1. Product Overview
The ASAIR ARD20 Series Digital Pressure Gauge is a versatile instrument designed for precise measurement of positive, negative, and vacuum pressure. It features a capacitance sensor, switching transducer output, and offers various output signals and connection types. This manual provides essential information for the safe and efficient operation of your device.

Figure 1: ASAIR ARD20 Series Digital Pressure Gauge with features: Positive & Negative / Vacuum Pressure, Accuracy: ±2%FS, Hysteresis/Window Simple, Switch/Analog signal output.
2. Safety Information
⚠ WARNING: This product can expose you to chemicals including lead, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov.
Always ensure the device is installed and operated by qualified personnel in accordance with local electrical codes and safety regulations. Disconnect power before performing any maintenance or wiring.
3. Product Specifications

Figure 2: Product dimensions and detailed specifications.
General Specifications:
- Theory: Capacitance Sensor
- Material: Polymer
- Type: Vacuum Sensor (also supports positive and negative pressure)
- Output: Switching Transducer (various options available)
- Use: Pressure Sensor
- High-concerned chemical: None
Available Models and Output Types:
| Pressure Range | Output Type | Model Number |
|---|---|---|
| Positive 0-250KPa | G1-8 2PNP 1-5V | 12000050482066396 |
| Positive 0-250KPa | G1-8 2PNP 4-20mA | 12000050482066397 |
| Positive 0-250KPa | G1-8 1NPN 485 | 12000050482066398 |
| Positive 0-1MPa | G1-8 2NPN 1-5V | 12000050482066399 |
| Positive 0-250KPa | G1-8 2NPN 1-5V | 12000050482066394 |
| Positive 0-250KPa | G1-8 2NPN 4-20mA | 12000050482066395 |
| Positive -0.1-1MPa | G1-8 2PNP 4-20mA | 12000052415260414 |
| Positive -0.1-1MPa | G1-8 1NPN 485 | 12000052415260415 |
| Positive -0.1-1MPa | G1-8 2NPN 4-20mA | 12000052415260412 |
| Positive -0.1-1MPa | G1-8 2PNP 1-5V | 12000052415260413 |
| Positive -0.1-1MPa | G1-8 2NPN 1-5V | 12000052415260411 |
| Positive -50-500KPa | G1-8 1NPN 485 | 12000052415260430 |
| Vacuum -101-0KPa | G1-8 2NPN 1-5V | 12000052415260431 |
| Positive -50-500KPa | G1-8 2PNP 1-5V | 12000052415260428 |
| Positive -50-500KPa | G1-8 2PNP 4-20mA | 12000052415260429 |
| Positive -50-500KPa | G1-8 2NPN 1-5V | 12000052415260426 |
| Positive -50-500KPa | G1-8 2NPN 4-20mA | 12000052415260427 |
| Positive 0-500KPa | G1-8 2PNP 4-20mA | 12000052415260424 |
| Positive 0-500KPa | G1-8 1NPN 485 | 12000052415260425 |
| Positive 0-500KPa | G1-8 2NPN 4-20mA | 12000052415260422 |
| Positive 0-500KPa | G1-8 2PNP 1-5V | 12000052415260423 |
| Positive 0-100KPa | G1-8 1NPN 485 | 12000052415260420 |
| Positive 0-500KPa | G1-8 2NPN 1-5V | 12000052415260421 |
| Positive 0-100KPa | G1-8 2PNP 1-5V | 12000052415260418 |
| Positive 0-100KPa | G1-8 2PNP 4-20mA | 12000052415260419 |
| Positive 0-100KPa | G1-8 2NPN 1-5V | 12000052415260416 |
| Positive 0-100KPa | G1-8 2NPN 4-20mA | 12000052415260417 |
| Vacuum -100-0KPa | G1-8 2NPN 1-5V | 12000050482066404 |
| Vacuum -100-0KPa | G1-8 2NPN 4-20mA | 12000050482066405 |
| Vacuum -100-0KPa | G1-8 2PNP 1-5V | 12000050482066406 |
| Vacuum -100-0KPa | G1-8 2PNP 4-20mA | 12000050482066407 |
| Positive 0-1MPa | G1-8 2NPN 4-20mA | 12000050482066400 |
| Positive 0-1MPa | G1-8 2PNP 1-5V | 12000050482066401 |
| Positive 0-1MPa | G1-8 2PNP 4-20mA | 12000050482066402 |
| Positive 0-1MPa | G1-8 1NPN 485 | 12000050482066403 |
| Mixed -100KPa-100KPa | G1-8 2PNP 4-20mA | 12000050482066412 |
| Mixed -100KPa-100KPa | G1-8 1NPN 485 | 12000050482066413 |
| Vacuum -100-0KPa | G1-8 1NPN 485 | 12000050482066408 |
| Vacuum -101-0KPa | G1-8 2PNP 4-20mA | 12000052415260434 |
| Mixed -100KPa-100KPa | G1-8 2NPN 1-5V | 12000050482066409 |
| Vacuum -101-0KPa | G1-8 1NPN 485 | 12000052415260435 |
| Mixed -100KPa-100KPa | G1-8 2NPN 4-20mA | 12000050482066410 |
| Vacuum -101-0KPa | G1-8 2NPN 4-20mA | 12000052415260432 |
| Mixed -100KPa-100KPa | G1-8 2PNP 1-5V | 12000050482066411 |
| Vacuum -101-0KPa | G1-8 2PNP 1-5V | 12000052415260433 |
4. Setup and Installation
4.1 Physical Installation
The ARD20 series pressure gauge is designed for easy installation. Ensure the mounting surface is stable and free from excessive vibration. Connect the pressure port securely to your system using the appropriate G1-8 fitting.

Figure 3: Installation method for the ARD20 Series Digital Pressure Gauge.
4.2 Electrical Wiring
Refer to the wiring diagram below for correct electrical connections. Ensure power is disconnected before wiring. Incorrect wiring can damage the device or lead to unsafe conditions.

Figure 4: Wiring Guide for the ARD20 Series. Consult this diagram for proper electrical connections based on your specific model and output type.
5. Operating Instructions
The ARD20 Series Digital Pressure Gauge features a user-friendly interface with three buttons: Up/Left key (▲), Confirm/Return key (S), and Down/Right key (▼).
5.1 Screen Button Description
Video 1: Tutorial for Screen Settings of the ARD20 Series Digital Pressure Gauge. This video demonstrates menu navigation and parameter adjustments.
- Up/Left Key (▲): Navigates up in menus or increases values.
- Confirm/Return Key (S): Confirms selections or returns to the previous menu.
- Down/Right Key (▼): Navigates down in menus or decreases values.
5.2 Main Menu Navigation
- Enter Main Menu: Press and hold the "S" setting key for 1 second.
- Navigate: Use the "▲" and "▼" keys to move between options (Main Menu, Param Setting, System Setting, Exit).
- Confirm/Enter: Press "S" to confirm a selection or enter a submenu.
- Exit: Select "Exit" and press "S" to return to the previous menu or main display.
5.3 Parameter Setting (Param Setting)
From the Main Menu, select "Param Setting" and press "S".
5.3.1 Channel 1/2 Setting
The setting steps for Channel 2 are identical to Channel 1.
- Select Channel: Choose "Channel 1 Set" or "Channel 2 Set" and press "S".
- Mode Setting:
- Select "Mode" and press "S".
- The system offers three pressure measurement modes: Simple, Hysteresis, and Window modes. The default is "Simple Mode".
- Use "▲" and "▼" to select your desired mode. Press "S" to confirm.
- Select "Exit" and press "S" to return to Channel 1/2 Setting.
- Threshold Value (Simple Mode):
- Select "Threshold Value" and press "S".
- Use "▲" and "▼" to adjust the set value (P). Adjustment accuracy is ±0.1kPa.
- Press "S" to confirm and return.
- Note: Peak and Valley values are not adjusted in Simple Mode.
- Peak Value (Hysteresis/Window Mode):
- Select "Peak Value" and press "S".
- Use "▲" and "▼" to adjust the peak value.
- Press "S" to confirm and return.
- Valley Value (Hysteresis/Window Mode):
- Select "Valley Value" and press "S".
- Use "▲" and "▼" to adjust the valley value.
- Press "S" to confirm and return.
- Hysteresis Value (Hysteresis Mode):
- Select "Hysteresis Value" and press "S".
- Use "▲" and "▼" to adjust the hysteresis value.
- Press "S" to confirm and return.
- Response Time:
- Select "Response Time" and press "S".
- Use "▲" and "▼" to adjust the response time.
- Press "S" to confirm and return.
- Output Mode:
- Select "Output Mode" and press "S".
- Choose between "Normally Open" or "Normally Close".
- Press "S" to confirm and return.
- Note: When output is ON: If set to normally open, output is ON. If set to normally closed, output is OFF. When output is OFF: If set to normally open, output is OFF. If set to normally closed, output is ON.
5.4 Unit Setting
From "Param Setting", select "Unit Setting" and press "S".
- Use "▲" and "▼" to select between "kPa" and "MPa".
- Press "S" to confirm. The system defaults to "kPa".
5.5 Zero Calibration and Reset
From "Param Setting", select "Zero Calibration" and press "S" to confirm. The device will perform a zero calibration.
To reset the zero pressure to factory standard, select "Zero Reset" and press "S". A "Setting Successful" message will confirm the reset.
5.6 System Setting
From the Main Menu, select "System Setting" and press "S".
5.6.1 Language Setting
Select "Language" and press "S".
- Two languages are available: Chinese and English.
- Press "S" to switch languages. The default language is Chinese.
5.6.2 Output Test
Select "Output Test" and press "S".
By turning on or off the OUT1/2 output ports, you can detect numerical changes to determine if the OUT1/2 output ports are normal. The test for Channel 2 is similar to Channel 1.
5.6.3 Font Colour
Select "Font Colour" and press "S".
- Six font colors are available: white, blue, orange, yellow, green, and purple.
- The default font color is green. Select a color and press "S" to confirm and return.
5.6.4 Factory Reset
Select "Factory Reset" and press "S".
- To start the factory reset process for all setting parameters, press the "▲" up key. This will restore the device to its factory default state.
- To cancel this operation, press the "▼" down key.
5.6.5 Version Information
Select "Version" and press "S" to view the software and hardware version information of the system. To exit this interface, press and hold the "S" setting key for 1 second.
6. Maintenance
- Cleaning: Use a soft, dry cloth to clean the device. Do not use abrasive cleaners or solvents.
- Inspection: Regularly inspect the pressure port and electrical connections for any signs of wear, damage, or corrosion.
- Calibration: If you suspect inaccurate readings, perform a "Zero Calibration" as described in section 5.5. For advanced calibration, consult a qualified technician.
- Storage: When not in use for extended periods, store the device in a dry, dust-free environment within its specified operating temperature range.
7. Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| No display/Device not powering on | No power supply or incorrect wiring. | Check power connections and ensure correct voltage. Refer to Section 4.2. |
| Inaccurate pressure readings | Sensor drift, incorrect unit setting, or environmental factors. | Perform "Zero Calibration" (Section 5.5). Verify unit setting (Section 5.4). Ensure stable operating conditions. |
| Output not functioning correctly | Incorrect output mode setting or faulty output port. | Check "Output Mode" (Section 5.3.1). Use "Output Test" (Section 5.6.2) to diagnose. Verify wiring. |
| Device unresponsive | Temporary software glitch. | Perform a "Factory Reset" (Section 5.6.4) as a last resort. |
8. User Tips
- Negative and Positive Scale: The device is capable of measuring both negative (vacuum) and positive pressure. Ensure your selected model supports the full range required for your application. The "Param Setting" menu allows for specific threshold and valley value adjustments for different pressure ranges.
- Language Setting: If the display language is not English, navigate to "System Setting" -> "Language" and press 'S' to switch to English.
9. Warranty and Support
For warranty information, technical support, or service inquiries, please contact ASAIR customer service or your local distributor. Keep your purchase receipt as proof of purchase.
Contact Information:
- For technical support, contact support@aosong.com.
- For customized service (e.g., cable length, connector, OEM, ODM), contact sales@aosong.com.