1. Introduction
The JUNTEK PSG9080 and PSG9060 are programmable dual-channel function/arbitrary waveform signal generators designed for high-precision signal generation. These instruments can produce a variety of waveforms, including sine, square, triangle, pulse, and arbitrary waves. The PSG9080 offers a frequency range from 1 nHz to 80 MHz, while the PSG9060 has a frequency range up to 60 MHz. Both models feature modulation, frequency sweep, measurement signal frequency, and programming functions.
The device can simultaneously display the output signal, amplitude, phase, duty cycle, and frequency. With excellent amplitude-frequency characteristics, multi-functionality, and high performance, these generators are suitable for education, research and development, production, testing, and other industrial applications. The instrument features an exquisite design and a 3.2-inch high-resolution color LCD screen for an enhanced visual experience.

Product Demonstration Video
2. Safety Requirements
Please review the following safety precautions carefully before operating the instrument to prevent personal injury or damage to the instrument and any connected products. Always follow these instructions for proper use.
- Use Proper Power Cord: Only use the exclusive power cord designed for the instrument and authorized for use in your local country.
- Connect the Probe Correctly: If using a probe, do not connect the ground lead to high voltage, as it shares the same potential as the ground.
- Observe All Terminal Ratings: To avoid fire or shock hazards, observe all ratings and markers on the instrument. Refer to this manual for more information before connecting.
- Use Proper Over-voltage Protection: Ensure no over-voltage (e.g., from lightning) reaches the product, as this could expose the operator to electric shock.
- Do Not Operate Without Covers: Do not operate the instrument with covers or panels removed.
- Do Not Insert Anything Into the Air Outlet: Avoid inserting objects into the air outlet to prevent damage.
- Avoid Circuit or Wire Exposure: Do not touch exposed junctions and components when the unit is powered on.
- Do Not Operate With Suspected Failures: If you suspect any damage, have the instrument inspected by JUNTEK authorized personnel before further operation. All maintenance, adjustments, or replacements must be performed by JUNTEK authorized personnel.
- Provide Adequate Ventilation: Inadequate ventilation can increase internal temperature, causing damage. Ensure the instrument is well-ventilated and regularly inspect the air outlet and fan.
- Do Not Operate in Wet Conditions: To avoid short circuits or electric shock, never operate the instrument in a humid environment.
- Do Not Operate in an Explosive Atmosphere: To avoid personal injuries or damage, never operate the instrument in an explosive atmosphere.
- Prevent Electrostatic Impact: Operate the instrument in an electrostatic discharge protective environment. Always ground both internal and external conductors of cables to release static before making connections.
- Handle with Caution: Handle with care during transportation to avoid damage to keys, knobs, interfaces, and other panel parts.
Important Notices
- Ensure the input power adapter is correct; this instrument uses a DC5V power adapter.
- The instrument's shell is fragile and prone to corrosion. Avoid impacts or proximity to chemicals.
- Operating temperature: 10~50℃; storage temperature: 20~70℃. Keep the instrument in a dry environment.
- Do not attempt to disassemble the instrument, as this will void the warranty. Repairs should only be made by designated repair outlets or the factory.
- Avoid placing unsafe items (e.g., lighted candles, cups of water, corrosive chemicals) on the instrument.
- The display screen is fragile. Do not touch or bump it. Keep children from playing with it. Clean the LCD surface carefully with a soft cloth.
- Do not move the instrument violently to avoid internal circuit damage. If the instrument malfunctions, contact the supplier.
3. Package Contents and Inspection
Upon receiving your new PSG9080/PSG9060 programmable signal generator, it is recommended to inspect the package and contents thoroughly.
3.1 Inspect the Packaging
If the packaging is damaged, retain all damaged packaging and cushioning materials until the shipment's completeness and functionality (electrical and mechanical) have been verified. The consignor or carrier is responsible for damage during shipment. JUNTEK is not responsible for free maintenance, rework, or replacement due to shipping damage.
3.2 Check the Contents
Verify the contents against the packing list below. If any items are damaged or missing, contact your JUNTEK sales representative immediately.

| Item | Description | Quantity |
|---|---|---|
| Host | PSG9080/PSG9060 programmable function/arbitrary waveform signal generator | 1 pc |
| Power cable | 1 pc | |
| USB cable | 1 pc | |
| Accessory | BNC to alligator clips test lead | 1 pc |
| BNC male plug test lead | 1 pc | |
| Certificate of conformity | 1 pc | |
| Quick start guide | 1 pc |
3.3 Inspect the Instrument
In case of any mechanical damage, missing parts, or failure in passing electrical and mechanical tests, contact your JUNTEK sales representative.
4. Product Overview
4.1 Front Panel Overview

- Power Key: Used to turn the generator on or off.
- Signal Input Connectors: Includes EXT.IN for external signal input (2Vpp~20Vpp) and MOD.IN for modulation input (0Vpp~3Vpp).
- Signal Output Connectors: CH1 and CH2 output connectors (BNC, 50Ω±10% nominal impedance). When enabled, these output waveforms according to the current configuration.
- Channel Control Area: CH1 and CH2 keys control the output of respective channels. The OUT key controls both CH1 and CH2 output switches.
- Direction Keys: Used to move the cursor for parameter editing or to delete numbers.
- Knob: Used to increase or decrease parameter values and quickly switch waveforms.
- Numeric Keyboard: (0-9, decimal point, sign key) for setting parameters.
- Function Keys: SYS (system settings/help), MOD (modulation mode/burst), MEAS (measurement/counting), LOAD (load parameters), SHIFT (modifier key), SWEEP (sweep mode), VCO (voltage control mode), FUNC (programming mode).
- Function Softkeys: Corresponding to the menu on the left of the display, these activate specific menu options.
4.2 Rear Panel Overview

- Sync Ports (IN/OUT): Allows synchronization of output signals from different signal generators.
- Communication Extension Port (TTL Ext.): A serial port with TTL level for secondary development, including GPI (digital signal input), GPO1/GPO2 (digital signal output), RXD (serial communication receive data), and TXD (serial communication send data).
- USB Interface: Connects the generator to a computer for remote control via PC software or programming.
- DC Power Socket: DC5V±0.5V 3A.
- AC Power Socket: 85-264V, 47-63Hz. Max input power less than 30W. Fuse: 250V, T3.15A.
- Power Switch: Turns the signal generator on or off.
- Cooling Holes: Dissipate internal heat.
- Chassis Ground: Connects to ground to prevent electric shock and ensure normal electrical system operation.
4.3 Display Interface
The 3.5-inch TFT color LCD screen (320x480 resolution, 64K true color) clearly displays the current function menu, parameter settings, and other information.

Key elements displayed include:
- Soft Key Menu: Displays the operation menu for the currently selected function.
- Finger Cursor: Indicates the selected channel (CH1 or CH2) for parameter changes.
- Current Channel Status: Shows "ON" or "OFF" for channel output.
- Waveform Display: Shows the waveform currently in the main channel.
- Parameters: Frequency, Amplitude, Offset, Phase, Duty Cycle, and an Information box for entered values and saved/loaded positions.
5. Setup
5.1 Power On and Initial Inspection
- Connect to Power: Connect the generator to an AC power supply using the provided power cord. The AC power supply input is 85-264V, 47-63Hz. A DC5V±0.5V 3A power adapter can also be used.
- Power-on: After connecting the power supply, press the power key on the front panel. The instrument will perform initialization and self-test, then display the default interface.
- Set System Language: Upon booting, you can select the desired system language using the soft keys. The instrument will remember your selection for future power-ons.
5.2 Output Basic Waveform
The PSG9080/PSG9060 can output basic waveforms (sine, square, triangle, pulse) from single or dual channels simultaneously. By default, dual channels output a sine waveform at 10kHz frequency and 5Vpp amplitude.
- Select Output Channel: Press the CH1 or CH2 key to select the desired channel. A finger cursor on the display indicates the active channel.
- Select Basic Waveforms: Press the Wave soft key. Use the knob or soft keys to choose the desired waveform type.
- Set Frequency: Press the Freq. soft key. Use the numeric keyboard or direction keys and knob to set the frequency (range: 0-80MHz for PSG9080, 0-60MHz for PSG9060). Press Freq. again to select the unit (MHz, KHz, Hz, mHz, μHz).
- Set Amplitude: Press the Ampl. soft key. Use the numeric keyboard or knob to set the amplitude (Vpp, MVpp). Note that amplitude range is limited by frequency.
- Set Offset: Press the Offset soft key. Use the numeric keyboard or knob to set the DC offset (V).
- Set Duty Cycle: Press the Duty soft key. Use the numeric keyboard or knob to adjust the duty cycle (0.01%-99.99% for pulse waves). Not adjustable for sine waves.
- Set Phase: Press the Phase soft key (on CH2 interface). Use direction keys and knob to adjust phase difference (0°-359.9°). Ensure CH1 and CH2 frequency synchronization is set in system settings first.
- Observe Output Waveforms: Connect CH1 and CH2 to an oscilloscope using BNC test leads to observe the waveforms.
6. Operating Instructions
6.1 Modulation Mode
The PSG9080/PSG9060 supports various modulation types: AM, FM, PM, ASK, FSK, PSK, Pulse, and Burst. Modulation involves adding information from a signal source to a carrier wave.

- Accessing Modulation: Press MOD to enter the modulation mode. Use the Type soft key or knob to select the desired modulation type.
- Carrier Waveform: Select sine, square, sawtooth, or arbitrary wave (DC signal wave cannot be used).
- Modulation Source: Choose internal or external. External modulation signals are input via the MOD connector on the front panel (0-20kHz, -5V~+5V).
- Parameters: Set modulation frequency, depth (AM), deviation (FM), polarity (ASK/FSK/PSK), phase (PSK), pulse width/period (Pulse), and burst parameters (number, trigger source, idle mode).
6.2 Measurement Mode
The instrument can measure frequency, positive/negative pulse width, period, and duty cycle of input signals.
- Accessing Measurement: Press MEAS to enter the measurement mode. Use the CNTR. soft key or knob to switch between measurement and counter functions.
- Measurement Range: Frequency 1Hz-100MHz, signal amplitude 2Vpp-20Vpp (input via Ext.IN).
- Settings: Configure coupling (DC/AC), gate time (0.001s-10s), and measurement mode (low/high frequency).
6.3 Frequency Sweep Mode
The sweep function allows for frequency, amplitude, or duty cycle sweeps.
- Accessing Sweep: Press SWEEP. Use the FUNC soft key to select frequency, amplitude, or duty cycle sweep.
- Sweep Channel: Select CH1 or CH2.
- Parameters: Set start/end frequency/amplitude/duty cycle, sweep time (0.01s-640s), sweep direction (round trip, increment, decrement), and sweep mode (linear, logarithmic).
- Start Sweep: Press the ON soft key to begin the sweep.
6.4 Voltage Control Mode
Control frequency, amplitude, and duty cycle using an external analog voltage signal.

- Accessing Voltage Control: Press VCO. Use the FUNC soft key to select frequency, amplitude, or duty cycle control.
- Voltage Control Channel: Select CH1 or CH2.
- Parameters: Set start/end values, calibrate min/max external input voltage, and choose control mode (linear/logarithmic).
- Enable Control: Press the ON soft key to activate voltage control.
6.5 Programming Mode
The instrument supports programmable sequences for automatic operation.

- Accessing Programming: Press FUNC to enter the programming mode interface.
- Modes: Switch between normal and debugging modes using the MODE soft key.
- Programming Steps: The instrument supports 100 groups (P00-P99) of programming, with each group containing 00-99 sets of instructions.
- Parameters: Each instruction includes parameters for Load (storage location), Time (running time), and Go To (next instruction).
- Cycle: Set the number of repetitions for a program sequence.
6.6 PC Software Control
The PSG9080/PSG9060 can be controlled via PC software, which provides a graphical interface for all instrument functions, including secondary development programming. The communication protocol is open.
- Software Download: The PC software and communication protocol can be downloaded from: http://68.168.132.244/PSG9000/material.html
- Installation:
- Install the USB driver by clicking "Install USB Drive" in the software.
- Install the main application by clicking "Main program" and following the on-screen instructions.
- Connection:
- Connect the instrument to your PC via the USB interface.
- In Device Manager, identify the assigned serial port (e.g., COM5).
- In the PC software, select the corresponding serial port and click "Open Port" then "Connect".
7. Specifications
The following tables detail the technical specifications for the PSG9080 and PSG9060 models.
7.1 General Specifications
| Parameter | PSG9080 | PSG9060 |
|---|---|---|
| Channel | 2 | |
| Sine wave frequency range | 1nHz-80MHz | 1nHz-60MHz |
| Sampling Rate | 300MSa/s | |
| Waveform vertical resolution | 14 bits | |
| Display | 3.5 inch TFT color LCD screen | |
| Resolution | 320*480 | |
| Color | 64K true color | |
| Storage (Groups) | 100 groups | |
| Interface | USB to serial interface, TTL serial port | |
| Communication rate | Standard 115200BPS | |
| Power Supply | 85V-264V, 47-63Hz (AC); DC5V±0.5V 3A (DC) | |
| Power Consumption | Less than 30W | |
| Fuse | 250VAC, T3.15A | |
| Operating Temperature | 0~40°C | |
| Humidity | <80% | |
| Dimensions (L x W x H) | 218.60mm x 103.30mm x 87.30mm | |
7.2 Waveform Characteristics
| Parameter | Specification |
|---|---|
| Wave length | 8192 points/channel |
| Waveform types | Sine, square, pulse (duty cycle, pulse width, cycle time adjustable), triangle, rising/falling sawtooth, CMOS, DC level, half/full wave, positive step, inverse staircase, exponential rise/fall, Lorentz pulse, multi-sound, irregular noise, trapezoidal pulse, sink pulse, AM/FM waveforms, 99 user-defined waveforms. |
| Square wave frequency range | 1nHz-30MHz |
| Triangle wave frequency range | 1nHz-50MHz |
| Pulse wave frequency range | 1nHz-30MHz |
| TTL digital wave frequency range | 1nHz-20MHz |
| Arbitrary wave frequency range | 1nHz-50MHz |
| Pulse width adjustment range | 5nS~4S |
| Pulse period adjustment range | 10nS~40S |
| Frequency minimum resolution | 1nHz |
| Frequency accuracy | ±5ppm 0 to 50°C |
| Frequency stability | ±3ppm per 1 year |
| Sine wave spectral purity (Total harmonic distortion) | ≤0.5% (@1kHz, 5Vpp) |
| Square wave rising/falling edge time | ≤10nS |
| Square wave overshoot | ≤10% |
| Square wave asymmetry | ≤0.1% |
| Ramp and triangle wave linearity | ≤1% (Less than 1MHz) |
| Ramp and triangle wave asymmetry | ≤1% (Less than 1MHz) |
| Pulse wave duty cycle | 0.01%-99.99% |
| Arbitrary wave minimum rise/fall time | ≤10nS |
| Arbitrary wave overshoot | ≤10% |
| Arbitrary waveform non-volatile storage number | 99 |
7.3 Output Characteristics
| Parameter | Specification |
|---|---|
| Amplitude range (Frequency ≤1MHz) | 2mVpp~25Vpp |
| Amplitude range (1MHz < Frequency < 11MHz) | 2mVpp~10Vpp |
| Amplitude range (11MHz < Frequency < 60MHz) | 2mVpp~5Vpp |
| Amplitude range (60MHz < Frequency < 80MHz) | 2mVpp~3.6Vpp (PSG9080 only) |
| Amplitude resolution | 1mVpp |
| Amplitude stability | ±1% ±1 mVpp (@ 1 kHz, >10 mVpp) |
| Amplitude flatness (<10MHz) | ±1% (0.1dB) |
| Amplitude flatness (10MHz~50MHz) | ±2% (0.2dB) |
| Amplitude flatness (50MHz~70MHz) | ±10% (0.9dB) |
| Amplitude flatness (70MHz~80MHz) | ±20% (1.9dB) (PSG9080 only) |
| DC bias offset adjustment range | -9.99V-15.00V |
| Offset resolution | 0.01V |
| Phase adjustment range | 0°-359.9° |
| Phase resolution | 0.1° |
| Output impedance | 50Ω (nominal) |
7.4 Modulation Characteristics
| Modulation Type | Carrier Wave | Modulation Source | Modulated Wave | Parameters |
|---|---|---|---|---|
| AM | Sine, square, sawtooth, arbitrary (except DC) | Internal/External | Sine, square, sawtooth, noise, arbitrary | Depth: 0-200%, Freq: 1mHz-1MHz |
| FM | Sine, square, sawtooth, arbitrary (except DC) | Internal/External | Sine, square, sawtooth, noise, arbitrary | Freq: 1mHz-1MHz, Deviation: 0.1Hz-10kHz |
| PM | Sine, square, sawtooth, arbitrary (except DC) | Internal/External | Sine, square, sawtooth, noise, arbitrary | Phase Deviation: 0°-360°, Freq: 1mHz-1MHz |
| ASK | Sine, square, sawtooth, arbitrary (except DC) | Internal/External | 50% duty cycle pulse wave | Amplitude: 0-200%, Polarity: Positive/Negative, Rate: 1mHz-1MHz, Freq Hopping: 0.1Hz-80MHz |
| FSK | Sine, square, sawtooth, arbitrary (except DC) | Internal/External | 50% duty cycle pulse wave | Polarity: Positive/Negative, Rate: 1mHz-1MHz, Freq Hopping: 0.1Hz-80MHz |
| PSK | Sine, square, sawtooth, arbitrary (except DC) | Internal/External | 50% duty cycle pulse wave | Polarity: Positive/Negative, Rate: 1mHz-1MHz, Phase: 0°-360° |
| PWM | Pulse wave | N/A | Regular and reverse | Pulse Period: 0.01μS-40S, Pulse Width: 0.001μS-4S |
| Burst | Sine, square, sawtooth, arbitrary (except DC) | Internal/External | N/A | Number: 1-1,000,000,000, Trigger: Manual, CH2, External (AC/DC), Idle Mode: Zero, Positive/Negative Max |
7.5 Measurement and Counting Function
| Parameter | Specification |
|---|---|
| Measurement functions | Frequency, positive/negative pulse width, period, duty cycle |
| Frequency measurement range | 1Hz~100MHz |
| Input signal amplitude range | 2Vpp~20Vpp |
| Period measurement | 10nS~4S |
| Pulse Width | 4S |
| Duty cycle measurement range | 0.01%~99.99% |
| Counter range | 0-4294967295 |
| Counter coupling method | DC and AC coupling |
| Counting method | Manual |
7.6 External Modulation Input Characteristics
| Parameter | Specification |
|---|---|
| Input signal amplitude range | 0~3Vpp |
| Input signal frequency range | DC~200kHz |
7.7 Sweep Characteristics
| Parameter | Specification |
|---|---|
| Sweep channel | CH1 or CH2 |
| Carrier wave | Sine, square, sawtooth, arbitrary (except DC) |
| Sweep function | Sweep frequency, sweep amplitude, sweep duty |
| Sweep mode | Linear, logarithmic |
| Sweep direction | Increment, decrement, round trip |
| Start/end frequency | Consistent with carrier frequency limits |
| Start/end amplitude | Consistent with carrier amplitude limits |
| Start/end duty cycle | Consistent with carrier duty cycle limits |
| Sweep time | 0.01S-640S |
7.8 Voltage Control Characteristics
| Parameter | Specification |
|---|---|
| Voltage control channel | CH1 or CH2 |
| Carrier wave | Sine, square, sawtooth, arbitrary (except DC) |
| Voltage control function | Frequency control, amplitude control, duty cycle control |
| Sweep mode | Linear, logarithmic |
| Voltage control voltage range | 0V-5V (start/end points arbitrarily settable) |
7.9 Programming Features
| Parameter | Specification |
|---|---|
| Running mode | Debugging, normal |
| Storage | P00-P19 |
| Programming number | 00-99 |
| Programming time (single serial number) | 0-99S |
8. Maintenance
Proper maintenance ensures the longevity and optimal performance of your JUNTEK signal generator.
- Keep Surfaces Clean and Dry: Regularly clean the instrument's surfaces to prevent dust and moisture from affecting performance. Use a soft, dry cloth. For the LCD screen, wipe carefully with a soft cloth.
- Adequate Ventilation: Ensure the instrument is placed in a well-ventilated area. Regularly inspect the heat dissipation holes and fan (if present) to ensure they are not obstructed.
- Environmental Conditions: Operate and store the instrument within the specified temperature ranges (operating: 10~50℃; storage: 20~70℃) and in a dry environment.
- Physical Care: The instrument's shell is fragile. Avoid impacts and keep it away from corrosive chemicals. Handle with care during movement to prevent damage to internal circuits, keys, knobs, and interfaces.
- Avoid Unsafe Items: Do not place lighted candles, cups with water, or corrosive chemicals on the instrument.
9. Troubleshooting
This section addresses common issues and their solutions. If a problem persists, contact JUNTEK support.
- Screen is dark (no display) after pressing the power key:
- Check if the power is correctly connected.
- Verify that the power key was pressed firmly.
- Restart the instrument after performing the above checks.
- If the issue continues, contact JUNTEK.
- Screen is too dark to read:
- Check if brightness and contrast settings are too low.
- Press the SYS key to enter system settings, navigate to brightness, and adjust using the knob.
- Settings are correct, but no waveform is generated:
- Ensure the BNC cable is tightly connected to the correct channel output terminal (CH1 or CH2).
- Check if the BNC cable has internal damage.
- Verify the BNC cable is tightly connected to the test instrument.
- Check if the backlight for CH1 or CH2 is on. If not, press the corresponding key to illuminate it.
- After checking the above, press and restart the instrument.
- If the issue persists, contact JUNTEK.
10. User Tips
- Professional Use: The PSG9080/PSG9060 is a robust and reliable instrument, frequently found in electronics workshops due to its performance and versatility.
- Model Selection: The primary difference between the PSG9060 and PSG9080 models is their maximum frequency output (60 MHz vs. 80 MHz, respectively). Choose the model that best suits your frequency requirements.
- Connectivity: While the device offers USB and serial (TTL) connectivity for PC control, it does not include a Wi-Fi module. Ensure you have the necessary cables for wired connections.
- Software Updates: Keep the PC software and instrument firmware updated for optimal performance and access to the latest features. Firmware updates can be performed via the PC software.
11. Warranty and Support
11.1 Warranty
JUNTEK warrants that the product mainframe and accessories will be free from defects in materials and workmanship within the warranty period. If a product proves defective within this period, JUNTEK guarantees free replacement or repair. This warranty does not cover damages caused by improper use or unauthorized modifications. There are no other warranties, expressed or implied, beyond what is expressly stated herein or in any applicable warranty card. JUNTEK is not liable for any consequential, indirect, ensuing, or special damages.
11.2 Contact and Support
For any problems or requirements regarding the product or this manual, please contact JUNTEK:
- E-mail: junce@junteks.com
- Website: www.junteks.com
Additional resources available:
- Operation Demo Video: Provides operational guidance.
- PC Software and Communication Protocol: Available for download.
- User Manual (PDF): A comprehensive PDF version of the user manual is available for download from the official website. (Reference: User manual (PDF))