1. Introduction
The ITECH IT-M3100D series is an ultra-compact, wide-range dual-channel DC power supply designed for various testing applications. This series offers high performance in a small footprint, making it suitable for laboratory, industrial, and educational environments. It features independent dual channels, flexible output configurations, and advanced control capabilities.
2. Safety Information
Please read all safety instructions carefully before operating the device. Failure to comply with these instructions may result in electric shock, fire, or damage to the product.
- Always connect the power supply to a grounded AC outlet.
- Do not operate the device in wet or damp conditions.
- Ensure proper ventilation to prevent overheating.
- Do not open the casing; there are no user-serviceable parts inside. Refer servicing to qualified personnel.
- Verify that the input voltage matches the power supply's rating before connecting.
- Avoid touching output terminals when the device is powered on.
3. Product Components and Overview
3.1 Front Panel
The front panel features the main display, control buttons, and a rotary encoder for setting parameters. The display provides real-time voltage and current readings for both channels.
3.2 Rear Panel
The rear panel includes the AC power input, Ethernet ports for network connectivity, and the output terminals for Channel 1 (CH1) and Channel 2 (CH2). It also features interface expansion slots for optional accessories.
4. Initial Setup
- Unpacking: Carefully remove the power supply from its packaging and inspect for any signs of damage.
- Placement: Place the unit on a stable, level surface with adequate ventilation around all sides. Do not block the ventilation openings.
- Power Connection: Connect the supplied AC power cord to the power input on the rear panel and then to a grounded AC power outlet.
- Load Connection: Connect your load to the CH1 and/or CH2 output terminals on the rear panel using appropriate wiring. Ensure correct polarity.
- Optional Interface Connection: If using optional accessories, insert them into the interface expansion slots on the rear panel and connect any necessary cables (e.g., USB, LAN, GPIB, RS-232/CAN, Analog/RS485).
4.1 Optional Accessories Interfaces
The rear panel supports various interface expansion modules:
| Model | Interface |
|---|---|
| IT-E1205 | GPIB interface |
| IT-E1206 | USB/LAN interface |
| IT-E1207 | RS-232/CAN interface |
| IT-E1208 | Analog interface/RS485 interface |
| IT-E1208D | Double channel analog/RS485 interface |
| IT-E1209 | USB interface |
5. Operation
5.1 Basic Operation
- Power On/Off: Press the power button on the front panel to turn the unit on or off.
- Setting Voltage/Current: Use the CH1/CH2 buttons to select the desired channel. Use the rotary encoder and 'Set' button to adjust voltage and current limits.
- Output On/Off: Use the 'On/Off' button to enable or disable the output to the load.
5.2 Advanced Features
5.2.1 CC & CV Priority Function
The CC (Constant Current) and CV (Constant Voltage) priority function helps to solve severe problems in long-term testing. You can select the CV priority mode to obtain a faster voltage rising speed, or choose CC priority mode to output current without overshoot. This function is beneficial for applications such as laser testing, IC testing, charge and discharge testing, power transient simulation, and characterization of automotive electronics.
5.2.2 Web Server Access
The IT-M3100D series has a built-in Web server. You can monitor and control it through your web browser. After the IT-M3100D is connected via LAN interface, enter the IP address of the power supply in the browser to access the web interface. This allows remote control and configuration of LAN parameters.
5.2.3 Synchronism (Link) Function
The IT-M3100D supports synchronization functionality for both single and multi-channel power supply systems. There are three synchronization modes: On/Off, Track, and Duplicate. These modes allow parameters to be automatically replicated or proportionally synced to other power supplies in the link. The ON/OFF function utilizes the ON/OFF delay function in the menu for synchronous or sequential power-on operations.
5.2.4 Multi-Protection Functions
The IT-M3100D dual-channel DC power supply includes various protection functions to safeguard the device and the connected load:
- OVP (Over Voltage Protection): Protects the load from excessive voltage.
- OCP (Over Current Protection): Protects the load from excessive current.
- OPP (Over Power Protection): Protects the power supply from excessive power output.
- U-Max/U-Min (Under/Over Voltage Limit): Sets upper and lower voltage limits.
- Sense (Remote Sense): Compensates for voltage drop in the load wires.
- Foldback: The output turns off when the power CV/CC is switched, protecting sensitive loads from overshoot.
6. Maintenance
- Cleaning: Use a soft, dry cloth to clean the exterior of the unit. Do not use abrasive cleaners or solvents.
- Ventilation: Regularly check that the ventilation openings are clear of dust and debris to ensure proper airflow.
- Storage: When not in use for extended periods, store the power supply in a cool, dry environment, away from direct sunlight and extreme temperatures.
- Calibration: Periodic calibration by qualified service personnel is recommended to maintain accuracy.
7. Troubleshooting
If you encounter issues with your IT-M3100D series power supply, refer to the following common problems and solutions:
| Problem | Possible Cause | Solution |
|---|---|---|
| No power/Unit does not turn on | Power cord not connected, power outlet faulty, internal fuse blown. | Check power cord connection. Test outlet with another device. If fuse is suspected, contact support. |
| No output voltage/current | Output disabled, OVP/OCP/OPP triggered, load disconnected or faulty. | Ensure output is enabled. Check protection settings and reset if necessary. Verify load connection and integrity. |
| Unstable output | Poor load connection, remote sense not properly configured, external interference. | Check load wiring for loose connections. Ensure remote sense is correctly wired if used. Isolate from strong electromagnetic fields. |
| Overheating | Blocked ventilation, operating in high ambient temperature, excessive load. | Ensure ventilation openings are clear. Operate within specified ambient temperature range. Reduce load if continuously overheating. |
8. Specifications
The following table details the technical specifications for the IT-M3141D and IT-M3142D models. Specifications for other models in the IT-M3100D series may vary.
| Parameter | IT-M3141D (CH1 & CH2) | IT-M3142D (CH1 & CH2) |
|---|---|---|
| Rated Output Value | ||
| Voltage | 0~30V | 0~60V |
| Current | 0~15A | 0~10A |
| Power | 0~400W | 0~400W |
| Power Regulation | ||
| Voltage (±% of Output+Offset) | ≤0.02%+0.005%FS | ≤0.02%+0.005%FS |
| Current (±% of Output+Offset) | ≤0.01%+0.01%FS | ≤0.01%+0.01%FS |
| Load Regulation | ||
| Voltage (±% of Output+Offset) | ≤0.005%+0.005%FS | ≤0.005%+0.005%FS |
| Current | ≤1mV | ≤1mV |
| Setup Resolution | ||
| Voltage | 1mV | 1mV |
| Current | 1mA | 1mA |
| Readback Resolution | ||
| Voltage | 1mV | 1mV |
| Current | 1mA | 1mA |
| Setting Accuracy | ||
| Voltage | ≤0.02%+0.02%FS | ≤0.02%+0.02%FS |
| Current | ≤0.1%+0.1%FS | ≤0.1%+0.1%FS |
| Resistance | ≤1%+1%FS | ≤1%+1%FS |
| Readback Accuracy | ||
| Voltage | ≤0.03%+0.02%FS | ≤0.03%+0.02%FS |
| Current | ≤0.1%+0.1%FS | ≤0.1%+0.1%FS |
| Ripple (20Hz-20MHz) | ||
| Peak value | ≤60mVpp | ≤60mVpp |
| Ripple (20Hz-20MHz) Current(RMS) | ||
| Current | ≤15mA | ≤15mA |
| Temperature Coefficients | ||
| Setting Temperature | ≤0.005%+0.5mV | ≤0.005%+0.5mV |
| Readback Temperature | ≤0.005%+0.5mV | ≤0.005%+0.5mV |
| Falling Time (full load) | ||
| Voltage | ≤30ms | ≤30ms |
| Falling Time (no load) | ||
| Voltage | ≤30ms | ≤30ms |
| Dynamic Mode | ||
| Voltage | 100VAC~240VAC (rated power) | 100VAC~240VAC (rated power) |
| AC Input | ||
| Frequency | 50/60Hz | 50/60Hz |
| Setup Stability-30min | ≤0.01%+0.01%FS | ≤0.01%+0.01%FS |
| Efficiency | 85% | 85% |
| Remote Sense Compensation Voltage | ≤2V | ≤2V |
| Constant Response Time | 1ms | 1ms |
| Power Factor | 0.98 | 0.98 |
| Maximum Input Current | 10A | 10A |
| Maximum Input Apparent Power | 10VA | 10VA |
| Net Weight | (5±0.5) kg | (5±0.5) kg |
*Net Weight is information subject to change without notice.
9. Warranty and Support