Hantek DSO5072P, DSO5102P, DSO5202P

Contents

Hantek DSO5072P/DSO5102P/DSO5202P Digital Storage Oscilloscope User Manual

Portable 2-Channel Digital Storage Oscilloscope with 70MHz, 100MHz, or 200MHz Bandwidth and 1GSa/s Sample Rate

1. Introduction

This manual provides essential information for the safe and effective operation of your Hantek DSO5072P, DSO5102P, or DSO5202P Digital Storage Oscilloscope. Please read this manual thoroughly before using the device and keep it for future reference.

Hantek DSO5072P, DSO5102P, DSO5202P Oscilloscope front view with model numbers highlighted
Figure 1: Hantek Digital Storage Oscilloscope models

2. Safety Information

WARNING: HAZARDOUS VOLTAGE INSIDE. DO NOT REMOVE THE COVER UNLESS BY SPECIFIED PERSONNEL.

Always ensure the device is connected to a power source within the specified voltage and frequency range (AC 100-240V, 45-440Hz, 30W MAX). Use only the provided power cable and accessories. Do not operate the device in wet conditions or near flammable materials. Refer all servicing to qualified personnel.

Rear panel of the oscilloscope showing power input and warning label
Figure 2: Rear panel with power input and safety warning

3. Key Features

  • Bandwidth Options: 70MHz (DSO5072P), 100MHz (DSO5102P), 200MHz (DSO5202P)
  • Channels: 2 Digital Channels
  • Real-Time Sample Rate: 1GSa/s
  • Display: Large 7.0-inch color WVGA (800x480) TFT LCD
  • Record Length: Up to 40K points
  • Trigger Modes: Edge, Pulse Width, Line Selectable Video, Slope, Overtime, etc.
  • Connectivity: USB Host and Device for PC connection and data storage
  • Measurements: Multiple automatic measurements
  • Math Functions: Four math functions, including FFT (Fast Fourier Transform)
  • PC Software: Provides software for PC real-time analysis and control

4. Package Contents

Verify that all items listed below are included in your package:

  • Hantek Digital Storage Oscilloscope Unit (DSO5072P, DSO5102P, or DSO5202P)
  • 2 x Oscilloscope Probes
  • 1 x USB Cable
  • 1 x Power Cable
  • 1 x CD (containing PC software and electronic manual)
  • 1 x Printed Quick Start Guide/Manual
  • Small accessories (e.g., probe color rings, adjustment tool)
Image showing all items included in the oscilloscope package: oscilloscope unit, two probes, USB cable, power cable, CD, and printed manual.
Figure 3: Standard package contents

5. Setup

5.1 Physical Placement

Place the oscilloscope on a stable, flat surface. Ensure adequate ventilation around the unit. The device features adjustable feet for optimal viewing angle.

Diagram showing front, rear, and side views of the oscilloscope, highlighting physical features and adjustable feet.
Figure 4: Oscilloscope physical views

5.2 Power Connection

  1. Connect the provided power cable to the AC input on the rear panel of the oscilloscope.
  2. Plug the other end of the power cable into a suitable AC power outlet (100-240V, 45-440Hz).

5.3 Probe Connection

  1. Connect the BNC connector of the oscilloscope probe to either the CH1 or CH2 input on the front panel.
  2. Ensure the probe is securely twisted into place.
  3. Attach the ground clip of the probe to the circuit's ground.
  4. Connect the probe tip to the test point on your circuit.

5.4 Initial Power-On

Press the power button located on the front panel to turn on the oscilloscope. The device will perform a self-test and display the main waveform interface.

6. Operation

6.1 Front Panel Overview

Front view of the Hantek DSO5102P oscilloscope, showing the screen, control knobs, and buttons.
Figure 5: Front Panel Controls

The front panel features a 7-inch display, various control knobs (Vertical, Horizontal, Trigger), and function buttons (Menu, Measure, Acquire, Display, Utility, Cursor, Save/Recall, Auto Set, Run/Stop, Single Seq, Trigger Level).

6.2 Basic Waveform Display

After connecting a probe to a signal source, press the AUTO SET button for automatic waveform optimization. Adjust the VERTICAL POSITION and VOLTS/DIV knobs to scale the vertical axis, and the HORIZONTAL POSITION and SEC/DIV knobs to scale the horizontal axis.

Oscilloscope display showing a clear waveform
Figure 6: Example Waveform Display

6.3 Trigger System

The trigger system stabilizes repetitive waveforms and captures single-shot events. Use the TRIGGER LEVEL knob to adjust the trigger threshold. Press the TRIGGER MENU button to access various trigger types:

  • Edge Trigger: Triggers on the rising or falling edge of a signal.
  • Pulse Width Trigger: Triggers on pulses of a specified width.
  • Video Trigger: Triggers on standard video signals.
  • Slope Trigger: Triggers on a specific slope of the waveform.
  • Overtime Trigger: Triggers when a signal remains high or low for a specified duration.

6.4 Measurements

The oscilloscope supports multiple automatic measurements and cursor measurements.

  • Automatic Measurements: Press the MEASURE button to display a menu of automatic measurements such as Frequency, Period, Max, Min, Peak-to-Peak Voltage (Pk-Pk), Rise Time, Fall Time, etc.
  • Cursor Measurements: Press the CURSOR button to activate cursors. Use the multi-function knob to position the cursors and read voltage (ΔV) and time (ΔT) differences directly on the screen.
Multiple oscilloscope screens showing various waveform types and measurement results.
Figure 7: Examples of various waveform displays and measurements

6.5 Math Functions

Press the MATH MENU button to access mathematical operations on waveforms, including addition, subtraction, multiplication, division, and FFT (Fast Fourier Transform) for frequency domain analysis.

6.6 USB Connectivity

The oscilloscope features a USB host port for saving waveforms and settings to a USB flash drive, and a USB device port for connecting to a PC for remote control and data analysis using the provided software.

Video 1: Overview of the Hantek DSO5202P oscilloscope, demonstrating physical features, menu navigation, and basic waveform measurement using cursors.

7. Specifications

7.1 General Specifications

FeatureDescription
Display7 inch 64K color TFT (diagonal liquid crystal), 800*480 pixels, Adjustable (16 gears) with the progress bar
Probe Compensator OutputAbout 5Vpp into >1MΩ load, 1kHz
Power Supply100-120VACRMS(±10%), 45Hz to 440Hz, CAT II; 220-240VACRMS(±10%), 45Hz to 65Hz, CAT II
Power Consumption<30W
Fuse2A, T rating, 250V
Operating Temperature32°F to 122°F (0°C to 50°C)
Nonoperating Temperature-40°F to 158.8°F (-40°C to +71°C)
Cooling MethodConvection
Humidity+104°F or below (+40°C or below): ≤90% RH; 104°F to 122°F (+41°C to 50°C): ≤60% RH
Operating AltitudeBelow 3,000m (10,000 feet)
Nonoperating AltitudeBelow 15,000m (50,000 feet)
Size (L×W×H)313mm x 108mm x 142mm
Weight2.08KG (Without Packing)
Table of general specifications for the oscilloscope, including display, power, environmental, and mechanical details.
Figure 8: General Specifications

7.2 Performance Specifications

FeatureDSO5202PDSO5102PDSO5072P
Sample RateReal-Time: 1GSa/s; Equivalent Sample: 25GS/s
Acquisition ModesNormal, Only data, High-frequency and random glitch capture
AverageWaveform Average, selectable 4, 8, 16, 32, 64, 128
Inputs CouplingAC, DC, GND
Inputs Impedance1MΩ±2% || 20pF±3pF
Probe Attenuation1X, 10X, 100X, 1000X
Supported Probe Attenuation Factor1X, 10X, 100X, 1000X
Maximum Input VoltageCAT I and CAT II: 300V RMS (10×); Installation Category: CAT III: 150V RMS (1×)
Sample Rate Range500MS/s - 1GS/s
Waveform Interpolation(sin X)/X
Record Length40K
SEC/DIV Range2ns/div to 80s/div
Bandwidth200MHz100MHz70MHz
Rise Time at BNC (Typical)1.8ns3.5ns5ns
Math+, -, *, /, FFT
FFT WindowsHanning, Flattop, Rectangular, Bartlett, Blackman; 1024 sample point
Bandwidth Limit20MHz
Low Frequency Response (-3db)≤10Hz at BNC
Trigger ModesAuto, Normal, Single
Trigger Coupling TypeDC, AC, Noise Reject, HF Reject, LF Reject
Trigger Frequency Counter6 digits
Frequency RangeAC coupled, from 4Hz minimum to rated bandwidth
Table of performance specifications for the oscilloscope, including sample rate, acquisition, inputs, and vertical settings.
Figure 9: Performance Specifications (Part 1)
Table of performance specifications for the oscilloscope, including horizontal, vertical, and trigger system details.
Figure 10: Performance Specifications (Part 2)

8. Troubleshooting

8.1 No Display After Power On

  • Ensure the power cable is securely connected to both the oscilloscope and the power outlet.
  • Verify the power outlet is functional.
  • Check the fuse on the rear panel (2A, T rating, 250V). Replace if blown.

8.2 No Waveform Displayed

  • Check if the probe is correctly connected to the input channel (CH1/CH2) and the signal source.
  • Ensure the probe's ground clip is connected to the circuit's ground.
  • Press the AUTO SET button to automatically adjust settings.
  • Adjust the VOLTS/DIV and SEC/DIV knobs to appropriate ranges.
  • Verify the trigger settings. If the trigger level is outside the waveform's amplitude, the waveform may not be stable. Adjust the TRIGGER LEVEL knob.
  • Ensure the input signal is within the oscilloscope's voltage limits.

8.3 Unstable Waveform

  • Adjust the trigger level and trigger type to match the input signal characteristics.
  • Check the probe compensation. Improper compensation can distort the waveform.
  • Ensure the signal source is stable.

9. Maintenance

  • Cleaning: Clean the exterior of the oscilloscope with a soft, damp cloth. Do not use abrasive cleaners or solvents. Ensure no liquid enters the device.
  • Storage: Store the oscilloscope in a dry, dust-free environment when not in use. Avoid extreme temperatures and humidity.
  • Probe Care: Handle probes carefully. Avoid bending or stressing the cable. Regularly check probe tips for damage.
  • Software Updates: Check the manufacturer's website periodically for firmware and software updates to ensure optimal performance and access to new features.

10. User Tips

  • Probe Compensation: For accurate measurements, always compensate your probes when first connecting them to the oscilloscope or if you suspect inaccuracies. Connect the probe to the probe compensator output (usually a square wave signal) and adjust the probe's compensation screw until the square wave is flat-topped.
  • Grounding: Proper grounding is crucial for accurate and safe measurements. Always connect the probe's ground clip to the circuit's ground reference.
  • USB Data Transfer: When saving waveforms or screenshots to a USB drive, ensure the drive is formatted to FAT32 for best compatibility.
  • PC Software: Utilize the PC software for advanced analysis, remote control, and easier data logging, especially for long-term observations or complex signal processing.

11. Warranty and Support

For warranty information, technical support, or service inquiries, please refer to the contact information provided with your purchase documentation or visit the official Hantek website. Keep your purchase receipt as proof of purchase for warranty claims.