1. General Safety Summary
Read the following safety precautions carefully to avoid injury and prevent damage to this product or any connected product. To avoid possible danger, be sure to use this product as specified.
1.1. Avoid Fire and Personal Injury
- Only professionally authorized personnel should perform repairs.
- Use the correct power cord specified for this product in your country.
- Before connecting the probe to the circuit under test, connect the probe to the oscilloscope. Before disconnecting the probe from the oscilloscope, disconnect the probe from the circuit under test.
- To avoid electric shock, the product is grounded through the grounding conductor of the power cord. The grounding conductor must be connected to ground. Before connecting the input or output of the product, ensure proper grounding.
- If using a probe, the probe ground must be connected to the ground. Do not connect the probe ground lead to high voltage, as this may cause dangerous voltage on the oscilloscope and probe connector, control equipment, or other surfaces.
- To avoid fire or excessive current, check all ratings and markings on the product. Consult the product manual for details on ratings before connecting.
- Ensure no overvoltages (e.g., from lightning) reach the product to prevent electric shock.
- Do not operate the product with the cover or panel open.
1.2. Maintain Proper Ventilation
Poor ventilation increases instrument temperature and can cause damage. Ensure good ventilation and regularly check vents and fans.1.3. Use a Suitable Fuse
Only use fuses specified for this product.
1.4. Avoid Exposed Circuits
Do not touch exposed connectors and components after power is turned on.
1.5. Do Not Operate if Malfunctioning
If you suspect the product is damaged, have it inspected by qualified service personnel.
1.6. Do Not Operate in Humid, Flammable, or Explosive Environments
Avoid operating the instrument in humid, flammable, or explosive environments to prevent short circuits, electric shock, or damage.
1.7. Keep Surface Clean and Dry
To prevent dust or moisture from affecting performance, keep the product surface clean and dry.
1.8. Anti-static Protection
Static electricity can damage the instrument. Test in an anti-static area if possible. Before connecting cables, briefly ground inner and outer conductors to discharge static electricity.
1.9. Attention to Handling Safety
During transportation, handle with care to avoid damage to buttons, knobs, or interfaces.
1.10. Safety Terms and Symbols
The following terms and symbols may appear on the product:
- DANGER: Indicates immediate damage to the user if action is taken.
- WARNING: Indicates potential harm to the user if action is taken.
- CAUTION: Indicates potential damage to the product or other property if action is taken.

Safety Symbols: Safety Warning, Protective Test, Ground Shell, Earth Terminal
1.11. Measurement Category
DPO6000 / MPO6000 series digital oscilloscopes perform measurements under Measurement Category I. This oscilloscope is only allowed to be used in the specified measurement category.
- Measurement Category I: Measurements on circuits not directly connected to the mains (e.g., internal protected circuits).
- Measurement Category II: Measurements on circuits directly connected to low-voltage equipment (e.g., home appliances, portable tools).
- Measurement Category III: Measurements in construction equipment (e.g., power distribution boards, circuit breakers, fixed equipment).
- Measurement Category IV: Measurements at the source of low-voltage equipment (e.g., electricity meters, major overcurrent protection devices).
2. Setup and Initial Inspection
2.1. General Inspection
2.1.1. Checking the Shipping Packaging
If the packaging carton or foam protective pad is seriously damaged, keep it until the entire machine and accessories pass electrical and mechanical tests.
2.1.2. Check the Whole Machine
If the instrument appears damaged, is not working properly, or fails performance tests, contact your dealer.
2.1.3. Check Random Accessories
If any accessory is missing or damaged, contact your dealer. Details of provided attachments are in Appendix A.

Accessories Table: Instrument, USB cable, Power Cord, Oscilloscope probe x4, Packing list.
2.2. Physical Dimensions
The oscilloscope has a length of 318 mm, a width of 140 mm, and a height of 150 mm.

Front and Top View of Oscilloscope with dimensions
2.3. Adjusting the Bracket
Properly adjust the stand to tilt the oscilloscope upward for better operation and display screen observation.
2.4. Connect the Power Cord
The AC power input specifications are: 100-120Vac, 50/60/400Hz; 100-240Vac, 50/60Hz; 50W max. Connect the oscilloscope to an AC power source using the provided power cord.

Rear view of the oscilloscope showing power jack
2.5. Startup Check
Press the power switch on the lower left corner of the front panel. The oscilloscope will perform self-tests, with keyboard indicators lighting up. After self-tests, the startup screen appears.
2.6. Probe Inspection and Connection
2.6.1. Safety
When using the probe, keep your fingers behind the safety ring. Do not touch metal parts when connected to high voltage. Connect the probe to the oscilloscope and ground terminal before measurements.

Oscilloscope probe with finger guard
2.6.2. Probe Attenuation Setting
Probes have different attenuation coefficients. Use the 'Vertical Menu' button (e.g., [CH1]) to select a probe option matching the probe attenuation coefficient. Ensure the 'Attenuation' switch on the probe matches the 'Probe' option in the oscilloscope (X1 or X10). X1 limits bandwidth to 6MHz; X10 provides full bandwidth.

Oscilloscope probe with attenuation switch
2.6.3. Connect the Passive Probe
- Connect the BNC end of the probe to the analog channel input on the front panel.
- Connect the probe ground alligator clip or ground spring to the circuit ground terminal, then connect the probe to the circuit under test.
2.6.4. Connect the Logic Probe
- Connect one end of the digital probe to the digital channel input interface on the front panel (no direction).
- Connect the other end of the digital probe to the signal under test. MPO6000 series comes standard with four LP104 digital probes.
2.6.5. Check the Probe Voltage
Perform a probe check each time you connect a probe to an input channel. Use the 'Vertical Menu' button (e.g., [CH1]) to set the probe attenuation coefficient.
2.6.6. Manual Probe Compensation
Adjust probe compensation when connecting for the first time to match the probe to the input channel. Uncompensated or overcompensated probes can cause measurement errors.
- Set probe option attenuation to 10X in the Channel Menu. Set the switch on the probe to 10X and connect to channel 1. Connect the probe tip to the ~2V @ 1K Hz connector and the reference wire to the 'probe element ground' connector. Press 'Auto Scale'.
- Check the shape of the displayed waveform.
- If necessary, use a non-metallic screwdriver to adjust the variable capacitance on the probe until the waveform is 'compensated correctly'.

Waveform compensation examples: Correctly compensated, Overcompensated, Undercompensated
3. Operation
3.1. Front Panel Overview
The front panel provides controls for various functions of the oscilloscope.

Front panel layout of the Hantek Oscilloscope
| SN | Description | SN | Description |
|---|---|---|---|
| 1 | LCD | 9 | Vertical control system |
| 2 | Menu show/hide keys | 10 | Signal source 2 output channel |
| 3 | Signal source button (signal source series) | 11 | CH1~CH4 signal input channel |
| 4 | Multifunction knob | 12 | Signal source 1 output channel |
| 5 | Menu function buttons | 13 | LA1~LA4 signal input channels (MPO6000 series) |
| 6 | Level control system | 14 | Probe compensation signal output/ground |
| 7 | Trigger control system | 15 | USB Host port |
| 8 | Shortcuts (Run/Stop, Single, Auto Scale) | 16 | Power On Key |
3.2. Rear Panel Overview
The rear panel provides various connection ports and controls.

Rear panel layout of the Hantek Oscilloscope
- Handle: Pull vertically to carry; press down when not needed.
- AC Power Jack: Input for AC power.
- AUX Port Trigger Output and Pass / Fail [Optional]:
- Trigger output: Outputs a pulse when the oscilloscope triggers, reflecting the current capture rate.
- Pass / Fail: Outputs a pulse when a failure is detected in pass/fail tests; continuously outputs low level if no failure.
- Signal Source 3 Output [MPO6000EDU Series]: For 3-channel signal source. Outputs signals according to current settings when turned on.
- USB Device: Connects the oscilloscope to a computer or printer.
- LAN / UART:
- LAN: Connects to network for remote control. Complies with LXI CORE 2011 DEVICE standard.
- UART [Optional]: Connects to a control system for remote control.
- HDMI Interface [Optional]: Connects to an HDMI display for a larger view.
3.3. Front Panel Functions
3.3.1. Level Control System
- Horizontal Offset: Modify horizontal displacement. Knob adjusts trigger point left/right. Press knob to reset to zero.
- Horizontal Time Base: Modify horizontal time base. Clockwise decreases, counter-clockwise increases. Press knob to switch between single and double window display modes.
3.3.2. Vertical Control System
- Analog Input Channel Switch: Press any button (1-4) to open/close the corresponding channel menu.
- Math: Press 'Math' button to open/close Math setup menu.
- Vertical Offset: Modify vertical shift of current channel waveform. Knob adjusts waveform up/down. Press knob to zero vertical displacement.
- Vertical Voltage: Modify vertical scale of current channel. Knob adjusts gears. Press knob to switch between 'Coarse' or 'Fine' adjustment.
3.3.3. Signal Source
- WaveGen: Press this key to enter the signal source setting interface.
Note: This function is only applicable to digital oscilloscopes with signal source channels.
3.3.4. Trigger Control
- Trigger Level: Modify trigger level. Knob adjusts level line up/down. Press knob to restore to zero.
- TrigMenu: Press to open trigger operation menu. Provides various trigger types.
- ForceTrig: Short press forces a trigger signal. Press and hold to open history waveform.
3.3.5. Automatic Waveform Display
- AutoScale: Press to enable automatic waveform setting. Adjusts vertical scale, horizontal time base, and trigger mode for best waveform display.
3.3.6. Operational Control
- Run/Stop: Press to 'run' (green backlight) or 'stop' (red backlight) waveform sampling.
- Single: Press to set trigger mode to 'Single'. Forces a trigger signal immediately when 'Force Trig' is pressed.
3.3.7. Multifunction Knob (V0)
Performs various settings: waveform brightness, trigger time, trigger type/source, signal source waveform, frequency, offset, amplitude, etc. Turn clockwise to increase, counter-clockwise to decrease. Press to select option.
3.3.8. Function Menu
- Default: Briefly press to execute default setting. Long press to enter Default function preset (factory, default, user settings).
- Help: Press to open/close built-in help system.
- Utility: Press to enter system function setting menu (interface, sound, language, advanced features like pass/fail tests).
- Cursor: Press to enter cursor measurement menu (manual, tracking modes).
- Measure: Press to enter measurement setting menu (data source, DVM on/off, all measurements, statistics). 'All Measurements' measures 42 waveform parameters.
- Save/Recall: Short press to enter file storage/recall interface (setup, waveform, reference, CSV). Supports internal/external storage. Long press saves picture in .bmp format to external storage.
- Display: Short press to enter display setting menu (waveform display type, afterglow time, waveform brightness, screen grid, grid brightness). Press and hold to clear afterglow, then collect/count again.
3.4. User Interface
The user interface displays various information about the oscilloscope's status and measurements.

Oscilloscope User Interface with labeled sections
- Hantek Trademark: Touch 'Touch Screen' in 'Utility', then Hantek logo to scan QR code for wizard interface.
- Running Status: AUTO (automatic mode, no triggers), READY (pre-trigger data acquired, ready for trigger), TD (triggered), ROLL (scroll mode, continuous acquisition), STOP (acquisition stopped), ARM (pre-trigger state).
- Main Time Base of Current Window: Length of time per division on horizontal axis.
- Current Sampling Rate, Sampling Points: Displays current sampling rate and memory depth. Changes with horizontal time base.
- Waveform Memory: Visual representation of waveform in memory.
- Horizontal Trigger Time: Adjust with horizontal offset knob. Press knob to reset.
- Trigger Level: Adjust with trigger level knob. Press knob to reset.
- Operation Menu Displays: Information for each function key.
- Time Display of Oscilloscope.
- WIFI Status Display.
- LAN Connection Status Indication: Lit when LAN is connected.
- USB External Storage Device Status Display: Lit when USB external storage device is connected.
- USB Host Computer Connection Status Display: Lit when USB host computer is connected.
- Logic Analyzer Status Display: Lit when LA channel is connected.
- Gen3 Status Display [MPO6000EDU series]: Lit when Gen3 channel is turned on.
- Gen2 Status Display [Signal Source Series]: Lit when Gen2 channel is turned on.
- Gen1 Status Display [Signal Source Series]: Lit when Gen1 channel is turned on.
- CH4 Status Display [4-channel series]: Lit when CH4 channel is turned on.
- CH3 Status Display [4-channel series]: Lit when CH3 channel is turned on.
- CH2 Status Display: Lit when CH2 channel is turned on.
- CH1 Status Display: Lit when CH1 channel is turned on.
- Channel Zero-Scale Position Display.
- Channel Trigger Level Position Display.
3.5. Sampling System
Understanding the sampling principle and the relationship between sampling rate and memory depth is crucial for oscilloscope waveform acquisition.
3.5.1. Sampling Principle
According to Nyquist, sampling frequency must be greater than twice the highest frequency (Fs > 2Fmax) to restore analog signals without distortion. Higher sampling rates yield more accurate waveforms but require faster conversion circuits.
3.5.2. Waveform Aliasing
If sampling conditions (Fs > 2Fmax) are not met, the reconstructed waveform frequency will be smaller than the actual signal. This can be avoided by increasing sampling frequency or using a low-pass (anti-aliasing) filter.
3.5.3. Waveform Distortion
Low sampling rates can lead to missing waveform details, causing the displayed waveform to differ from the actual signal.
3.5.4. Waveform Miss
When the sampling rate is too low, the reconstructed waveform may not reflect all actual signals.
3.5.5. Sampling Rate and Memory Depth
The highest sampling rate is 1GSa/s. The actual sampling rate is determined by the horizontal time base scale and storage depth. Memory depth is the number of waveform points stored in one triggered acquisition (max 128Mpts).
Relationship: Sampling time = memory depth / (sampling rate (Sa/s))
3.5.6. Waveform Interpolation
Digital oscilloscopes use interpolation to increase signal visibility by connecting discrete samples. Methods include sine, linear, and step interpolation.
- Linear Interpolation: Connects adjacent points with straight lines. Suitable for straight-edge signals.
- Step Interpolation: Inserts the value of the previous sampling point between adjacent samples.
- Sine Interpolation: Uses mathematical processing to create a more realistic waveform shape. Recommended when sampling rate is 3-5 times system bandwidth.

Various waveform displays (YT single/double window, unaveraged/averaged, XY single/double window)
3.6. Operational Control
Use 'Run/Stop' and 'Single' buttons to control sampling.
- Run/Stop: Green for running, red for stopped.
- Single: Captures and displays a single acquisition.
3.6.1. Collection Method
Access via 'Utility -> Acquisition -> Acquisition Mode'.
- Normal: Samples at equal time intervals, best for most waveforms.
- Peak: Collects max/min values to capture signal envelope or narrow pulses. Avoids aliasing but increases noise.
- Average: Averages multiple samples to reduce random noise and improve vertical resolution.
- High Resolution: Uses oversampling to average neighboring points, reducing noise and smoothing waveforms. Improves measurement accuracy by sacrificing bandwidth.
3.7. Horizontal Control System
3.7.1. Time Base Mode
Access via 'Utility -> Acquisition -> Display Mode'. Includes YT, XY, and Roll modes.
- YT Mode: Y-axis is voltage, X-axis is time. Enters scan mode if horizontal time base is 100ms/div or slower.
- XY Mode: Converts two input channels from voltage-time to voltage-voltage display. X-axis/Y-axis combinations can be CH1-CH2 or CH3-CH4. Useful for Lissajous figures to measure phase difference.
- Roll Mode: Waveform scrolls from right to left. Horizontal gear range 100ms to 100s.
3.8. Vertical Control System
3.8.1. Enable Analog Channel
DPO6000/MPO6000 series have 4 analog input channels (CH1-CH4). Press the corresponding button on the front panel to turn a channel on/off.
3.8.2. Channel Coupling
Filters out unwanted signals. Options: DC (passes both DC and AC), AC (blocks DC), GND (blocks both DC and AC). Select via '1 -> Coupling' or F1 [V0] key.
3.8.3. Bandwidth Limitation
Reduces noise by limiting bandwidth. Default is off. Set to 20 MHz to attenuate high-frequency components. Switch via '1 key' then F2 key.
3.8.4. Waveform Inversion
Press '1 -> Invert' to turn waveform inversion on/off. Inverted mode flips the waveform voltage value.
3.8.5. Probe Ratio
Manually set probe attenuation ratio via '1 -> Probe'.
3.8.6. Vertical Scale
Adjusts voltage value per division (V/div). Rotate SCALE knob. 'Coarse' (1-2-5 steps) or 'Fine' adjustment modes available.
3.9. Analog Channel Delay Setting
Corrects zero offset caused by probe cable propagation delay. Adjust with multifunction knob V0 (range -100ns to 100ns). Press knob to reset to 0.00s.
3.10. Trigger System
Triggers define when the oscilloscope starts acquiring and displaying data.
3.10.1. Trigger Source
Select via 'Trig Menu'. Analog channels CH1-CH4, digital channels D1.0-D4.3 can be trigger sources. An invalid trigger occurs if the selected channel is not on.
3.10.2. Acquisition Process
The acquisition memory is divided into pre-trigger and post-trigger buffers. Data is continuously stored in the pre-trigger buffer until a trigger condition is met, then post-trigger data is stored.
3.10.3. Trigger Mode
- Automatic Mode: Allows free running acquisitions without a valid trigger.
- Normal Mode: Updates waveform only with a valid trigger.
- Single Trigger: Stops after a single valid trigger.
3.10.4. Trigger Holdoff
Time difference between detecting one trigger and preparing for another. Used for stable display of complex waveforms. Access via 'Trig Menu -> Hysteresis'.
3.10.5. Trigger Types
DPO6000/MPO6000 series have up to 16 trigger functions. Access via 'Trig Menu'.
- Edge Trigger: Triggers on a specified edge (rising, falling, or both) and voltage level.
- Pulse Width Trigger: Triggers on positive or negative pulses of a specified width.
- Video Trigger: Captures complex waveforms of standard analog and HD video signals (NTSC, PAL, HDTV).
- Slope Trigger: Triggers on positive or negative slope from one level to another within a specified time.
- Overtime Trigger: Triggers when the time interval from a rising/falling edge passing the trigger level to the adjacent falling/rising edge passing the trigger level exceeds a set overtime period.
- Window Trigger: Provides high and low trigger levels. Triggers when input signal passes high or low level.
- Pattern Trigger: Identifies trigger conditions by looking for specific patterns (high, low, or inactive) across channels.
- Interval Trigger: Triggers when the interval between two consecutive rising/falling edges meets a set time condition.
- Under Amp Trigger: Triggers on pulses that cross one trigger level but not another (positive or negative runt pulses).
- Delay Trigger: Triggers when the time difference between edges from two data sources meets a preset time limit.
- Setup/Hold Trigger: Triggers when setup time (data stable before clock edge) or hold time (data stable after clock edge) is less than preset.
- UART Trigger [Optional]: Triggers on frame start, stop bit, data, parity error, or data bit error of UART signal.
- LIN Trigger [Optional]: Triggers on LIN bus signal message frames (interval field, sync field, ID field, identifier, data).
- CAN Trigger [Optional]: Triggers on CAN bus frames (start bit, remote frame ID, data frame ID, error frame, overload frame).
- SPI Trigger [Optional]: Triggers when specified data is found in SPI communication, based on SCL clock and SDA data sources.
- IIC Trigger [Optional]: Triggers on IIC bus signals (start condition, stop condition, no response, address, restart, address and data).
3.11. Mathematics
The oscilloscope supports various mathematical operations on waveforms.
- Addition: Adds waveform values of two data sources point by point.
- Subtraction: Subtracts waveform values of two data sources point by point.
- Multiplication: Multiplies waveform values of two data sources point by point.
- Division: Divides waveform values of two data sources point by point.
- FFT (Fast Fourier Transform): Converts time domain signal to frequency components (spectrum). Used to measure harmonic content, noise characteristics, vibration analysis.
- Integral: Calculates integral of a single data source.
- Differential: Calculates discrete-time derivative of a single data source.
- Square Root: Calculates square root of a single data source waveform point by point.
- Expression: User-defined operation expressions using analog channel waveforms.
3.12. Measurement
The oscilloscope provides automatic measurement of 42 waveform parameters and statistics.
3.12.1. Measurement Parameter Description
Includes time parameters (Period, Frequency, Rise Time, Fall Time, Pulse Width, Duty Cycle, Burst Length, Max/Min Time), count value parameters (Positive/Negative Pulse Count, Rise/Fall Pulse Count, Trigger Count), delay and phase parameters (FRR, FFF, FRF, FFR, LRR, LRF, LFR, LFF, Phase), and voltage parameters (Maximum, Minimum, Peak-Peak, Top, Bottom, Amplitude, Middle, Average, Period Average, RMS, Overshoot, Preshoot, Fall Overshoot, Rise Pre-shoot, Variance).
3.12.2. All Measurements
Press 'Meas' on the front panel, then 'All Measurements' to measure all 42 parameters with one key. Results displayed at bottom of screen with color markers.
3.12.3. Measurement Statistics
Displays current, average, maximum, minimum, root mean square error, and count values for selected waveform parameters.
3.12.4. Digital Voltmeter (DVM)
Built-in 5-digit DVM and 6-digit frequency meter. Measures AC RMS, DC, AC+DC RMS voltage and frequency. Access via 'Meas -> Digital Voltmeter -> Enable'.
3.12.5. Grid Scale Measurement
Estimates frequency and voltage amplitude using grid divisions and scale factor.
3.12.6. Automatic Measurement
After connecting and inputting a signal, press 'Auto Scale'. Options: Single cycle, Multi-cycle, Auto range, Data source.
3.12.7. Cursor Measurement
Uses amplitude (horizontal dashed line) and time (vertical dashed line) cursors. Modes: Manual (adjust cursors manually) and Tracking (cross cursor automatically positioned on waveform). Access via 'Cursors' button.
3.13. Bode Plots (Signal Source Series)
Semi-logarithmic plot of transfer function versus frequency. Displays frequency response of a system. Access via 'Utility -> Bode Plots'.
- Enable Bode Plots: 'Utility -> Bode Plots -> Function [On, Off]'.
- Data Source Settings: Input source (CH1-CH4), Output source (CH1-CH4).
- Chart Settings: Gain gear, Gain Offset, Phase scale, Phase Offset.
- Operation: 'Utility -> Bode Plots -> Run'.
3.14. Digital Channel
MPO6000 series has 4 analog and 16 digital channels. DPO6000 series can be upgraded with LP104 digital probe. Compares voltage to logic threshold (1 if greater, 0 otherwise).
- Select Digital Channel: 'Utility -> Logic Analyzer -> Data Source [D1, D2, D3, D4]'.
- Enable Digital Channels: Connect digital probe, logic analyzer enable switch turns on automatically.
- Set Digital Channel Thresholds Type: 'Utility -> Logic Analyzer -> Type'. Options include 5.0V CMOS, 3.3V CMOS, 2.5V CMOS, 1.8V CMOS, ECL, PECL, LVDS, 0V, User-defined.
3.15. Waveform Generator
DPO6000C/MPO6000D have two built-in channels (MPO6000EDU has three) and a 25 MHz signal source. Can output sine, square, pulse, triangle, noise, DC, sampling, exponent, semi-distortion, Lorentz, dual-tone multi-frequency, Gauss, ECG signals (13 basic types) and 4 arbitrary waveforms.
- Output Basic Waveform: Select waveform, data source, channel, frequency, amplitude, offset voltage, phase, load impedance.
- Modulation: Supports AM, FM, PM, PWM.
- Burst: Outputs waveforms with a specified number of cycles.
- Edit Arbitrary Waveform: Use 'Wave Editor Setup.exe' software.
3.16. Pass / Fail Test
Evaluates signal quality by testing waveforms in real-time against set rules.
- Enable Pass / Fail Test: 'Utility -> Pass / Fail -> Enable test -> On'. Failure area is shaded.
- Data Source: Select channel (CH1-CH4).
- Rule: Customize rules for pass/fail tests using 'Horizontal' and 'Vertical' soft keys.
- Test and Output: Set output format of test results. Options: Message (display results), Stop On Output (On/Off), Output to DDS (signal source output).
- Segmented Acquisition: Stores data from multiple triggers in segments.
- History Waveform: Stores up to 80,000 frames. Allows observation and analysis of previously stored waveforms.
- Display Control: Sets waveform display type (Vector/Dot), persist time, waveform brightness, grid type, grid brightness, color grade.
4. Maintenance
4.1. Daily Maintenance and Cleaning
- When storing or placing the oscilloscope, do not expose the LCD monitor to direct sunlight for a long time.
- Check the oscilloscope and probe frequently.
- Use a lint-free cloth to remove dust. Be careful not to scratch the display filter.
- Clean with a soft cloth dampened with water. For thorough cleaning, use 75% isopropyl alcohol in water.
CAUTION: Do not use corrosive or chemical cleaning agents.
4.2. Calibration
Perform calibration to ensure accurate measurements. Execute at any time, especially if ambient temperature changes by more than 5°C. Warm up for over 30 minutes, disconnect all input channels, then press 'Utility -> Calibration'.
5. Troubleshooting
5.1. Common Faults and Solutions
- The oscilloscope does not turn on after pressing the power switch:
- Check if the power cord is connected normally.
- Check if the power switch is pressed.
- After checking, try turning on the instrument again. If it still doesn't turn on, contact Hantek Technical Support.
- No waveform is displayed on the input signal screen after the oscilloscope is turned on:
- Check if the oscilloscope probe is correctly connected to the BNC connector of the signal input channel.
- Check if the channel switch is on (CH1~CH4 menu button).
- Check if the input signal is correctly connected to the probe.
- Check if there is signal output from the measured circuit.
- If it is a DC signal with large amplitude, increase the amplitude.
- Try pressing the automatic measurement button to detect the signal. If still no waveform, contact Hantek Technical Support.
- The input signal waveform distortion is serious:
- Check if the oscilloscope probe is connected to the channel BNC connector.
- Check if the probe is well connected to the measured object.
- Check if the oscilloscope probe is calibrated properly. If not, calibrate according to the manual.
- The waveform continuously scrolls on the screen and cannot be triggered:
- Check if the trigger data source is consistent with the signal input channel.
- Check if the trigger level is adjusted correctly. Press the trigger level knob to return to the signal center point.
- Check if the trigger mode is correct. Select the appropriate trigger mode for different input signals.
6. Technical Specifications
All technical specifications apply to the DPO6000 / MPO6000 series digital storage oscilloscopes. To verify specifications, the oscilloscope must have been working continuously for more than 20 minutes within the specified operating temperature, and a self-calibration must be performed if the temperature changes by more than 5°C.
6.1. Acquire
| Parameter | Specification |
|---|---|
| Acquisition method | Normal, average, peak, high resolution |
| Real-time sampling rate | 1 GSa/s (single channel), 500 MSa/s (dual channel), 250 MSa/s (three/four channels) |
| Peak Detect | Analog Channel 4ns |
| Averaging mode | N samples for analog channels (N values: 2, 4, 8, 16, 32, 64, 128, 256, 512, and 1024) |
| High resolution | 12bit |
| Minimum detection pulse width | 8ns |
| Memory Depth | Single Channel 64M / 128M [EDU], Dual channel 32M / 64M [EDU], Three or four channels 16M / 32M [EDU] |
6.2. Input
| Parameter | Specification |
|---|---|
| Number of channels | 4 analog channels, 16 (4 * 4) digital channels (total opened channels < 4) |
| Input coupling | DC, AC or GND (Digital channels not supported) |
| Input resistance | Analog channel: 1MΩ ± 2%, 25pF ± 3pF; Digital channels: 300KΩ ± 2%, 8 pF ± 3 pF |
| Voltage rating | 300V CAT II |
| Input voltage | Analog channels 300VRMS (10X); Digital channel -25V~25V |
6.3. Horizontal
| Parameter | Specification |
|---|---|
| Waveform Interpolation | Sine, Linear, Step |
| Horizontal scale | 2ns/div~100s/div 1,2,5 steps |
| Time base mode | Y-T, X-Y, Roll |
| X-Y number | 1, 2 channel combination 1 XY; 3, 4 channels combined 1 XY |
| Zero offset | ± 0.5 div x minimum time base scale |
| Sampling rate and delay time accuracy | ±25ppm |
| Clock drift | ≤±5 ppm/year |
| Incremental time measurement accuracy (Full bandwidth) | single shot, "sampling" mode: ± (1 sampling interval + 100ppm x reading + 0.6ns); > 16 times average: ± (1 sampling interval + 100ppm × reading + 0.4ns) |
6.4. Vertical
| Parameter | DPO6084 | DPO6104 | DPO6204 |
|---|---|---|---|
| Bandwidth (-3db) | 80MHz | 100MHz | 200MHz |
| Vertical resolution | Analog channel 8bit, Digital channel 1bit | ||
| Vertical scale | 500µV/div~10V/div at input BNC | ||
| Position range | 500µV/div to 128mV/div, ±1V; 129mV/div to 1.28V/div, ±10V; 1.29V/div to 10V/div, ±50V | ||
| Selectable analog bandwidth limit, typically | 20MHz | ||
| Low frequency response (-3db) | ≤10Hz at BNC | ||
| Rise time at BNC, typical | ≤4.4ns | ≤3.5ns | ≤1.8ns |
| Vertical gain accuracy | In "normal" or "average" acquisition mode, 10V/div to 10mV/div is ± 3%; 5mV/div to 500uV/div is ± 4%. | ||
| DC offset accuracy | ± 0.1div ± 2 mV ± 1% offset value | ||
| Channel Isolation | DC to Maximum Bandwidth: > 40 dB | ||
6.4.1. Vertical (Digital Channel: for MPO6000D / MPO600EDU or DPO6000B / DPO6000C purchase digital probe LP104)
| Parameter | Specification |
|---|---|
| Threshold | Adjustable threshold for 4 channels as a group |
| Threshold selection | TTL (1.4 V), 5.0 V CMOS (+2.5 V), 3.3 V CMOS (+1.65 V), 2.5 V CMOS (+1.25 V), 1.8 V CMOS (+0.9 V), ECL (-1.3 V), PECL (+3.7 V), LVDS (+1.2 V), 0V, Custom |
| Threshold range | ±7.0V, 10Mv steps |
| Threshold accuracy | ± (Threshold setting of 100mV + 3%) |
| Dynamic range | ± 5.0V + threshold |
| Minimum voltage swing | 500 mVpp |
| Vertical resolution | 1 bit |
6.5. Trigger
| Parameter | Specification |
|---|---|
| Trigger level range | ± 4 divisions from the center of the screen |
| Trigger mode | Auto, Normal, Single |
| Holdoff range | 8ns~10s |
| Trigger level accuracy | CH1~CH4 0.2 divisions x volts / division within ± 4 divisions from the center display |
| Edge trigger Slope | Rising, Falling, Rising or Falling |
| Edge trigger Data source | CH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3 |
| Pulse width trigger Polarity | Positive, Negative |
| Pulse width trigger Conditions | <, >, !=, = |
| Pulse width trigger Data source | CH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3 |
| Pulse width range | 8ns~10s |
| Video trigger Signal standard | NTSC, PAL, HDTV720P, HDTV1080P, HDTV1080i |
| Video trigger data source | CH1~CH4 |
| Video trigger Synchronize | Scan line, line number, odd field, even field, all fields |
| Video trigger Scope | Rising, falling |
| Video trigger Condition | <, >, !=, = |
| Video trigger Data source | CH1~CH4 |
| Video trigger Time range | 8ns~10s |
| Overtime trigger Data source | CH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3 |
| Overtime trigger Polarity | Positive, Negative |
| Overtime trigger Time range | 8ns~10s |
| Window trigger Data source | CH1~CH4 |
| Pattern trigger Pattern | 0: low level; 1: high level; X: ignore; |
| Pattern trigger Data source | CH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3 |
| Interval trigger Slope | Rising, falling |
| Interval trigger Condition | <, >, !=, = |
| Interval trigger Data source | CH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3 |
| Interval trigger Time range | 8ns~10s |
| Delay trigger Edge type | Rising, falling |
| Delay trigger Data source | CH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3 |
| Delay trigger Condition | <, >, !=, = |
| Delay trigger Time range | 8ns~10s |
| Setup/hold trigger Edge type | Rising, falling |
| Setup/hold trigger Data source | CH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3 |
| Setup/hold trigger Time range | 8ns~10s |
| Under Amp trigger Polarity | Positive, Negative |
| Under Amp trigger Condition | <, >, !=, = |
| Under Amp trigger Data source | CH1~CH4 |
| Under Amp trigger Time range | 8ns~10s |
| UART trigger Condition | Start, stop, data, parity, receive error |
| UART trigger Data source (RX/TX) | CH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3 |
| UART trigger Data format | Hex |
| UART trigger Data length | 1 byte |
| UART trigger Data bits | 5 bit, 6 bit, 7 bit, 8 bit |
| UART trigger Parity | None, Odd, Even |
| UART trigger Idle level | high, low |
| UART trigger Baud rate (optional) | 110/300/600/1200/2400/4800/9600/14400/19200/38400/57600/115200/230400/380400/460400 bit/s |
| UART trigger Baud rate (custom) | 300bit/s~334000bit/s |
| LIN trigger Condition | Interval field, sync field, ID field, sync error, identifier, ID, and data |
| LIN trigger Data source | CH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3 |
| LIN trigger Data format | Hex |
| LIN trigger Baud rate (optional) | 110/300/600/1200/2400/4800/9600/14400/19200/38400/57600/115200/230400/380400/460400 bit/s |
| LIN trigger Baud rate (custom) | 300bit/s~334000bit/s |
| CAN trigger Condition | Start bit, remote frame ID, data frame ID, frame ID, data frame data, error frame, all errors, response error, overload frame |
| CAN trigger Data source | CH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3 |
| CAN trigger Data format | Hex |
| CAN trigger Baud rate (optional) | 10000, 20000, 33300, 500000, 62500, 83300, 100000, 125000, 250000, 500000, 800000, 1000000 |
| CAN trigger Baud rate (custom) | 5kbit/s~1Mbit/s |
| SPI trigger Data source | CH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3 |
| SPI trigger Data format | Hex |
| SPI trigger Data bits | 4 bit, 8 bit, 16 bit, 24 bit |
| IIC trigger Data source (SDA/SCL) | CH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3 |
| IIC trigger Data format | Hex |
| IIC trigger Data index | 0~7 |
| IIC trigger Condition | Start bit, stop bit, no response, address, address and data, restart |
6.6. Display
| Parameter | Specification |
|---|---|
| Display Type | 7 "TFT LCD Diagonal |
| Display resolution | 800 (horizontal) * 480 (vertical) pixels |
| Display color | 16 million colors (24 bit true color) |
| Persist time | Min, 1 s, 5 s, 10 s, 30s, infinite |
| Display type | Point, vector |
| Display mode | Color grade |
| Display brightness | Adjustable |
| Grid Type | Optional |
| Grid brightness | Adjustable |
| Waveform brightness | Adjustable |
6.7. Arbitrary Signal Generator (with Signal Source Series)
| Parameter | Specification |
|---|---|
| Number of channels | 2/3 channels |
| Sampling rate | 200MSa/s |
| Vertical resolution | 12 bits |
| Frequency | 25 MHz |
| Standard waveform | sine wave, square wave, pulse, triangle wave, noise, DC, Sinc, Exp, Haver Sine, Lorentz, dual tone multi-frequency, Gauss, ECG signal |
| Arbitrary waveform | Arb1, Arb2, Arb3, Arb4 |
| Frequency characteristics (Sine wave) | 1uHz~25MHz |
| Frequency characteristics (Square wave) | 1uHz~15MHz |
| Frequency characteristics (Pulse) | 1uHz~5MHz |
| Frequency characteristics (Triangle wave) | 1uHz~1MHz |
| Frequency characteristics (Noise, Sinc, Exp, Haver Sine, Lorentz, dual tone multi-frequency, Gauss, ECG) | 1uHz~5MHz |
| Arbitrary wave Length | 8KSa |
| Arbitrary wave Range | 1uHz~75MHz |
| Arbitrary wave Frequency Accuracy | (less than 10 kHz) 50 ppm (more than 10 kHz) |
| Arbitrary wave Resolution | 0.1 Hz or 4 bits, whichever is greater |
| Output range | 10mV~7Vpp (high impedance), 5mV~3.5Vpp(50Ω) |
| DC offset Range | ±7 V, high impedance |
| Probe compensator output Voltage, typical | 2Vpp input ≥1MΩ load |
| Probe compensator output Frequency, typical | 1kHz |
6.8. General Technical Specifications
| Parameter | Specification |
|---|---|
| Power supply voltage | 100-120VACRMS (± 10%), 45Hz to 440Hz, CAT II; 120-240VACRMS (± 10%), 45Hz to 66Hz, CAT II |
| Power consumption | less than 30W |
| Fuse | T, 3.15A, 250V, 5x20mm |
| Operating temperature | 0~50 °C (32~122 °F) |
| Storage temperature | -40~+71 °C (-40~159.8 °F) |
| Humidity | ≤ + 40 °C: ≤90% relative humidity; +41 °C ~ 50 °C: ≤60% relative humidity |
| Cooling method | Convection |
| Altitude | with and without operation: 3000m (10,000 feet) |
| Random vibration | 0.31 g RMS from 50Hz to 500Hz, 10 minutes per axis |
| Mechanical shock (when not operating) | 2.46g RMS from 5Hz to 500Hz, 10 minutes per axis |
| Mechanical shock (during operation) | 50g, 11ms, half sine wave |
| Oscilloscope size | 318 x 140 x 150mm (length x width x height) |
| Weight | 2900g |
7. Warranty and Support
7.1. Copyright Statement
Qingdao Hantek Electronic Co., Ltd. reserves the right to amend this document without prior notice. The information provided is believed to be correct and reliable, but infallibility is not guaranteed. Ensure the latest and valid versions of technical documents are used. Responsibility for obtaining third-party consent/authorization for their products/patents lies with the user's company.
7.2. Technical Support
For questions or ambiguities, contact support through the following ways:
- Contact your local dealer.
- Contact the local office directly under Qingdao Hantek Electronic Co., Ltd.
- Contact the headquarters of Qingdao Hantek Electronic Co., Ltd.
Contact Method:
Qingdao Hantek Electronic Co., Ltd.
Website: http://www.hantek.com/en/index.html
Address: 35# Building No.780 Baoyuan Road, High-tech District, Qingdao China
Zip Code: 266114
Telephone: 0532-88705792
Fax: 0532-88705691
Email: service@Hantek.com
Technical Support:
Telephone: 0532-88703687
Email: support@Hantek.com
7.3. Equipment Recycling
This equipment uses natural resources. Improper disposal may harm the environment or human health. Recycle appropriately to ensure most materials are reused.
7.4. Hazardous Substances or Elements
The product contains certain hazardous substances or elements as per SJ / T 11363-2006 standard. 'X' indicates content exceeds limit, '0' indicates content is below limit.
| Part Name | Pb | Hg | Cd | Cr(Vi) | PBB | PBDE |
|---|---|---|---|---|---|---|
| Product shell, chassis | X | 0 | 0 | X | 0 | 0 |
| Display module | X | X | 0 | 0 | 0 | 0 |
| Circuit board components | X | 0 | 0 | X | 0 | 0 |
| Power components | X | 0 | 0 | X | 0 | 0 |
| Wire and cable assembly | X | 0 | 0 | 0 | 0 | 0 |
| Connector | X | 0 | 0 | X | 0 | 0 |
| Fasteners and mounting hardware | X | 0 | X | X | 0 | 0 |
| Other attachments (Including probe) | X | 0 | 0 | X | 0 | 0 |
| Other | 0 | 0 | 0 | 0 | 0 | 0 |
8. User Tips
- For optimal performance, allow the oscilloscope to warm up for at least 30 minutes before use.
- If you experience waveform distortion, ensure your probes are properly calibrated. Refer to the 'Manual Probe Compensation' section for instructions.
- When troubleshooting, always start by checking basic connections and power before delving into complex settings.
- Utilize the 'Auto Scale' function for a quick and optimized display of your input signal.
- If you need to analyze specific signal characteristics, explore the various trigger types available in the 'Trig Menu' to isolate the desired events.