ZOYI ZT-703S

Contents

ZOYI ZT-703S Two-Channel Oscilloscope Multimeter: 3-in-1 Electrical Meter User Manual

Model: ZT-703S | Brand: ZOYI

1. Introduction and Overview

The ZOYI ZT-703S is a versatile 3-in-1 electrical testing device, integrating a digital oscilloscope, a true RMS multimeter, and a signal generator into a single, compact unit. Designed for high-precision measurements, it features a 3.5-inch IPS color screen, dual-channel input, and a 50MHz bandwidth oscilloscope. This manual provides essential information for the safe and effective operation, maintenance, and troubleshooting of your ZT-703S device.

ZOYI ZT-703S 3-in-1 device compared to traditional separate equipment
Figure 1.1: ZT-703S 3-in-1 functionality versus traditional separate equipment, highlighting its integrated oscilloscope, signal generator, multimeter, IPS color screen, data storage, and compact size.
ZOYI ZT-703S highlighting its 3-in-1 features
Figure 1.2: Key features of the ZT-703S, including multimeter, signal generator, post-glow mode, dual-channel measurement, and 50MHz bandwidth.

2. Setup

2.1 Battery Installation and Charging

The ZT-703S comes with a rechargeable lithium-ion battery (3400mAh) included. Ensure the battery is properly installed before first use. To charge the device, connect it to a 5V/2A Type-C charger. A full charge ensures optimal performance and portability.

ZOYI ZT-703S charging with a 3400mAh battery
Figure 2.1: The ZT-703S device being charged, illustrating its 3400mAh rechargeable lithium-ion battery.

2.2 Connecting Probes

Depending on the function (oscilloscope or multimeter), connect the appropriate probes to the input terminals. For oscilloscope measurements, use the BNC connectors. For multimeter functions, use the standard banana plug inputs (10A, mA, COM, VΩHz). Ensure all connections are secure before taking measurements.

2.3 Powering On/Off

Press and hold the power button (usually marked with a circle and line) to turn the device on or off. The 3.5-inch IPS color screen will illuminate upon startup.

ZOYI ZT-703S displaying a waveform on its 3.5-inch TFT color screen
Figure 2.2: A close-up view of the ZT-703S's 3.5-inch TFT color screen displaying a waveform.

3. Operating Instructions

3.1 Mode Selection

Use the 'MODE' button to switch between the Oscilloscope, Multimeter, and Signal Generator functions. The current mode will be indicated on the screen.

3.2 Oscilloscope Operation

The ZT-703S features a 50MHz bandwidth oscilloscope with dual-channel synchronous input. It supports real-time sampling and various trigger modes.

  • Dual-Channel Input: Connect two probes to simultaneously view and compare two different waveforms.
  • Waveform Display: The device can display various waveforms, including sine, square, and triangle waves.
  • Trigger Modes: Utilize multiple trigger modes (Automatic, Normal, Single) and trigger edges (Rising edge, Falling edge) to stabilize and capture specific waveform events.
  • Post-Glow Mode: This feature allows for the display of waveform persistence, useful for observing transient signals or signal variations over time.
  • Automatic Measurement: The device can automatically measure parameters such as Period, Frequency, Max value, Min value, Peak-to-peak value, Effective value, Duty cycle, and Coefficient.
ZOYI ZT-703S displaying various waveforms in real-time
Figure 3.1: Real-time display of various waveforms on the ZT-703S screen.
ZOYI ZT-703S showing dual-channel input capability
Figure 3.2: Illustration of the dual-channel input capability of the ZT-703S, comparing single versus dual channel displays.
ZOYI ZT-703S demonstrating multiple trigger modes
Figure 3.3: The ZT-703S displaying various trigger modes for waveform capture.
ZOYI ZT-703S showing post-glow mode for waveform persistence
Figure 3.4: The ZT-703S in post-glow mode, illustrating waveform persistence.

3.3 Multimeter Operation

The ZT-703S functions as a 25000 counts True RMS multimeter, capable of measuring a wide range of electrical parameters with automatic ranging.

  • Measurements: Supports measurement of DC/AC Voltage, DC/AC Current, Resistance, Capacitance, Frequency, Diode, and Continuity.
  • Automatic Ranging: The 'AUTO RANGE' button allows the device to automatically select the appropriate measurement range for convenience.
  • Data Hold: Use the 'HOLD SAVE' button to freeze the current reading on the display.
ZOYI ZT-703S performing voltage measurement
Figure 3.5: The ZT-703S in use, demonstrating its multimeter function for voltage measurement.

3.4 Signal Generator Operation

The integrated signal generator can output various waveforms for testing and calibration purposes.

  • Waveform Output: Generate sine wave, square wave, triangle wave, and sawtooth wave signals.
  • Parameter Adjustment: Adjust parameters such as voltage peak-to-peak (VPP) and frequency (FREQ) as needed.
ZOYI ZT-703S displaying signal generator output options
Figure 3.6: The ZT-703S screen showing options for signal generator output, including sine, square, and triangle waves.

3.5 Data Management

  • Data Saving: The 'HOLD SAVE' button can also be used to save measurement data or waveforms.
  • PC Connection: Connect the device to a computer via USB for data transfer and further analysis.
ZOYI ZT-703S connected to a computer for data saving
Figure 3.7: The ZT-703S connected to a laptop, illustrating its capability for data saving and transfer to a computer.

3.6 General Controls

  • REL (Relative Measurement): Used for relative measurements.
  • AUTO RANGE: Toggles automatic ranging on/off.
  • HOLD SAVE: Holds the current reading or saves data.
  • MODE: Switches between main functions (Oscilloscope, Multimeter, Signal Generator).
  • MENU: Accesses device settings and options.
  • Navigation Buttons: Used to navigate menus and adjust parameters.
  • Language Switching: The device supports switching between different languages.

4. Features

The ZOYI ZT-703S offers a comprehensive set of features for electrical testing:

  • Integrated Oscilloscope, Signal Generator, and Multimeter functions.
  • Voltage Measurement capabilities.
  • Upgradable firmware.
  • 3.5-inch IPS color screen with full viewing angle.
  • Diode and Continuity testing.
  • Measurement of multiple waveform types.
  • Waveform saving and data saving functions.
  • Intelligent overheat protection.
  • Automatic adjustment (Auto Range).
  • Data transfer to PC.
  • Language switching.
  • True RMS measurement.
Icon grid of ZOYI ZT-703S functional characteristics
Figure 4.1: An iconographic representation of the ZT-703S's functional characteristics, including oscilloscope, signal generator, multimeter, voltage measurement, upgradability, 3.5-inch screen, diode, multiple waveform measurement, full viewing angle, waveform saving, intelligent overheat protection, data saving, automatic adjustment, data transfer, language switching, and True RMS.

5. Specifications

5.1 General Characteristics

FeatureSpecification
Display25000 counts
Measurement RangeAutomatic / Manual
ChargeType-C (5V / 2A)
Battery Capacity3400 mAh
Supported FunctionsMultimeter, Oscilloscope, Signal Generator
MaterialABS / TPE
Dimensions177 * 89 * 40 mm
Weight380g

5.2 Oscilloscope Parameters

ParameterSpecification
Bandwidth50 MHz, two channels
Sampling MethodReal-time sampling
Sampling Rate200 MSa/s
Input CouplingDC, AC
Input Impedance1 MΩ, @16pF
AttenuationX1, X10
Record LengthMinimum 500, 15s, 10s, infinite
Time Base Range10ns/div - 20s/div
Time Base Accuracy20 ppm
Record DepthUp to 128 Kbit
Sensitivity20mV/div - 10V/div
Offset Range4 grids
Analog Bandwidth50 MHz
Trigger FrequencyLess than 10Hz
Automatic MeasurementPeriod, Frequency, Max value, Min value, Peak-to-peak value, Effective value, Duty cycle, Coefficient
Trigger ModeAutomatic, Normal, Single
Trigger EdgeRising edge, Falling edge
Signal Generator OutputSine wave, Square wave, Triangle wave, Sawtooth wave
Operating ModesYT, XY, Roll mode

5.3 Multimeter Parameters

MeasurementRangeAccuracy
DC Voltage1000.0V / 250.00V / 25.000V / 2.5000V±(0.5%+3)
250.00mV±(0.5%+3)
AC Voltage750.0V / 250.00V / 25.000V / 2.5000V±(0.5%+3)
250.00mV±(0.5%+3)
DC Current10.000A / 2.5000A±(0.5%+3)
250.00mA / 25.000mA±(0.5%+3)
AC Current10.000A / 2.5000A±(0.5%+3)
250.00mA / 25.000mA±(0.5%+3)
Resistance250.00KΩ / 2.5000KΩ / 2.5000Ω±(0.2%+3)
250.00MΩ / 2.5000MΩ±(1%+3)
250.0MΩ±(5.0%+5)
Capacitance999.9μF / 99.99μF / 9.999μF±(5.0%+20)
999.9nF / 99.99nF±(5.0%+5)
Frequency999.9Hz / 99.99Hz / 9.999kHz / 99.99kHz / 999.9kHz±(0.1%+2)
9.999MHz±(0.1%+2)
DiodeYesN/A
ContinuityYesN/A
DC Current AccuracyN/A±3%
Detailed electrical and multimeter specifications for ZOYI ZT-703S
Figure 5.1: Comprehensive tables detailing the electrical characteristics and multimeter parameters of the ZOYI ZT-703S.

6. Maintenance

6.1 Cleaning

Regularly clean the device with a soft, dry cloth. Avoid using abrasive cleaners or solvents that could damage the casing or screen. Ensure the device is powered off before cleaning.

6.2 Battery Care

To prolong battery life, avoid fully discharging the device frequently. Recharge the 3400mAh lithium-ion battery when the low battery indicator appears. If storing the device for an extended period, ensure it is partially charged (around 50%) and recharge it periodically.

6.3 Fuse Information

The ZT-703S is equipped with intelligent overheat protection, which includes internal fuses to safeguard the device. If the device malfunctions due to an overload, the fuses may need inspection or replacement. Refer to a qualified technician for fuse replacement to ensure proper protection and functionality.

Internal view of ZOYI ZT-703S showing fuses for overheat protection
Figure 6.1: An internal view of the ZT-703S, highlighting the fuses that provide intelligent overheat protection.

7. Troubleshooting

If you encounter issues with your ZOYI ZT-703S, consider the following general troubleshooting steps:

  • Device not powering on: Ensure the battery is charged. Connect the device to a Type-C charger and try again.
  • Incorrect readings: Check probe connections. Ensure the correct mode (Oscilloscope, Multimeter) is selected for the measurement. Verify that the measurement range is appropriate (use AUTO RANGE if unsure).
  • No waveform display: In oscilloscope mode, check probe connections, input coupling (AC/DC), and trigger settings. Adjust the time base and sensitivity settings.
  • Signal generator not outputting: Verify the signal generator mode is active and output parameters are correctly set.
  • Device unresponsive: Try performing a soft reset by holding down the power button for an extended period (e.g., 10-15 seconds) until the device restarts or powers off.

For persistent issues, please refer to the support section or contact the seller for assistance.

8. Warranty and Support

Your ZOYI ZT-703S device is backed by certain guarantees and support options:

  • Free Returns: Enjoy free returns within 90 days of purchase.
  • Fast Delivery Guarantee: In case of delayed delivery, a $1.00 coupon code may be provided. Refunds are available if the package is lost or items are damaged during transit, or if there is no delivery within 30 days.
  • Technical Support: For any technical queries, operational difficulties not covered in this manual, or warranty claims, please contact the seller directly. Provide your purchase details and a clear description of the issue for prompt assistance.