1. Introduction

The DM91A Digital Multimeter is an innovative, high-performance, highly reliable, full-featured, large-screen display, handheld, 4-digit true RMS digital multimeter. The meter sets the kernel on integrated circuits and an A/D converter, and is provided with an overload protection circuit and fuse disconnection prompt circuit. It can be used to measure DC and AC voltage, AC and DC current, resistance, capacitance, frequency, live wire, battery voltage, non-contact AC voltage sensing (NCV), diode, and circuit continuity. This multimeter can be widely used in schools, research institutes, factories, companies, and other social areas.

2. Safety Information

The DM91A Digital Multimeter is designed according to IEC61010-1 600V (CATIII) and Level 2 Pollution. To ensure correct and safe use of the meter, please read this User Manual carefully. To avoid personal injury caused by electric shock and damage to the meter, please pay attention to the following safety tips:

  • Do not measure any voltage that exceeds the measuring range specified for this meter.
  • Despite the internal protection circuit for the resistance measurement field, do not add a voltage of 100V or more to the input terminal in the field of resistance measurement.
  • Check the integrity of the probes.
  • Avoid using the meter under direct sunlight or extreme high-temperature conditions as far as possible.
  • Avoid electric shock hazard when measuring voltages exceeding 30V AC or 60V DC.
  • Before measuring current, turn off the power supply, and disconnect it from its circuit before measuring it.
  • Pay attention to the polarity of the battery when replacing it.

Electrical Symbols:

  • Dangerous voltage
  • Grounding
  • AC (alternating current)
  • DC (direct current)
  • AC or DC
  • Double insulation
  • Fuse
  • Warning, check the Manual for relevant instructions.

3. Product Overview

3.1 Product Appearance and Description

The Mestek DM91A Digital Multimeter features a compact design with a clear display and intuitive controls. Below is an illustration of the multimeter's panel layout and its key components.

Mestek DM91A Digital Multimeter front view with probes and screwdriver
Figure 1: Mestek DM91A Digital Multimeter with accessories. This image shows the front of the DM91A multimeter, including its display, rotary switch, function buttons, and input jacks. Accompanying accessories like test leads and a screwdriver are also visible.
DM91 Multimeter labeled parts and counting parameters table
Figure 2: DM91 Multimeter labeled parts and counting parameters. This diagram highlights the various components of the DM91 model, including NCV detection, alarm indicator, LCD, flashlight switch, function range selection knob, 10A input terminal, COM terminal, and V/Ω/mA/µA/Cap/Battery/Live input terminal. A table of counting parameters for DC voltage, AC voltage, DC current, resistance, and on-off buzzer/diode is also included.
DM91A Multimeter labeled parts and counting parameters table
Figure 3: DM91A Multimeter labeled parts and counting parameters. This diagram highlights the various components of the DM91A model, including NCV detection, alarm indicator, LCD, SEL/VFC button, measuring range, function range selection knob, 10A input terminal, COM terminal, and V/Ω/mA/µA/Cap/Battery/Live input terminal. A table of counting parameters for DC voltage, AC voltage, DC current, resistance, capacitance, and on-off/diode is also included.
DM91S Multimeter labeled parts and counting parameters table
Figure 4: DM91S Multimeter labeled parts and counting parameters. This diagram highlights the various components of the DM91S model, including NCV detection, alarm indicator, LCD, SEL function select button, NCV detection, function range selection knob, 10A input terminal, COM terminal, and V/Ω/mA/µA/Cap/Battery/Live input terminal. A table of counting parameters for DC voltage, AC voltage, DC current, resistance, capacitance, and smart recognition is also included.

Panel Layout:

  1. NCV, continuity and live wire indicator
  2. Liquid Crystal Display (LCD)
  3. "SEL/V.F.C" function selection and variable-frequency voltage measurement button
  4. "RANGE" range switching button
  5. "HZ/" frequency switching and flashlight control button
  6. "HOLD/" data hold and backlight control button
  7. Function range selection knob
  8. "V/Ω/mA/µA/Cap/Battery/Live" input terminal
  9. "COM" terminal
  10. 10A input terminal

3.2 Function Buttons

Button Function
SEL/V.F.C Press “SEL” button to switch the function and measurement unit; in AC voltage field, hold down this button for 2 seconds will make the meter enter into V.F.C measurement mode. When the knob is at OFF position, hold down the SEL button and rotate the knob to turn on the meter, this cancels the automatic shutdown function.
RANGE Press “RANGE” button to select auto/manual range. In auto range mode, press this button to enter manual range mode; when pressing each time, a higher range is activated, but when reaching the highest range, it will go back to the lowest range, then it cycles in the same order. When this button is held down for 2 seconds in manual range mode, it returns to auto range mode.
Hz/ In AC voltage or AC current field, press this button to display the measured voltage or the frequency of the current; Press again to exit the frequency display. Hold down this button for about 2 seconds to turn on the flashlight, which turns off automatically after 15 seconds; it can be turned off by holding down the button again.
HOLD/ To maintain the reading during the measurement, you can press the “HOLD” button, then the measurement reading will be locked on the display. Press the "HOLD" button again, the reading-hold state is released. Hold down this button for about 2 seconds to turn on the backlight, which turns off automatically after 15 seconds; it can be turned off by holding down the button again.

3.3 LCD Display

The LCD provides clear readings and indicators for various functions. Refer to the table below for symbol explanations.

DM91A Multimeter display showing 9999
Figure 5: DM91A Multimeter display. This image shows the digital display of the DM91A multimeter, indicating a reading of "9.999" and various mode indicators like "Auto" and "nF".
Symbol Description
DCDirect current
ACAlternating current
Diode symbolDiode
Continuity symbolContinuity
NCVNon-contact AC voltage sensing
V.F.CVariable-frequency voltage measurement
LiveLive wire test
Auto-off symbolAutomatic shut-down
FUSE symbolFuse disconnected
Battery symbolLow battery voltage will trigger this symbol, prompting to replace the battery right soon.
Ω, KΩ, MΩResistance unit
nF, µF, mFCapacitance unit
AutoAutomatic range
ManualManual range
mV, VVoltage unit
µA, mA, ACurrent unit
Minus signIndicates "negative"

4. Specifications

4.1 General Features

  • Display: 9999 4-digit, true RMS
  • Operating Mode: Auto Range
  • Fuse disconnection alarm function
  • Maximum voltage to ground at the voltage input terminal: 600V CAT III
  • Data hold function
  • AC voltage/current frequency measurement
  • Over-range indication: OL
  • Low battery voltage indication: (battery symbol)
  • Automatic shutdown time: 15 Minutes
  • Working environment: 0 ~ 40°C (32 ~104 °F); < 80% RH
  • Storage conditions: -10 ~ 50°C (14 ~ 122 °F); <70% RH
  • Power supply: 2×1.5V AAA batteries (not included)
  • Safety level: IEC61010-1, CAT III 600V
  • Dimensions: 147 x 71 x 45mm
  • Weight: About 195g

Accessories:

  • User Manual - 1 copy
  • Probes - 1 set
  • 1.5V AAA Battery - 2 pieces (not included)

4.2 Accuracy Indicators

Accuracy: % Reading digits, one year warranty since the date of shipment. Ambient temperature: 18°C to 28°C. Environment humidity: < 80%.

4.2.1 DC Voltage

Overload protection: PTC 600V DC or AC RMS.

Range Resolution Accuracy
999.9mV0.1mV±(0.8% Reading + 2 digits)
9.999V0.001V±(0.7% Reading + 3 digits)
99.99V0.01V±(0.7% Reading + 3 digits)
600V0.1V±(0.7% Reading + 3 digits)

4.2.2 AC Voltage

Input impedance about 10MΩ, displays true RMS, frequency response: 45~400HZ. Maximum input voltage: 600V AC (effective value).

Range Resolution Accuracy
0.999V0.001V±(1.2% Reading + 3 digits)
9.999V0.01V±(1% Reading + 3 digits)
99.99V0.1V±(1% Reading + 3 digits)
600V0.1V±(1% Reading + 3 digits)

4.2.3 Resistance

Range Resolution Accuracy
999.9Ω0.1Ω±(1% Reading + 2 digits)
9.999KΩ0.001kΩ±(1% Reading + 2 digits)
99.99kΩ0.01kΩ±(1% Reading + 2 digits)
999.9kΩ0.1kΩ±(1% Reading + 2 digits)
9.999MΩ0.001MΩ±(2.0% Reading + 3 digits)
99.99MΩ0.01MΩ±(2.0% Reading + 3 digits)

4.2.4 Live Wire Test

Function Description
LiveInsert only one probe into the voltage input terminal, and use it to test the power socket to determine the neutral and live wires.

4.2.5 Diode and Circuit Continuity

Open circuit voltage is about 3.3V.

Function Description
Diode / ContinuityContinuity test and diode measurement are executed in smart mode without the need to press any switching button; when the measured resistance is less than about 30 ohm, it displays the on-resistance value, while the internal buzzer beeps, and the continuity indicator (green) lights; When measuring a diode, it displays the approximate forward voltage of the diode.

4.2.6 DC Current

Overload protection: µA/mA range: F600mA/250V fuse (fast fusing-off), 10A range: F10A/250V fuse (fast fusing-off). Maximum input current: mA jack: 600mA, 10A jack: 10A.

Range Resolution Accuracy
999.9µA0.1µA±(1.2% Reading + 3 digits)
9999µA1µA±(0.8% Reading + 3 digits)
99.99mA0.01mA±(0.8% Reading + 3 digits)
600.0mA0.1mA±(0.8% Reading + 3 digits)
10A0.01A±(0.8% Reading + 3 digits)

4.2.7 AC Current

Overload protection: µA/mA range: F600mA/250V fuse (fast fusing-off), 10A range: F10A/250V fuse (fast fusing-off). Maximum input current: mA jack: 600mA, 10A jack: 10A. Frequency response: 45~400HZ.

Range Resolution Accuracy
999.9µA0.1µA±(1.5% Reading + 3 digits)
9999µA1µA±(1.0% Reading + 3 digits)
99.99mA0.01mA±(1.0% Reading + 3 digits)
600.0mA0.1mA±(1.0% Reading + 3 digits)
10A0.01A±(1.0% Reading + 3 digits)

4.2.8 Capacitance

Range Resolution Accuracy
9.999nF0.001nF±(4.0% Reading + 30 digits)
99.99nF0.01nF±(4.0% Reading + 30 digits)
999.9nF0.1nF±(4.0% Reading + 30 digits)
9.999µF0.001µF±(4.0% Reading + 3 digits)
99.99µF0.01µF±(4.0% Reading + 3 digits)
999.9µF0.1µF±(4.0% Reading + 3 digits)
9.999mF0.001mF±(5.0% Reading + 30 digits)
99.99mF0.01mF±(5.0% Reading + 30 digits)

4.2.9 NCV Non-Contact AC Voltage Sensing

Range Description
Low fieldDisplay --L, NCV indicator (green) lights, and the buzzer gives out the alarm.
High fieldDisplay --H, NCV indicator (two red LEDs) lights, while the buzzer gives out the alarm.

4.2.10 Battery Voltage Measurement

Range Description
1.5VThe load resistance is about 30 ohm, and it shows the battery voltage.
9VThe load resistance is about 300 ohm, and it shows the battery voltage.
Feature comparison table for DM91, DM91A, DM91S
Figure 15: Feature Comparison Table. This table provides a detailed comparison of features and specifications across the DM91, DM91A, and DM91S models, including DC voltage, AC voltage, current, resistance, capacitance, and special functions like True RMS, Live FireWire Identification, VFC, and Fuse blow alarm.

5. Initial Setup

Before using the multimeter, ensure batteries are correctly installed. Refer to the "Battery Replacement" section under Maintenance for detailed instructions on installing or replacing batteries.

6. Measurement Operation

6.1 AC and DC Voltage Measurement

Warning: Do not measure any voltage higher than DC 600V or AC 600V RMS to prevent electric shock and/or damage to the meter. Do not apply more than DC 600V or AC 600V RMS voltage between the common terminal and earth to prevent electric shock and/or damage to the meter.

  1. Turn the rotary switch to the V field.
  2. Connect the black probe to the "COM" input socket and the red probe to the "V" input socket, respectively.
  3. The default position in the field is DC voltage. Press the SEL button to switch to AC voltage. In the AC voltage field, hold down the SEL/VFC button for about 2 seconds; the meter will enter into the variable-frequency voltage measurement mode, which allows measuring the variable-frequency voltage stably. Press the button again, it will exit the VFC mode.
  4. Use the two probe tips to measure the voltage of the circuit under test (in parallel with the circuit under test).
  5. Get the measured voltage value on the display. When measuring DC voltage, the display will show the polarity of the voltage to which the red probe is connected. When measuring AC voltage, press the HZ button to display the voltage frequency value.

Note: In the DC mV or AC 9.999V range, even if there is no input or connection to the probes, sometimes the meter will display several digits, which is normal. In this case, by short-circuiting the “V” and “COM” terminals, the meter can be reset to zero.

DM91A Multimeter showing DC Voltage, DC Current, AC Voltage measurements
Figure 6: DC Voltage, DC Current, and AC Voltage Measurement. This image illustrates how to perform DC voltage measurement using a 9V battery, DC current measurement on a circuit, and AC voltage measurement from a wall outlet.

Video 1: Product demonstration showing various measurement functions including voltage, current, resistance, and NCV.

6.2 Resistance and Capacitance Measurement

Warning: To avoid damage to the meter or the device under test, disconnect the power supply of the circuit under test and fully discharge all high-voltage capacitors before measuring the resistance.

Resistance Measurement:

  1. Turn the rotary switch to the Ω field. The meter displays “OL” at the beginning, showing that the input is open-circuit, i.e., no resistance has been connected.
  2. Connect the black probe to the "COM" input socket and the red probe to the "V" input socket, respectively.
  3. Use the probe tips to measure the resistance of the circuit under test.
  4. Read the measured resistance value on the display.

Note: When measuring low resistance, to ensure accuracy, please short-circuit the two probes to get the short-circuit resistance value, then it shall be reduced from the measured resistance value. When measuring the resistance in the 99.99MΩ field, in certain cases, it will take several seconds to get the reading stabilized. This is normal for high resistance measurements.

DM91A Multimeter showing resistance measurement
Figure 7: Resistance Measurement. The image shows the DM91A multimeter measuring resistance, with test leads connected to a standard source outputting 1 ohm.

Capacitance Measurement:

  1. Turn the rotary switch to the Capacitance field. Press the SEL key to switch to capacitance measurement mode.
  2. Connect the black probe to the "COM" input socket and the red probe to the "V" input socket, respectively.
  3. Use the probe tips to measure the capacitance of the capacitor under test.
  4. Read the measured capacitance value on the display.

Note: In the capacitive measurement mode, it will display "OL" when the maximum range is exceeded. When measuring a large-capacity capacitor, it is normal to take several seconds to measure it. Before measuring the capacitor, it is necessary to release the residual charge, which is especially important for capacitors with high voltage in order to avoid damage to the meter and personal hazards.

DM91A Multimeter with probes, showing capacitance measurement
Figure 8: Capacitance Measurement. The image shows the DM91A multimeter measuring the capacitance of a large capacitor, with test leads connected to the capacitor terminals.

6.3 AC and DC Current Measurement

Warning: Do not attempt measuring the current on the circuit when the voltage between the open circuit voltage and earth exceeds 250V. If the fuse is blown during measurement, it may cause damage to the meter or personal injury. When the fuse is disconnected, to avoid damage to the meter or equipment, please replace the fuse with the same one before measuring. Use the correct input socket and function field for the measurement. When the probe is plugged into the current input socket, do not connect the probe tip to any circuit in parallel mode.

  1. Cut off the power of the circuit under test. Discharge all high voltage capacitors on the circuit under test.
  2. Rotate the rotary switch to the appropriate current field. Use the "SEL" button to select DC or AC measurement mode.
  3. Insert the black probe to the "COM" input socket and the red probe to the corresponding current input socket.
  4. Disconnect the circuit under test. Connect the black probe to the low potential end of the circuit that has been disconnected, and connect the red probe to the high potential end of the circuit that has been disconnected. (Reversing the probes will give you a negative reading value, but it will not damage the meter.)
  5. Turn on the power of the circuit under test and then get the displayed reading. If the display only shows "OL", it means that the input exceeds the range.
  6. Cut off the power of the circuit under test. Discharge all high-voltage capacitors. Remove the meter probes and restore the circuit to its original condition.

Note: When measuring a large current of 5-10A, the power-on time should not exceed 10 seconds so as to avoid unstable test data due to the heat. When measuring multiple times, the interval between two measurements should be 3-5 minutes.

DM91A Multimeter showing AC Current, Resistance, Diode/Continuity measurements
Figure 9: AC Current Measurement. This image shows the DM91A multimeter measuring AC current in a circuit, with test leads connected in parallel.

6.4 Diode and Continuity Test

Warning: To avoid damage to the meter or the device under test, it is necessary to disconnect the power of the circuit under test and discharge all high-voltage capacitors before measuring the diodes.

  1. Turn the rotary switch to the Diode/Continuity field and connect the probe tips to both ends of the measured object.
  2. Connect the black probe to the "COM" input socket and the red probe to the "V" input socket, respectively.
  3. If the resistance of the measured object is less than about 30 ohm, the meter will automatically switch to the Continuity field. The green indicator of the on/off indicator lights up and the buzzer sounds, indicating that the two points are connected, while the LCD displays the resistance value.
  4. If the object to be measured is a diode, the meter will automatically switch to the diode field when it is in forward continuity, and the LCD screen displays the approximate forward voltage of the diode. When the diode is in open circuit or in reverse polarity, the screen will display “OL”.

Note: A normal diode in the circuit should still produce a forward voltage drop of 0.5V to 0.8V; however, the reverse bias reading varies from the resistance of the other channels between the two probes.

DM91A Multimeter with probes, showing diode measurement
Figure 10: Diode Measurement. The image shows the DM91A multimeter performing a diode test, with test leads connected to a diode.

6.5 Live Wire Test

  1. Turn the rotary switch to the Live field. The screen displays "-----".
  2. Only connect the red probe to the V input socket.
  3. Plug the red probe tip into the power socket L jack, or near the live wire. If the meter detects AC voltage, the screen will show LIVE; the two red LEDs on the right top side of the panel will light up, and the buzzer will give out the alarm, prompting the line is a live wire.

6.6 NCV (Non-Contact Voltage) Test

Note: The detection operation may be affected by factors such as socket design, insulation thickness, and type. Even if it is not indicated as Live, voltage may still exist. Do not rely solely on non-contact voltage detector to determine if a voltage is present on the wire. When the input voltage is led to the meter input terminal, the voltage sensing indicator may light up due to the presence of induced voltage. External interference sources (such as flashlights, motors, etc.) may accidentally trigger non-contact voltage detection.

  1. Set the function range switch to the NCV field. To determine the presence of an AC voltage or electromagnetic field on the object, place the probe with the mark “NCV” on the front of the meter near the object being measured.
  2. When AC voltage is sensed, the screen, NCV indicator, and buzzer will indicate the voltage level simultaneously.
  3. When the induced voltage is low, the display shows “--L”. The green LED on the left side of the NCV indicator lights up and the buzzer gives out an alarm continuously.
  4. When the induced voltage is high, the display shows “--H”. The two red LEDs on the left side of the NCV indicator light up and the buzzer gives out an alarm continuously at a higher frequency.
DM91A Multimeter showing NCV and LED flashlight
Figure 11: NCV and LED Flashlight. This image shows the DM91A multimeter detecting non-contact voltage (NCV) with its display showing "-L" and the LED flashlight activated.
DM91A Multimeter showing NCV detection
Figure 12: NCV Detection. This image shows the DM91A multimeter performing NCV detection near a wall outlet, with the display indicating voltage presence.

6.7 Battery Measurement

  1. Plug the black probe into the COM jack and the red probe into the "VmA" jack.
  2. Set the function range switch to the 9V or 1.5V battery measurement field according to the measured battery type.
  3. Use the red probe to touch the “+” pole of the measured battery, and the black probe to touch the “-” pole.
  4. Read the displayed battery voltage from the LCD screen.
DM91A Multimeter showing battery measurement
Figure 13: Battery Measurement. This image shows the DM91A multimeter measuring the voltage of a 9V battery, with the display showing "9.28V".

7. Maintenance

If the battery symbol appears on the LCD during use of the meter, the battery must be replaced to prevent the meter from working abnormally.

7.1 Battery Replacement

  1. Unplug the test lines. Turn off the power.
  2. Use a screwdriver to open the battery cover on the back, then remove the battery.
  3. Place a similar battery (2x 1.5V AAA batteries) and fix the cover.

7.2 Fuse Replacement

Warning: When the fuse is disconnected, to avoid damage to the meter or equipment, please replace the fuse with the same one before measuring.

  1. First remove the test lines and shut down the meter.
  2. Use a screwdriver to open the back cover, then remove the blown fuse.
  3. Insert a similar fuse (F600mA/250V for µA/mA range, F10A/250V for 10A range), relocate the back cover, and fix it with the screws.
DM91A Multimeter showing Double Fuse protection
Figure 14: Double Fuse Protection. This image shows the internal circuit board of the multimeter, highlighting the double fuse design for overload protection.

7.3 General Care

If necessary, use a soft cloth to wipe off the surface of the meter. Do not use organic solvents or abrasives that can corrode or dissolve the case.

8. Troubleshooting

  • Fuse Disconnection Alarm: The meter has a fuse disconnection alarm function. If the rotary switch is turned to the 9V battery, 1.5V battery, µA or mA fields when the 600mA/250V fuse on the V/mA/µA input end is disconnected, the screen will display "FUSE", "mA" and " " symbols, while the buzzer gives out the alarm, prompting that the fuse at the mA input end is disconnected. Now it is the time to replace with a similar fuse in order to conduct measurement. If the rotary switch is turned to the 10A field when the 10A/250V fuse on the 10A input end is disconnected, the screen displays “FUSE”, “A” and “ ” symbols, and the buzzer gives out the alarm, prompting that the fuse on the 10A current input end is disconnected. It is also necessary to replace with a similar fuse in order to conduct the measurement.
  • "OL" Display: If the display shows "OL" during a measurement, it indicates that the input exceeds the current measurement range. Adjust the range or check the circuit.
  • Low Battery Indication: If the battery symbol appears on the LCD, replace the 2x 1.5V AAA batteries immediately to ensure proper operation.
  • Unstable Resistance Readings: When measuring high resistance values (e.g., in the 99.99MΩ range), it may take several seconds for the reading to stabilize. This is normal behavior.
  • Unstable Capacitance Readings: When measuring large-capacity capacitors, it may take several seconds for the measurement to complete. This is normal behavior. Ensure the capacitor is fully discharged before measurement.
  • Voltage Sensing Indicator Lights Up Without Live Wire: The voltage sensing indicator may light up due to the presence of induced voltage or external interference sources (like flashlights, motors). Do not rely solely on NCV for voltage presence confirmation.

9. User Tips

  • For accurate low resistance measurements, first short-circuit the probes to determine their inherent resistance, then subtract this value from your measurement.
  • When performing current measurements, especially for high currents (5-10A), limit the power-on time to under 10 seconds to prevent heat-induced instability. Allow 3-5 minutes between repeated high-current measurements.
  • To disable the automatic shutdown function, hold down the SEL button while turning the rotary knob from OFF to ON.
  • Utilize the built-in flashlight (long press Hz button) and backlight (long press HOLD button) for improved visibility in dark environments.

10. Warranty and Support

This product comes with a one-year warranty from the date of shipment, covering manufacturing defects. For technical support, service, or warranty claims, please contact your retailer or the manufacturer directly. Keep your purchase receipt as proof of purchase.

Ask a question about this manual