Renke Digital Aquaculture Water DO Sensor Online Optical Dissolved Oxygen Meter

Contents

Renke Digital Aquaculture Water DO Sensor

Online Optical Dissolved Oxygen Meter Instruction Manual

1. Introduction

This manual provides detailed instructions for the installation, operation, maintenance, and troubleshooting of the Renke Digital Aquaculture Water DO Sensor. This advanced optical dissolved oxygen meter is designed for precise and reliable water quality monitoring in aquaculture and various environmental applications. It features a fluorescent method for dissolved oxygen measurement, integrated temperature compensation, RS485 output, and IP68 protection level.

Renke Dissolved Oxygen Probe in water with IP68 protection level
Figure 1: Renke Dissolved Oxygen Probe, showcasing its IP68 protection level for submerged use.
Key features of the Renke DO sensor including fluorescent method, temperature compensation, RS485 output, 5m cable, and IP68 protection.
Figure 2: Overview of sensor features: Integrated fluorescent method, temperature compensation, freshwater regular type, RS485 output, 5m customizable wire, and IP68 protection.

2. Specifications

The following table details the technical specifications of the Renke Digital Aquaculture Water DO Sensor:

Attribute Specification
Product Name Aquaculture Water DO Sensor Online Optical Dissolved Oxygen Meter
Brand Name Renke
Measurement Range 0~20mg/L
Accuracy ±3%FS
Resolution 0.01mg/L
Power Supply DC 10~30V
Output Signal RS485
Working Environment 0~40℃
Storage Conditions -10~60℃
Wire Length Default 5m (customizable)
Shell Material Polyoxymethylene (POM), Stainless Steel
Protection Level IP68
Dimensions 171mm (Length) x 30mm (Diameter)
Dimensions of the Renke DO sensor: 171mm length and 30mm diameter.
Figure 3: Sensor dimensions, showing a length of 171mm and a diameter of 30mm.

3. Package Contents

Upon receiving your Renke Digital Aquaculture Water DO Sensor, please verify that all items listed below are included:

  • 1 x Dissolved Oxygen Sensor
  • 1 x 5m Cable (default length, customizable)
  • 1 x Warranty Card
  • 1 x Certificate of Conformity
Contents of the sensor package: one sensor, a 5m cable, a warranty card, and a certificate of conformity.
Figure 4: Package contents including the sensor, 5m cable, warranty card, and certificate of conformity.

4. Setup and Connection

The Renke DO Sensor utilizes an RS485 output signal for data transmission. It can be integrated into various monitoring systems. Ensure the power supply is within the DC 10-30V range.

4.1. Optical Measurement Principle

The sensor operates on the principle of fluorescence quenching. Blue light excites a fluorescent dye on the measuring foil, which then emits red light. The presence of oxygen quenches this fluorescence, reducing the intensity and lifetime of the emitted red light. By measuring these changes, the dissolved oxygen concentration is determined.

Diagram illustrating the optical dissolved oxygen measurement principle with blue and red LEDs, measuring foil, and photoreceptor.
Figure 5: Optical dissolved oxygen measurement principle, showing blue and red LEDs, light filters, sapphire window, measuring foil, and photoreceptor.

4.2. Connection Methods for Data Monitoring

The sensor offers flexible connection options for real-time data monitoring:

  1. Method One: Environmental Monitoring Host & Cloud Platform

    Connect the dissolved oxygen sensor to a computer room environment host. The host can then upload data to a free cloud platform, allowing for mobile, computer, or tablet monitoring anytime and anywhere.

    Diagram for Method One: Dissolved oxygen sensor connects to a computer room environment host, which then uploads data to a free cloud platform for mobile and computer viewing.
    Figure 6: Data monitoring via environmental monitoring host and cloud platform.
  2. Method Two: NFC Collector & Cloud Platform

    Connect the NFC dissolved oxygen sensor to an NFC collector. Data can be uploaded to the free cloud platform through NFC configuration software, enabling monitoring via mobile phone, computer, or tablet.

    Diagram for Method Two: NFC dissolved oxygen sensor connects to an NFC collector, which uploads data to a free cloud platform for mobile and computer viewing.
    Figure 7: Data monitoring via NFC collector and cloud platform.
  3. Method Three: USB to 485 Module

    The device can be uploaded to a computer through a USB to 485 module to realize real-time monitoring directly on a PC.

    Diagram for Method Three: Dissolved oxygen sensor connects to a USB to 485 module, which then connects to a computer for real-time monitoring.
    Figure 8: Data monitoring via USB to 485 module.
  4. Method Four: Single-Chip Microcomputer or PLC

    The equipment can be uploaded to a single-chip microcomputer or PLC through a 485-to-IC/232 module (provided by the customer) for data viewing and integration into custom control systems.

    Diagram for Method Four: Dissolved oxygen sensor connects to a 485-to-IC/232 module, which then connects to a single-chip microcomputer or PLC for data viewing.
    Figure 9: Data monitoring via single-chip microcomputer or PLC.

5. Operation

Once the sensor is properly connected to your chosen monitoring system and powered on, it will begin measuring dissolved oxygen levels. The RS485 output provides digital data that can be read and interpreted by compatible devices or software.

5.1. Initial Power-Up

After connecting the sensor to a stable DC 10-30V power source, allow a few minutes for the sensor to stabilize before taking readings. Refer to your specific data acquisition system's manual for instructions on reading RS485 data.

5.2. Data Interpretation

The sensor provides dissolved oxygen readings in mg/L. The integrated temperature compensation ensures accurate readings across varying water temperatures within the specified working environment of 0~40℃.

Video 1: A brief overview of the Renke Digital Aquaculture Water DO Sensor.

6. Maintenance

The Renke DO Sensor is designed for low maintenance, featuring a robust construction and an optical measurement method that eliminates the need for frequent calibration or electrolyte replacement.

6.1. Cleaning the Sensor Head

The sensor's design, including its 316 stainless steel membrane head and ultra-high transparent glass, makes it easy to clean. Regularly inspect the measuring foil and glass for any fouling (e.g., algae, debris) and gently clean with a soft cloth and clean water. Avoid abrasive materials or harsh chemicals that could damage the optical components.

Features highlighting low maintenance: no oxygen consumption, no electrolyte needed, easy to clean, and no frequent calibration.
Figure 10: Sensor benefits: no oxygen consumption, no electrolyte needed, easy to clean, and no frequent calibration.
Sensor construction showing corrosion-resistant material, ultra-high transparent glass, and 316 stainless steel membrane head design.
Figure 11: Sensor construction details, featuring corrosion-resistant materials, ultra-high transparent glass, and a 316 stainless steel membrane head.

6.2. Storage

When not in use, store the sensor in a dry environment within the specified storage conditions of -10~60℃. Ensure the sensor head is clean and protected from physical damage.

7. Troubleshooting

This section provides solutions to common issues you might encounter with your DO sensor.

Problem Possible Cause Solution
No reading or unstable readings
  • Incorrect power supply
  • Loose cable connection
  • Fouling on sensor head
  • Incompatible data acquisition system
  • Verify power supply is DC 10-30V
  • Check all cable connections
  • Clean the sensor head as described in Section 6.1
  • Ensure your system is compatible with RS485 output
Inaccurate readings
  • Sensor not fully submerged
  • Fouling on sensor head
  • Environmental interference
  • Ensure sensor is fully immersed in the water
  • Clean the sensor head
  • Relocate sensor away from strong electrical interference

8. User Tips

  • Calibration: Although the optical method reduces the need for frequent calibration, periodic checks against a known standard can ensure long-term accuracy, especially in critical applications.
  • Sensor Placement: For optimal results, place the sensor in an area with good water circulation to ensure representative dissolved oxygen measurements. Avoid stagnant zones.
  • Cable Management: Secure the sensor cable to prevent damage or accidental disconnection, especially in dynamic aquatic environments.

9. Warranty and Support

The Renke Digital Aquaculture Water DO Sensor comes with a 1-year warranty from the date of purchase. This warranty covers defects in materials and workmanship under normal use. For technical support or warranty claims, please contact your vendor or the manufacturer directly.

The manufacturer offers customized support including OEM, ODM, and OBM services for specific project requirements.

© 2024 Renke. All rights reserved.