NJTY DT-6013

Contents

NJTY DT-6013 Digital Capacitance Meter User Manual

Model: DT-6013 | Brand: NJTY

Introduction

The NJTY DT-6013 is a compact and durable digital capacitance meter designed for high-precision measurements up to 100,000 uF. It features a 6000-count LCD digital display, data retention, backlighting, and a zero adjustment function, making it a versatile tool for various applications in home appliance inspection, electrician technical schools, and factories.

NJTY DT-6013 Digital Capacitance Meter
Figure 1: NJTY DT-6013 Digital Capacitance Meter

Setup

1. Battery Installation

  1. The device is powered by two 1.5V AA batteries.
  2. Open the battery compartment cover on the back of the meter.
  3. Insert the batteries, ensuring correct polarity as indicated inside the compartment.
  4. Close the cover securely.
Battery compartment design
Figure 2: Battery compartment with batteries installed

2. Connecting Test Leads

  1. Connect the red test clip to the red jack (CX) and the black test clip to the black jack (CX) on the meter.
  2. Ensure connections are firm for accurate readings.
Test lead connection points
Figure 3: Test lead input jacks

Operating Instructions

1. Power On/Off

  • To power on the device, turn the central gear knob to any desired capacitance measurement range.
  • To power off, turn the gear knob to the 'OFF' position.

2. Range Selection

  • Rotate the central gear knob to select the appropriate capacitance measurement range (e.g., 600pF, 6nF, 60nF, 6uF, 60uF, 600uF, 6000uF, 60mF, 100mF).
  • Always start with a higher range if the capacitance value is unknown, then decrease the range for more precise readings.

3. Zero Adjustment

  • Before measuring, short the test leads (touch the red and black clips together).
  • Press the 'ZERO' button to clear any residual capacitance and ensure accurate readings.

4. Taking Measurements

  1. Ensure the capacitor to be measured is fully discharged before connecting it to the meter.
  2. Connect the test clips to the capacitor terminals.
  3. The capacitance value will be displayed on the LCD screen.
Meter measuring 100,000 uF large capacitance
Figure 4: Measuring a large capacitance (100,000 uF)
Meter measuring 35uF capacitor
Figure 5: Measuring a 35uF capacitor
Meter measuring 100uF capacitor
Figure 6: Measuring a 100uF capacitor
Meter measuring 250uF capacitor
Figure 7: Measuring a 250uF capacitor

5. Data Hold

  • Press the 'HOLD/*' button to freeze the current reading on the display.
  • Press it again to release the data hold function.

6. Backlighting

  • Press and hold the 'HOLD/*' button to activate the screen backlight for improved visibility in low-light conditions.
  • Press and hold again to turn off the backlight.

7. Auto Power Off

  • The meter features an auto power-off function to conserve battery life if inactive for a period.
DT-6013 Capacitance Meter Panel Features
Figure 8: DT-6013 Capacitance Meter Panel Features
Video 1: Product demonstration showing various measurements.

Maintenance

1. Battery Replacement

  • When the low voltage indicator appears on the display, replace the batteries promptly to ensure continued accurate operation.
  • Refer to the 'Battery Installation' section for detailed instructions.

2. Cleaning

  • Wipe the meter with a dry, soft cloth.
  • Do not use abrasive cleaners or solvents, as they may damage the casing or display.

Troubleshooting

1. Meter turns on but does not work / No reading

  • Check battery level and replace if necessary.
  • Ensure test leads are securely connected to the meter and the component being tested.
  • Verify the gear knob is set to the correct measurement range.
  • Perform zero adjustment before measurement.

2. Incorrect Readings

  • Ensure the capacitor is fully discharged before testing.
  • Check for loose connections or damaged test leads.
  • Verify the correct range is selected; if the value is too high or too low for the selected range, the reading may be inaccurate or show 'OL' (Overload).

3. Display 'OL'

  • This indicates an over-range condition. Select a higher capacitance range or ensure the component is within the meter's measurable limits.

Specifications

Capacitance Measurement Ranges

Measurement RangePrecisionResolving Power
600pF±2%+100.1pF
6nF±0.6%+100.001nF
60nF±0.6%+100.01nF
6uF±0.6%+100.001uF
60uF±0.6%+100.01uF
600uF±0.6%+100.1uF
6000uF±2%+100.001mF
60mF±4%+100.01mF
100mF±5%+100.1mF
Technical specifications table
Figure 9: Detailed Technical Specifications

General Features and Specifications

  • Display Type: 4-Digit LCD Digital Display (6000 Counts)
  • Measurement Method: Double integral A/D conversion
  • Sampling Rate: Approx. 3 seconds per second
  • On-off buzzer: Yes
  • Data Hold Function: Supported
  • Screen Backlighting: Supported
  • Zero Adjustment Function: Supported
  • Auto Power Off: Supported
  • Low Voltage Display: Yes
  • Over-range Display: The highest position shows "1"
  • Battery: 2 x 1.5V AA (No. 5 battery)
  • Dimensions: 148 x 70 x 35 mm
  • Certifications: CE, RoHS
Capacitance meter size
Figure 10: Capacitance Meter Dimensions

User Tips

  • Always discharge capacitors before testing to prevent damage to the meter or injury.
  • For small capacitors, use the zero adjustment function to eliminate lead capacitance for more accurate readings.
  • When measuring large capacitors, allow a few seconds for the reading to stabilize.
  • Ensure the test leads are in good condition and properly connected for reliable measurements.

Warranty and Support

For technical support, service, or warranty inquiries, please contact the manufacturer or your point of purchase. Keep your purchase receipt as proof of purchase for any warranty claims.