Introduction
This manual provides detailed instructions for the setup, operation, maintenance, and troubleshooting of your AmScope T570 Series Trinocular Compound Microscope. This research-grade instrument features an infinity-corrected optical system with semi-plan achromatic lenses and a Koehler illumination system, making it ideal for clinics, laboratories, and higher education.
The trinocular design allows for comfortable viewing and easy attachment of a camera for image capture. With a magnification range from 40X to 3000X, this microscope is designed for precise observation of minute details in biological specimens.
1. Setup and Assembly
1.1 Unpacking and Component Identification
Carefully unpack all components from the shipping carton and verify against the packing list:
- One Trinocular Compensation-Free Head
- One Microscope Body with Frame, Base, and Koehler Illumination System
- Four High Quality Semi-Plan Objectives: 4X, 10X, 40X, and 100X
- One Pair of Extreme Widefield Eyepieces: WF10X
- One Pair of Extreme Widefield Eyepieces: WF30X
- One Dust Cover
1.2 Assembly Instructions
- Attach the Trinocular Head: Loosen the head locking screw on the microscope body. Carefully place the trinocular head onto the top of the microscope body, ensuring it seats properly. Tighten the head locking screw to secure it.
- Insert Eyepieces: Remove the protective caps from the eyepiece tubes. Insert the WF10X eyepieces into the two ocular tubes on the trinocular head.
- Install Objectives: Gently screw each objective lens (4X, 10X, 40X, 100X) into the revolving nosepiece in increasing order of magnification, ensuring they are securely fastened.
- Connect Power: Locate the power input port at the rear of the microscope base. Connect the provided power cord to the microscope and then to a suitable electrical outlet (85-230VAC 50/60Hz).

Figure 1: Front-left view of the AmScope T570D microscope, showing eyepieces, objectives, stage, and focus knobs.

Figure 2: Rear-right view of the AmScope T570D microscope, highlighting the power input port and main power switch.
2. Operating Instructions
2.1 Basic Operation
- Power On: Flip the main power switch located on the base of the microscope. The illumination lamp should turn on. Adjust the brightness using the dimmer control.
- Place Specimen: Place a prepared microscope slide onto the mechanical stage. Use the stage clips to hold it gently in place. Use the X-Y translation knobs to center the specimen area of interest under the objective lens.
- Adjust Interpupillary Distance: While looking through the eyepieces, adjust the distance between the two eyepiece tubes until you see a single, circular field of view.
- Focusing:
- Rotate the revolving nosepiece to select the lowest power objective (e.g., 4X).
- Using the coarse focus knob, raise the stage until the objective is just above the slide, then slowly lower it while looking through the eyepieces until the specimen comes into rough focus.
- Use the fine focus knob for precise focusing.
- Once focused with the 4X objective, you can switch to higher power objectives. Minor fine focus adjustments may be needed.
- Adjusting Magnification: Rotate the revolving nosepiece to select the desired objective lens (4X, 10X, 40X, 100X). The total magnification is the eyepiece magnification (e.g., 10X or 30X) multiplied by the objective magnification.

Figure 3: Close-up view of the mechanical stage, stage clips, and objective lenses in position.

Figure 4: Close-up of the coarse and fine focus knobs, along with the X-Y translation knobs for stage movement.
2.2 Koehler Illumination System
The Koehler illumination system provides even and optimal illumination for high-quality imaging.
- Adjust Field Diaphragm: Close the field diaphragm (located at the base of the microscope, below the stage) until its image is visible in the field of view. Use the condenser centering screws to center the diaphragm image. Open the field diaphragm just enough to illuminate the entire field of view.
- Adjust Condenser Aperture Diaphragm: The condenser (located directly below the stage) has an iris diaphragm. Adjust this diaphragm to control contrast and resolution. A good starting point is to close it until it is about 70-80% open relative to the numerical aperture (NA) of the objective in use.
- Adjust Condenser Height: Use the condenser focus knob to raise or lower the condenser for optimal illumination. The condenser should generally be near its highest position for most observations.
2.3 Using Oil Immersion (100X Objective)
The 100X objective is designed for use with immersion oil to achieve maximum resolution.
- Focus on your specimen with the 40X objective.
- Rotate the nosepiece halfway between the 40X and 100X objectives.
- Place a small drop of immersion oil directly onto the center of the specimen slide.
- Carefully rotate the 100X objective into the oil drop. Ensure the objective makes contact with the oil.
- Look through the eyepieces and use the fine focus knob to bring the specimen into sharp focus.
- After use, always clean the 100X objective lens and the slide immediately with lens paper moistened with lens cleaning solution to remove all traces of oil.

Figure 5: Close-up view of the objective lenses, showing their magnifications (4X, 10X, 40X, 100X Oil).
2.4 Trinocular Port Usage
The trinocular port allows for the attachment of a microscope camera (sold separately).
- Remove the protective cap from the trinocular port.
- Attach your camera adapter and camera to the port.
- Adjust the height of the photo-port tube (if applicable) to achieve parfocality between the eyepieces and the camera view.
- Consult your camera's manual for specific software and operation instructions.
3. Maintenance
3.1 Cleaning
- Optical Components: Use only specialized lens cleaning solution and lens paper or cotton swabs for cleaning eyepieces and objective lenses. Gently wipe in a circular motion from the center outwards. Never use abrasive materials or solvents.
- Microscope Body: Clean the painted and plastic surfaces with a soft, damp cloth. For stubborn stains, a mild detergent can be used. Avoid getting liquids into electrical components.
- Immersion Oil: Always clean immersion oil from the 100X objective immediately after use.
3.2 Storage
- When not in use, cover the microscope with the provided dust cover to protect it from dust and debris.
- Store the microscope in a cool, dry place, away from direct sunlight, high humidity, and corrosive chemicals.
- If storing for extended periods, remove eyepieces and objectives and store them in a dry, sealed container with desiccant.
4. Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| No illumination | Power cord not connected; power switch off; dimmer set to minimum. | Check power connections; turn on switch; increase dimmer setting. |
| Image is blurry or unclear | Improper focus; objective or eyepiece dirty; condenser not adjusted correctly; slide upside down. | Adjust fine focus; clean lenses; adjust condenser and diaphragms; check slide orientation. |
| Dark spots in field of view | Dust on eyepiece, objective, or slide. | Clean eyepieces, objectives, and slide. Rotate eyepiece to determine if dust moves with it. |
| Cannot achieve sharp focus at high magnification | Immersion oil not used with 100X objective; objective dirty. | Apply immersion oil for 100X objective; clean objective lens. |
5. Specifications
| Optical System | Infinity-corrected |
| Head | Trinocular, 30° incline, 360° rotatable |
| Interpupillary Adjustment | Compensation-free gemel, 48mm-75mm |
| Ocular Tube Diameter | 30mm |
| Eyepieces | 10X FN 22mm, 30X |
| Photo-port | Height-adjustable 23mm |
| Nosepiece | Quadruple reverse |
| Objective Lenses | Semi-plan achromatic 4X, 10X, 40X (spring), 100X (spring, oil) |
| Focusing System | Coaxial coarse and fine focus |
| Stage | Double-layer 180mm x 160mm |
| X-Y Translation Range | 75mm x 50mm |
| Fine-focus Precision | 0.002mm |
| Sub-stage Condenser-holder | Rack-and-pinion, centerable |
| Condenser | Achromatic Abbe condenser, NA1.25, iris diaphragm |
| Transmitted Illumination | High power Koehler compatible |
| Transmitted Light Field Lens | Iris diaphragm |
| Power | 85-230VAC 50/60Hz |
5.1 Objective Lenses
| Magnification | Corrections | NA | Cover-glass Thickness |
| 4X | Semi-plan achromatic | 0.10 | 0.17mm |
| 10X | Semi-plan achromatic | 0.25 | 0.17mm |
| 40X | Semi-plan achromatic | 0.65 | 0.17mm |
| 100X | Semi-plan achromatic | 1.25 | 0.17mm |
5.2 Magnification Chart
| Eyepiece | Objective | |||
| 4X | 10X | 40X | 100X | |
| 10X | 40X | 100X | 400X | 1000X |
| 30X | 120X | 300X | 1200X | 3000X |
6. User Tips
- Always start observation with the lowest magnification objective (4X) to locate your area of interest and then gradually increase magnification.
- When switching to higher power objectives, only use the fine focus knob for adjustments to avoid damaging the slide or objective.
- Ensure your specimen slides are clean and free of air bubbles for optimal viewing.
- For best results with Koehler illumination, take the time to properly adjust the field and aperture diaphragms for each objective.
- Use a dedicated microscope cleaning kit for all optical components to prevent scratches and maintain image quality.
7. Warranty and Support
For warranty information, technical support, or service inquiries, please contact your authorized AmScope dealer or visit the official AmScope website. Please have your model number (T570D) and purchase information ready when contacting support.