instrument with covers or panels removed.
  • Do Not Insert Anything Into the Air Outlet: Avoid inserting objects into the air outlet to prevent damage.
  • Avoid Circuit or Wire Exposure: Do not touch exposed junctions and components when the unit is powered on.
  • Do Not Operate With Suspected Failures: If you suspect any damage, have the instrument inspected by JUNTEK authorized personnel before further operation. All maintenance, adjustments, or replacements must be performed by JUNTEK authorized personnel.
  • Provide Adequate Ventilation: Inadequate ventilation can increase internal temperature, causing damage. Ensure the instrument is well-ventilated and regularly inspect the air outlet and fan.
  • Do Not Operate in Wet Conditions: To avoid short circuits or electric shock, never operate the instrument in a humid environment.
  • Do Not Operate in an Explosive Atmosphere: To avoid personal injuries or damage, never operate the instrument in an explosive atmosphere.
  • Prevent Electrostatic Impact: Operate the instrument in an electrostatic discharge protective environment. Always ground both internal and external conductors of cables to release static before making connections.
  • Handle with Caution: Handle with care during transportation to avoid damage to keys, knobs, interfaces, and other panel parts.

Important Notices

  • Ensure the input power adapter is correct; this instrument uses a DC5V power adapter.
  • The instrument's shell is fragile and prone to corrosion. Avoid impacts or proximity to chemicals.
  • Operating temperature: 10~50℃; storage temperature: 20~70℃. Keep the instrument in a dry environment.
  • Do not attempt to disassemble the instrument, as this will void the warranty. Repairs should only be made by designated repair outlets or the factory.
  • Avoid placing unsafe items (e.g., lighted candles, cups of water, corrosive chemicals) on the instrument.
  • The display screen is fragile. Do not touch or bump it. Keep children from playing with it. Clean the LCD surface carefully with a soft cloth.
  • Do not move the instrument violently to avoid internal circuit damage. If the instrument malfunctions, contact the supplier.

3. Package Contents and Inspection

Upon receiving your new PSG9080/PSG9060 programmable signal generator, it is recommended to inspect the package and contents thoroughly.

3.1 Inspect the Packaging

If the packaging is damaged, retain all damaged packaging and cushioning materials until the shipment's completeness and functionality (electrical and mechanical) have been verified. The consignor or carrier is responsible for damage during shipment. JUNTEK is not responsible for free maintenance, rework, or replacement due to shipping damage.

3.2 Check the Contents

Verify the contents against the packing list below. If any items are damaged or missing, contact your JUNTEK sales representative immediately.

JUNTEK PSG9080 Programmable Signal Generator and included accessories in packaging.
Figure 3.1: PSG9080/PSG9060 unit with standard accessories and packaging.
Table 3.1: Standard Package Contents
ItemDescriptionQuantity
HostPSG9080/PSG9060 programmable function/arbitrary waveform signal generator1 pc
Power cable1 pc
USB cable1 pc
AccessoryBNC to alligator clips test lead1 pc
BNC male plug test lead1 pc
Certificate of conformity1 pc
Quick start guide1 pc

3.3 Inspect the Instrument

In case of any mechanical damage, missing parts, or failure in passing electrical and mechanical tests, contact your JUNTEK sales representative.

4. Product Overview

4.1 Front Panel Overview

Front panel of PSG9080 with labeled components: 3.5-inch LCD screen, function soft keys, number keys, knob, direction keys, switch, input interface, output interface, channel control keys.
Figure 4.1: Front Panel Layout of the PSG9080/PSG9060.
  • Power Key: Used to turn the generator on or off.
  • Signal Input Connectors: Includes EXT.IN for external signal input (2Vpp~20Vpp) and MOD.IN for modulation input (0Vpp~3Vpp).
  • Signal Output Connectors: CH1 and CH2 output connectors (BNC, 50Ω±10% nominal impedance). When enabled, these output waveforms according to the current configuration.
  • Channel Control Area: CH1 and CH2 keys control the output of respective channels. The OUT key controls both CH1 and CH2 output switches.
  • Direction Keys: Used to move the cursor for parameter editing or to delete numbers.
  • Knob: Used to increase or decrease parameter values and quickly switch waveforms.
  • Numeric Keyboard: (0-9, decimal point, sign key) for setting parameters.
  • Function Keys: SYS (system settings/help), MOD (modulation mode/burst), MEAS (measurement/counting), LOAD (load parameters), SHIFT (modifier key), SWEEP (sweep mode), VCO (voltage control mode), FUNC (programming mode).
  • Function Softkeys: Corresponding to the menu on the left of the display, these activate specific menu options.

4.2 Rear Panel Overview

Rear panel of PSG9080 with labeled components: Ground terminal, SYNC IN/OUT, TTL Ext. interface, DC5V power input, USB port, AC power socket, Power switch, Heat dissipation holes.
Figure 4.2: Rear Panel Layout of the PSG9080/PSG9060.
  • Sync Ports (IN/OUT): Allows synchronization of output signals from different signal generators.
  • Communication Extension Port (TTL Ext.): A serial port with TTL level for secondary development, including GPI (digital signal input), GPO1/GPO2 (digital signal output), RXD (serial communication receive data), and TXD (serial communication send data).
  • USB Interface: Connects the generator to a computer for remote control via PC software or programming.
  • DC Power Socket: DC5V±0.5V 3A.
  • AC Power Socket: 85-264V, 47-63Hz. Max input power less than 30W. Fuse: 250V, T3.15A.
  • Power Switch: Turns the signal generator on or off.
  • Cooling Holes: Dissipate internal heat.
  • Chassis Ground: Connects to ground to prevent electric shock and ensure normal electrical system operation.

4.3 Display Interface

The 3.5-inch TFT color LCD screen (320x480 resolution, 64K true color) clearly displays the current function menu, parameter settings, and other information.

Close-up of the PSG9080 display showing dual channel sine wave outputs with frequency, amplitude, offset, and duty cycle parameters.
Figure 4.3: Main Display Interface.

Key elements displayed include:

  • Soft Key Menu: Displays the operation menu for the currently selected function.
  • Finger Cursor: Indicates the selected channel (CH1 or CH2) for parameter changes.
  • Current Channel Status: Shows "ON" or "OFF" for channel output.
  • Waveform Display: Shows the waveform currently in the main channel.
  • Parameters: Frequency, Amplitude, Offset, Phase, Duty Cycle, and an Information box for entered values and saved/loaded positions.

5. Setup

5.1 Power On and Initial Inspection

  1. Connect to Power: Connect the generator to an AC power supply using the provided power cord. The AC power supply input is 85-264V, 47-63Hz. A DC5V±0.5V 3A power adapter can also be used.
  2. Power-on: After connecting the power supply, press the power key on the front panel. The instrument will perform initialization and self-test, then display the default interface.
  3. Set System Language: Upon booting, you can select the desired system language using the soft keys. The instrument will remember your selection for future power-ons.

5.2 Output Basic Waveform

The PSG9080/PSG9060 can output basic waveforms (sine, square, triangle, pulse) from single or dual channels simultaneously. By default, dual channels output a sine waveform at 10kHz frequency and 5Vpp amplitude.

  1. Select Output Channel: Press the CH1 or CH2 key to select the desired channel. A finger cursor on the display indicates the active channel.
  2. Select Basic Waveforms: Press the Wave soft key. Use the knob or soft keys to choose the desired waveform type.
  3. Set Frequency: Press the Freq. soft key. Use the numeric keyboard or direction keys and knob to set the frequency (range: 0-80MHz for PSG9080, 0-60MHz for PSG9060). Press Freq. again to select the unit (MHz, KHz, Hz, mHz, μHz).
  4. Set Amplitude: Press the Ampl. soft key. Use the numeric keyboard or knob to set the amplitude (Vpp, MVpp). Note that amplitude range is limited by frequency.
  5. Set Offset: Press the Offset soft key. Use the numeric keyboard or knob to set the DC offset (V).
  6. Set Duty Cycle: Press the Duty soft key. Use the numeric keyboard or knob to adjust the duty cycle (0.01%-99.99% for pulse waves). Not adjustable for sine waves.
  7. Set Phase: Press the Phase soft key (on CH2 interface). Use direction keys and knob to adjust phase difference (0°-359.9°). Ensure CH1 and CH2 frequency synchronization is set in system settings first.
  8. Observe Output Waveforms: Connect CH1 and CH2 to an oscilloscope using BNC test leads to observe the waveforms.

6. Operating Instructions

6.1 Modulation Mode

The PSG9080/PSG9060 supports various modulation types: AM, FM, PM, ASK, FSK, PSK, Pulse, and Burst. Modulation involves adding information from a signal source to a carrier wave.

Modulation interface of PSG9080 showing settings for Type, Sharp, Source, Mod.Freq, Mod.Depth for CH1 and CH2.
Figure 6.1: Modulation Interface.
  • Accessing Modulation: Press MOD to enter the modulation mode. Use the Type soft key or knob to select the desired modulation type.
  • Carrier Waveform: Select sine, square, sawtooth, or arbitrary wave (DC signal wave cannot be used).
  • Modulation Source: Choose internal or external. External modulation signals are input via the MOD connector on the front panel (0-20kHz, -5V~+5V).
  • Parameters: Set modulation frequency, depth (AM), deviation (FM), polarity (ASK/FSK/PSK), phase (PSK), pulse width/period (Pulse), and burst parameters (number, trigger source, idle mode).

6.2 Measurement Mode

The instrument can measure frequency, positive/negative pulse width, period, and duty cycle of input signals.

  • Accessing Measurement: Press MEAS to enter the measurement mode. Use the CNTR. soft key or knob to switch between measurement and counter functions.
  • Measurement Range: Frequency 1Hz-100MHz, signal amplitude 2Vpp-20Vpp (input via Ext.IN).
  • Settings: Configure coupling (DC/AC), gate time (0.001s-10s), and measurement mode (low/high frequency).

6.3 Frequency Sweep Mode

The sweep function allows for frequency, amplitude, or duty cycle sweeps.

  • Accessing Sweep: Press SWEEP. Use the FUNC soft key to select frequency, amplitude, or duty cycle sweep.
  • Sweep Channel: Select CH1 or CH2.
  • Parameters: Set start/end frequency/amplitude/duty cycle, sweep time (0.01s-640s), sweep direction (round trip, increment, decrement), and sweep mode (linear, logarithmic).
  • Start Sweep: Press the ON soft key to begin the sweep.

6.4 Voltage Control Mode

Control frequency, amplitude, and duty cycle using an external analog voltage signal.

Voltage Control Function interface showing V.Control Frequency settings for CH1 and CH2.
Figure 6.2: Voltage Control Function Interface.
  • Accessing Voltage Control: Press VCO. Use the FUNC soft key to select frequency, amplitude, or duty cycle control.
  • Voltage Control Channel: Select CH1 or CH2.
  • Parameters: Set start/end values, calibrate min/max external input voltage, and choose control mode (linear/logarithmic).
  • Enable Control: Press the ON soft key to activate voltage control.

6.5 Programming Mode

The instrument supports programmable sequences for automatic operation.

Screenshots of the PC software interface for PSG9080, showing control panel, arbitrary waveform editor, and programming mode.
Figure 6.3: PC Software Interface for Programming.
  • Accessing Programming: Press FUNC to enter the programming mode interface.
  • Modes: Switch between normal and debugging modes using the MODE soft key.
  • Programming Steps: The instrument supports 100 groups (P00-P99) of programming, with each group containing 00-99 sets of instructions.
  • Parameters: Each instruction includes parameters for Load (storage location), Time (running time), and Go To (next instruction).
  • Cycle: Set the number of repetitions for a program sequence.

6.6 PC Software Control

The PSG9080/PSG9060 can be controlled via PC software, which provides a graphical interface for all instrument functions, including secondary development programming. The communication protocol is open.

  • Software Download: The PC software and communication protocol can be downloaded from: http://68.168.132.244/PSG9000/material.html
  • Installation:
    1. Install the USB driver by clicking "Install USB Drive" in the software.
    2. Install the main application by clicking "Main program" and following the on-screen instructions.
  • Connection:
    1. Connect the instrument to your PC via the USB interface.
    2. In Device Manager, identify the assigned serial port (e.g., COM5).
    3. In the PC software, select the corresponding serial port and click "Open Port" then "Connect".

7. Specifications

The following tables detail the technical specifications for the PSG9080 and PSG9060 models.

7.1 General Specifications

ParameterPSG9080PSG9060
Channel2
Sine wave frequency range1nHz-80MHz1nHz-60MHz
Sampling Rate300MSa/s
Waveform vertical resolution14 bits
Display3.5 inch TFT color LCD screen
Resolution320*480
Color64K true color
Storage (Groups)100 groups
InterfaceUSB to serial interface, TTL serial port
Communication rateStandard 115200BPS
Power Supply85V-264V, 47-63Hz (AC); DC5V±0.5V 3A (DC)
Power ConsumptionLess than 30W
Fuse250VAC, T3.15A
Operating Temperature0~40°C
Humidity<80%
Dimensions (L x W x H)218.60mm x 103.30mm x 87.30mm

7.2 Waveform Characteristics

ParameterSpecification
Wave length8192 points/channel
Waveform typesSine, square, pulse (duty cycle, pulse width, cycle time adjustable), triangle, rising/falling sawtooth, CMOS, DC level, half/full wave, positive step, inverse staircase, exponential rise/fall, Lorentz pulse, multi-sound, irregular noise, trapezoidal pulse, sink pulse, AM/FM waveforms, 99 user-defined waveforms.
Square wave frequency range1nHz-30MHz
Triangle wave frequency range1nHz-50MHz
Pulse wave frequency range1nHz-30MHz
TTL digital wave frequency range1nHz-20MHz
Arbitrary wave frequency range1nHz-50MHz
Pulse width adjustment range5nS~4S
Pulse period adjustment range10nS~40S
Frequency minimum resolution1nHz
Frequency accuracy±5ppm 0 to 50°C
Frequency stability±3ppm per 1 year
Sine wave spectral purity (Total harmonic distortion)≤0.5% (@1kHz, 5Vpp)
Square wave rising/falling edge time≤10nS
Square wave overshoot≤10%
Square wave asymmetry≤0.1%
Ramp and triangle wave linearity≤1% (Less than 1MHz)
Ramp and triangle wave asymmetry≤1% (Less than 1MHz)
Pulse wave duty cycle0.01%-99.99%
Arbitrary wave minimum rise/fall time≤10nS
Arbitrary wave overshoot≤10%
Arbitrary waveform non-volatile storage number99

7.3 Output Characteristics

ParameterSpecification
Amplitude range (Frequency ≤1MHz)2mVpp~25Vpp
Amplitude range (1MHz < Frequency < 11MHz)2mVpp~10Vpp
Amplitude range (11MHz < Frequency < 60MHz)2mVpp~5Vpp
Amplitude range (60MHz < Frequency < 80MHz)2mVpp~3.6Vpp (PSG9080 only)
Amplitude resolution1mVpp
Amplitude stability±1% ±1 mVpp (@ 1 kHz, >10 mVpp)
Amplitude flatness (<10MHz)±1% (0.1dB)
Amplitude flatness (10MHz~50MHz)±2% (0.2dB)
Amplitude flatness (50MHz~70MHz)±10% (0.9dB)
Amplitude flatness (70MHz~80MHz)±20% (1.9dB) (PSG9080 only)
DC bias offset adjustment range-9.99V-15.00V
Offset resolution0.01V
Phase adjustment range0°-359.9°
Phase resolution0.1°
Output impedance50Ω (nominal)

7.4 Modulation Characteristics

Modulation TypeCarrier WaveModulation SourceModulated WaveParameters
AMSine, square, sawtooth, arbitrary (except DC)Internal/ExternalSine, square, sawtooth, noise, arbitraryDepth: 0-200%, Freq: 1mHz-1MHz
FMSine, square, sawtooth, arbitrary (except DC)Internal/ExternalSine, square, sawtooth, noise, arbitraryFreq: 1mHz-1MHz, Deviation: 0.1Hz-10kHz
PMSine, square, sawtooth, arbitrary (except DC)Internal/ExternalSine, square, sawtooth, noise, arbitraryPhase Deviation: 0°-360°, Freq: 1mHz-1MHz
ASKSine, square, sawtooth, arbitrary (except DC)Internal/External50% duty cycle pulse waveAmplitude: 0-200%, Polarity: Positive/Negative, Rate: 1mHz-1MHz, Freq Hopping: 0.1Hz-80MHz
FSKSine, square, sawtooth, arbitrary (except DC)Internal/External50% duty cycle pulse wavePolarity: Positive/Negative, Rate: 1mHz-1MHz, Freq Hopping: 0.1Hz-80MHz
PSKSine, square, sawtooth, arbitrary (except DC)Internal/External50% duty cycle pulse wavePolarity: Positive/Negative, Rate: 1mHz-1MHz, Phase: 0°-360°
PWMPulse waveN/ARegular and reversePulse Period: 0.01μS-40S, Pulse Width: 0.001μS-4S
BurstSine, square, sawtooth, arbitrary (except DC)Internal/ExternalN/ANumber: 1-1,000,000,000, Trigger: Manual, CH2, External (AC/DC), Idle Mode: Zero, Positive/Negative Max

7.5 Measurement and Counting Function

ParameterSpecification
Measurement functionsFrequency, positive/negative pulse width, period, duty cycle
Frequency measurement range1Hz~100MHz
Input signal amplitude range2Vpp~20Vpp
Period measurement10nS~4S
Pulse Width4S
Duty cycle measurement range0.01%~99.99%
Counter range0-4294967295
Counter coupling methodDC and AC coupling
Counting methodManual

7.6 External Modulation Input Characteristics

ParameterSpecification
Input signal amplitude range0~3Vpp
Input signal frequency rangeDC~200kHz

7.7 Sweep Characteristics

ParameterSpecification
Sweep channelCH1 or CH2
Carrier waveSine, square, sawtooth, arbitrary (except DC)
Sweep functionSweep frequency, sweep amplitude, sweep duty
Sweep modeLinear, logarithmic
Sweep directionIncrement, decrement, round trip
Start/end frequencyConsistent with carrier frequency limits
Start/end amplitudeConsistent with carrier amplitude limits
Start/end duty cycleConsistent with carrier duty cycle limits
Sweep time0.01S-640S

7.8 Voltage Control Characteristics

ParameterSpecification
Voltage control channelCH1 or CH2
Carrier waveSine, square, sawtooth, arbitrary (except DC)
Voltage control functionFrequency control, amplitude control, duty cycle control
Sweep modeLinear, logarithmic
Voltage control voltage range0V-5V (start/end points arbitrarily settable)

7.9 Programming Features

ParameterSpecification
Running modeDebugging, normal
StorageP00-P19
Programming number00-99
Programming time (single serial number)0-99S

8. Maintenance

Proper maintenance ensures the longevity and optimal performance of your JUNTEK signal generator.

  • Keep Surfaces Clean and Dry: Regularly clean the instrument's surfaces to prevent dust and moisture from affecting performance. Use a soft, dry cloth. For the LCD screen, wipe carefully with a soft cloth.
  • Adequate Ventilation: Ensure the instrument is placed in a well-ventilated area. Regularly inspect the heat dissipation holes and fan (if present) to ensure they are not obstructed.
  • Environmental Conditions: Operate and store the instrument within the specified temperature ranges (operating: 10~50℃; storage: 20~70℃) and in a dry environment.
  • Physical Care: The instrument's shell is fragile. Avoid impacts and keep it away from corrosive chemicals. Handle with care during movement to prevent damage to internal circuits, keys, knobs, and interfaces.
  • Avoid Unsafe Items: Do not place lighted candles, cups with water, or corrosive chemicals on the instrument.

9. Troubleshooting

This section addresses common issues and their solutions. If a problem persists, contact JUNTEK support.

  • Screen is dark (no display) after pressing the power key:
    • Check if the power is correctly connected.
    • Verify that the power key was pressed firmly.
    • Restart the instrument after performing the above checks.
    • If the issue continues, contact JUNTEK.
  • Screen is too dark to read:
    • Check if brightness and contrast settings are too low.
    • Press the SYS key to enter system settings, navigate to brightness, and adjust using the knob.
  • Settings are correct, but no waveform is generated:
    • Ensure the BNC cable is tightly connected to the correct channel output terminal (CH1 or CH2).
    • Check if the BNC cable has internal damage.
    • Verify the BNC cable is tightly connected to the test instrument.
    • Check if the backlight for CH1 or CH2 is on. If not, press the corresponding key to illuminate it.
    • After checking the above, press and restart the instrument.
    • If the issue persists, contact JUNTEK.

10. User Tips

  • Professional Use: The PSG9080/PSG9060 is a robust and reliable instrument, frequently found in electronics workshops due to its performance and versatility.
  • Model Selection: The primary difference between the PSG9060 and PSG9080 models is their maximum frequency output (60 MHz vs. 80 MHz, respectively). Choose the model that best suits your frequency requirements.
  • Connectivity: While the device offers USB and serial (TTL) connectivity for PC control, it does not include a Wi-Fi module. Ensure you have the necessary cables for wired connections.
  • Software Updates: Keep the PC software and instrument firmware updated for optimal performance and access to the latest features. Firmware updates can be performed via the PC software.

11. Warranty and Support

11.1 Warranty

JUNTEK warrants that the product mainframe and accessories will be free from defects in materials and workmanship within the warranty period. If a product proves defective within this period, JUNTEK guarantees free replacement or repair. This warranty does not cover damages caused by improper use or unauthorized modifications. There are no other warranties, expressed or implied, beyond what is expressly stated herein or in any applicable warranty card. JUNTEK is not liable for any consequential, indirect, ensuing, or special damages.

11.2 Contact and Support

For any problems or requirements regarding the product or this manual, please contact JUNTEK:

Additional resources available:

  • Operation Demo Video: Provides operational guidance.
  • PC Software and Communication Protocol: Available for download.
  • User Manual (PDF): A comprehensive PDF version of the user manual is available for download from the official website. (Reference: User manual (PDF))