Hti HT-1891

Contents

HTI HT-1891 Handheld Digital Manometer Air Pressure Meter

Model: HT-1891 | Brand: Hti

Introduction

The HTI HT-1891 Digital Manometer is a professional instrument designed for measuring gauge and differential pressure. It offers high accuracy and a wide range of applications, including HVAC systems, clean rooms, furnace and boiler maintenance, and air conditioning systems. This manual provides detailed instructions for the safe and effective use of your manometer.

HTI HT-1891 Manometer showing various applications like HVAC, clean rooms, furnace, and air conditioning.
Figure 1: HTI HT-1891 Manometer in various application scenarios.

Package Contents

Please check the package contents upon receipt to ensure all items are present and undamaged:

  • 1 x HTI HT-1891 Digital Manometer
  • 2 x Connection Hoses
  • 1 x Quick Connector
  • 1 x 9V Battery
  • 1 x Carrying Case
  • 1 x User Manual (this document)
Full package contents including the manometer, battery, manual, quick connector, hoses, and carrying case.
Figure 2: Contents of the HTI HT-1891 Manometer package.

Product Overview

The HT-1891 features a robust design with a clear dual LCD display and intuitive controls.

Front and back view of the HTI HT-1891 Manometer, showing buttons and battery compartment.
Figure 3: Front and back view of the HTI HT-1891 Digital Manometer.

Key Components:

  • Dual LCD Display: Shows current pressure readings and other information.
  • UNITS Button: Cycles through 11 selectable pressure units.
  • HOLD Button: Freezes the current reading on the display.
  • DIF Button: Activates differential pressure measurement mode.
  • Power Button: Turns the device ON/OFF.
  • Backlight Button: Activates the LCD backlight for low-light conditions.
  • REC Button: Records Max/Min/Average readings.
  • Pressure Ports: Two ports for connecting hoses for differential pressure measurement.
Front and side view of the HTI HT-1891 Manometer with dimensions (170mm length, 54.3mm width, 34mm depth).
Figure 4: Dimensions of the HTI HT-1891 Manometer.

Specifications

FunctionRangeResolutionFunctionRangeResolution
InH2O±138.30.1Ozin2±80.001
Psi±5.0000.001ftH2O±11.530.01
Mbar±344.70.1cmH2O±351.50.1
Kpa±34.470.001Kgcm2±0.3510.001
inHg±1.0180.001Bar±0.3440.001
mmHg±258.50.1
  • Display: Dual LCD
  • Accuracy: 0.3% FSO (25°C)
  • Repeatability: 0.2% (Max 0.5% FSO)
  • Linearity/Hysteresis: 0.29% FSO
  • Maximum Pressure: 10 PSI (Note: This specification might refer to a similar model, HT-1890, as indicated in the source data. Please verify with product labeling.)
  • Response Time: 0.5s Typical
  • Low Range Indicator: Err.1
  • Under Range Indicator: Err.2
  • Operating Conditions: 32°F to 122°F (0°C to 50°C)
Detailed specifications table for the HTI HT-1891 manometer.
Figure 5: Detailed specifications of the HTI HT-1891 Manometer.

Setup

  1. Battery Installation: Open the battery compartment on the back of the device. Insert the included 9V battery, ensuring correct polarity. Close the compartment securely.
  2. Connecting Hoses: For differential pressure measurements, connect the two provided hoses to the pressure ports on top of the manometer. Ensure a snug fit to prevent leaks.
  3. Power On: Press the red power button to turn on the device. The dual LCD display will illuminate.
  4. Unit Selection: Press the 'UNITS' button to cycle through the 11 available measurement units (in H2O, psi, bar, mbar, Kpa, inHg, mmHg, ozin2, ftH2O, kgcm2, Bar) until your desired unit is displayed.

Operating Instructions

1. Single Point Pressure Measurement:

  1. Ensure only one hose is connected to a pressure port, or measure ambient pressure directly.
  2. Turn on the device and select the desired unit.
  3. The display will show the current pressure reading.

2. Differential Pressure Measurement:

  1. Connect both hoses to the pressure ports.
  2. Connect the other ends of the hoses to the two points where differential pressure needs to be measured.
  3. Press the 'DIF' button to activate differential pressure mode. The display will show the difference between the two pressure inputs.

3. Data Hold:

  • Press the 'HOLD' button to freeze the current reading on the display.
  • Press 'HOLD' again to release and resume live readings.

4. Max/Min/Average Recording:

  • Press the 'REC' button once to start recording maximum, minimum, and average values.
  • Press 'REC' repeatedly to cycle through MAX, MIN, and AVG readings.
  • Press and hold 'REC' to exit recording mode.

5. Backlight:

  • Press the backlight button (usually indicated by a light bulb icon) to turn the LCD backlight on or off.
HTI HT-1891 Manometer with its LCD backlight activated.
Figure 6: Manometer display with backlight for improved visibility.

6. Zeroing and Data Compensation:

  • The device supports zeroing, offset, and data compensation functions for accurate measurements. Refer to the on-screen prompts or specific sections in the full manual for detailed procedures on these advanced features.

Maintenance

  • Cleaning: Use a soft, dry cloth to clean the device. Do not use abrasive cleaners or solvents.
  • Battery Replacement: When the low battery indicator appears on the display, replace the 9V battery promptly to ensure accurate readings.
  • Storage: Store the manometer in its carrying case in a cool, dry place when not in use. Avoid extreme temperatures and humidity.
  • Hose Care: Inspect the connection hoses regularly for cracks or damage. Replace them if necessary to maintain measurement accuracy.

Troubleshooting

  • Device does not power on: Check if the 9V battery is correctly installed and has sufficient charge. Replace if necessary.
  • Inaccurate readings: Ensure hoses are securely connected and not kinked. Perform zeroing/calibration if available (refer to advanced features). Check for environmental factors affecting pressure.
  • 'Err.1' (Low Range Indicator): The measured pressure is below the device's measurable range.
  • 'Err.2' (Under Range Indicator): The measured pressure is significantly below the device's measurable range, possibly indicating a vacuum or a faulty connection.
  • Display is dim: Activate the backlight using the backlight button. If still dim, replace the battery.

User Tips

  • Always perform a zero adjustment before critical measurements, especially for differential pressure, to ensure baseline accuracy.
  • When measuring differential pressure, ensure both ports are exposed to the same ambient pressure before connecting to the system to establish a true zero reference.
  • For prolonged storage, remove the battery to prevent leakage and damage to the device.
  • Keep the pressure ports clean and free from debris to avoid blockages and inaccurate readings.

Safety Information

  • Do not exceed the maximum pressure rating of the device (10 PSI for HT-1890, verify for HT-1891). Over-pressurization can damage the sensor and void the warranty.
  • This device is not a personal safety device. Do not use it for applications where personal safety depends on its accuracy.
  • Avoid dropping the device or subjecting it to strong impacts.
  • Do not expose the device to corrosive gases or liquids.
  • Keep out of reach of children.
  • Dispose of batteries according to local regulations.

Manufacturing and Quality Assurance

The HTI HT-1891 Manometer is manufactured under strict quality control processes to ensure reliability and performance. The following video provides insight into the production and testing facilities:

Video 1: Overview of HTI's manufacturing and quality assurance processes.

Warranty and Support

This HTI product is covered by a standard manufacturer's warranty against defects in materials and workmanship. For warranty claims or technical support, please contact your retailer or the manufacturer directly. Keep your purchase receipt as proof of purchase.


Documents and downloads

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.