JUNCTEK PSG9060

Contents

JUNCTEK PSG9060 Programmable DDS Signal Generator User Manual

Model: PSG9060

1. Introduction

The JUNCTEK PSG9060 programmable dual-channel function/arbitrary waveform signal generator is a versatile instrument capable of generating various waveforms including sine, square, triangle, pulse, and arbitrary waves. It features a frequency range from 1nHz to 60MHz (for sine wave) and includes functions such as modulation, frequency sweep, measurement signal frequency, and programming.

The instrument can display output signal, amplitude, phase, duty cycle, and frequency simultaneously. Its design emphasizes excellent amplitude-frequency characteristics, multi-functionality, high performance, and portability, making it suitable for education, research and development, production, and testing environments. The PSG9060 is equipped with a 3.5-inch high-resolution color LCD screen for a clear visual experience.

Product overview video demonstrating key features and functions.

2. Safety Requirements

Please review the following safety precautions carefully before operating the instrument to prevent personal injury or damage to the device. Always follow the instructions in this manual for proper use.

General Safety Summary

  • Use Proper Power Cord: Only use the exclusive power cord designed for the instrument and authorized for use in your local country.
  • Connect the Probe Correctly: If a probe is used, do not connect the ground lead to high voltage as it has isobaric potential with the ground.
  • Observe All Terminal Ratings: To avoid fire or shock hazards, observe all ratings and markers on the instrument and consult this manual for more information before connecting.
  • Use Proper Over-voltage Protection: Ensure no over-voltage (e.g., from lightning) reaches the product to prevent electric shock.
  • Do Not Operate Without Covers: Do not operate the instrument with covers or panels removed.
  • Do Not Insert Anything Into the Air Outlet: Avoid inserting objects into the air outlet to prevent damage.
  • Avoid Circuit or Wire Exposure: Do not touch exposed junctions and components when the unit is powered on.
  • Do Not Operate With Suspected Failures: If damage is suspected, have the instrument inspected by JUNTEK authorized personnel before further operations. All maintenance, adjustments, or replacements must be performed by authorized personnel.
  • Provide Adequate Ventilation: Inadequate ventilation can cause overheating and damage. Keep the instrument well-ventilated and regularly inspect the air outlet and fan.
  • Do Not Operate in Wet Conditions: To avoid short circuits or electric shock, never operate the instrument in a humid environment.
  • Do Not Operate in an Explosive Atmosphere: Never operate the instrument in an explosive atmosphere to prevent personal injury or damage.
  • Keep Instrument Surfaces Clean and Dry: Maintain clean and dry surfaces to prevent dust or moisture from affecting performance.
  • Prevent Electrostatic Impact: Operate in an electrostatic discharge protective environment. Ground internal and external cable conductors to release static before connections.
  • Handle with Caution: Handle with care during transportation to avoid damage to keys, knobs, interfaces, and other panel parts.

Notices

  • Ensure the input power is correct.
  • The instrument shell is fragile; avoid hitting it or exposing it to chemicals.
  • Operating temperature: 10~50℃; storage temperature: 20~70℃. Keep in a dry environment.
  • Do not attempt to disassemble the instrument; it voids the warranty. Repairs must be made by designated repair outlets or the factory.
  • Avoid placing unsafe items (e.g., lighted candles, cups with water, corrosive chemicals) on the instrument surface.
  • The display screen is fragile; do not touch or bump it. Clean the LCD surface carefully with a soft cloth.
  • Do not move the instrument violently to avoid internal circuit damage. Contact the supplier if it malfunctions.

3. Setup and Package Contents

3.1. Inspecting the Packaging and Contents

Upon receiving your PSG9060, it is recommended to inspect the instrument following these steps:

  1. Inspect the Packaging: If the packaging is damaged, do not discard it until the shipment has been checked for completeness and passed electrical and mechanical tests. The consigner or carrier is liable for shipping damage.
  2. Check the Contents: Verify the contents against the packing list. If any items are damaged or missing, contact your JUNTEK sales representative.

Package Contents:

Item Quantity
PSG9060 Programmable Function/Arbitrary Waveform Signal Generator (Host) 1 pc
Power cable 1 pc
USB cable 1 pc
BNC to alligator clips test lead 1 pc
BNC male plug test lead 1 pc
Certificate of conformity 1 pc
Quick guide 1 pc
PSG9060 Signal Generator with included accessories: power cable, USB cable, BNC to alligator clips test lead, BNC male plug test lead, and quick start guide.

The PSG9060 signal generator and its standard accessories.

3.2. Power On and Initial Inspection

  1. Connect to Power: Connect the generator to an AC power supply using the provided power cord. The AC input specification is 85-264V, 47-63Hz. A DC5V±0.5V 3A power adapter can also be used.
  2. Power-on: Press the power key on the front panel to turn on the generator. The instrument will perform initialization and self-test. The default interface will be displayed. If the instrument does not start normally, refer to the "Troubleshooting" section.
  3. Set the System Language: After booting, press the corresponding soft key to select your desired system language. The language setting is saved for future power-ons.

4. Instrument Overview

4.1. Front Panel Overview

Front panel of the PSG9060 signal generator with labels for controls and display.

Front view of the PSG9060 signal generator.

Front Panel Components:

  • Power Key: Turns the generator on/off.
  • Signal Input Connectors (EXT.IN, MOD.IN): EXT.IN for external signal input (2Vpp~20Vpp), MOD.IN for modulation input (0Vpp~3Vpp).
  • Signal Output Connectors (CH1, CH2): BNC connectors for CH1 and CH2 output, 50Ω±10% nominal impedance.
  • Channel Control Area (CH1, CH2, OUT keys): Controls output for CH1 and CH2. CH1/CH2 keys toggle channel output (backlight indicates ON). OUT key controls both CH1 and CH2 output switches.
  • Direction Keys: Moves cursor for digit selection during parameter setting, deletes numbers.
  • Knob: Increases/decreases parameter values, quickly switches waveforms.
  • Numeric Keyboard (0-9, ., +/-): Used to set parameters.
  • Function Keys (SYS, MOD, MEAS, LOAD, SHIFT, SWEEP, VCO, FUNC): Access various modes and functions.
  • Menu Softkeys: Activates corresponding menu options displayed on the LCD.
  • LCD (3.5 inches TFT, 480x320): Displays current function menu, parameter settings, and waveforms.

4.2. Rear Panel Overview

Rear panel of the PSG9060 signal generator with labels for ports and power input.

Rear view of the PSG9060 signal generator.

Rear Panel Components:

  • Sync Ports: Synchronizes phases of output signals from different signal generators.
  • Communication Extension Port (TTL Ext.): TTL level serial port for secondary development (GPI, GPO1, GPO2, RXD, TXD, +5V, GND).
  • USB Interface: Connects to a computer for remote control via PC software or programming.
  • DC Power Socket: DC5V±0.5V 3A.
  • AC Power Socket: 85-264V, 47-63Hz, max 30W input, fuse 250V, T3.15A.
  • Switch: Turns the signal generator on or off.
  • Cooling Holes: Dissipates internal heat.
  • Chassis Ground: Connects to ground for electrical safety and normal operation.

4.3. Display Interface

Detailed view of the PSG9060 display interface showing waveform parameters and controls.

Example of the PSG9060 display interface.

Display Elements:

  • Soft Key Menu: Displays operation menu for the currently selected function/interface.
  • Finger Cursor: Indicates the currently selected channel (CH1 or CH2) for parameter changes.
  • Current Channel Status: Shows ON/OFF status of channels.
  • Phase: Displays current waveform phase.
  • Waveform Display: Shows the waveform currently in the main channel.
  • Frequency: Displays and allows adjustment of waveform frequency.
  • Amplitude: Displays and allows adjustment of waveform amplitude.
  • Offset: Displays and allows adjustment of DC offset.
  • Duty Cycle: Displays and allows adjustment of waveform duty cycle (not adjustable for sine wave).
  • Information Box: Displays entered values and save/load positions.

5. Operating Instructions

5.1. Outputting Basic Waveforms

The PSG9060 can output basic waveforms (Sine, Square, Triangle, Pulse) from one or both channels. By default, both channels output a sine waveform at 10kHz frequency and 5Vpp amplitude.

  1. Select Output Channel: Press the CH1 or CH2 key to select the desired channel. The finger cursor will indicate the active channel.
  2. Select Basic Waveforms: Press the Wave soft key. Use the knob or soft keys to select the desired waveform type.
  3. Set Frequency: Press the Freq. soft key. Use the numeric keyboard or direction keys and knob to adjust the frequency (range: 0-80MHz). Press Freq. again to select the unit (MHz, kHz, Hz, mHz, μHz).
  4. Set Amplitude: Press the Ampl. soft key. Use the numeric keyboard or direction keys and knob to adjust the amplitude. The amplitude range is frequency-dependent (refer to specifications).
  5. Set Offset: Press the Offset soft key. Use the direction keys and knob or numeric keyboard to adjust the DC offset.
  6. Set Duty Cycle: Press the Duty soft key. Use the direction keys and knob or numeric keyboard to adjust the duty cycle (0.01-99.99% for pulse wave; not adjustable for sine wave).
  7. Set Phase: On the CH2 interface, press the Phase soft key. Use arrow keys and knob/numeric keyboard to adjust phase difference (0°-359.99°). Ensure CH1 and CH2 frequency synchronization is set in system settings first.
  8. Observe Output Waveforms: Connect CH1 and CH2 to an oscilloscope using BNC test leads to observe the waveforms.

5.2. Modulation Mode Interface and Parameter Settings

The PSG9060 supports various modulation types: AM, FM, PM, ASK, FSK, PSK, PWM, and BURST. Modulation processes a signal by adding it to a carrier wave for transmission.

Diagram showing various modulation functions like FM, PM, AM, FSK, PSK and their parameters.

Overview of modulation functions and their parameters.

General Modulation Steps:

  1. Press the MOD key to enter the modulation mode interface.
  2. Press the Type soft key to select the desired modulation type (AM, FM, PM, ASK, FSK, PSK, PWM, BURST).
  3. Configure carrier waveform, modulation source (internal/external), modulation frequency/rate, depth/deviation, polarity, etc., as required for the specific modulation type.

For detailed parameter settings for each modulation type, refer to the user manual PDF section 3.3 (pages 25-33).

5.3. Measurement Mode Interface

The instrument can measure frequency, period, positive/negative pulse width, and duty cycle of input signals.

Screenshot of the PSG9060 measurement mode interface showing coupling, gate time, and measurement results.

Measurement mode interface.

Measurement Function:

  • Frequency Range: 1Hz-100MHz.
  • Signal Amplitude Range: 2Vpp-20Vpp (input via Ext.IN).
  • Coupling Setting: Use soft keys and knob to switch between AC or DC coupling.
  • Gate Time Setting: Adjust gate time from 0.001s to 10s (default 1s).
  • Measurement Mode: Switch between low frequency (<2kHz) or high frequency (>2kHz) modes.

Counter Function:

Calculates the number of input signal periods in real-time.

  • Coupling Setting: Same as measurement function.
  • After setting, press ON to start counting, OFF to stop, RST to restore defaults, and SAVE to save parameters.

5.4. Frequency Sweep Mode Interface

The PSG9060 supports frequency sweep, amplitude sweep, and duty cycle sweep functions.

Screenshot of the PSG9060 sweep frequency interface showing start/end frequency, sweep time, and sweep mode.

Sweep frequency interface.

Frequency Sweep:

  • Press SWEEP, then FUNC to select sweep type.
  • Sweep Channel: Switch between CH1 or CH2.
  • Start/End Frequency: Set the upper and lower limits of the sweep.
  • Sweep Time: Adjust scan time from 0.01s to 640s.
  • Sweep Direction: Select increment, decrement, or round trip.
  • Sweep Mode: Choose linear or logarithmic.
  • Press ON to start frequency sweep.

Amplitude Sweep & Duty Cycle Sweep:

Similar settings apply for amplitude and duty cycle sweeps, including channel selection, start/end values, sweep time, direction, and mode. Press ON to activate.

5.5. Voltage Control Mode Interface

The PSG9060 offers voltage control for frequency, amplitude, and duty cycle.

Screenshot of the PSG9060 voltage control frequency interface.

Voltage control frequency interface.

Voltage Control Frequency:

  • Press VCO, then FUNC to select frequency control.
  • Voltage Control Channel: Switch between CH1 or CH2.
  • Start/End Frequency: Set the range.
  • Min/Max Voltage Calibration: Calibrate external input signal voltage.
  • Voltage Control Mode: Choose linear or logarithmic.
  • Press ON to activate voltage control frequency.

Voltage Control Amplitude & Duty Cycle:

Similar settings apply for amplitude and duty cycle voltage control, including channel selection, start/end values, and calibration. Press ON to activate.

5.6. Programming Mode Interface

The PSG9060 allows for programmable waveform output, enabling automatic work sequences.

Screenshot of the PSG9060 programming mode interface showing program steps and parameters.

Programming mode interface.

  • Press FUNC to enter programming mode. Press MODE to switch between normal and debugging modes.
  • P00-P19: 20 storage locations for programs.
  • Serial Number (00-99): Up to 100 instructions per program.
  • Load: Indicates parameter storage location.
  • Time: Current running time of the serial number.
  • Go to: Transfers control to a specified serial number.
  • Cycle: Sets the number of repetitions for a serial number.
  • Press SAVE to save program data.

5.7. System Settings

Access system settings by pressing the SYS key.

  • Clear Storage: Clears parameters in the current position.
  • Sound Setting: Toggles sound ON/OFF.
  • Brightness Adjustment: Adjusts screen brightness.
  • Language Setting: Selects English or Chinese.
  • Built-in Wave Number: Sets the number of built-in waves (00-21).
  • Arbitrary Wave Number: Sets the number of arbitrary waves (01-99).
  • Waveform Loading Methods: Selects Auto or Fast.
  • System Information: Displays product model, hardware/software/FPGA versions, official website, and QR code.
  • Firmware Update: Press Update soft key to enter update mode, then connect to PC software.
  • Synchronization: Waveform, frequency, amplitude, and duty cycle synchronization can be turned ON/OFF for CH1/CH2.
  • External Synchronization: Toggles external synchronization ON/OFF.
  • Frequency Fine-tuning: Adjusts frequency for waveform synchronization (range 0-99).
  • Restore Factory Settings: Press ON, then Yes to restore.
  • System Color: Selects one of four system colors.
  • Auto Power-on: Toggles auto power-on ON/OFF.
  • CH1/CH2 Amplitude Fine-tuning: Calibrates amplitude for accuracy.

5.8. PC Software Control Output

The PSG9060 can be controlled via PC software. The software has both Chinese and English interfaces.

Diagram showing PC control and APP control options for the PSG9060.

PC and APP control options.

Software Setup:

  1. Download the PC software from the provided link: PSG9080_EN_Setup.zip.
  2. Run the installer. Select the desired language.
  3. Click Install USB Drive to install the serial port driver.
  4. Click Main program to install the application. Follow the prompts to complete installation.

Connecting to PC:

  1. Right-click Computer → Properties → Device Manager to observe the serial port assigned by the computer.
  2. In the PC software, select the corresponding serial port and click connect.
Detailed view of the new PC software interface for PSG9060, showing four function interfaces.

New PC software interface with four function interfaces.

6. Specifications

Parameter Specification
ModelPSG9060
Channel2
Sine wave frequency range1nHz-60MHz
Square wave frequency range1nHz-30MHz
Triangle wave frequency range1nHz-50MHz
Pulse wave frequency range1nHz-30MHz
TTL digital wave frequency range1nHz-20MHz
Arbitrary wave frequency range1nHz-50MHz
Frequency resolution1nHz
Sampling Rate300MSa/S
Wave length8192 points/channel
Waveform vertical resolution14 bits
Basic waveformsSine, square, pulse, ramp, CMOS, DC, partial sine, noise wave
Built-in waveformsSine, square, pulse, triangle, ramp, CMOS, DC level, partial sine, half, full, positive step, inverse step, noise, exponential rise/fall, logarithm ascending/descending, sinker pulse, multi-tone, lorenz
Pulse width adjustment range5nS~4S
Pulse period adjustment range10nS~40S
Frequency accuracy±5ppm 0 to 50℃
Frequency stability±3ppm per 1 year
Total harmonic distortion (Sine)≤0.5% (@1kHz, 5Vpp)
Square wave Rise/Fall Time≤10nS
Square wave Overshoot≤10%
Square wave Non-symmetry≤0.1%
Ramp/Triangle wave Non-linearity≤1% (Less than 1MHz)
Ramp/Triangle wave Non-symmetry≤1% (Less than 1MHz)
Pulse wave Duty cycle0.01%-99.99%
Pulse wave Leading/Trailing Edge Time≤10nS
Pulse wave Overshoot≤10%
Amplitude range (<1MHz)2mVpp~25Vpp
Amplitude range (1MHz-11MHz)2mVpp~10Vpp
Amplitude range (11MHz-60MHz)2mVpp~5Vpp
Amplitude range (60MHz-80MHz)2mVpp~3.6Vpp
Amplitude resolution1mVpp
Amplitude stability±1% ±1 mVpp (@ 1 kHz, >10 mVpp)
Amplitude flatness (<10MHz)±1% (0.1dB)
Amplitude flatness (10MHz-50MHz)±2% (0.2dB)
Amplitude flatness (50MHz-70MHz)±10% (0.9dB)
Amplitude flatness (70MHz-80MHz)±20% (1.9dB)
DC Offset Range-9.99V-12.00V
DC Offset Resolution0.01V
Phase adjustment range0°-359.99°
Phase resolution0.01°
Output impedance50Ω (typical)
ProtectionShort-circuit protection
Modulation typeAM, FM, PM, ASK, FSK, PSK, PWM, BURST
Trigger source (Burst)Manual burst, CH2 burst, external burst (AC), external burst (DC)
Measurement functionFrequency, positive/negative pulse width, period, duty cycle
Counter range0-4294967295
Sweep functionSweep frequency, sweep amplitude, sweep duty
Sweep modeLinear, logarithmic
VCO functionFrequency control, amplitude control, and duty cycle control
Program modeDebug and routinely store 100 groups of programs, each with 100 steps
Product dimension247*243*102mm
Net weight1038g
Package dimension340*280*140mm
Gross weight1900g

7. Troubleshooting

This section lists common issues and their solutions. If problems persist after following these steps, contact JUNTEK support.

7.1. Screen is dark (no display) after pressing the power key:

  • Check if the power is correctly connected.
  • Verify the power key is fully pressed.
  • Restart the instrument after checking the above.
  • If it still doesn't work, contact JUNTEK.

7.2. Screen is too dark to see contents clearly:

  • Check if brightness and contrast settings are too low.
  • Press the SYS key to enter system settings, navigate to brightness, and adjust with the knob.

7.3. Settings are correct but no waveform is generated:

  • Check if the BNC cable is tightly connected to the corresponding channel output terminal (CH1 or CH2).
  • Inspect the BNC cable for internal damage.
  • Check if the BNC cable is tightly connected to the test instrument.
  • Verify if the backlight of CH1 or CH2 is on. If not, press the corresponding key to illuminate it.
  • After checking the above, press and restart the instrument.
  • If it still doesn't work, contact JUNTEK.

8. User Tips

No specific user tips were found in the provided reviews or Q&A content.

9. Warranty and Support

9.1. Warranty

JUNTEK warrants that the product mainframe and accessories will be free from defects in materials and workmanship within the warranty period. If a product proves defective within this period, JUNTEK guarantees free replacement or repair. There are no other warranties, expressed or implied, except as expressly set forth herein or other applicable warranty card. JUNTEK is not liable for any consequential, indirect, ensuing, or special damages for any breach of warranty.

9.2. Support and Further Information

For more product information, including machine model and factory number, press the SYS key on the instrument. You can also refer to relevant manuals by visiting the official JUNTEK website (www.junteks.com) to download them.

Additional resources available:

  • "PSG9060 Operation Demo Video": Provides operational videos.
  • "PSG9060 PC Software and Communication Protocol": Provides details on PC software and communication protocols.
  • "PSG9060 User Manual": Includes technical specifications, functions, operation methods, possible failures, and solutions.

If you have any problems or requirements when using the product or this manual, please contact JUNTEK via email at junce@junteks.com.

A PDF version of the user manual is available for download: User manual (PDF).