Hantek CC65, CC650, CC800

Contents

Hantek AC/DC Current Clamp Probe User Manual

Models: CC65, CC650, CC800

1. Introduction

The Hantek AC/DC Current Clamp Probe is a versatile tool designed for accurate current measurement in conjunction with oscilloscopes and multimeters. It utilizes Permalloy and Hall components for non-contact measurement based on the principle of electromagnetic induction, ensuring the integrity of the circuit under test. This manual provides detailed instructions for the safe and effective use of your current clamp probe.

Hantek AC/DC Current Clamp Probe models CC65, CC650, CC800 with 'Official Certification' logo.

Figure 1: Hantek AC/DC Current Clamp Probe models CC65, CC650, CC800.

2. Safety Information

Always observe the following safety precautions to prevent injury and avoid damage to the instrument:

  • Do not exceed the maximum voltage rating of 300VDC or 240V RMS AC when measuring. Exceeding this limit can cause overheating and damage to the instrument.
  • Ensure the current clamp's direction is consistent with the current flow direction for accurate readings.
  • When not in use, unplug the battery to prevent discharge.
  • Always connect the probe to the oscilloscope or multimeter before clamping onto a live wire.

3. Setup

3.1. Battery Installation

  1. The current clamp probe is powered by a 9V block battery (6F22). This battery is not included in the standard package and must be purchased separately.
  2. Open the battery compartment cover located on the back of the unit.
  3. Connect the 9V battery to the battery terminals, ensuring correct polarity.
  4. Place the battery into the compartment and close the cover securely.

Note: When not in use, the battery should be unplugged to prevent it from being in a standby state and consuming power. A red 'LO BAT' indicator light signifies a low battery warning.

Hantek CC65 Current Clamp Probe with battery compartment open, showing battery terminals.

Figure 2: Battery compartment of the CC65 Current Clamp Probe.

3.2. Connecting to Measurement Device

The Hantek Current Clamp Probe features a BNC interface for direct connection to an oscilloscope. For multimeter use, an optional multimeter adapter is required.

3.2.1. Oscilloscope Connection

  • Connect the BNC plug of the current clamp probe directly to a BNC input channel on your oscilloscope (e.g., Channel 1).
  • For DC current measurements, ensure the oscilloscope is set to DC coupling.

3.2.2. Multimeter Connection (Optional)

  • If using a multimeter, connect the optional universal DMM adapter to the current clamp probe.
  • Insert the multimeter's COM end and V-Ω end into the adapter.
  • Ensure the input impedance of the multimeter is at least 10k ohms.
Diagram showing connection of current clamp probe to both an oscilloscope and a multimeter via an adapter.

Figure 3: Connection setup for oscilloscope and multimeter.

4. Operating Instructions

4.1. Power On and Zero Adjustment

  1. Turn on the current clamp probe. A green 'POWER' indicator light will illuminate if the battery is sufficient.
  2. Before measurement, perform a zero adjustment. Press and hold the 'ZERO' button until the oscilloscope's waveform returns to the reference (zero) position. Then release the button.

4.2. Selecting Measurement Range (Gear Switch)

The current clamp probe has a gear switch to select the appropriate measurement range:

  • 10mV/A: For measuring higher currents.
  • 100mV/A: For measuring lower currents, offering higher sensitivity.

Select the gear suitable for your measurement parameters. Refer to the parameter accuracy tables in the Specifications section for guidance. For oscilloscope use, set the corresponding attenuation multiple (e.g., 10X for 100mV/A gear, 100X for 10mV/A gear).

Hantek CC65 Current Clamp Probe showing gear switch, zero button, and correct/incorrect clamping methods.

Figure 4: Controls and clamping examples.

4.3. Performing a Measurement

  1. Ensure the current clamp is properly connected to your oscilloscope or multimeter and zero-adjusted.
  2. Open the clamp jaws using the wrench button.
  3. Carefully clamp the probe onto the conductor carrying the current you wish to measure. Ensure the conductor is centered within the jaws.
  4. Important: The direction of the current clamp must be consistent with the direction of the current flow for accurate readings. The front side of the label indicates forward current (positive waveform), and the label down indicates reverse current.
  5. Observe the current waveform and value on your oscilloscope or multimeter.

Video 1: Demonstration of current measurement using the Hantek Current Clamp Probe with an oscilloscope.

4.4. Interpreting Results

The oscilloscope will display the current waveform, allowing for analysis of instantaneous current, starting current, and other characteristics. The multimeter will show the RMS or DC current value depending on its settings.

Oscilloscope screen displaying current waveforms for both DC and AC measurements.

Figure 5: Example current waveforms on an oscilloscope screen.

5. Maintenance

  • Keep the probe clean and free from dust and moisture. Use a soft, dry cloth for cleaning.
  • Store the probe in a dry environment when not in use.
  • Always remove the 9V battery if the probe will not be used for an extended period to prevent battery leakage and damage to the device.
  • Avoid dropping or subjecting the probe to strong impacts.

6. Troubleshooting

  • No Power / LO BAT Indicator: Check the 9V battery. Replace if low or depleted. Ensure it is correctly installed.
  • Inaccurate Readings:
    • Perform zero adjustment before measurement.
    • Ensure the correct gear (10mV/A or 100mV/A) is selected for the expected current range.
    • Verify the probe ratio setting on the oscilloscope matches the clamp's sensitivity (e.g., 100X for 10mV/A).
    • Check that the current clamp is properly closed around a single conductor.
    • Ensure the direction of the current clamp aligns with the current flow.
    • Avoid strong electromagnetic interference in the measurement environment.
  • Over-range Indication: If the current exceeds the selected range, the instrument may heat up or be damaged. Select a higher current range (e.g., switch from 100mV/A to 10mV/A if measuring higher currents).

7. Specifications

7.1. General Specifications

  • Brand: Hantek
  • Power Supply: 9V block battery (6F22)
  • Output Interface: BNC Plug
  • Conductor Size:
    • CC65: Max diameter 9mm
    • CC650/CC800: Max diameter 30mm
Comparison image showing the maximum conductor diameter for CC65 (9mm) and CC650/CC800 (30mm) models.

Figure 6: Conductor size comparison for different models.

7.2. Measurement Parameters and Accuracy (CC65)

Oscilloscope attenuation: 10x for 100mV/A gear, 100x for 10mV/A gear.

Machine GearMeasuring RangeAccuracy Range
Direct Current100mV/A10mA-20A±(1.5%+5mA)
10mV/A100mA-40A±(2%+20mA)
40A-65A±(4%+0.3A)
Alternating Current100mV/A100mA-10A (40Hz-2KHz)±(2%+30mA)
100mA-10A (2KHz-10KHz)±(4%+30mA)
10A-15A (40Hz-20KHz)±(8%+30mA)
10mV/A100mA-40A (40Hz-1KHz)±(2%+30mA)
100mA-40A (1KHz-2KHz)±(4%+30mA)
100mA-40A (3KHz-5KHz)±(6%+30mA)
40A-65A (40Hz-20KHz)±(8%+0.3A)
Load ResistanceTypical value is 10K Ω
Temperature Coefficient0.1X (precision specified) temperature coefficient. (0°C to 18°C, 28°C to 50°C)

Figure 7: CC65 Measurement Parameters and Accuracy.

7.3. Measurement Parameters and Accuracy (CC650 / CC800)

Oscilloscope attenuation: 100x for 10mV/A gear, 1000x for 1mV/A gear. The voltage of the measured conductor must be less than 300V DC or 240V RMS AC.

Model NumberMachine GearMeasuring RangeAccuracy Range
CC650 Direct Current10mV/A1A - 40A±(1.5%+0.5 A)
1mV/A1A - 650A±(1.5%+0.5 A)
CC800 Direct Current10mV/A1A - 40A±(1.5%+0.5 A)
1mV/A1A - 800A±(1.5%+0.5 A)
CC650/CC800 Alternating Current10mV/A1A - 40A (50Hz - 500Hz)±(2%+0.5A)
1mV/A1A - 500A (50Hz - 5KHz)±(2%+0.5A)
Load ResistanceTypical value is 10K Ω
Temperature Coefficient0.1X (precision specified) temperature coefficient. (0°C to 18°C, 28°C to 50°C)

Figure 8: CC650/CC800 Measurement Parameters and Accuracy.

8. User Tips

  • For the most accurate readings, ensure the conductor is positioned centrally within the clamp jaws.
  • Always perform a zero adjustment before each measurement, especially when switching between DC and AC modes or changing ranges.
  • When using with an oscilloscope, remember to adjust the vertical scale (Volts/Div) and probe attenuation settings to match the current clamp's output sensitivity (mV/A) and probe ratio for correct current display.
  • If measuring very small currents, use the 100mV/A range for higher sensitivity, and ensure minimal electromagnetic interference.