1. Introduction
This manual provides essential information for the proper installation, operation, and maintenance of your Industrial pH Sensor (S-pH-01B). This sensor is designed for measuring hydrogen-ion activity in water-based solutions, indicating acidity or basicity (pH level). It is suitable for various applications including soil and water quality monitoring.
The S-pH-01B features MODBUS-RTU RS485 and 0-2V Analog Voltage output, an IP68-rated electrode, and an IP65-rated transmitter, ensuring robust performance in professional and demanding environments.
Key Features:
- High Performance: Offers high accuracy, fast response, and good repeatability.
- Universal Protocol: Supports MODBUS-RTU RS485 and 0~2V Analog Voltage output.
- Wide-range Power Supply: Operates with a supply voltage between 3.6V and 30V.
- Strong Robustness: Designed for long-term immersion in soil or water, with anti-corrosion properties.
- Maintenance-Free Design: Features a unique junction that prevents clogging, simplifying maintenance.
2. Safety Information
Warning: Children need to use this product under adult supervision.
- Do not use in humid or high-temperature environments beyond specified operating ranges.
- Do not short-circuit the power interface.
- Keep away from children to prevent accidental ingestion or damage.
- Do not disassemble the product without authorization.
- Use only appropriate power supply and compatible accessories.
- Slight heating during operation is normal and not a cause for concern.
3. Package Contents
- Industrial pH Sensor with waterproof aviation connector: 1 piece
- Basic accessories (if applicable, actual contents may vary depending on the model)
4. Setup and Installation
4.1. Product Overview
4.2. Wiring Diagram
Connect the sensor according to the following wiring diagram for RS485 communication:
4.3. System Integration
The pH sensor can be integrated with a Sensor Hub (4G Data Logger) and pH Transmitter for comprehensive data collection and management, as shown in the system diagram below.
4.4. General Installation
- Electrode: Utilizes 3/4"NPT screw threads for secure mounting.
- Transmitter: Designed for mounting with appropriate holes.
- Ensure all connections are secure and waterproof, especially the aviation connector.
- For detailed technical information and specific installation procedures, refer to the official website or product page.
5. Operating Instructions
- Pre-use Check: Please check the product thoroughly before use for any visible damage or defects.
- Connection: Connect the sensor according to the provided wiring diagram and normal electronic operation procedures.
- Power Supply: Use a suitable power supply (3.6V ~ 30V DC) and compatible accessories.
- Data Monitoring: Utilize the SenseCAP Portal and SenseCAP Mate for remote data management and monitoring.
5.1. Calibration
The device is factory calibrated prior to shipment. For re-calibration, please refer to the detailed calibration document:
6. Maintenance
6.1. Handling Crystals on Probe
If you observe white solid granular crystals around the probe shroud, this is a normal occurrence caused by temperature changes, environmental factors, or redox processes during delivery. It does not affect the internal solution or the sensor's normal operation.
To clean, gently wipe off the crystals with a wet paper towel. Ensure children do not have access to the crystals to avoid accidental ingestion.
6.2. General Care
- The sensor is designed to be maintenance-free with its unique junction preventing clogging.
- Regularly inspect cables and connectors for any signs of wear or damage.
- Keep the sensor clean from excessive debris or buildup that might affect readings.
7. Troubleshooting
7.1. pH Measurement Issues
- Q: Is the sensor factory calibrated?
- A: Yes, the device is fully calibrated prior to shipment from the factory. For re-calibration, refer to the PH Calibration by Modbus Command document.
7.2. Temperature Output Inaccuracy
- Q: Why is the temperature output from the sensor inaccurate?
- A: The pH electrode itself does not contain a temperature sensor; the temperature reading is from a plate temperature sensor. If precise temperature sensing is required, you can integrate an NTC (10K, 3950) thermistor temperature sensor.
7.3. Soil Measurement
- Q: Can the sensor be used in soil? How to install?
- A: Yes, the sensor can be used in soil. The principle involves measuring pH in soil moisture, so readings will be affected by soil moisture content; drier soil may lead to greater error. For installation in soil: Dig a pit, fill it with some water, stir the soil into a slurry, and then carefully place the sensor glass probe. Be cautious not to apply excessive force and avoid hard objects like stones, which could damage the probe.
8. Specifications
8.1. pH Measurement Specifications
| Parameter | Value |
|---|---|
| Range | 0 ~ 14pH |
| Accuracy | ± 0.1 pH |
| Resolution | 0.01 pH |
8.2. General Parameters
| Parameter | S-pH-01A (Analog Voltage) | S-pH-01B (RS485) |
|---|---|---|
| Product Model | S-pH-01A | S-pH-01B |
| Output Interface | Analog Voltage 0-2V (Output resistance ~0ohm) | RS-485 MODBUS-RTU |
| Power Supply | 3.9-30V/DC | 3.9-30V/DC |
| Power Consumption | 35mA@24V DC | 35mA@24V DC |
| Start-up time | < 2 seconds | |
| pH Measurement |
High impedance and isolated input; Range: 0 - 14PH; Resolution: 0.01PH; Accuracy: +/- 0.1PH; Can be used for a solution or semi-solid substrate |
|
| Temperature Measurement |
Range: -40~80°C; Resolution: 0.1°C; Accuracy: +/-0.5°C |
|
| IP Rating |
Electrode: IP68; Transmitter: IP65 |
|
| Operating Temperature | -40℃ ~ 85℃ | |
| Installation |
Electrode: 3/4"NPT screw threads; Transmitter: Mounting hole |
|
| Cable Length |
Power and Signal Cable: 2 meters (Customizable) Electrode Cable: 5 meters |
|
| Dimension |
Electrode: Width*Diameter 160*30mm; 3/4"NPT screw threads Transmitter: 140mm*65mm*50mm |
|
Note: Version 1 of this pH Meter/Sensor supported RS485 protocol and 4~20mA current (S-pH-01A), which has been replaced by Version 2 (S-pH-01B) supporting RS485 protocol and 0~2V analog voltage.
9. User Tips
- When measuring pH in soil, ensure the soil is adequately moist to minimize measurement errors. Creating a slurry can improve accuracy.
- For applications requiring precise temperature compensation, consider integrating an external NTC thermistor as the internal temperature sensor is for the plate, not the pH electrode itself.
- Regularly check the official documentation for updated calibration procedures or technical notes.
10. Storage Conditions
- Store the product in a dry and well-ventilated place.
- Avoid direct sunlight and moisture.
11. Warranty and Support
This product is covered by a limited warranty against manufacturing defects. Damage caused by misuse, modification, or improper operation is not covered.
For support or detailed technical information, please refer to the official website or product page, or contact the official store directly.
Additional documents and guides are available: