1. Introduction
This manual provides instructions for the safe and effective use of the ZOYI P2060 BNC Oscilloscope Probe Kit. The P2060 is a high-precision probe designed for use with oscilloscopes, featuring integrated injection molding for durability and reliable signal transmission. It supports 1X and 10X attenuation ratios and is suitable for various electrical measurement tasks.

2. Package Contents
Verify that all items are present and in good condition upon unpacking:
- P2060 Oscilloscope Probe
- Grounding Alligator Clip
- Probe Hook
- Adjustment Tool
- Color Rings (for channel identification)

3. Setup Instructions
Connecting the Probe to the Oscilloscope
Connect the BNC connector of the P2060 probe to the input channel of your oscilloscope. Ensure a secure connection to reduce signal interference and ensure high-frequency signal transmission.

Figure 3: BNC Interface Connection Attaching Probe Accessories
Attach the desired probe tip (e.g., probe hook) and the grounding alligator clip to the probe body. The grounding clip should be connected to the circuit's ground reference.

Figure 4: Probe Components and Features
4. Operating Instructions
Selecting Attenuation Ratio (1X/10X)
The P2060 probe features a sliding switch on its body to select between 1X and 10X attenuation. The 1X setting provides a direct signal, while the 10X setting attenuates the signal by a factor of 10, allowing for measurement of higher voltages and reducing probe loading effects. Ensure your oscilloscope's channel setting matches the probe's attenuation ratio (e.g., if the probe is set to 10X, the oscilloscope input should also be set to 10X).
Capacitance Compensation Adjustment
To ensure accurate waveform reproduction, the probe's capacitance compensation should be adjusted to match the input capacitance of your oscilloscope. Use the provided adjustment tool to turn the screw on the probe's BNC connector end while observing a square wave signal (typically from the oscilloscope's probe compensation output). Adjust until the square wave appears flat-topped, without overshoot or undershoot.

Figure 5: Capacitance Compensation Adjustment Making Measurements
Once connected and compensated, touch the probe tip to the test point in your circuit and ensure the ground clip is securely connected to the circuit's ground. Observe the waveform on your oscilloscope.
5. Specifications
The following table details the technical characteristics of the P2060 Oscilloscope Probe:

| Characteristic | Value |
|---|---|
| Model | P2060 |
| Attenuation Ratio | 1X & 10X |
| Bandwidth | 60MHz |
| Rise time | 5.8ns |
| Input Resistance | 1MΩ/10MΩ ±2% |
| Input Capacitance | 10X: 14pF~18pF |
| Max. Input Voltage | 1X: 150V RMS CAT II 10X: 300V RMS CAT II |
| Compensation Range | 15pF~45pF |
| Operation Temperature | -10°C ~ +55°C |
| Humidity | -40°C or below ≤90% +41°C to +50°C ≤60% |
| Altitude | Operating 3000m Nonoperating 15000m |
| Cable Length | 110 ±2cm |
| Weight | About 55g |
| Safety | IEC61010-031 |
6. Maintenance
Cleaning
Gently wipe the probe body and cable with a soft, dry cloth. For stubborn dirt, a slightly damp cloth with mild soap can be used, ensuring no moisture enters the BNC connector or switch mechanism. Do not use abrasive cleaners or solvents.
Storage
Store the probe in a clean, dry environment, away from direct sunlight and extreme temperatures. Avoid bending the cable sharply or placing heavy objects on it to prevent damage.
7. Troubleshooting
No Signal on Oscilloscope
Ensure the BNC connector is securely attached to the oscilloscope. Check that the probe tip is making good contact with the test point and the ground clip is properly connected to the circuit's ground. Verify the oscilloscope channel is enabled and correctly configured.
Distorted or Incorrect Waveform
Perform capacitance compensation adjustment as described in Section 4.2. Ensure the probe's attenuation switch (1X/10X) matches the setting on your oscilloscope. Check for damaged cables or connectors.
Incorrect Amplitude Reading
Confirm that the oscilloscope's vertical scale (Volts/Div) is correctly set, taking into account the probe's attenuation ratio. For example, if the probe is set to 10X, the actual voltage is 10 times the reading on the oscilloscope screen (or the oscilloscope should automatically adjust if its probe setting is correct).
8. User Tips
- Always connect the ground clip to the circuit's ground before connecting the probe tip to a live signal.
- When measuring high-frequency signals, use the 10X attenuation setting to minimize probe loading and improve signal integrity.
- Regularly check and adjust capacitance compensation, especially when switching between different oscilloscopes or after long periods of storage.
- Utilize the color rings to easily identify different probes when using multiple channels on your oscilloscope.
9. Warranty and Support
For warranty information, technical support, or service inquiries, please contact the seller or manufacturer directly. Keep your purchase receipt as proof of purchase.