FUZRR ES6030

Contents

FUZRR ES6030 Transformer Turns Ratio Tester

User Manual

Model: ES6030 | Brand: FUZRR

1. Introduction

The FUZRR ES6030 is a portable, high-precision Transformer Turns Ratio (TTR) tester designed for accurate measurement of transformer ratio and group. It is suitable for testing single-phase, three-phase, Z-type transformers, as well as PT and CT transformers. Equipped with a 5-inch full-color touchscreen and intuitive controls, the ES6030 offers reliable performance for various industrial and laboratory applications.

2. Safety Information

Please read and understand all safety instructions before operating the device. Failure to follow these instructions may result in injury or damage to the instrument.

  • Always ensure the transformer under test is de-energized and safely isolated before connecting the tester.
  • Use only the provided test leads and accessories.
  • Do not operate the instrument in wet conditions or explosive atmospheres.
  • Ensure proper connection of HV (High Voltage) and LV (Low Voltage) terminals to avoid damage.
  • The instrument is designed for CAT III 600V, Pollution Grade 2 environments.
  • Do not attempt to open or repair the instrument; refer servicing to qualified personnel.

3. Package Contents

Verify that all items are present in the package:

  • FUZRR ES6030 Transformer Turns Ratio Tester
  • Data Cable (USB)
  • Software CD
  • User Manual (this document)
  • Warranty Certification
  • Test Lines (2 sets, with red, black, green, yellow clips)
  • Charger
  • Carrying Bag
FUZRR ES6030 TTR Tester and all included accessories
Figure 3.1: ES6030 Tester and Accessories

4. Product Overview

The ES6030 features a robust design with clearly labeled ports and an intuitive user interface.

Front view of the FUZRR ES6030 TTR Tester
Figure 4.1: Front View of ES6030 Tester
Back view of the FUZRR ES6030 TTR Tester
Figure 4.2: Back View of ES6030 Tester
Left side view of the FUZRR ES6030 TTR Tester
Figure 4.3: Left Side View of ES6030 Tester
Right side view of the FUZRR ES6030 TTR Tester
Figure 4.4: Right Side View of ES6030 Tester

4.1 Meter Structure and Components

Diagram showing the numbered parts of the ES6030 TTR Tester
Figure 4.5: ES6030 Meter Structure
  1. HV side A phase terminal
  2. HV side B phase terminal
  3. HV side C phase terminal
  4. LV side A phase terminal
  5. LV side B phase terminal
  6. LV side C phase terminal
  7. USB Port
  8. Power button
  9. Confirmation key
  10. Return key
  11. Test key
  12. Color touchscreen display
  13. Multi-function key
  14. Right navigation button
  15. Left navigation button
  16. Down navigation button
  17. Up navigation button
  18. Charging port

5. Specifications

Feature Description
Function Transformer ratio and group test for three-phase, single-phase, Z-type transformers, PT, CT. Supports blind testing of three-phase transformers.
Type of Test Product Single-phase transformer, three-phase transformer, Z-type transformer, PT, CT
Variable Ratio Range 0.9 ~ 10000
Overload Protection Yes
Power DC9.6V 3800mAh Large capacity lithium battery
Charging Function Yes
Display Mode 5-inch touchscreen (854x480 dots) full-color LCD display
Interactive Mode Touch screen & physical buttons
LCD Size Length × Width: 108 mm × 65 mm
Instrument Size Length × Width × Height: 229 mm × 166 mm × 74 mm
Mobile APP Connectivity Yes, via Bluetooth connection
Computer Connectivity Yes, via USB cable connection
Data Storage 500 groups
Data Access Data access function available
Battery Voltage Display Real-time display of battery power; indicates when charging is needed.
Auto Shut Off Configurable from 1 to 30 minutes
Power Dissipation Standby: approx. 2W (100% brightness); Measured at: 12W Max
Weight Instrument: 1.5kg (including battery); Test cables: 1.7kg
Working Temperature & Humidity -10℃ ~ 40℃; below 70%RH
Storage Temperature & Humidity -20℃ ~ 60℃; below 70%RH
Insulation Resistance Above 10 MΩ (500V between circuit and housing)
Withstand Voltage AC 3700V / rms (between circuit and enclosure)
Electromagnetic Characteristics IEC61010-4-3, Wireless frequency electromagnetic field ≤ 1V/m
Safety Regulations IEC61010-1, CAT III 600V, Pollution grade 2

5.1 Measurement Range and Accuracy

Measure the Ratio Range Resolution Ratio Accuracy
0.9 ≤ TTR < 10 0.0001 ±0.2% ± 2dgt
10 ≤ TTR < 100 0.001 ±0.2% ± 2dgt
100 ≤ TTR < 1000 0.01 ±0.2% ± 2dgt
1000 ≤ TTR < 2000 0.1 ±0.2% ± 2dgt
2000 ≤ TTR < 10000 0.1 ±0.5% ± 2dgt
TTR ≥ 10000 1 --

6. Setup

6.1 Charging the Battery

The ES6030 is powered by a DC9.6V 3800mAh lithium battery. Before first use, ensure the battery is fully charged.

  1. Connect the provided charger to the charging port (18) on the instrument.
  2. Plug the charger into a suitable power outlet.
  3. The charging indicator on the device will show the charging status.
  4. A full charge typically takes several hours. The battery voltage is displayed in real-time on the screen.

6.2 Connecting Test Leads

Proper connection of test leads is crucial for accurate and safe measurements.

  1. Ensure the transformer under test is completely de-energized and isolated.
  2. Connect the red, yellow, and green clips of the HV (High Voltage) test leads to the corresponding HV side A, B, and C phase terminals (1, 2, 3) on the ES6030.
  3. Connect the red, yellow, and green clips of the LV (Low Voltage) test leads to the corresponding LV side A, B, and C phase terminals (4, 5, 6) on the ES6030.
  4. Attach the other end of the HV test leads to the high voltage windings of the transformer.
  5. Attach the other end of the LV test leads to the low voltage windings of the transformer.
  6. Double-check all connections for secure contact and correct phasing.
Example of test lead connection to a transformer
Figure 6.1: Example of Test Lead Connection

7. Operating Instructions

7.1 Powering On/Off

  • To power on, press and hold the Power button (8) until the screen illuminates.
  • To power off, press and hold the Power button (8) until the shutdown prompt appears, then confirm.

7.2 Navigating the Interface

The ES6030 features a 5-inch touchscreen (12) and physical navigation buttons (14, 15, 16, 17) for interaction.

  • Use the touchscreen to select menu options directly.
  • Alternatively, use the Up (17), Down (16), Left (15), and Right (14) buttons to navigate through menus.
  • Press the Confirmation key (9) to select an option or confirm an action.
  • Press the Return key (10) to go back to the previous screen or cancel an action.
Close-up of the ES6030 screen and control buttons
Figure 7.1: Screen and Button Interaction

7.3 Performing a Test

  1. From the main menu, select the appropriate test type: "Three phase", "Single phase", or "Z-type".
  2. Follow the on-screen prompts to configure any necessary parameters (e.g., winding configuration).
  3. Ensure all test leads are correctly and securely connected to the transformer.
  4. Press the "TEST" button (11) to initiate the measurement.
  5. The results, including turns ratio and group, will be displayed on the screen.
  6. The instrument supports blind testing for three-phase transformers.

7.4 Data Management

The ES6030 can store up to 500 groups of measurement data.

  • Saving Data: After a test, the instrument may prompt to save the results. Follow the on-screen instructions.
  • Viewing Data: Select the "View" option from the main menu to access stored data. You can browse through saved test results.
  • Connecting to PC: Use the provided USB cable to connect the ES6030 to a computer via the USB port (7). Install the software from the included CD to download and analyze data.
  • Mobile APP: The instrument supports Bluetooth connection for mobile app integration. Pair your device with the ES6030 via Bluetooth to use the mobile application for data access and control.
ES6030 TTR Tester connected to a transformer with mobile app displaying results
Figure 7.2: Testing with Mobile App Connectivity

8. Maintenance

8.1 Cleaning

  • Wipe the instrument's exterior with a soft, damp cloth. Do not use abrasive cleaners or solvents.
  • Ensure no moisture enters the ports or openings.

8.2 Storage

  • When not in use, store the ES6030 in its carrying bag in a cool, dry place, away from direct sunlight and extreme temperatures.
  • Ensure the battery is partially charged (around 50%) before long-term storage to prolong its lifespan.
  • Refer to storage temperature and humidity specifications in Section 5.

8.3 Battery Care

  • Charge the battery when the low voltage indicator appears on the screen.
  • Avoid completely draining the battery frequently.
  • Use only the original charger provided with the instrument.

9. Troubleshooting

Problem Possible Cause Solution
Instrument does not power on. Low battery or faulty power button. Charge the battery. If issue persists, contact support.
Inaccurate readings. Loose test lead connections, incorrect test type selected, or transformer not de-energized. Check all connections, ensure correct test mode, and verify transformer is de-energized.
Cannot connect to PC/Mobile App. Incorrect cable, driver not installed (PC), Bluetooth not enabled/paired (Mobile). Ensure correct USB cable is used, install PC software/drivers, enable Bluetooth on mobile and pair with device.
Screen unresponsive. Software glitch or hardware issue. Try restarting the device. If unresponsive, contact support.

10. User Tips

  • For optimal battery life, avoid storing the device in extreme temperatures.
  • Regularly back up your measurement data to a PC to prevent loss.
  • Keep the software on your PC and mobile app updated for the best performance and features.

11. Warranty and Support

The FUZRR ES6030 Transformer Turns Ratio Tester comes with a standard manufacturer's warranty. Please refer to the included Warranty Certification for detailed terms and conditions.

For technical support, service, or warranty claims, please contact your local distributor or the manufacturer directly. Contact information can typically be found on the warranty card or the manufacturer's official website.


Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.