Hantek DPO6000/MPO6000 Series

Contents

Hantek DPO6000/MPO6000 Series Digital Fluorescent Oscilloscope User Manual

Models: DPO6104B, DPO6204B, DPO6104C, DPO6204C, MPO6104D, MPO6204D

1. General Safety Summary

Read the following safety precautions carefully to avoid injury and prevent damage to this product or any connected product. To avoid possible danger, be sure to use this product as specified.

1.1. Avoid Fire and Personal Injury

  • Only professionally authorized personnel should perform repairs.
  • Use the correct power cord specified for this product in your country.
  • Before connecting the probe to the circuit under test, connect the probe to the oscilloscope. Before disconnecting the probe from the oscilloscope, disconnect the probe from the circuit under test.
  • To avoid electric shock, the product is grounded through the grounding conductor of the power cord. The grounding conductor must be connected to ground. Before connecting the input or output of the product, ensure proper grounding.
  • If using a probe, the probe ground must be connected to the ground. Do not connect the probe ground lead to high voltage, as this may cause dangerous voltage on the oscilloscope and probe connector, control equipment, or other surfaces.
  • To avoid fire or excessive current, check all ratings and markings on the product. Consult the product manual for details on ratings before connecting.
  • Ensure no overvoltages (e.g., from lightning) reach the product to prevent electric shock.
  • Do not operate the product with the cover or panel open.

1.2. Maintain Proper Ventilation

Poor ventilation increases instrument temperature and can cause damage. Ensure good ventilation and regularly check vents and fans.1.3. Use a Suitable Fuse

Only use fuses specified for this product.

1.4. Avoid Exposed Circuits

Do not touch exposed connectors and components after power is turned on.

1.5. Do Not Operate if Malfunctioning

If you suspect the product is damaged, have it inspected by qualified service personnel.

1.6. Do Not Operate in Humid, Flammable, or Explosive Environments

Avoid operating the instrument in humid, flammable, or explosive environments to prevent short circuits, electric shock, or damage.

1.7. Keep Surface Clean and Dry

To prevent dust or moisture from affecting performance, keep the product surface clean and dry.

1.8. Anti-static Protection

Static electricity can damage the instrument. Test in an anti-static area if possible. Before connecting cables, briefly ground inner and outer conductors to discharge static electricity.

1.9. Attention to Handling Safety

During transportation, handle with care to avoid damage to buttons, knobs, or interfaces.

1.10. Safety Terms and Symbols

The following terms and symbols may appear on the product:

  • DANGER: Indicates immediate damage to the user if action is taken.
  • WARNING: Indicates potential harm to the user if action is taken.
  • CAUTION: Indicates potential damage to the product or other property if action is taken.

Safety Symbols: Safety Warning, Protective Test, Ground Shell, Earth Terminal
Safety Symbols: Safety Warning, Protective Test, Ground Shell, Earth Terminal

1.11. Measurement Category

DPO6000 / MPO6000 series digital oscilloscopes perform measurements under Measurement Category I. This oscilloscope is only allowed to be used in the specified measurement category.

  • Measurement Category I: Measurements on circuits not directly connected to the mains (e.g., internal protected circuits).
  • Measurement Category II: Measurements on circuits directly connected to low-voltage equipment (e.g., home appliances, portable tools).
  • Measurement Category III: Measurements in construction equipment (e.g., power distribution boards, circuit breakers, fixed equipment).
  • Measurement Category IV: Measurements at the source of low-voltage equipment (e.g., electricity meters, major overcurrent protection devices).

2. Setup and Initial Inspection

2.1. General Inspection

2.1.1. Checking the Shipping Packaging

If the packaging carton or foam protective pad is seriously damaged, keep it until the entire machine and accessories pass electrical and mechanical tests.

2.1.2. Check the Whole Machine

If the instrument appears damaged, is not working properly, or fails performance tests, contact your dealer.

2.1.3. Check Random Accessories

If any accessory is missing or damaged, contact your dealer. Details of provided attachments are in Appendix A.

Accessories Table: Instrument, USB cable, Power Cord, Oscilloscope probe x4, Packing list.
Accessories Table: Instrument, USB cable, Power Cord, Oscilloscope probe x4, Packing list.

2.2. Physical Dimensions

The oscilloscope has a length of 318 mm, a width of 140 mm, and a height of 150 mm.

Front and Top View of Oscilloscope with dimensions
Front and Top View of Oscilloscope with dimensions

2.3. Adjusting the Bracket

Properly adjust the stand to tilt the oscilloscope upward for better operation and display screen observation.

2.4. Connect the Power Cord

The AC power input specifications are: 100-120Vac, 50/60/400Hz; 100-240Vac, 50/60Hz; 50W max. Connect the oscilloscope to an AC power source using the provided power cord.

Rear view of the oscilloscope showing power jack
Rear view of the oscilloscope showing power jack

2.5. Startup Check

Press the power switch on the lower left corner of the front panel. The oscilloscope will perform self-tests, with keyboard indicators lighting up. After self-tests, the startup screen appears.

2.6. Probe Inspection and Connection

2.6.1. Safety

When using the probe, keep your fingers behind the safety ring. Do not touch metal parts when connected to high voltage. Connect the probe to the oscilloscope and ground terminal before measurements.

Oscilloscope probe with finger guard
Oscilloscope probe with finger guard

2.6.2. Probe Attenuation Setting

Probes have different attenuation coefficients. Use the 'Vertical Menu' button (e.g., [CH1]) to select a probe option matching the probe attenuation coefficient. Ensure the 'Attenuation' switch on the probe matches the 'Probe' option in the oscilloscope (X1 or X10). X1 limits bandwidth to 6MHz; X10 provides full bandwidth.

Oscilloscope probe with attenuation switch
Oscilloscope probe with attenuation switch

2.6.3. Connect the Passive Probe

  1. Connect the BNC end of the probe to the analog channel input on the front panel.
  2. Connect the probe ground alligator clip or ground spring to the circuit ground terminal, then connect the probe to the circuit under test.

2.6.4. Connect the Logic Probe

  1. Connect one end of the digital probe to the digital channel input interface on the front panel (no direction).
  2. Connect the other end of the digital probe to the signal under test. MPO6000 series comes standard with four LP104 digital probes.

2.6.5. Check the Probe Voltage

Perform a probe check each time you connect a probe to an input channel. Use the 'Vertical Menu' button (e.g., [CH1]) to set the probe attenuation coefficient.

2.6.6. Manual Probe Compensation

Adjust probe compensation when connecting for the first time to match the probe to the input channel. Uncompensated or overcompensated probes can cause measurement errors.

  1. Set probe option attenuation to 10X in the Channel Menu. Set the switch on the probe to 10X and connect to channel 1. Connect the probe tip to the ~2V @ 1K Hz connector and the reference wire to the 'probe element ground' connector. Press 'Auto Scale'.
  2. Check the shape of the displayed waveform.
  3. If necessary, use a non-metallic screwdriver to adjust the variable capacitance on the probe until the waveform is 'compensated correctly'.

Waveform compensation examples: Correctly compensated, Overcompensated, Undercompensated
Waveform compensation examples: Correctly compensated, Overcompensated, Undercompensated

3. Operation

3.1. Front Panel Overview

The front panel provides controls for various functions of the oscilloscope.

Front panel layout of the Hantek Oscilloscope
Front panel layout of the Hantek Oscilloscope

SNDescriptionSNDescription
1LCD9Vertical control system
2Menu show/hide keys10Signal source 2 output channel
3Signal source button (signal source series)11CH1~CH4 signal input channel
4Multifunction knob12Signal source 1 output channel
5Menu function buttons13LA1~LA4 signal input channels (MPO6000 series)
6Level control system14Probe compensation signal output/ground
7Trigger control system15USB Host port
8Shortcuts (Run/Stop, Single, Auto Scale)16Power On Key

3.2. Rear Panel Overview

The rear panel provides various connection ports and controls.

Rear panel layout of the Hantek Oscilloscope
Rear panel layout of the Hantek Oscilloscope

  1. Handle: Pull vertically to carry; press down when not needed.
  2. AC Power Jack: Input for AC power.
  3. AUX Port Trigger Output and Pass / Fail [Optional]:
    • Trigger output: Outputs a pulse when the oscilloscope triggers, reflecting the current capture rate.
    • Pass / Fail: Outputs a pulse when a failure is detected in pass/fail tests; continuously outputs low level if no failure.
  4. Signal Source 3 Output [MPO6000EDU Series]: For 3-channel signal source. Outputs signals according to current settings when turned on.
  5. USB Device: Connects the oscilloscope to a computer or printer.
  6. LAN / UART:
    • LAN: Connects to network for remote control. Complies with LXI CORE 2011 DEVICE standard.
    • UART [Optional]: Connects to a control system for remote control.
  7. HDMI Interface [Optional]: Connects to an HDMI display for a larger view.

3.3. Front Panel Functions

3.3.1. Level Control System

  • Horizontal Offset: Modify horizontal displacement. Knob adjusts trigger point left/right. Press knob to reset to zero.
  • Horizontal Time Base: Modify horizontal time base. Clockwise decreases, counter-clockwise increases. Press knob to switch between single and double window display modes.

3.3.2. Vertical Control System

  • Analog Input Channel Switch: Press any button (1-4) to open/close the corresponding channel menu.
  • Math: Press 'Math' button to open/close Math setup menu.
  • Vertical Offset: Modify vertical shift of current channel waveform. Knob adjusts waveform up/down. Press knob to zero vertical displacement.
  • Vertical Voltage: Modify vertical scale of current channel. Knob adjusts gears. Press knob to switch between 'Coarse' or 'Fine' adjustment.

3.3.3. Signal Source

  • WaveGen: Press this key to enter the signal source setting interface.

Note: This function is only applicable to digital oscilloscopes with signal source channels.

3.3.4. Trigger Control

  • Trigger Level: Modify trigger level. Knob adjusts level line up/down. Press knob to restore to zero.
  • TrigMenu: Press to open trigger operation menu. Provides various trigger types.
  • ForceTrig: Short press forces a trigger signal. Press and hold to open history waveform.

3.3.5. Automatic Waveform Display

  • AutoScale: Press to enable automatic waveform setting. Adjusts vertical scale, horizontal time base, and trigger mode for best waveform display.

3.3.6. Operational Control

  • Run/Stop: Press to 'run' (green backlight) or 'stop' (red backlight) waveform sampling.
  • Single: Press to set trigger mode to 'Single'. Forces a trigger signal immediately when 'Force Trig' is pressed.

3.3.7. Multifunction Knob (V0)

Performs various settings: waveform brightness, trigger time, trigger type/source, signal source waveform, frequency, offset, amplitude, etc. Turn clockwise to increase, counter-clockwise to decrease. Press to select option.

3.3.8. Function Menu

  • Default: Briefly press to execute default setting. Long press to enter Default function preset (factory, default, user settings).
  • Help: Press to open/close built-in help system.
  • Utility: Press to enter system function setting menu (interface, sound, language, advanced features like pass/fail tests).
  • Cursor: Press to enter cursor measurement menu (manual, tracking modes).
  • Measure: Press to enter measurement setting menu (data source, DVM on/off, all measurements, statistics). 'All Measurements' measures 42 waveform parameters.
  • Save/Recall: Short press to enter file storage/recall interface (setup, waveform, reference, CSV). Supports internal/external storage. Long press saves picture in .bmp format to external storage.
  • Display: Short press to enter display setting menu (waveform display type, afterglow time, waveform brightness, screen grid, grid brightness). Press and hold to clear afterglow, then collect/count again.

3.4. User Interface

The user interface displays various information about the oscilloscope's status and measurements.

Oscilloscope User Interface with labeled sections
Oscilloscope User Interface with labeled sections

  1. Hantek Trademark: Touch 'Touch Screen' in 'Utility', then Hantek logo to scan QR code for wizard interface.
  2. Running Status: AUTO (automatic mode, no triggers), READY (pre-trigger data acquired, ready for trigger), TD (triggered), ROLL (scroll mode, continuous acquisition), STOP (acquisition stopped), ARM (pre-trigger state).
  3. Main Time Base of Current Window: Length of time per division on horizontal axis.
  4. Current Sampling Rate, Sampling Points: Displays current sampling rate and memory depth. Changes with horizontal time base.
  5. Waveform Memory: Visual representation of waveform in memory.
  6. Horizontal Trigger Time: Adjust with horizontal offset knob. Press knob to reset.
  7. Trigger Level: Adjust with trigger level knob. Press knob to reset.
  8. Operation Menu Displays: Information for each function key.
  9. Time Display of Oscilloscope.
  10. WIFI Status Display.
  11. LAN Connection Status Indication: Lit when LAN is connected.
  12. USB External Storage Device Status Display: Lit when USB external storage device is connected.
  13. USB Host Computer Connection Status Display: Lit when USB host computer is connected.
  14. Logic Analyzer Status Display: Lit when LA channel is connected.
  15. Gen3 Status Display [MPO6000EDU series]: Lit when Gen3 channel is turned on.
  16. Gen2 Status Display [Signal Source Series]: Lit when Gen2 channel is turned on.
  17. Gen1 Status Display [Signal Source Series]: Lit when Gen1 channel is turned on.
  18. CH4 Status Display [4-channel series]: Lit when CH4 channel is turned on.
  19. CH3 Status Display [4-channel series]: Lit when CH3 channel is turned on.
  20. CH2 Status Display: Lit when CH2 channel is turned on.
  21. CH1 Status Display: Lit when CH1 channel is turned on.
  22. Channel Zero-Scale Position Display.
  23. Channel Trigger Level Position Display.

3.5. Sampling System

Understanding the sampling principle and the relationship between sampling rate and memory depth is crucial for oscilloscope waveform acquisition.

3.5.1. Sampling Principle

According to Nyquist, sampling frequency must be greater than twice the highest frequency (Fs > 2Fmax) to restore analog signals without distortion. Higher sampling rates yield more accurate waveforms but require faster conversion circuits.

3.5.2. Waveform Aliasing

If sampling conditions (Fs > 2Fmax) are not met, the reconstructed waveform frequency will be smaller than the actual signal. This can be avoided by increasing sampling frequency or using a low-pass (anti-aliasing) filter.

3.5.3. Waveform Distortion

Low sampling rates can lead to missing waveform details, causing the displayed waveform to differ from the actual signal.

3.5.4. Waveform Miss

When the sampling rate is too low, the reconstructed waveform may not reflect all actual signals.

3.5.5. Sampling Rate and Memory Depth

The highest sampling rate is 1GSa/s. The actual sampling rate is determined by the horizontal time base scale and storage depth. Memory depth is the number of waveform points stored in one triggered acquisition (max 128Mpts).

Relationship: Sampling time = memory depth / (sampling rate (Sa/s))

3.5.6. Waveform Interpolation

Digital oscilloscopes use interpolation to increase signal visibility by connecting discrete samples. Methods include sine, linear, and step interpolation.

  • Linear Interpolation: Connects adjacent points with straight lines. Suitable for straight-edge signals.
  • Step Interpolation: Inserts the value of the previous sampling point between adjacent samples.
  • Sine Interpolation: Uses mathematical processing to create a more realistic waveform shape. Recommended when sampling rate is 3-5 times system bandwidth.

Various waveform displays (YT single/double window, unaveraged/averaged, XY single/double window)
Various waveform displays (YT single/double window, unaveraged/averaged, XY single/double window)

3.6. Operational Control

Use 'Run/Stop' and 'Single' buttons to control sampling.

  • Run/Stop: Green for running, red for stopped.
  • Single: Captures and displays a single acquisition.

3.6.1. Collection Method

Access via 'Utility -> Acquisition -> Acquisition Mode'.

  • Normal: Samples at equal time intervals, best for most waveforms.
  • Peak: Collects max/min values to capture signal envelope or narrow pulses. Avoids aliasing but increases noise.
  • Average: Averages multiple samples to reduce random noise and improve vertical resolution.
  • High Resolution: Uses oversampling to average neighboring points, reducing noise and smoothing waveforms. Improves measurement accuracy by sacrificing bandwidth.

3.7. Horizontal Control System

3.7.1. Time Base Mode

Access via 'Utility -> Acquisition -> Display Mode'. Includes YT, XY, and Roll modes.

  • YT Mode: Y-axis is voltage, X-axis is time. Enters scan mode if horizontal time base is 100ms/div or slower.
  • XY Mode: Converts two input channels from voltage-time to voltage-voltage display. X-axis/Y-axis combinations can be CH1-CH2 or CH3-CH4. Useful for Lissajous figures to measure phase difference.
  • Roll Mode: Waveform scrolls from right to left. Horizontal gear range 100ms to 100s.

3.8. Vertical Control System

3.8.1. Enable Analog Channel

DPO6000/MPO6000 series have 4 analog input channels (CH1-CH4). Press the corresponding button on the front panel to turn a channel on/off.

3.8.2. Channel Coupling

Filters out unwanted signals. Options: DC (passes both DC and AC), AC (blocks DC), GND (blocks both DC and AC). Select via '1 -> Coupling' or F1 [V0] key.

3.8.3. Bandwidth Limitation

Reduces noise by limiting bandwidth. Default is off. Set to 20 MHz to attenuate high-frequency components. Switch via '1 key' then F2 key.

3.8.4. Waveform Inversion

Press '1 -> Invert' to turn waveform inversion on/off. Inverted mode flips the waveform voltage value.

3.8.5. Probe Ratio

Manually set probe attenuation ratio via '1 -> Probe'.

3.8.6. Vertical Scale

Adjusts voltage value per division (V/div). Rotate SCALE knob. 'Coarse' (1-2-5 steps) or 'Fine' adjustment modes available.

3.9. Analog Channel Delay Setting

Corrects zero offset caused by probe cable propagation delay. Adjust with multifunction knob V0 (range -100ns to 100ns). Press knob to reset to 0.00s.

3.10. Trigger System

Triggers define when the oscilloscope starts acquiring and displaying data.

3.10.1. Trigger Source

Select via 'Trig Menu'. Analog channels CH1-CH4, digital channels D1.0-D4.3 can be trigger sources. An invalid trigger occurs if the selected channel is not on.

3.10.2. Acquisition Process

The acquisition memory is divided into pre-trigger and post-trigger buffers. Data is continuously stored in the pre-trigger buffer until a trigger condition is met, then post-trigger data is stored.

3.10.3. Trigger Mode

  • Automatic Mode: Allows free running acquisitions without a valid trigger.
  • Normal Mode: Updates waveform only with a valid trigger.
  • Single Trigger: Stops after a single valid trigger.

3.10.4. Trigger Holdoff

Time difference between detecting one trigger and preparing for another. Used for stable display of complex waveforms. Access via 'Trig Menu -> Hysteresis'.

3.10.5. Trigger Types

DPO6000/MPO6000 series have up to 16 trigger functions. Access via 'Trig Menu'.

  • Edge Trigger: Triggers on a specified edge (rising, falling, or both) and voltage level.
  • Pulse Width Trigger: Triggers on positive or negative pulses of a specified width.
  • Video Trigger: Captures complex waveforms of standard analog and HD video signals (NTSC, PAL, HDTV).
  • Slope Trigger: Triggers on positive or negative slope from one level to another within a specified time.
  • Overtime Trigger: Triggers when the time interval from a rising/falling edge passing the trigger level to the adjacent falling/rising edge passing the trigger level exceeds a set overtime period.
  • Window Trigger: Provides high and low trigger levels. Triggers when input signal passes high or low level.
  • Pattern Trigger: Identifies trigger conditions by looking for specific patterns (high, low, or inactive) across channels.
  • Interval Trigger: Triggers when the interval between two consecutive rising/falling edges meets a set time condition.
  • Under Amp Trigger: Triggers on pulses that cross one trigger level but not another (positive or negative runt pulses).
  • Delay Trigger: Triggers when the time difference between edges from two data sources meets a preset time limit.
  • Setup/Hold Trigger: Triggers when setup time (data stable before clock edge) or hold time (data stable after clock edge) is less than preset.
  • UART Trigger [Optional]: Triggers on frame start, stop bit, data, parity error, or data bit error of UART signal.
  • LIN Trigger [Optional]: Triggers on LIN bus signal message frames (interval field, sync field, ID field, identifier, data).
  • CAN Trigger [Optional]: Triggers on CAN bus frames (start bit, remote frame ID, data frame ID, error frame, overload frame).
  • SPI Trigger [Optional]: Triggers when specified data is found in SPI communication, based on SCL clock and SDA data sources.
  • IIC Trigger [Optional]: Triggers on IIC bus signals (start condition, stop condition, no response, address, restart, address and data).

3.11. Mathematics

The oscilloscope supports various mathematical operations on waveforms.

  • Addition: Adds waveform values of two data sources point by point.
  • Subtraction: Subtracts waveform values of two data sources point by point.
  • Multiplication: Multiplies waveform values of two data sources point by point.
  • Division: Divides waveform values of two data sources point by point.
  • FFT (Fast Fourier Transform): Converts time domain signal to frequency components (spectrum). Used to measure harmonic content, noise characteristics, vibration analysis.
  • Integral: Calculates integral of a single data source.
  • Differential: Calculates discrete-time derivative of a single data source.
  • Square Root: Calculates square root of a single data source waveform point by point.
  • Expression: User-defined operation expressions using analog channel waveforms.

3.12. Measurement

The oscilloscope provides automatic measurement of 42 waveform parameters and statistics.

3.12.1. Measurement Parameter Description

Includes time parameters (Period, Frequency, Rise Time, Fall Time, Pulse Width, Duty Cycle, Burst Length, Max/Min Time), count value parameters (Positive/Negative Pulse Count, Rise/Fall Pulse Count, Trigger Count), delay and phase parameters (FRR, FFF, FRF, FFR, LRR, LRF, LFR, LFF, Phase), and voltage parameters (Maximum, Minimum, Peak-Peak, Top, Bottom, Amplitude, Middle, Average, Period Average, RMS, Overshoot, Preshoot, Fall Overshoot, Rise Pre-shoot, Variance).

3.12.2. All Measurements

Press 'Meas' on the front panel, then 'All Measurements' to measure all 42 parameters with one key. Results displayed at bottom of screen with color markers.

3.12.3. Measurement Statistics

Displays current, average, maximum, minimum, root mean square error, and count values for selected waveform parameters.

3.12.4. Digital Voltmeter (DVM)

Built-in 5-digit DVM and 6-digit frequency meter. Measures AC RMS, DC, AC+DC RMS voltage and frequency. Access via 'Meas -> Digital Voltmeter -> Enable'.

3.12.5. Grid Scale Measurement

Estimates frequency and voltage amplitude using grid divisions and scale factor.

3.12.6. Automatic Measurement

After connecting and inputting a signal, press 'Auto Scale'. Options: Single cycle, Multi-cycle, Auto range, Data source.

3.12.7. Cursor Measurement

Uses amplitude (horizontal dashed line) and time (vertical dashed line) cursors. Modes: Manual (adjust cursors manually) and Tracking (cross cursor automatically positioned on waveform). Access via 'Cursors' button.

3.13. Bode Plots (Signal Source Series)

Semi-logarithmic plot of transfer function versus frequency. Displays frequency response of a system. Access via 'Utility -> Bode Plots'.

  • Enable Bode Plots: 'Utility -> Bode Plots -> Function [On, Off]'.
  • Data Source Settings: Input source (CH1-CH4), Output source (CH1-CH4).
  • Chart Settings: Gain gear, Gain Offset, Phase scale, Phase Offset.
  • Operation: 'Utility -> Bode Plots -> Run'.

3.14. Digital Channel

MPO6000 series has 4 analog and 16 digital channels. DPO6000 series can be upgraded with LP104 digital probe. Compares voltage to logic threshold (1 if greater, 0 otherwise).

  • Select Digital Channel: 'Utility -> Logic Analyzer -> Data Source [D1, D2, D3, D4]'.
  • Enable Digital Channels: Connect digital probe, logic analyzer enable switch turns on automatically.
  • Set Digital Channel Thresholds Type: 'Utility -> Logic Analyzer -> Type'. Options include 5.0V CMOS, 3.3V CMOS, 2.5V CMOS, 1.8V CMOS, ECL, PECL, LVDS, 0V, User-defined.

3.15. Waveform Generator

DPO6000C/MPO6000D have two built-in channels (MPO6000EDU has three) and a 25 MHz signal source. Can output sine, square, pulse, triangle, noise, DC, sampling, exponent, semi-distortion, Lorentz, dual-tone multi-frequency, Gauss, ECG signals (13 basic types) and 4 arbitrary waveforms.

  • Output Basic Waveform: Select waveform, data source, channel, frequency, amplitude, offset voltage, phase, load impedance.
  • Modulation: Supports AM, FM, PM, PWM.
  • Burst: Outputs waveforms with a specified number of cycles.
  • Edit Arbitrary Waveform: Use 'Wave Editor Setup.exe' software.

3.16. Pass / Fail Test

Evaluates signal quality by testing waveforms in real-time against set rules.

  • Enable Pass / Fail Test: 'Utility -> Pass / Fail -> Enable test -> On'. Failure area is shaded.
  • Data Source: Select channel (CH1-CH4).
  • Rule: Customize rules for pass/fail tests using 'Horizontal' and 'Vertical' soft keys.
  • Test and Output: Set output format of test results. Options: Message (display results), Stop On Output (On/Off), Output to DDS (signal source output).
  • Segmented Acquisition: Stores data from multiple triggers in segments.
  • History Waveform: Stores up to 80,000 frames. Allows observation and analysis of previously stored waveforms.
  • Display Control: Sets waveform display type (Vector/Dot), persist time, waveform brightness, grid type, grid brightness, color grade.

4. Maintenance

4.1. Daily Maintenance and Cleaning

  • When storing or placing the oscilloscope, do not expose the LCD monitor to direct sunlight for a long time.
  • Check the oscilloscope and probe frequently.
  • Use a lint-free cloth to remove dust. Be careful not to scratch the display filter.
  • Clean with a soft cloth dampened with water. For thorough cleaning, use 75% isopropyl alcohol in water.

CAUTION: Do not use corrosive or chemical cleaning agents.

4.2. Calibration

Perform calibration to ensure accurate measurements. Execute at any time, especially if ambient temperature changes by more than 5°C. Warm up for over 30 minutes, disconnect all input channels, then press 'Utility -> Calibration'.

5. Troubleshooting

5.1. Common Faults and Solutions

  1. The oscilloscope does not turn on after pressing the power switch:
    1. Check if the power cord is connected normally.
    2. Check if the power switch is pressed.
    3. After checking, try turning on the instrument again. If it still doesn't turn on, contact Hantek Technical Support.
  2. No waveform is displayed on the input signal screen after the oscilloscope is turned on:
    1. Check if the oscilloscope probe is correctly connected to the BNC connector of the signal input channel.
    2. Check if the channel switch is on (CH1~CH4 menu button).
    3. Check if the input signal is correctly connected to the probe.
    4. Check if there is signal output from the measured circuit.
    5. If it is a DC signal with large amplitude, increase the amplitude.
    6. Try pressing the automatic measurement button to detect the signal. If still no waveform, contact Hantek Technical Support.
  3. The input signal waveform distortion is serious:
    1. Check if the oscilloscope probe is connected to the channel BNC connector.
    2. Check if the probe is well connected to the measured object.
    3. Check if the oscilloscope probe is calibrated properly. If not, calibrate according to the manual.
  4. The waveform continuously scrolls on the screen and cannot be triggered:
    1. Check if the trigger data source is consistent with the signal input channel.
    2. Check if the trigger level is adjusted correctly. Press the trigger level knob to return to the signal center point.
    3. Check if the trigger mode is correct. Select the appropriate trigger mode for different input signals.

6. Technical Specifications

All technical specifications apply to the DPO6000 / MPO6000 series digital storage oscilloscopes. To verify specifications, the oscilloscope must have been working continuously for more than 20 minutes within the specified operating temperature, and a self-calibration must be performed if the temperature changes by more than 5°C.

6.1. Acquire

ParameterSpecification
Acquisition methodNormal, average, peak, high resolution
Real-time sampling rate1 GSa/s (single channel), 500 MSa/s (dual channel), 250 MSa/s (three/four channels)
Peak DetectAnalog Channel 4ns
Averaging modeN samples for analog channels (N values: 2, 4, 8, 16, 32, 64, 128, 256, 512, and 1024)
High resolution12bit
Minimum detection pulse width8ns
Memory DepthSingle Channel 64M / 128M [EDU], Dual channel 32M / 64M [EDU], Three or four channels 16M / 32M [EDU]

6.2. Input

ParameterSpecification
Number of channels4 analog channels, 16 (4 * 4) digital channels (total opened channels < 4)
Input couplingDC, AC or GND (Digital channels not supported)
Input resistanceAnalog channel: 1MΩ ± 2%, 25pF ± 3pF; Digital channels: 300KΩ ± 2%, 8 pF ± 3 pF
Voltage rating300V CAT II
Input voltageAnalog channels 300VRMS (10X); Digital channel -25V~25V

6.3. Horizontal

ParameterSpecification
Waveform InterpolationSine, Linear, Step
Horizontal scale2ns/div~100s/div 1,2,5 steps
Time base modeY-T, X-Y, Roll
X-Y number1, 2 channel combination 1 XY; 3, 4 channels combined 1 XY
Zero offset± 0.5 div x minimum time base scale
Sampling rate and delay time accuracy±25ppm
Clock drift≤±5 ppm/year
Incremental time measurement accuracy (Full bandwidth)single shot, "sampling" mode: ± (1 sampling interval + 100ppm x reading + 0.6ns); > 16 times average: ± (1 sampling interval + 100ppm × reading + 0.4ns)

6.4. Vertical

ParameterDPO6084DPO6104DPO6204
Bandwidth (-3db)80MHz100MHz200MHz
Vertical resolutionAnalog channel 8bit, Digital channel 1bit
Vertical scale500µV/div~10V/div at input BNC
Position range500µV/div to 128mV/div, ±1V; 129mV/div to 1.28V/div, ±10V; 1.29V/div to 10V/div, ±50V
Selectable analog bandwidth limit, typically20MHz
Low frequency response (-3db)≤10Hz at BNC
Rise time at BNC, typical≤4.4ns≤3.5ns≤1.8ns
Vertical gain accuracyIn "normal" or "average" acquisition mode, 10V/div to 10mV/div is ± 3%; 5mV/div to 500uV/div is ± 4%.
DC offset accuracy± 0.1div ± 2 mV ± 1% offset value
Channel IsolationDC to Maximum Bandwidth: > 40 dB

6.4.1. Vertical (Digital Channel: for MPO6000D / MPO600EDU or DPO6000B / DPO6000C purchase digital probe LP104)

ParameterSpecification
ThresholdAdjustable threshold for 4 channels as a group
Threshold selectionTTL (1.4 V), 5.0 V CMOS (+2.5 V), 3.3 V CMOS (+1.65 V), 2.5 V CMOS (+1.25 V), 1.8 V CMOS (+0.9 V), ECL (-1.3 V), PECL (+3.7 V), LVDS (+1.2 V), 0V, Custom
Threshold range±7.0V, 10Mv steps
Threshold accuracy± (Threshold setting of 100mV + 3%)
Dynamic range± 5.0V + threshold
Minimum voltage swing500 mVpp
Vertical resolution1 bit

6.5. Trigger

ParameterSpecification
Trigger level range± 4 divisions from the center of the screen
Trigger modeAuto, Normal, Single
Holdoff range8ns~10s
Trigger level accuracyCH1~CH4 0.2 divisions x volts / division within ± 4 divisions from the center display
Edge trigger SlopeRising, Falling, Rising or Falling
Edge trigger Data sourceCH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3
Pulse width trigger PolarityPositive, Negative
Pulse width trigger Conditions<, >, !=, =
Pulse width trigger Data sourceCH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3
Pulse width range8ns~10s
Video trigger Signal standardNTSC, PAL, HDTV720P, HDTV1080P, HDTV1080i
Video trigger data sourceCH1~CH4
Video trigger SynchronizeScan line, line number, odd field, even field, all fields
Video trigger ScopeRising, falling
Video trigger Condition<, >, !=, =
Video trigger Data sourceCH1~CH4
Video trigger Time range8ns~10s
Overtime trigger Data sourceCH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3
Overtime trigger PolarityPositive, Negative
Overtime trigger Time range8ns~10s
Window trigger Data sourceCH1~CH4
Pattern trigger Pattern0: low level; 1: high level; X: ignore;
Pattern trigger Data sourceCH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3
Interval trigger SlopeRising, falling
Interval trigger Condition<, >, !=, =
Interval trigger Data sourceCH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3
Interval trigger Time range8ns~10s
Delay trigger Edge typeRising, falling
Delay trigger Data sourceCH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3
Delay trigger Condition<, >, !=, =
Delay trigger Time range8ns~10s
Setup/hold trigger Edge typeRising, falling
Setup/hold trigger Data sourceCH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3
Setup/hold trigger Time range8ns~10s
Under Amp trigger PolarityPositive, Negative
Under Amp trigger Condition<, >, !=, =
Under Amp trigger Data sourceCH1~CH4
Under Amp trigger Time range8ns~10s
UART trigger ConditionStart, stop, data, parity, receive error
UART trigger Data source (RX/TX)CH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3
UART trigger Data formatHex
UART trigger Data length1 byte
UART trigger Data bits5 bit, 6 bit, 7 bit, 8 bit
UART trigger ParityNone, Odd, Even
UART trigger Idle levelhigh, low
UART trigger Baud rate (optional)110/300/600/1200/2400/4800/9600/14400/19200/38400/57600/115200/230400/380400/460400 bit/s
UART trigger Baud rate (custom)300bit/s~334000bit/s
LIN trigger ConditionInterval field, sync field, ID field, sync error, identifier, ID, and data
LIN trigger Data sourceCH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3
LIN trigger Data formatHex
LIN trigger Baud rate (optional)110/300/600/1200/2400/4800/9600/14400/19200/38400/57600/115200/230400/380400/460400 bit/s
LIN trigger Baud rate (custom)300bit/s~334000bit/s
CAN trigger ConditionStart bit, remote frame ID, data frame ID, frame ID, data frame data, error frame, all errors, response error, overload frame
CAN trigger Data sourceCH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3
CAN trigger Data formatHex
CAN trigger Baud rate (optional)10000, 20000, 33300, 500000, 62500, 83300, 100000, 125000, 250000, 500000, 800000, 1000000
CAN trigger Baud rate (custom)5kbit/s~1Mbit/s
SPI trigger Data sourceCH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3
SPI trigger Data formatHex
SPI trigger Data bits4 bit, 8 bit, 16 bit, 24 bit
IIC trigger Data source (SDA/SCL)CH1~CH4, D1.0~D1.3, D2.0~D2.3, D3.0~D3.3, D4.0~D4.3
IIC trigger Data formatHex
IIC trigger Data index0~7
IIC trigger ConditionStart bit, stop bit, no response, address, address and data, restart

6.6. Display

ParameterSpecification
Display Type7 "TFT LCD Diagonal
Display resolution800 (horizontal) * 480 (vertical) pixels
Display color16 million colors (24 bit true color)
Persist timeMin, 1 s, 5 s, 10 s, 30s, infinite
Display typePoint, vector
Display modeColor grade
Display brightnessAdjustable
Grid TypeOptional
Grid brightnessAdjustable
Waveform brightnessAdjustable

6.7. Arbitrary Signal Generator (with Signal Source Series)

ParameterSpecification
Number of channels2/3 channels
Sampling rate200MSa/s
Vertical resolution12 bits
Frequency25 MHz
Standard waveformsine wave, square wave, pulse, triangle wave, noise, DC, Sinc, Exp, Haver Sine, Lorentz, dual tone multi-frequency, Gauss, ECG signal
Arbitrary waveformArb1, Arb2, Arb3, Arb4
Frequency characteristics (Sine wave)1uHz~25MHz
Frequency characteristics (Square wave)1uHz~15MHz
Frequency characteristics (Pulse)1uHz~5MHz
Frequency characteristics (Triangle wave)1uHz~1MHz
Frequency characteristics (Noise, Sinc, Exp, Haver Sine, Lorentz, dual tone multi-frequency, Gauss, ECG)1uHz~5MHz
Arbitrary wave Length8KSa
Arbitrary wave Range1uHz~75MHz
Arbitrary wave Frequency Accuracy(less than 10 kHz) 50 ppm (more than 10 kHz)
Arbitrary wave Resolution0.1 Hz or 4 bits, whichever is greater
Output range10mV~7Vpp (high impedance), 5mV~3.5Vpp(50Ω)
DC offset Range±7 V, high impedance
Probe compensator output Voltage, typical2Vpp input ≥1MΩ load
Probe compensator output Frequency, typical1kHz

6.8. General Technical Specifications

ParameterSpecification
Power supply voltage100-120VACRMS (± 10%), 45Hz to 440Hz, CAT II; 120-240VACRMS (± 10%), 45Hz to 66Hz, CAT II
Power consumptionless than 30W
FuseT, 3.15A, 250V, 5x20mm
Operating temperature0~50 °C (32~122 °F)
Storage temperature-40~+71 °C (-40~159.8 °F)
Humidity≤ + 40 °C: ≤90% relative humidity; +41 °C ~ 50 °C: ≤60% relative humidity
Cooling methodConvection
Altitudewith and without operation: 3000m (10,000 feet)
Random vibration0.31 g RMS from 50Hz to 500Hz, 10 minutes per axis
Mechanical shock (when not operating)2.46g RMS from 5Hz to 500Hz, 10 minutes per axis
Mechanical shock (during operation)50g, 11ms, half sine wave
Oscilloscope size318 x 140 x 150mm (length x width x height)
Weight2900g

7. Warranty and Support

Qingdao Hantek Electronic Co., Ltd. reserves the right to amend this document without prior notice. The information provided is believed to be correct and reliable, but infallibility is not guaranteed. Ensure the latest and valid versions of technical documents are used. Responsibility for obtaining third-party consent/authorization for their products/patents lies with the user's company.

7.2. Technical Support

For questions or ambiguities, contact support through the following ways:

  • Contact your local dealer.
  • Contact the local office directly under Qingdao Hantek Electronic Co., Ltd.
  • Contact the headquarters of Qingdao Hantek Electronic Co., Ltd.

Contact Method:
Qingdao Hantek Electronic Co., Ltd.
Website: http://www.hantek.com/en/index.html
Address: 35# Building No.780 Baoyuan Road, High-tech District, Qingdao China
Zip Code: 266114
Telephone: 0532-88705792
Fax: 0532-88705691
Email: service@Hantek.com

Technical Support:
Telephone: 0532-88703687
Email: support@Hantek.com

7.3. Equipment Recycling

This equipment uses natural resources. Improper disposal may harm the environment or human health. Recycle appropriately to ensure most materials are reused.

7.4. Hazardous Substances or Elements

The product contains certain hazardous substances or elements as per SJ / T 11363-2006 standard. 'X' indicates content exceeds limit, '0' indicates content is below limit.

Part NamePbHgCdCr(Vi)PBBPBDE
Product shell, chassisX00X00
Display moduleXX0000
Circuit board componentsX00X00
Power componentsX00X00
Wire and cable assemblyX00000
ConnectorX00X00
Fasteners and mounting hardwareX0XX00
Other attachments (Including probe)X00X00
Other000000

8. User Tips

  • For optimal performance, allow the oscilloscope to warm up for at least 30 minutes before use.
  • If you experience waveform distortion, ensure your probes are properly calibrated. Refer to the 'Manual Probe Compensation' section for instructions.
  • When troubleshooting, always start by checking basic connections and power before delving into complex settings.
  • Utilize the 'Auto Scale' function for a quick and optimized display of your input signal.
  • If you need to analyze specific signal characteristics, explore the various trigger types available in the 'Trig Menu' to isolate the desired events.