Hantek TO1112/TO1112C/TO1112D

Contents

Hantek TO1000 Series Handheld Oscilloscope User Manual

Models: TO1112, TO1112C, TO1112D

1. Product Overview

The Hantek TO1000 series is a new generation of handheld flat oscilloscopes, featuring a 7-inch industrial specification TFT-LCD (800x480 resolution) multi-point capacitive touch screen for intuitive operation. These devices integrate multiple functions, offering a versatile solution for various testing needs.

The series includes the following models:

  • TO1112: Digital Storage Oscilloscope function only.
  • TO1112C: Digital Storage Oscilloscope + High-precision Multimeter functions.
  • TO1112D: Digital Storage Oscilloscope + High-precision Multimeter + Arbitrary Function Signal Generator functions.

Key features across the series include:

  • Channels: 2-channel (TO1112, TO1112C, TO1112D) or 4-channel (other TO1000 series models not explicitly listed here).
  • Bandwidth: Up to 110MHz.
  • Sampling Rate: 250MSa/s (for 2-channel models).
  • Storage Depth: 8Mpts (single channel).
  • Display: 7-inch industrial TFT-LCD with 800x480 resolution, multi-point full touch capacitive screen.
  • Battery: Built-in 10000mAh detachable lithium battery (for 2-channel models).
  • Charging: Supports multiple fast charging protocols, standard 18W charger, Type-C interface.
  • Design: Slim, lightweight, easy to carry, unique suspended foot design for hands-free use, anti-fall and anti-seismic sturdy construction.
  • Connectivity: Standard USB port for data transfer and PC connection.

The oscilloscope offers a user-friendly and easy-to-use UI for a good interaction experience.

Video: Hantek TO1000 Series Handheld Oscilloscope product overview, showcasing its multi-functionality, touch screen interface, portability, and durability.

Hantek TO1112 Handheld Oscilloscope with key features

Image: Hantek TO1112 Handheld Oscilloscope highlighting its sampling rate (250MSa/s), 2 channels, and 110MHz bandwidth.

Hantek TO1112D Oscilloscope with integrated functions

Image: Hantek TO1112D Oscilloscope displaying its integrated functions: Digital storage oscilloscope, Arbitrary function signal generator, High-precision multimeter, Data recorder, FFT spectrum analyzer, and Frequency meter.

2. Safety Information

Read the following safety precautions carefully to avoid injury and to prevent damage to this product or any connected equipment. Use this product in accordance with regulations.

2.1 General Security Issues

  • Repairs should only be performed by professionally authorized personnel.
  • Use the correct power cable approved for the country where the product is used.
  • Connect and disconnect probes correctly: Connect the probe to the oscilloscope first, then to the circuit. Disconnect from the circuit first, then from the oscilloscope.
  • Ground the product properly to avoid electric shocks.
  • Connect the probe properly: The ground wire of the probe is at ground potential. Do not connect it to high voltage. Keep fingers behind the safety ring.
  • Observe all terminal rating values to avoid fire or excessive current.
  • Do not operate with the cover open.
  • Avoid circuit exposure; do not touch exposed connectors or components when power is on.
  • If the product is suspected to be faulty or damaged, contact qualified maintenance personnel.

2.2 Working Conditions

  • Operating Temperature: 0°C to 50°C
  • Storage Temperature: -20°C to 70°C
  • Humidity: ≤90% RH (at ≤40°C), ≤60% RH (at 41°C to 50°C).
  • Altitude: 3,000m (in and out of operation).
  • Vibration: 0.31g RMS (50Hz to 500Hz) for 10 minutes in each axial direction during operation; 2.46g RMS (5Hz to 500Hz) for 10 minutes in each axial direction when not operating.

Warning: To avoid short circuit or electric shock, do not operate the device in a damp environment.

2.3 Pollution Level

The device is rated for Pollution Degree 2, meaning generally only dry non-conductive pollution occurs. Transient conduction due to condensation may sometimes occur (typical for general indoor environments).

3. Setup and Initial Preparation

3.1 Package Contents

Upon receiving your oscilloscope, verify the package contents:

  • Oscilloscope unit
  • X1/X10 Probe
  • Crocodile clip testing line
  • USB computer communication cable
  • Power plug (18W charger, Type-C)
  • Instrument package
  • For TO1112C and TO1112D: Additional multimeter test line
  • For TO1112D: Additional crocodile clip testing line

If any accessories are missing or damaged, contact your dealer.

Hantek TO1112C Oscilloscope with included accessories

Image: Hantek TO1112C Oscilloscope and its standard accessories, including the main unit, carrying case, probe, HT324 cable, 18W charger with Type-C cable, and multimeter probes.

TO1000 Series Accessories Table

Image: A visual representation of the TO1000 Series accessories, including the oscilloscope, carrying case, probe, HT324 cable, 18W charger, and multimeter probes (for TO1002C/D series standard).

3.2 Powering On and Charging

The oscilloscope is equipped with 4x 3.7V 2600 mAh batteries. Ensure batteries have sufficient power before use.

  • To Power On/Off: Press the power button located at the bottom of the right panel. A buzzer will sound upon startup.
  • Battery Status: A blank battery icon on the screen indicates low battery. The oscilloscope will prompt "shut down after 30s" if the battery level drops below 3.5V.
  • Charging: Connect the oscilloscope to a power socket using the provided USB data cable and 18W power adapter. The Type-C interface is used for charging. The power button indicator light will be on during charging when the machine is off. If the machine is on, the battery icon on the screen will indicate charging status (slow charge / quick charge).

3.3 Language Settings

The oscilloscope supports Chinese and English. From the main interface, tap [HOME] > [Settings] > Language to select your preferred language. The default language is English.

3.4 Functional Check

To enter the oscilloscope interface, tap [OSC] from the main interface. To restore factory default configurations, tap Utility > Default in the lower left corner of the screen.

3.5 Probe Compensation

Probe compensation ensures accurate measurements. Perform this adjustment when connecting a probe for the first time or if measurements are inaccurate.

  • Connection: Connect one end of the probe to the CH1 channel and the other end to the AFG channel.
  • Adjustment: In the oscilloscope interface, tap Utility > Probe calibrate. Set the probe switch to 10X as prompted.
  • Waveform Check: Observe the waveform. If it's overcompensated or undercompensated, use a non-metallic screwdriver to adjust the variable capacitance on the probe until the waveform is "compensated correctly."

4. Operation

4.1 Main Interface and Navigation

The main interface provides quick access to various functions:

  • Oscilloscope Icon: Tap to enter the oscilloscope page.
  • Multimeter Icon: Tap to enter the multimeter page (if supported by model).
  • Signal Source Icon: Tap to enter the signal source page (if supported by model).
  • File Browsing: Tap to view saved pictures, waveforms, and settings.
  • Settings: Access language, system information, sound, brightness, upgrade, auto lock, and USB sharing.
  • Gallery: View saved pictures.
  • Battery: Displays real-time battery level.
  • Power Switch: Shuts down the device.
  • USB Disk Icon: Appears when a USB flash drive is inserted.
  • Hantek Icon: Displays product information.

4.2 Touch Screen Gestures

The device is primarily operated via its capacitive touch screen, supporting multi-touch gestures:

  • Touching: Tap menus, functions, virtual keyboards (character/numeric) to select or input.
  • Kneading: Use two fingers to zoom in/out on waveforms (spread fingers to enlarge, pinch to shrink). Vertical kneading adjusts vertical gain, horizontal kneading adjusts horizontal time base.
  • Dragging: Drag targets (e.g., waveforms, cursors, triggers) to change their position or displacement.

4.3 Setting Parameters

Tap on the parameter to be set, and a virtual keyboard (character or numeric) will pop up for input. Long press up/down arrows to change values continuously.

4.4 Horizontal System

The horizontal system controls the time base and displacement of waveforms.

  • Adjust Horizontal Displacement: Drag the waveform or the horizontal trigger position icon to move the waveform left/right relative to the screen center.
  • Adjust Horizontal Time Base (Time/Grid): Tap the +/- buttons at the bottom of the screen or use kneading gestures to change the time base (s/div).
  • Pan or Zoom: After stopping waveform collection, you can pan and zoom the displayed data using touch gestures.

4.5 Vertical System

The vertical system controls the sensitivity, offset, and coupling of each channel.

  • Open/Close Analog Channel: Tap the channel icon to toggle between open, open with vertical menu, and closed. Yellow indicates open, gray indicates closed.
  • Adjust Vertical Sensitivity (V/div): Tap the +/- buttons next to the channel icon or use vertical kneading gestures to adjust the waveform amplitude.
  • Adjust Vertical Offset: Drag the waveform or use touch gestures to move the waveform up/down relative to the zero point.
  • Channel Coupling: Set to DC (both AC/DC components pass), AC (DC component blocked), or Grounded (both AC/DC components blocked).
  • 20M Bandwidth Limit: Enable/disable to reduce noise in the display waveform.
  • Waveform Inversion: Inverts the displayed waveform.
  • Probe: Set the probe attenuation ratio (e.g., 1X, 10X) for accurate measurements.
  • Amplitude Unit: Select between Voltage (default) or Current.

4.6 Mathematical Operations

The oscilloscope supports various mathematical operations on analog channel waveforms.

  • Algebraic Operations: Addition (+), subtraction (-), multiplication (*), and division (/) between two data sources (CH1-CH4). Access via Next > MATH.
  • FFT Operation: Converts time-domain signals to frequency-domain. Access via Next > MATH > FFT. Configurable parameters include data source, vertical units (VRms or dB), and window functions (Hanning, Hamming, Blackman, Bartlett, Flat top, Rectangle).

4.7 Reference Channel

You can save and recall reference waveforms.

  • Save: Tap Utility > Save > Ref wave, select data sources (CH1-CH4), and tap save.
  • Recall: From the oscilloscope interface, tap Next > Reference waveform. Select a saved file to display.
  • Movement/Scaling: Reference waveforms can be moved and scaled independently of channels.

4.8 Trigger Oscilloscope

Triggering sets conditions for capturing and displaying waveforms.

  • Trigger Level: Adjust the signal voltage corresponding to the trigger point.
  • Trigger Data Source: Select CH1-CH4 as the trigger source.
  • Trigger Mode: Automatic (acquires without valid trigger), Normal (acquires only with valid trigger), Single (stops after one acquisition).
  • Trigger Hold-off: Sets the time the oscilloscope waits before re-triggering, useful for complex waveforms.
  • Trigger Types: Edge-triggered (rising, falling, double edge), Pulse Width Trigger (positive/negative pulses of specified width), Video Trigger (NTSC, PAL, HDTV standards), Slope Trigger (time between two voltage levels), Timeout Trigger (time interval between edges).

4.9 Measurements

The oscilloscope offers various measurement methods.

  • Grid Scale Measurement: Estimate frequency and voltage amplitude using the grid.
  • Automatic Measurements: Up to 42 waveform parameters can be measured automatically. Tap Measure, select source (CH1-CH4), and choose parameters (e.g., Peak-Peak, Max, Min, Frequency, Period, Rise Time, Fall Time, Duty Cycle, etc.).
  • Cursor Measurement: Use vertical (time) and horizontal (amplitude) cursors to measure specific points on the waveform.

4.10 Acquisition

Acquisition settings determine how waveforms are sampled and displayed.

  • Acquisition Mode: Normal (samples at equal intervals), Average (reduces random noise), Peak Detection (captures narrow pulses), High Resolution (reduces random noise for smoother waveforms).
  • Average Times: Set the number of averages (2-256) in Average mode.
  • Display Mode: YT (voltage vs. time, default), XY (voltage vs. voltage, for phase difference analysis), Scroll (waveform refresh from right to left, for low frequencies).
  • Storage Depth: Number of waveform points stored (4K, 20K, 200K, 2M, 4M, 8M). Higher depth ensures higher sampling rate at a given time base.
  • Interpolation Methods: Sine wave, Step, Linear interpolation to reconstruct waveforms from discrete samples.

4.11 Multimeter Function (TO1112C/D only)

Access the multimeter function by tapping the [Multimeter] icon on the main interface or [DMM] virtual key.

  • Measurements: DC voltage (V-, mV-), AC voltage (V~), Resistance (Ω), Diode, On/Off measurement, DC current (A-, mA-), AC current (A~, mA-), Capacitance.
  • Display: Real-time numerical display, trend diagrams, and numerical tables.
  • Connection: Use the multimeter probes. Insert black probe into COM, red probe into V/Ω/C (for voltage/resistance/capacitance/diode) or mA/A (for current).

4.12 Signal Generator Function (TO1112D only)

Access the signal generator by tapping the [Signal Source] icon or [AFG] virtual key.

  • Waveform Types: Sine wave, square wave, triangle wave, noise, DC, and arbitrary waveforms.
  • Parameters: Set frequency, period, amplitude, high/low level, offset, phase, duty cycle (for square wave), symmetry (for triangle wave), output load impedance.
  • Modulation: Supports Amplitude Modulation (AM) and Frequency Modulation (FM). Set shape, frequency, depth, and offset of the modulated wave.
  • Burst: Generate a burst of pulses with configurable count.
  • Edit Arbitrary Waveform: Use the WaveEditor software on a PC to create and download arbitrary waveforms to the device via USB.

4.13 File Management

The device allows saving and browsing various data types.

  • Save Options: Pictures (.bmp), binary waveforms (.lwf), settings (.pho), reference waveforms (.ref), and CSV data. Saved to internal memory or USB flash drive.
  • File Browse: Access saved files from the main interface. Pictures can also be viewed in the Gallery.
  • Delete/Rename: Files can be deleted or renamed through the file browse menu.

4.14 Remote Control

The oscilloscope can be controlled via a PC using Keysight IO Libraries Suite software. Connect the device to a PC via USB, install the necessary drivers, and use the Interactive IO interface to send commands.

5. Specifications

5.1 General Oscilloscope Parameters

ParameterTO1112 Series (250M Sampling)
SeriesTO1000 Series (1G Sampling)
Horizontal Scale Range2ns/div~100s/div (1, 2, 5 sequence)
Incremental Time Measurement Accuracy±(1 sampling interval + 100ppm × reading + 0.6ns)
Measurement Accuracy (Full Bandwidth)≥16 times on average
A/D Converter8-bit resolution, each channel sampled simultaneously
Vertical Scale Range10mV/div~10V/div at input BNC
Vertical Resolution8bit
Vertical Sensitivity Range10mV/div~10V/div
Offset Range±1Vmax (100mV/div) to ±50Vmax (10V/div)
Dynamic Range±5div (8bit)
Maximum Input Voltage300VRMS (10X)
Digital Channels2
Record Length8M
Display Resolution800*480 Pixels
Display Size7 Inches & Above
Battery IncludedYes
IS Battery DemountableYes
CertificationCE
DIY SuppliesElectrical
Oscilloscope Parameter Table

Image: Detailed table of oscilloscope parameters including basic parameters, horizontal scale, measurement accuracy, vertical settings, and dynamic range.

5.2 Multimeter Parameters (TO1112C/D only)

Measurement ItemRangePrecisionResolution
DC Voltage400mV± (1%+2)100 μV
4.000V1mV
40.00V10mV
400.0V100mV
600.0V1V
Overload protection: 400mV: 250V, other: 600Vrms.
AC Voltage4.000V± (1.2%+5)1mV
40.00V10mV
400.0V100mV
600.0V1V
Frequency: 40Hz~400Hz
DC Current40.00mA± (1%+2)10 μA
200.0mA100 μA
Overload protection: self restoring fuse: 200mA/250V, 4A and 10A range no fuse.
AC Current4.000A± (3%+2)1mA
10.00A10mA
Frequency: 40Hz~400Hz
Self restoring fuse: 200mA/250V, 4A and 10A range no fuse.
Resistance400.0 Ω± (1.3%+2)0.1 Ω
4.000 KΩ1 Ω
60.00 KΩ10 Ω
400.0 KΩ100 Ω
4.000 MΩ1K Ω
Resistance40.00 MΩ± (1.5%±3)10K Ω
Overload protection: 220Vrms
Capacitance40.00nF± (3%+5)10pF
400.0nF100pF
4.000μF1nF
40.00μF10nF
100.0μF100nF
Diode0V~1.0V
On off Measurement< 50 Ω
Multimeter Parameters Table

Image: Detailed table of multimeter parameters including DC/AC Voltage, DC/AC Current, Resistance, Capacitance, Diode, and On/Off Measurement ranges, precision, and resolution.

5.3 Arbitrary Waveform Generator Parameters (TO1112D only)

ParameterValue
Channel Number1
Sampling Rate200MSa/s
Vertical Resolution12bit
Highest Frequency25MHz
Standard WaveformSine wave, Square wave, Triangle wave, Noise, DC
Sine Wave Frequency Range0.1Hz~25MHz
Square Wave/Pulse Frequency Range0.1Hz~10MHz
Square Wave/Pulse Duty Ratio0%~100%
Triangle Wave Frequency Range0.1Hz~1MHz
Triangle Wave Symmetry0%~100%
Noise Bandwidth> 25MHz
DC Offset (50 Ω)±1.5V
DC Offset (high resistance)±3.0V
DC Resolution100 μV
DC Precision (1KHz)2%
Arbitrary Wave Frequency Range0.1Hz~10MHz
Arbitrary Wave SupportUpper computer download, external memory calling out
Amplitude (50 Ω)0mV~3Vpp
Amplitude (high resistance)0mV~6Vpp
Frequency Resolution0.1 Hz or 4 bits (greater of the two)
Wavelength4KSa
Frequency Accuracy100 ppm (< 10 kHz), 50 ppm (> 10 kHz)
FM ModulationModulation Waveform: Sine, Square, Triangle
Modulation Frequency: 1Hz~50KHz
Modulation Deviation: 0.1Hz~1KHz
AM ModulationModulation Waveform: Sine, Square, Triangle
Modulation Frequency: 1Hz~50KHz
Modulation Depth: 0%~120%
Burst TypeMulti-period, Infinite
Burst Cycle Number1~2000000000
Burst Trigger SourceManual
Arbitrary Wave Generator Parameters Table

Image: Detailed table of Arbitrary Wave Generator parameters including channel, sampling rate, resolution, frequency, waveforms, and modulation settings.

5.4 Physical Dimensions

The device measures approximately 248 mm (width) x 176 mm (height).

Hantek TO1112D Oscilloscope with integrated functions and dimensions

Image: Hantek TO1112D Oscilloscope with integrated functions and an illustration of its physical dimensions.

6. Maintenance and Cleaning

6.1 General Maintenance

  • Avoid exposing the LCD to direct sunlight for prolonged periods during storage or placement.
  • Perform self-calibration if the temperature changes by more than 5°C to maintain measurement accuracy. Preheat the oscilloscope for 30 minutes before self-calibration. Disconnect all probes and cables from input connectors before calibration.

6.2 Cleaning Instructions

Regularly inspect and clean the outer surface of the instrument and probes:

  • Use a lint-free cloth to remove dust from the oscilloscope and probes. Be careful not to scratch the display filter material.
  • Clean the oscilloscope with a soft cloth dampened with water. For a more thorough cleaning, use a 75% isopropyl alcohol aqueous solution.
  • Note: Do not use any corrosive or chemical cleaning agents to avoid damaging the surface.
  • Ensure the device is completely dry before powering it on again to prevent electrical short circuits or personal injury due to moisture.

7. Troubleshooting

7.1 No Display After Power Button Press

  • Check if the battery is charged.
  • Ensure the power button is pressed firmly.
  • If the issue persists after checking and restarting, contact Hantek support.

7.2 No Waveform Displayed on Input Signal Screen

  • Verify the probe is properly connected to the BNC connector of the signal input channel.
  • Check if the channel switch is enabled (CH1 to CH4).
  • Confirm the input signal is correctly connected to the probe.
  • Verify that the measured circuit has signal output.
  • If it's a DC signal with high amplitude, adjust the amplitude gear to a higher value.
  • Press the automatic measurement button to auto-detect the signal.
  • If no waveform is still displayed, contact Hantek for technical support.

7.3 Serious Distortion of Input Signal Waveform

  • Check if the oscilloscope probe is properly connected to the BNC connector.
  • Check if the probe is properly connected to the object under test.
  • Ensure the probe is well calibrated. If not, perform probe calibration as described in the manual.

7.4 Waveform Constantly Scrolling, Cannot Be Triggered

  • Check if the trigger source is consistent with the signal input channel.
  • Verify the trigger level is properly adjusted.
  • Check if the trigger mode is appropriate. The default is edge trigger; select the correct trigger for different input signals.

8. User Tips

  • Check Dimensions: For portability and fitting into workspaces, always verify the device dimensions (approx. 248mm x 176mm) before use or transport.
  • Battery Life: Utilize the fast charging feature and the 10000mAh battery for extended use. Keep a Type-C charger handy, as it's compatible with many mobile phones.
  • Hands-Free Operation: Make use of the unique suspended foot design to position the oscilloscope conveniently on a workbench, freeing your hands for other tasks.
  • Multi-functionality: Remember that models TO1112C and TO1112D offer multimeter and waveform generator functions respectively, in addition to the oscilloscope. Switch between modes easily via the main interface icons.

9. Warranty and Support

9.1 Warranty Summary

Qingdao Hantek Electronics Co., LTD. (Hantek) warrants that the host and accessories of this product are free from material and process defects during the warranty period. If the product is found to be defective during this period, Hantek will repair or replace it free of charge.

For detailed warranty regulations, repair services, or full warranty instructions, please refer to the Hantek official website or contact a Hantek repair center or local office.

Hantek disclaims all other warranties, express or implied, including but not limited to, any implied warranties of merchantability and fitness for a special purpose. Hantek is not liable for indirect, special, or consequential damages.

9.2 Contact Information

If you have any questions or require service and support, please contact Hantek through the following channels:

  • Email: service@hantek.com, support@hantek.com
  • Website: http://www.hantek.com
  • Address: #35 Building, No. 780 Baoyuan Road, High-tech Zone, Qingdao, Shandong, China 266114
  • Switchboard: 400-036-7077
  • Tel: (0086)532-55678770 & 55678772 & 55678773