1. Introduction
The JUNCTEK PSG9060 programmable dual-channel function/arbitrary waveform signal generator is a versatile instrument capable of generating various waveforms including sine, square, triangle, pulse, and arbitrary waves. It features a frequency range from 1nHz to 60MHz (for sine wave) and includes functions such as modulation, frequency sweep, measurement signal frequency, and programming.
The instrument can display output signal, amplitude, phase, duty cycle, and frequency simultaneously. Its design emphasizes excellent amplitude-frequency characteristics, multi-functionality, high performance, and portability, making it suitable for education, research and development, production, and testing environments. The PSG9060 is equipped with a 3.5-inch high-resolution color LCD screen for a clear visual experience.
Product overview video demonstrating key features and functions.
2. Safety Requirements
Please review the following safety precautions carefully before operating the instrument to prevent personal injury or damage to the device. Always follow the instructions in this manual for proper use.
General Safety Summary
- Use Proper Power Cord: Only use the exclusive power cord designed for the instrument and authorized for use in your local country.
- Connect the Probe Correctly: If a probe is used, do not connect the ground lead to high voltage as it has isobaric potential with the ground.
- Observe All Terminal Ratings: To avoid fire or shock hazards, observe all ratings and markers on the instrument and consult this manual for more information before connecting.
- Use Proper Over-voltage Protection: Ensure no over-voltage (e.g., from lightning) reaches the product to prevent electric shock.
- Do Not Operate Without Covers: Do not operate the instrument with covers or panels removed.
- Do Not Insert Anything Into the Air Outlet: Avoid inserting objects into the air outlet to prevent damage.
- Avoid Circuit or Wire Exposure: Do not touch exposed junctions and components when the unit is powered on.
- Do Not Operate With Suspected Failures: If damage is suspected, have the instrument inspected by JUNTEK authorized personnel before further operations. All maintenance, adjustments, or replacements must be performed by authorized personnel.
- Provide Adequate Ventilation: Inadequate ventilation can cause overheating and damage. Keep the instrument well-ventilated and regularly inspect the air outlet and fan.
- Do Not Operate in Wet Conditions: To avoid short circuits or electric shock, never operate the instrument in a humid environment.
- Do Not Operate in an Explosive Atmosphere: Never operate the instrument in an explosive atmosphere to prevent personal injury or damage.
- Keep Instrument Surfaces Clean and Dry: Maintain clean and dry surfaces to prevent dust or moisture from affecting performance.
- Prevent Electrostatic Impact: Operate in an electrostatic discharge protective environment. Ground internal and external cable conductors to release static before connections.
- Handle with Caution: Handle with care during transportation to avoid damage to keys, knobs, interfaces, and other panel parts.
Notices
- Ensure the input power is correct.
- The instrument shell is fragile; avoid hitting it or exposing it to chemicals.
- Operating temperature: 10~50℃; storage temperature: 20~70℃. Keep in a dry environment.
- Do not attempt to disassemble the instrument; it voids the warranty. Repairs must be made by designated repair outlets or the factory.
- Avoid placing unsafe items (e.g., lighted candles, cups with water, corrosive chemicals) on the instrument surface.
- The display screen is fragile; do not touch or bump it. Clean the LCD surface carefully with a soft cloth.
- Do not move the instrument violently to avoid internal circuit damage. Contact the supplier if it malfunctions.
3. Setup and Package Contents
3.1. Inspecting the Packaging and Contents
Upon receiving your PSG9060, it is recommended to inspect the instrument following these steps:
- Inspect the Packaging: If the packaging is damaged, do not discard it until the shipment has been checked for completeness and passed electrical and mechanical tests. The consigner or carrier is liable for shipping damage.
- Check the Contents: Verify the contents against the packing list. If any items are damaged or missing, contact your JUNTEK sales representative.
Package Contents:
| Item | Quantity |
|---|---|
| PSG9060 Programmable Function/Arbitrary Waveform Signal Generator (Host) | 1 pc |
| Power cable | 1 pc |
| USB cable | 1 pc |
| BNC to alligator clips test lead | 1 pc |
| BNC male plug test lead | 1 pc |
| Certificate of conformity | 1 pc |
| Quick guide | 1 pc |
The PSG9060 signal generator and its standard accessories.
3.2. Power On and Initial Inspection
- Connect to Power: Connect the generator to an AC power supply using the provided power cord. The AC input specification is 85-264V, 47-63Hz. A DC5V±0.5V 3A power adapter can also be used.
- Power-on: Press the power key on the front panel to turn on the generator. The instrument will perform initialization and self-test. The default interface will be displayed. If the instrument does not start normally, refer to the "Troubleshooting" section.
- Set the System Language: After booting, press the corresponding soft key to select your desired system language. The language setting is saved for future power-ons.
4. Instrument Overview
4.1. Front Panel Overview
Front view of the PSG9060 signal generator.
Front Panel Components:
- Power Key: Turns the generator on/off.
- Signal Input Connectors (EXT.IN, MOD.IN): EXT.IN for external signal input (2Vpp~20Vpp), MOD.IN for modulation input (0Vpp~3Vpp).
- Signal Output Connectors (CH1, CH2): BNC connectors for CH1 and CH2 output, 50Ω±10% nominal impedance.
- Channel Control Area (CH1, CH2, OUT keys): Controls output for CH1 and CH2. CH1/CH2 keys toggle channel output (backlight indicates ON). OUT key controls both CH1 and CH2 output switches.
- Direction Keys: Moves cursor for digit selection during parameter setting, deletes numbers.
- Knob: Increases/decreases parameter values, quickly switches waveforms.
- Numeric Keyboard (0-9, ., +/-): Used to set parameters.
- Function Keys (SYS, MOD, MEAS, LOAD, SHIFT, SWEEP, VCO, FUNC): Access various modes and functions.
- Menu Softkeys: Activates corresponding menu options displayed on the LCD.
- LCD (3.5 inches TFT, 480x320): Displays current function menu, parameter settings, and waveforms.
4.2. Rear Panel Overview
Rear view of the PSG9060 signal generator.
Rear Panel Components:
- Sync Ports: Synchronizes phases of output signals from different signal generators.
- Communication Extension Port (TTL Ext.): TTL level serial port for secondary development (GPI, GPO1, GPO2, RXD, TXD, +5V, GND).
- USB Interface: Connects to a computer for remote control via PC software or programming.
- DC Power Socket: DC5V±0.5V 3A.
- AC Power Socket: 85-264V, 47-63Hz, max 30W input, fuse 250V, T3.15A.
- Switch: Turns the signal generator on or off.
- Cooling Holes: Dissipates internal heat.
- Chassis Ground: Connects to ground for electrical safety and normal operation.
4.3. Display Interface
Example of the PSG9060 display interface.
Display Elements:
- Soft Key Menu: Displays operation menu for the currently selected function/interface.
- Finger Cursor: Indicates the currently selected channel (CH1 or CH2) for parameter changes.
- Current Channel Status: Shows ON/OFF status of channels.
- Phase: Displays current waveform phase.
- Waveform Display: Shows the waveform currently in the main channel.
- Frequency: Displays and allows adjustment of waveform frequency.
- Amplitude: Displays and allows adjustment of waveform amplitude.
- Offset: Displays and allows adjustment of DC offset.
- Duty Cycle: Displays and allows adjustment of waveform duty cycle (not adjustable for sine wave).
- Information Box: Displays entered values and save/load positions.
5. Operating Instructions
5.1. Outputting Basic Waveforms
The PSG9060 can output basic waveforms (Sine, Square, Triangle, Pulse) from one or both channels. By default, both channels output a sine waveform at 10kHz frequency and 5Vpp amplitude.
- Select Output Channel: Press the
CH1orCH2key to select the desired channel. The finger cursor will indicate the active channel. - Select Basic Waveforms: Press the
Wavesoft key. Use the knob or soft keys to select the desired waveform type. - Set Frequency: Press the
Freq.soft key. Use the numeric keyboard or direction keys and knob to adjust the frequency (range: 0-80MHz). PressFreq.again to select the unit (MHz, kHz, Hz, mHz, μHz). - Set Amplitude: Press the
Ampl.soft key. Use the numeric keyboard or direction keys and knob to adjust the amplitude. The amplitude range is frequency-dependent (refer to specifications). - Set Offset: Press the
Offsetsoft key. Use the direction keys and knob or numeric keyboard to adjust the DC offset. - Set Duty Cycle: Press the
Dutysoft key. Use the direction keys and knob or numeric keyboard to adjust the duty cycle (0.01-99.99% for pulse wave; not adjustable for sine wave). - Set Phase: On the CH2 interface, press the
Phasesoft key. Use arrow keys and knob/numeric keyboard to adjust phase difference (0°-359.99°). Ensure CH1 and CH2 frequency synchronization is set in system settings first. - Observe Output Waveforms: Connect CH1 and CH2 to an oscilloscope using BNC test leads to observe the waveforms.
5.2. Modulation Mode Interface and Parameter Settings
The PSG9060 supports various modulation types: AM, FM, PM, ASK, FSK, PSK, PWM, and BURST. Modulation processes a signal by adding it to a carrier wave for transmission.
Overview of modulation functions and their parameters.
General Modulation Steps:
- Press the
MODkey to enter the modulation mode interface. - Press the
Typesoft key to select the desired modulation type (AM, FM, PM, ASK, FSK, PSK, PWM, BURST). - Configure carrier waveform, modulation source (internal/external), modulation frequency/rate, depth/deviation, polarity, etc., as required for the specific modulation type.
For detailed parameter settings for each modulation type, refer to the user manual PDF section 3.3 (pages 25-33).
5.3. Measurement Mode Interface
The instrument can measure frequency, period, positive/negative pulse width, and duty cycle of input signals.
Measurement mode interface.
Measurement Function:
- Frequency Range: 1Hz-100MHz.
- Signal Amplitude Range: 2Vpp-20Vpp (input via Ext.IN).
- Coupling Setting: Use soft keys and knob to switch between AC or DC coupling.
- Gate Time Setting: Adjust gate time from 0.001s to 10s (default 1s).
- Measurement Mode: Switch between low frequency (<2kHz) or high frequency (>2kHz) modes.
Counter Function:
Calculates the number of input signal periods in real-time.
- Coupling Setting: Same as measurement function.
- After setting, press
ONto start counting,OFFto stop,RSTto restore defaults, andSAVEto save parameters.
5.4. Frequency Sweep Mode Interface
The PSG9060 supports frequency sweep, amplitude sweep, and duty cycle sweep functions.
Sweep frequency interface.
Frequency Sweep:
- Press
SWEEP, thenFUNCto select sweep type. - Sweep Channel: Switch between CH1 or CH2.
- Start/End Frequency: Set the upper and lower limits of the sweep.
- Sweep Time: Adjust scan time from 0.01s to 640s.
- Sweep Direction: Select increment, decrement, or round trip.
- Sweep Mode: Choose linear or logarithmic.
- Press
ONto start frequency sweep.
Amplitude Sweep & Duty Cycle Sweep:
Similar settings apply for amplitude and duty cycle sweeps, including channel selection, start/end values, sweep time, direction, and mode. Press ON to activate.
5.5. Voltage Control Mode Interface
The PSG9060 offers voltage control for frequency, amplitude, and duty cycle.
Voltage control frequency interface.
Voltage Control Frequency:
- Press
VCO, thenFUNCto select frequency control. - Voltage Control Channel: Switch between CH1 or CH2.
- Start/End Frequency: Set the range.
- Min/Max Voltage Calibration: Calibrate external input signal voltage.
- Voltage Control Mode: Choose linear or logarithmic.
- Press
ONto activate voltage control frequency.
Voltage Control Amplitude & Duty Cycle:
Similar settings apply for amplitude and duty cycle voltage control, including channel selection, start/end values, and calibration. Press ON to activate.
5.6. Programming Mode Interface
The PSG9060 allows for programmable waveform output, enabling automatic work sequences.
Programming mode interface.
- Press
FUNCto enter programming mode. PressMODEto switch between normal and debugging modes. - P00-P19: 20 storage locations for programs.
- Serial Number (00-99): Up to 100 instructions per program.
- Load: Indicates parameter storage location.
- Time: Current running time of the serial number.
- Go to: Transfers control to a specified serial number.
- Cycle: Sets the number of repetitions for a serial number.
- Press
SAVEto save program data.
5.7. System Settings
Access system settings by pressing the SYS key.
- Clear Storage: Clears parameters in the current position.
- Sound Setting: Toggles sound ON/OFF.
- Brightness Adjustment: Adjusts screen brightness.
- Language Setting: Selects English or Chinese.
- Built-in Wave Number: Sets the number of built-in waves (00-21).
- Arbitrary Wave Number: Sets the number of arbitrary waves (01-99).
- Waveform Loading Methods: Selects Auto or Fast.
- System Information: Displays product model, hardware/software/FPGA versions, official website, and QR code.
- Firmware Update: Press
Updatesoft key to enter update mode, then connect to PC software. - Synchronization: Waveform, frequency, amplitude, and duty cycle synchronization can be turned ON/OFF for CH1/CH2.
- External Synchronization: Toggles external synchronization ON/OFF.
- Frequency Fine-tuning: Adjusts frequency for waveform synchronization (range 0-99).
- Restore Factory Settings: Press
ON, thenYesto restore. - System Color: Selects one of four system colors.
- Auto Power-on: Toggles auto power-on ON/OFF.
- CH1/CH2 Amplitude Fine-tuning: Calibrates amplitude for accuracy.
5.8. PC Software Control Output
The PSG9060 can be controlled via PC software. The software has both Chinese and English interfaces.
PC and APP control options.
Software Setup:
- Download the PC software from the provided link: PSG9080_EN_Setup.zip.
- Run the installer. Select the desired language.
- Click
Install USB Driveto install the serial port driver. - Click
Main programto install the application. Follow the prompts to complete installation.
Connecting to PC:
- Right-click Computer → Properties → Device Manager to observe the serial port assigned by the computer.
- In the PC software, select the corresponding serial port and click connect.
New PC software interface with four function interfaces.
6. Specifications
| Parameter | Specification |
|---|---|
| Model | PSG9060 |
| Channel | 2 |
| Sine wave frequency range | 1nHz-60MHz |
| Square wave frequency range | 1nHz-30MHz |
| Triangle wave frequency range | 1nHz-50MHz |
| Pulse wave frequency range | 1nHz-30MHz |
| TTL digital wave frequency range | 1nHz-20MHz |
| Arbitrary wave frequency range | 1nHz-50MHz |
| Frequency resolution | 1nHz |
| Sampling Rate | 300MSa/S |
| Wave length | 8192 points/channel |
| Waveform vertical resolution | 14 bits |
| Basic waveforms | Sine, square, pulse, ramp, CMOS, DC, partial sine, noise wave |
| Built-in waveforms | Sine, square, pulse, triangle, ramp, CMOS, DC level, partial sine, half, full, positive step, inverse step, noise, exponential rise/fall, logarithm ascending/descending, sinker pulse, multi-tone, lorenz |
| Pulse width adjustment range | 5nS~4S |
| Pulse period adjustment range | 10nS~40S |
| Frequency accuracy | ±5ppm 0 to 50℃ |
| Frequency stability | ±3ppm per 1 year |
| Total harmonic distortion (Sine) | ≤0.5% (@1kHz, 5Vpp) |
| Square wave Rise/Fall Time | ≤10nS |
| Square wave Overshoot | ≤10% |
| Square wave Non-symmetry | ≤0.1% |
| Ramp/Triangle wave Non-linearity | ≤1% (Less than 1MHz) |
| Ramp/Triangle wave Non-symmetry | ≤1% (Less than 1MHz) |
| Pulse wave Duty cycle | 0.01%-99.99% |
| Pulse wave Leading/Trailing Edge Time | ≤10nS |
| Pulse wave Overshoot | ≤10% |
| Amplitude range (<1MHz) | 2mVpp~25Vpp |
| Amplitude range (1MHz-11MHz) | 2mVpp~10Vpp |
| Amplitude range (11MHz-60MHz) | 2mVpp~5Vpp |
| Amplitude range (60MHz-80MHz) | 2mVpp~3.6Vpp |
| Amplitude resolution | 1mVpp |
| Amplitude stability | ±1% ±1 mVpp (@ 1 kHz, >10 mVpp) |
| Amplitude flatness (<10MHz) | ±1% (0.1dB) |
| Amplitude flatness (10MHz-50MHz) | ±2% (0.2dB) |
| Amplitude flatness (50MHz-70MHz) | ±10% (0.9dB) |
| Amplitude flatness (70MHz-80MHz) | ±20% (1.9dB) |
| DC Offset Range | -9.99V-12.00V |
| DC Offset Resolution | 0.01V |
| Phase adjustment range | 0°-359.99° |
| Phase resolution | 0.01° |
| Output impedance | 50Ω (typical) |
| Protection | Short-circuit protection |
| Modulation type | AM, FM, PM, ASK, FSK, PSK, PWM, BURST |
| Trigger source (Burst) | Manual burst, CH2 burst, external burst (AC), external burst (DC) |
| Measurement function | Frequency, positive/negative pulse width, period, duty cycle |
| Counter range | 0-4294967295 |
| Sweep function | Sweep frequency, sweep amplitude, sweep duty |
| Sweep mode | Linear, logarithmic |
| VCO function | Frequency control, amplitude control, and duty cycle control |
| Program mode | Debug and routinely store 100 groups of programs, each with 100 steps |
| Product dimension | 247*243*102mm |
| Net weight | 1038g |
| Package dimension | 340*280*140mm |
| Gross weight | 1900g |
7. Troubleshooting
This section lists common issues and their solutions. If problems persist after following these steps, contact JUNTEK support.
7.1. Screen is dark (no display) after pressing the power key:
- Check if the power is correctly connected.
- Verify the power key is fully pressed.
- Restart the instrument after checking the above.
- If it still doesn't work, contact JUNTEK.
7.2. Screen is too dark to see contents clearly:
- Check if brightness and contrast settings are too low.
- Press the
SYSkey to enter system settings, navigate to brightness, and adjust with the knob.
7.3. Settings are correct but no waveform is generated:
- Check if the BNC cable is tightly connected to the corresponding channel output terminal (
CH1orCH2). - Inspect the BNC cable for internal damage.
- Check if the BNC cable is tightly connected to the test instrument.
- Verify if the backlight of
CH1orCH2is on. If not, press the corresponding key to illuminate it. - After checking the above, press and restart the instrument.
- If it still doesn't work, contact JUNTEK.
8. User Tips
No specific user tips were found in the provided reviews or Q&A content.
9. Warranty and Support
9.1. Warranty
JUNTEK warrants that the product mainframe and accessories will be free from defects in materials and workmanship within the warranty period. If a product proves defective within this period, JUNTEK guarantees free replacement or repair. There are no other warranties, expressed or implied, except as expressly set forth herein or other applicable warranty card. JUNTEK is not liable for any consequential, indirect, ensuing, or special damages for any breach of warranty.
9.2. Support and Further Information
For more product information, including machine model and factory number, press the SYS key on the instrument. You can also refer to relevant manuals by visiting the official JUNTEK website (www.junteks.com) to download them.
Additional resources available:
- "PSG9060 Operation Demo Video": Provides operational videos.
- "PSG9060 PC Software and Communication Protocol": Provides details on PC software and communication protocols.
- "PSG9060 User Manual": Includes technical specifications, functions, operation methods, possible failures, and solutions.
If you have any problems or requirements when using the product or this manual, please contact JUNTEK via email at junce@junteks.com.
A PDF version of the user manual is available for download: User manual (PDF).