NJTY T21L

Contents

NJTY T21L Inductance Digital Multimeter

Model: T21L

Brand: NJTY

1. Introduction

The NJTY T21L is a multi-purpose digital multimeter designed for high measurement accuracy, fast response speed, and a high safety level. It features a special chip with up to 6000 counts, composed of high-precision AD and a high-speed digital processor. This ensures accurate measurements, high resolution, and fast calculation speed, capable of high-speed true effective value calculation of AC 1KHz. The device is software calibrated, maintaining accuracy over long-term use.

Its ergonomic design makes it suitable for various industrial applications. The multimeter offers numerous measurement functions and a user-friendly interface, catering to professional and maintenance engineers. It can measure DC and AC voltage, DC and AC current, resistance, capacitance, inductance, temperature, diode, and perform circuit continuity testing. For convenience in dark environments, it is equipped with a backlight and an illumination lamp.

Video: NJTY T21L Multimeter in action, demonstrating various measurement functions.

2. Safety Information

2.1. Safety Instructions

  • Always comply with standard safety regulations regarding electric shocks and proper instrument usage.
  • Use only the provided test probes, ensuring they are intact before each use.

2.2. Safety Signs

Symbol Description
⚠️Warning
~AC (alternative current)
---DC (direct current)
⏚Ground
☐Double insulation
CEEuropean union standard
⚡High voltage warning
CATIIII category overvoltage protection

2.3. Precautions for Use

  • Avoid using the meter near devices with large electromagnetic disturbances.
  • Do not use if the meter or test probes are physically damaged.
  • Incorrect use may compromise safety functions.
  • Exercise caution around bare conductors or buses.
  • Do not use near explosive gas, vapor, or dust.
  • Always select the correct input function and range.
  • Do not switch ranges while measuring voltage or current.
  • Ensure input values do not exceed specified limits to prevent damage.
  • Do not touch unused input terminals when connected to a circuit under test.
  • Use caution to prevent electric shock when measuring voltages exceeding 60 VDC or 30 VAC.
  • When using test probes, keep fingers behind the guard ring.
  • Before changing the range, ensure test probes are disconnected from the circuit.
  • Before measuring resistance, diode, capacitance, or continuity, power off the circuit and discharge all high-voltage capacitors.
  • Do not measure resistance on a live circuit or perform buzzer tests on live circuits.
  • Before current measurement, check the meter's fuse. Turn off the circuit power before connecting the meter.
  • When repairing TVs or measuring power conversion circuits, be aware of high-amplitude voltage pulses to avoid meter damage.

3. Panel Diagram

Familiarize yourself with the components of your multimeter:

NJTY T21L Multimeter Panel Diagram

Figure 3.1: Front Panel Layout of the NJTY T21L Multimeter.

  1. Illumination
  2. NCV sensing area
  3. LED display
  4. Buzzer
  5. Backlight/illumination button
  6. "FUNC/Hz" selection button
  7. Gear rotary switch
  8. VΩMAL C input socket
  9. COM input socket
  10. 10A input socket
  11. Lock keys
  12. NCV indicator

4. Monitor Symbols Explanation

The display uses various symbols to indicate the current status and measurement mode:

Symbol Description
Battery low voltage display
Auto power off
-Negative polarity input indication
ACAlternative input indication
DCDirect input indication
🔊In continuity test mode
▶️In diode test mode
LInductance measurement
AUTOAuto ranging mode
HOLDData hold mode
°C, °FTemperature
%Duty cycle
NCVNon-contact voltage detection mode
V, mVVolt, unit of Voltage
A, mA, μAAmpere, unit of current
Ω, KΩ, MΩOhm, unit of resistor
Hz, KHz, MHzHz, hertz, unit of frequency
mF, μF, nF, pFFarah, unit of capacitance
H, mH, μHInductance unit

5. Setup

5.1. Battery Installation

The instrument uses three AAA 1.5V batteries as the power supply. Ensure batteries are installed correctly in the battery compartment.

To replace the battery:

  1. Turn off the instrument.
  2. Pull all test probes out of the input sockets.
  3. Use a screwdriver to loosen the screws fixing the battery cover.
  4. Remove the battery cover.
  5. Remove the old batteries.
  6. Replace with new 3x1.5V AAA batteries.
  7. Install the battery cover and close the screws.

Note: When the low voltage symbol (battery icon) appears on the display, replace the batteries immediately. Insufficient battery power can lead to incorrect readings and potential safety hazards.

Back of multimeter showing battery compartment

Figure 5.1: Rear view of the multimeter with the battery compartment.

6. Operating Instructions

6.1. On/Off Operation

Turn the gear rotary switch to the desired function to power on the device. If there is no operation for about 15 minutes, the instrument will enter sleep mode (indicated by a beep sound). Press any button or rotate the gear switch to restart.

6.2. Button Function Description

  • HOLD Key (Data Hold):
    1. During measurement, short press to freeze the current display value. Press again or rotate the gear switch to cancel.
    2. Press and hold the HOLD button while turning on the meter to disable the automatic shutdown function (three beeps will confirm). The device will reset to automatic shutdown upon restart.
  • FUNC/Hz Key (Function Selection):
    1. In AC/DC voltage and current measurement modes, use to switch between current and voltage levels. For AC voltage and current, long press to switch to frequency measurement (0-1KHz signals with voltage < 250V).
    2. In the Resistance/Continuity/Diode/Capacitance range, short press to cycle through Resistance, Continuity, Diode, and Capacitance measurement modes.
  • Backlight and Lighting Functions:
    1. Short press the backlight/flashlight button to turn on the backlight. Short press again to turn it off. Backlight automatically turns off after 30 seconds of inactivity.
    2. Long press the backlight/flashlight button to turn on the flashlight and backlight simultaneously. Short press again to turn both off. Both automatically turn off after 30 seconds of inactivity.
    3. If the backlight is on, long pressing the button for 2 seconds will also turn on the flashlight. Both will turn off after 30 seconds of inactivity.
Multimeter with backlight and flashlight on

Figure 6.1: Multimeter demonstrating backlight and flashlight features.

6.3. ACV and DCV Measurement

⚠️ Do not measure any voltage greater than AC600V or DC600V to prevent electric shock or damage to the instrument.

  1. Turn the rotary switch to the "mV" or "V" position. Press the "FUNC/Hz" key to select DC or AC.
  2. Connect the black test lead to the COM input socket and the red test lead to the V input socket.
  3. Use the test probes to measure the voltage across the circuit under test (connect in parallel).
  4. The display will show the measured voltage. For DC voltage, a "-" symbol indicates negative polarity if the red test lead is connected to the negative pole.

Note: In DC or AC 600mV and 6V ranges, if no test probes are connected, the meter may show readings. Short-circuit the "V-Ω" and "COM" terminals to zero the display.

Multimeter measuring DC voltage

Figure 6.2: Multimeter displaying a DC voltage measurement.

6.4. Resistance Measurement

⚠️ To avoid damage, cut off power to the circuit and discharge all high-voltage capacitors before measuring resistance.

  1. Turn the rotary switch to the Resistance/Continuity/Diode/Capacitance position. Press "FUNC/Hz" to select the "Ω" range.
  2. Connect the black test lead to the COM input socket and the red test lead to the "V/Ω" input socket.
  3. Use the test probes to measure the resistance of the circuit under test.
  4. The display will show the measured resistance value.

Note: Measured resistance on a circuit may differ from rated values due to parallel components. For accurate low resistance measurements, short-circuit the test leads first and subtract this value from the measurement. If "OL" is displayed, the value exceeds the range; switch to a higher range.

6.5. Diode Measurement

⚠️ To avoid damage, cut off power to the circuit and discharge all high-voltage capacitors before measuring diodes.

  1. Turn the rotary switch to the Resistance/Continuity/Diode/Capacitance position. Press "FUNC/Hz" to select the Diode range.
  2. Connect the black test lead to the COM input socket and the red test lead to the "V/Ω" input socket.
  3. Connect the black test probe to the cathode and the red test probe to the anode of the diode.
  4. The meter will display the forward conduction voltage. If polarity is reversed, "OL" will be displayed.

6.6. Buzzer Test (Continuity)

⚠️ To avoid damage, cut off power to the circuit and discharge all high-voltage capacitors before performing continuity tests.

  1. Turn the rotary switch to the Resistance/Continuity/Diode/Capacitance position. Press "FUNC/Hz" to select the Continuity range.
  2. Connect the black test lead to the COM input socket and the red test lead to the "V/Ω" input socket.
  3. Use the test probes to measure the resistance of the circuit. If resistance is less than 60Ω, the indicator light will illuminate, the buzzer will sound continuously, and the resistance value will be displayed.

6.7. Capacitance Measurement

⚠️ To avoid damage, cut off power to the circuit and discharge all high-voltage capacitors before measuring capacitance.

  1. Turn the rotary switch to the Resistance/Continuity/Diode/Capacitance position. Press "FUNC/Hz" to select the Capacitance range.
  2. Connect the black test lead to the COM input socket and the red test lead to the "V/Ω" input socket.
  3. Touch the two terminal pins of the capacitor with the test probes.
  4. The display will show the measured capacitance value.

Note: Large capacitance measurements may take time to stabilize. When measuring polarized capacitors, observe correct polarity to prevent damage and ensure accuracy.

6.8. Frequency Measurement

⚠️ Do not test any voltage higher than 250V to prevent electric shock or damage to the instrument.

  1. Turn the rotary switch to the "Hz%" position.
  2. Connect the black test lead to the COM input socket and the red test lead to the "Hz" input socket.
  3. Use the test probes to measure the frequency of the circuit under test.
  4. Press the "FUNC/Hz" key to display frequency and duty cycle respectively.

6.9. Current Measurement (AC/DC)

⚠️ Never attempt to make current measurements on a circuit when the open circuit voltage to ground exceeds 250V. If the fuse blows, the meter may be damaged or you may be injured. Check the meter's fuse before measuring current. Do not connect test probes in parallel to any circuit when plugged into current input sockets (μA, mA, A).

  1. Turn the rotary switch to the "μA", "mA", or "A" position. Press "FUNC/Hz" to select AC or DC.
  2. Connect the black test lead to the COM input socket. For currents less than 600mA, connect the red test lead to the VΩMA input socket. For currents between 600mA and 10A, connect the red test lead to the 10A input socket.
  3. Disconnect the circuit to be tested. Connect the black test lead to the lower voltage end of the disconnected circuit and the red test lead to the higher voltage end.
  4. Connect power to the circuit and read the display. If "OL" is displayed, the input exceeds the selected range. Power off and switch to a higher range.
  5. In ACA measurement state, long press "FUNC/Hz" to display the frequency of the AC signal (0~1KHz).
Red and black test probes

Figure 6.3: Red and black test probes for current measurement.

6.10. NCV Test (Non-contact Voltage Detection)

  1. Rotate the rotary switch to the "NCV" position.
  2. Place the top of the meter (NCV sensing area) close to the conductor.
  3. If AC voltage is detected, the indicator light will illuminate based on signal strength, and the buzzer will emit alarm sounds at different frequencies.

Note: Voltage may still be present even if not indicated. Do not rely solely on NCV detectors. External interference (flashlights, motors) may trigger false positives.

Multimeter performing NCV test

Figure 6.4: NCV (Non-contact Voltage Detection) in use.

6.11. Temperature Measurement

  1. Turn the rotary switch to the "°C (°F)" position. The instrument will display the ambient temperature.
  2. Remove the test leads and insert the thermocouple into the "COM" and "VΩMA" input terminals, observing correct polarity.
  3. The meter will display the approximate temperature value from the thermocouple.
Multimeter performing temperature test

Figure 6.5: Multimeter measuring temperature with a thermocouple.

6.12. Inductance Measurement

  1. Turn the rotary switch to the "L" position.
  2. Connect the black test lead to the COM input socket and the red test lead to the "VΩMAL" input socket.
  3. Use the test probes to touch both ends of the inductance device.
  4. The display will show the measured inductance value.
Multimeter performing inductance measurement

Figure 6.6: Multimeter measuring inductance.

7. Maintenance

7.1. General Maintenance

  • To avoid electric shock or damage, do not get the inside of the meter wet.
  • Before opening the case or battery cover, turn off the power and ensure test probes are disconnected.
  • Clean the instrument case regularly with a damp cloth and a small amount of detergent. Do not use abrasives or chemical solvents.
  • Dirty or damp input sockets can cause oxidation, affecting readings.

7.2. Battery and Fuse Replacement

Refer to the Setup section (5.1 Battery Installation) for battery replacement instructions.

Fuse Replacement:

Only use designated fuses (600mA/250V and 10A/250V fast-blow fuses). The spare fuse is located under the battery cover.

  1. Turn off the instrument and remove all test probes.
  2. Open the battery cover as described for battery replacement.
  3. Carefully remove the old or damaged fuse.
  4. Replace with a new, designated fuse.
  5. Install the battery cover and close the screws.

Note: When the low voltage symbol (battery icon) appears, replace the battery immediately.

8. Specifications

8.1. Comprehensive Indicators

  • Operating conditions: 600V CAT II, Pollution grade: 2, Height: under 2000m
  • Working temperature: 0-40 °C (<80%RH)
  • Storage temperature: -10-60 °C (<70%RH, remove battery)
  • Test/calibrate temperature: 25 °C ±2°C
  • Max voltage between measurement end and ground: 600V
  • Fuse protection: μA/mA fuse 600mA/250V; A fuse 10A/250V
  • Conversion rate: ~3s/second
  • Display: 6000 counts LED display
  • Overload: 'OL' displayed
  • Low voltage display: Battery icon displayed when working voltage is low
  • Input polarity indicator: '-' shown automatically
  • Battery: 3 x 1.5V AAA
  • Dimensions: 147mm (L) x 71mm (W) x 45mm (H)
  • Weight: ~220g (battery excluded)

8.2. Accuracy Index

Accuracy: ± (%reading + digit), one year warranty from manufacture date. Conditions: surroundings temperature from 18 °C to 28 °C, <80%RH.

Direct Current Voltage (DCV)

Range Resolution Accuracy
60mV0.01mV± (0.5%Readings+5digits)
600mV0.1mV
6V0.001V
60V0.01V
600V0.1V

Input resistance: 10MΩ

Maximum input voltage: 600VDC

Alternative Current Voltage (TRMS)

Range Resolution Accuracy
60mV0.01mV± (1%Readings+4digits)
600mV0.1mV
6V0.001V
60V0.01V
600V0.1V

Input resistance: 10MΩ

Maximum input voltage: AC600V

Frequency response: 40HZ to 1KHZ true RMS

Resistance

Range Resolution Accuracy
600Ω0.1Ω± (0.8%Readings+5digits)
6KΩ0.001KΩ
60KΩ0.01KΩ
600KΩ0.1KΩ
6MΩ0.001MΩ
60MΩ0.01MΩ

Overload protection: 250V DC/AC

Open circuit voltage: 0.52V

Diode and Buzzer

Function Testing condition Open-circuit voltage
DiodeDisplays approximate forward voltage of the diode3.2V
BuzzerBuzzer sounds and red indicator lights up when internal resistance is < 60Ω1V

Overload protection: 250V DC/AC

Direct Current (DCA)

Range Resolution Accuracy
600μA0.1μA± (1.2%Readings+5digits)
6000μA1μA
60mA0.01mA± (1.2%Readings+5digits)
600mA0.1mA
6A0.001A± (3%Readings+5digits)
10A0.01A

Overload protection: μA/mA range fuse (600mA/250V); 10A range fuse (10A/250V).

Note: For currents > 10A, continuous measurement time should not exceed 10 seconds to prevent solder joint melting. Allow 1 minute rest after measurement.

Alternative Current (ACA)

Range Resolution Accuracy
600μA0.1μA± (1.2%Readings+5digits)
6100μA1μA
60mA0.01mA± (1.2%Readings+5digits)
600mA0.1mA
6A0.001A± (3%Readings+5digits)
10A0.01A

Overload protection: μA/mA range fuse (600mA/250V); 10A range fuse (10A/250V).

Maximum input current: mA level: 600mA AC effective value; 10A level: 10A AC effective value.

Frequency response: 40Hz-1KHz, true RMS

Frequency

Range Resolution Accuracy
9.999Hz0.001Hz± (1.5%Readings+5digits)
99.99Hz0.01Hz
999.9Hz0.1Hz
9.999KHz0.001KHz
99.99KHz0.01KHz
999.9KHz0.1KHz
9.999MHz0.001MHz

Input voltage range: 200mV-10V AC RMS

Overload protection: 250V DC/AC

Capacitance

Range Resolution Accuracy
6000pF1pF± (4.0%Readings+5digits)
60nF0.01nF
600nF0.1nF
6µF0.001µF
60μF0.01 µF
600μF0.1μF
6mF0.001mF
60mF0.01mF
100mF0.1mF

Overload protection: 250V DC/AC

Temperature

Range Resolution Accuracy
°C1°C± (1.0%+3) Readings (-20°C~1000°C)
°F1°F± (1.0%+3) Readings (-4°F~1832°F)

Overload protection: 250V DC/AC

Inductance

Range Resolution Accuracy
600μH0.1μH± (4.0%Readings+5digits)
6mH0.001mH
60mH0.01mH
600mH0.1mH
6H0.001H
60H0.01H

Overload protection: 250V DC/AC

9. User Tips

  • Packaging: If you are concerned about shipping, consider requesting extra protective packaging like bubble wrap.
  • Power Source: The multimeter operates on 3 AAA batteries, which are convenient to replace.
  • Probes: While the included probes are functional, some users suggest considering replacement probes for higher precision or specific applications.
  • Display: The large LED screen provides clear and easy-to-read measurements from various angles.
  • Flashlight: The built-in flashlight is quite bright and useful for working in dimly lit areas.
  • Magnetic Design: The magnetic back allows for convenient hands-free operation by attaching the multimeter to metal surfaces.
  • Inductance Measurement: This model specifically supports inductance measurement, a key feature for many electrical tasks.
  • Spare Fuses: The product typically includes spare glass fuses, which are located under the battery cover.
Multimeter showing magnetic design

Figure 9.1: Multimeter demonstrating its magnetic design for hands-free use.

10. Troubleshooting

  • No Display/Power On: Check battery installation and ensure batteries are not depleted. Replace if necessary.
  • "OL" Displayed: This indicates an overload or that the measured value exceeds the selected range. Switch to a higher range or check for an open circuit.
  • Inaccurate Readings: Ensure test probes are properly connected and not damaged. Verify the correct measurement function and range are selected. Check battery level.
  • Meter Enters Sleep Mode: This is normal after 15 minutes of inactivity. Press any button or rotate the gear switch to resume operation.
  • Fuse Blown: If the meter stops functioning during current measurement, the fuse may be blown. Replace it with a designated spare fuse (see Maintenance section).

11. Warranty & Support

The NJTY T21L Digital Multimeter comes with a one-year warranty from the date of manufacture, covering defects in materials and workmanship under normal use. For technical support, service, or warranty claims, please contact your retailer or the manufacturer directly.

A user manual in PDF format is available for download: User Manual (PDF)