1. Introduction
The KAIWEETS KC602 is a smart digital clamp meter designed for accurate and versatile electrical measurements. Featuring 6000 counts True-RMS, auto-ranging capabilities, and a range of functions including AC/DC current, AC/DC voltage, resistance, continuity, capacitance, diode, NCV detection, inrush current, and temperature, it is an essential tool for electricians, technicians, and DIY enthusiasts. Its intelligent mode simplifies operation by automatically identifying measurement signals.
2. Safety Information
WARNING: To avoid possible electric shock, fire, or personal injury, please read all safety information before you use the product.
- Always ensure the meter is in the correct function and range before making measurements.
- Do not apply more than the rated voltage, as marked on the meter, between the terminals or between any terminal and earth ground.
- Use caution when working with voltages above 30V AC RMS, 42V peak, or 60V DC. Such voltages pose a shock hazard.
- Before measuring current, ensure the circuit is de-energized and the clamp jaw is properly closed around a single conductor.
- Replace the batteries immediately when the low battery indicator appears to ensure accurate readings.
- Do not operate the meter if it appears damaged or if the case is open.
- Adhere to local and national safety codes.
3. Package Contents
Verify that all items are present in your package:
- 1 x KAIWEETS KC602 Clamp Meter
- 2 x Test Leads (Red and Black)
- 1 x Storage Bag
- 3 x AAA Batteries (included)
- 1 x User Manual (this document)

4. Product Features
- Upgraded Intelligent Mode: Automatically identifies AC/DC voltage/current, resistance, and continuity measurement signals for simplified operation.
- Multifunctions: Measures AC/DC current/voltage, resistance, continuity, capacitance, diode, NCV, firewire, and temperature.
- Large HD Color Screen: Provides clear readings, even in dimly lit environments, and is easier to read than traditional meters.
- Self-test Function: Automatically performs a self-test and displays "CAL" upon power-on to ensure measurement accuracy.
- Inrush Measurement: Measures inrush current (5-600A) to help diagnose startup problems in motors or circuits (AC current mode only).
- D-shaped Jaws: Designed to securely hold test leads and enhance portability.
- Additional Features: MAX/MIN value hold, data hold, flashlight, and auto-off (after 15 minutes of inactivity).
- Safety Rating: CAT.IV 600V.


5. Specifications
| Category | Description/Value |
|---|---|
| Brand | KAIWEETS |
| Type | Clamp Meter |
| Model | KC602 |
| Color | Red |
| Maximum Display | 6000 counts |
| Batteries | 3 AAA batteries required (included) |
| DC Voltage | 600mV/6V/60V/600V |
| AC Voltage | 600mV/6V/60V/600V |
| DC Current | 60A/600A |
| AC Current | 60A/600A |
| Resistance | 600Ω/6kΩ/60kΩ/600kΩ/6MΩ/60MΩ |
| Capacitance | 6nF/60nF/600nF/6uF/60uF/600uF/6mF/60mF |
| Frequency | 100Hz/1kHz/10kHz/100kHz/1MHz/10MHz |
| Temperature | Yes |
| NCV (Non-Contact Voltage) | Yes |
| Flashlight/Backlight | Yes |
| Color Screen | Yes |
| Product Weight | 0.52kg |
| Package Weight | 0.9kg |
| Package Size (L x W x H) | 22.5 x 11.5 x 5 cm |
6. Setup
6.1. Battery Installation
- Locate the battery compartment cover on the back of the meter.
- Use a screwdriver to loosen the screw(s) and remove the cover.
- Insert three (3) AAA batteries, observing the correct polarity (+/-) as indicated inside the compartment.
- Replace the battery cover and secure it with the screw(s).
6.2. Connecting Test Leads
- Insert the black test lead into the "COM" (common) input jack.
- Insert the red test lead into the "INPUT" jack for voltage, resistance, capacitance, frequency, diode, and continuity measurements.
- Ensure the connections are firm before use.

7. Operating Instructions
7.1. Power On/Off
Press the red power button to turn the meter ON or OFF. The meter will perform a self-test and display "CAL" upon power-on.

7.2. Smart Mode (Auto-recognition)
When the meter is turned on, it defaults to intelligent mode. In this mode, the meter automatically detects and measures AC/DC voltage, AC/DC current (via clamp), resistance, and continuity without manual function selection.

7.3. Function Selection (Manual Mode)
Press the "FUNC" button to cycle through different measurement functions if you need to manually select a specific mode not automatically detected by Smart Mode, or to switch between AC/DC for certain measurements.
7.4. AC/DC Voltage Test
- Connect the test leads to the circuit in parallel.
- The meter will automatically detect AC or DC voltage in Smart Mode. If in manual mode, select the appropriate voltage function.
- Read the voltage value on the display.

7.5. AC/DC Current Test (Clamp Function)
- Ensure the circuit is de-energized if possible, or exercise extreme caution.
- Open the clamp jaw and enclose only ONE conductor of the circuit.
- Close the clamp jaw completely.
- The meter will display the AC or DC current.

7.6. Resistance, Continuity, Capacitance, Diode Test
- Ensure the circuit or component under test is completely de-energized before connecting the test leads.
- Select the desired function (Resistance, Continuity, Capacitance, or Diode) using the "FUNC" button if not in Smart Mode.
- Connect the test leads across the component.
- Read the measurement on the display. For continuity, an audible beep indicates a continuous circuit.

7.7. NCV (Non-Contact Voltage) Detection
- Select the NCV function.
- Bring the top of the clamp meter near the conductor or outlet.
- The meter will indicate the presence of AC voltage through visual and/or audible alerts.

7.8. Inrush Current Test
This function is available only in AC current mode.
- Select the AC current mode.
- Press the "INRUSH" button (if available, or cycle through functions until Inrush is selected).
- Clamp around the single conductor of the device you wish to test during its startup phase.
- The meter will capture and display the peak inrush current.

7.9. Temperature Measurement
If your model includes a temperature probe, connect it to the appropriate input jacks and select the temperature function to measure ambient or surface temperature.
7.10. Flashlight and Backlight
The meter features a built-in flashlight and backlight for the display, making it easy to work in dimly lit areas. Refer to the meter's buttons for activation (often a dedicated button or a long press of another function button).

7.11. Auto Shut-off
To conserve battery life, the meter will automatically power off after approximately 15 minutes of inactivity. Press any button or cycle the power to turn it back on.

8. Maintenance
8.1. Cleaning
Wipe the meter's case with a damp cloth and mild detergent. Do not use abrasives or solvents. Keep the input terminals free of dirt and moisture.
8.2. Battery Replacement
When the low battery indicator appears on the display, replace the batteries as described in Section 6.1. Remove batteries if the meter is not used for an extended period.
8.3. Storage
Store the meter in a cool, dry place, away from direct sunlight and extreme temperatures. Use the provided storage bag to protect it from dust and physical damage.
9. Troubleshooting
- Meter does not power on: Check battery installation and ensure batteries are not depleted. Replace if necessary.
- Inaccurate readings: Ensure test leads are properly connected. Check battery level. Verify the correct function is selected (if not in Smart Mode). Ensure the clamp jaw is fully closed around a single conductor for current measurements.
- No display/flickering display: Low battery. Replace batteries.
- "OL" or "OVER" displayed: The measured value exceeds the selected range. Switch to a higher range if available, or verify the circuit is within the meter's capabilities.
- No NCV detection: Ensure the NCV function is active and the sensor is close enough to the AC voltage source.
10. Warranty & Support
For warranty information, technical support, or service inquiries, please refer to the contact information provided by your retailer or visit the official KAIWEETS website. Keep your purchase receipt as proof of purchase.