1. Overview
Thank you for purchasing the Professional Series COM-300 Meter. The COM-300 is an advanced waterproof combo meter with superior accuracy. This meter measures four parameters: pH, Electrical Conductivity (EC), Total Dissolved Solids (TDS), and temperature. It can be used to test the acidity or alkalinity of almost any liquid, provided the liquid contains less than 50% alcohol and a conductivity of at least 10 µS (5 ppm TDS (0.5 scale)). The COM-300 features 2 different conductivity conversion factors to convert EC to TDS using the 0.5 (NaCl) or 0.7 (442TM) as well as selectable modes for measurements in the µS, mS, and ppm scales. The meter incorporates automatic temperature compensation (ATC) to accommodate for discrepancies in temperature that would naturally occur. The meter also features digital calibration for easy and precise calibration that will yield better results.
Product demonstration video for the HM COM-300 Combo Meter.
2. Specifications
| Parameter | Range | Resolution | Accuracy |
|---|---|---|---|
| pH | 0.0 – 14.0 pH | 0.1 pH | +/- 0.1 pH |
| EC | 0 - 9990 µS; 0 - 9.99 mS | 0-999: 1 µS; 1000-9990: 10 µS; 0-9.99 mS: 0.01 mS | +/- 2% |
| TDS | 0 - 8560 ppm (mg/L) (0.5 NaCl scale; 0.7 442 scale) | 0-999: 1 ppm; 1000-9990: 10 ppm | +/- 2% |
| Temperature | 0.1 - 50°C; 32 - 122°F | 0.1 °C/°F | +/- 1 °C/°F |
- EC-to-TDS Conversion Factor: Pre-programmed, non-linear conversions for 0.5 (NaCl) or 0.7 (442™) solutions, selected by the user.
- Temperature Compensation: Automatic (ATC) to 25°C (ATC range: 1-50°C)
- Calibration: pH: Push Button Auto-calibration to 4, 7 or 10 pH for a fine tune digital calibration. EC/TDS: Digital calibration by push button (to any point within the range)
- Auto Shut-Off: After 5 minutes
- Probe: Detachable electrodes
- Display: LCD panel
- Housing: IP-67 Waterproof (Submersible, floats)
- Power source: 2 x AAA batteries (not included)
- Battery Life: Approximately 100 hours of usage (high ranges use greater power consumption)
- Dimensions: 18.5 x 3.4 x 3.4 cm (7.3 x 1.3 x 1.3 inches)
- Weight: 124g (4.4 oz) with batteries

Detailed specifications for the HM COM-300 Combo Meter.
3. Housing and Screen Description
Housing Components:
- Lanyard hook
- Battery compartment
- LCD display
- Button panel
- Power/Enter button
- PH/EC mode, calibration mode, calibration-up button
- EC/TDS mode, hold, calibration-down button
- Sensor locking ring
- Detachable sensor (Part # SP-C3)
LCD Display Indicators:
- EC mode (µS)
- EC mode (mS)
- TDS mode (ppm)
- pH mode
- pH / EC / TDS measurement
- Calibration solution / conversion factor selected
- Hold mode (appears when HOLD is pressed)
- Temperature measurement
- Celsius / Fahrenheit mode
- Low battery indicator
- Warning symbol

Labeled diagram of the COM-300 meter and its LCD display.
4. Setup and Preparation
Before First Use:
- The meter's electrode is shipped with a wetted sponge of Potassium Chloride (KCl) storage solution. If the sponge is dry (solutions can evaporate), use the included Mylar pack of storage solution to saturate the probe before first use. Let it sit in solution for a minimum of 30 minutes.
- For new meters, for your first few tests, lightly swirl the meter in the water or solution you are testing 3 times and allow the meter to sit while adjusting to the reading.
- If you need to test the pH of low conductivity/TDS water (below 50 µS/25 ppm), you should do so only after using the meter a few times in a pH buffer solution or higher conductivity water.
- Never touch the glass sensor or reference tube with your fingers. Be sure not to hit the sensor or reference tube against a glass or beaker, to avoid cracking it.
- Always make sure that the cap is on tight after usage.
Battery Installation:
The HM COM-300 uses 2 x AAA batteries (not included).
- Twist open the end cap on the top of the meter.
- Twist the battery cap to the open position.
- Remove any old batteries.
- Insert new batteries with the "+" side facing up.
- Close the battery cap and turn to the close position.
- Close and tighten the end cap to maintain waterproofness.
NOTE: Do not reverse the polarity of the batteries. This may short-circuit the meter.
5. Operating Instructions
5.1. Switching Between pH and EC/TDS Modes
- To switch between the pH and EC/Conductivity modes, press the 'PH/EC' (middle button) on the meter.
- The meter will retain this setting until changed again.
5.2. pH Measurements
- Remove the cap.
- Press the Power button. The display will activate.
- If the meter is in EC/TDS mode, press the 'PH/EC' button to switch to pH mode.
- Dip the meter into the water sample, liquid or solution to be tested.
- Lightly swirl the meter to ensure the removal of trapped air bubbles or electric charges. Do not tap it against the glass.
- The meter will display a reading almost immediately. Keep the meter in the liquid until the reading stabilizes (approx. 5-30 seconds) for an accurate reading.
NOTE - Minor fluctuations are a normal function of pH in certain environmental situations.
- To view the reading out of the liquid, press the HOLD button while the meter is in the liquid. This will freeze the reading on the screen. Pressing the HOLD button again will release it.
- Press the Power button to turn the meter off.
- Shake any excess water off the meter. For best results, rinse with distilled or deionized water after each usage and wipe with a tissue. Put the cap back on. Store the meter standing upright to ensure complete saturation of the sensor.
NOTE - The COM-300 is extremely sensitive in low conductivity water (below 50 µS). It is not recommended to use this meter in water below 10 µS. If you do use the meter in such water, it is better to test in flowing water, or by swirling the meter continuously while obtaining the reading. The reading should stabilize in 30-45 seconds.

Measuring water quality with the COM-300 meter.
5.3. EC/TDS Measurements
- Remove the cap.
- Press the Power button. The display will activate.
- To change the mode, press and hold the "HOLD / MODE" button. The unit will cycle through the four possible modes:
- EC - µS
- TDS - ppm (with 0.5 conversion factor and temperature coefficient)
- TDS - ppm (with 0.7 conversion factor and temperature coefficient)
- EC - mS
- Release the “HOLD / MODE" button when the display shows the desired mode.
- Dip the Meter into the water sample or solution to be tested.
- Lightly swirl the meter and tap it against the bottom of the beaker to ensure the removal of trapped air bubbles or electric charges.
- The meter will display a reading almost immediately. Keep the meter in the water until the reading stabilizes (approx. 30 seconds) for an accurate reading.
NOTE - Newer meters may take up to 2 minutes to fully stabilize. This time will decrease with usage as the sensor adapts.
- To view the reading out of the water sample or solution, quickly press the “HOLD / MODE" button while the meter is in the water. This will hold the reading on the screen. Quickly pressing the "HOLD / MODE" button again will release it.
- Press the Power button to turn the meter off.
- Shake any excess water off the meter. For best results, rinse with distilled or deionized water after each usage and wipe with a tissue. Put the cap back on. Store the meter standing upright to ensure complete saturation of the sensor.
5.4. Temperature Measurements
The temperature reading is always displayed on the LCD panel during measurement mode and is shown simultaneously for pH and EC/TDS readings. It is not shown when the meter is in calibration mode. The default reading for the meter is in Celsius.
- Remove the cap.
- Click the Power button. The display will activate.
- The temperature reading is always displayed on the LCD screen (except in calibration and reset mode) and can be used to take air or liquid temperature measurements.
- To measure the temperature of a water sample or solution, dip the meter into it. The temperature reading will change immediately (unless the liquid is at room temperature). For very hot or cold liquids, the reading may take slightly longer to stabilize.
NOTE - Because of the sensitivity of the pH sensor and reference tube, it is not recommended to use your meter as a thermometer in very hot or very cold liquids. Do not keep the meter in very hot or cold liquids for extended periods of time.
5.5. Switching Temperature Mode (Celsius/Fahrenheit)
- To switch between Celsius and Fahrenheit, press and hold the 'HOLD/MODE' button in pH mode.
6. Calibration
The COM-300 features digital calibration for both EC/TDS and pH. Always calibrate to a fresh solution.
6.1. EC/TDS Calibration
The COM-300 is factory calibrated to a 1413 µS KCl solution. Recalibration may be necessary for specific applications or if the calibration has shifted over time. It is recommended to calibrate as close as possible to the range that will be measured.
- Turn the meter on by pressing the Power button.
- Make sure the meter is in the mode that matches the calibration solution (e.g., ppm 0.7 for a 0.7 solution).
- Dip the meter into a laboratory-certified EC or TDS calibration solution. Lightly stir and tap the meter on the bottom of the glass to remove any air bubbles.
NOTE - If the measurement matches the calibration solution value, your COM-300 is already properly calibrated. Stop here!
- Press and hold the 'PH/EC/CAL' button. The temperature display will change to a 'CAL' image.
- The meter will automatically adjust to a reading within a range of the calibration solution.
- Change the reading so that it matches the calibration solution using the 'UP' (PH/EC/CAL) or 'DOWN' (HOLD/MODE) buttons.
- To set the calibration, press the Power button once. A flashing "CAL" image will appear, followed by "END". If "END" does not flash, the meter has not been properly calibrated.
- Your meter is now recalibrated.
6.2. Temperature Calibration
- You must have a correct thermometer next to the COM-300.
- With the COM-300 on and the cap off and the meter in the air, press and hold the Power and 'PH/EC/CAL' buttons together for 3 seconds. The meter will enter temperature calibration mode.
- Adjust to the correct temperature by pressing the 'UP' or 'DOWN' buttons.
- Once correct, click the Power button. The temperature will flash and the screen will flash C...CA...CAL.
- The screen will briefly display "End" and then revert to measurement mode. It is now calibrated.
6.3. pH Calibration (Automatic)
The COM-300 is factory calibrated to pH 7.0. Recalibration is recommended at least once per month, or more frequently if testing wide ranges.
- Turn the meter on by clicking the Power button.
- Make sure the meter is in pH mode.
- Insert the meter into a pH buffer solution of 4.0, 7.0 or 10.0. Press and hold the 'CAL' button for 5 seconds.
- The temperature reading will change to a flashing 'CAL' image for 10 seconds. Do not press any buttons. The meter will automatically recognize the solution it is in (or the closest to it). This number will flash and the letters of 'CAL' will flash as 'C...CA...CAL' indicating progress. Allow 5 - 60 seconds during this stage. Do not press any buttons and do not move the meter.
- When the meter is calibrated, 'End' will flash briefly and the screen will revert to measurement mode. Your meter is now recalibrated.
6.4. pH Calibration (Manual / Fine Tuning)
- Turn the meter on by clicking the Power button.
- Make sure the meter is in pH mode.
- Insert the meter into a known pH buffer solution.
- Press and hold the CAL button for 5 seconds.
- The temperature will change to a flashing 'CAL' image for 10 seconds before entering automatic calibration. During this time, click either the up button (PH/EC/CAL) or down button (HOLD/MODE) to remain in manual calibration mode.
- Change the reading so that it matches the value of the calibration solution. To increase the reading, click the up button and to decrease the reading click the down button. Press and hold either button for rapid advancement.
- To set the calibration, click the Power button once. 'C...CA...CAL' will flash briefly, followed by 'End' and the temperature will return to the screen. Your meter is now recalibrated.
7. Care and Maintenance
- Do not store the unit in high temperature or direct sunlight.
- Do not touch the electrodes. Skin oils may adversely affect the reading. If you do touch the electrodes, clean immediately with alcohol or distilled water.
- After repeated usage in high TDS water, it is advised to clean the electrodes to prevent residue build-up.
- For best results, always stir or tap the meter in the water sample to dislodge any air bubbles or remove any lingering electrical charges.
- Water volume, the positioning of the electrode in the water sample, and temperature may affect the reading.
- Do not keep the meter in very hot water for extended periods of time.
- If testing two water samples in a wide range (e.g., 15ppm and 3000ppm), make sure to rinse the electrodes with DI or distilled water or alcohol after each test to ensure accurate readings and prevent a build-up of TDS on the electrodes.
- The COM-300 is waterproof. However, prior to completely submersing the meter into water, always ensure the black probe gasket ring and battery compartment are secured tightly.
7.1. Cleaning
- To clean the unit, use a soft rag or towel. Wipe with water and a mild soap.
- To clean the electrodes, use rubbing alcohol and a cotton swab. Lightly clean the electrodes. Rinse with DI or distilled water. Air dry.
7.2. Electrode Replacement
If your electrode has been damaged, you can purchase a new one without needing to purchase a new meter. To replace the electrode:
- Remove the black electrode locking ring by twisting it counter-clockwise.
- Gently pull the electrode off the unit.
- Gently insert the new electrode into the unit. Be sure to align the grooves and eight pins properly. Never force the electrode into the unit!
- Make sure the rubber ring is properly positioned on the electrode.
- Screw the black locking ring back onto the unit by twisting it clockwise. Tighten.

Detachable electrode probe for easy replacement.
8. Troubleshooting
| Problem | Error Message | Reason(s) and/or Potential Solution(s) |
|---|---|---|
| The meter will not power on. | - | 1. Change the batteries. 2. Double-check the polarity of the batteries. |
| Incorrect Readings. | - | 1. Recalibrate the meter. 2. Switch modes. |
| The TDS/EC level is out of range (oor). | 00r | (oor) The meter must be used within the specified range limits. |
| The sensor has been disconnected or damaged internally. | Err | 1. (Error) Double-check the sensor is connected, or you may need a new sensor. |
| The temperature is out of range. | 00r | (Error) The meter must be used within the specified temperature limits. |
| The pH measurement is unstable. | ! | 1. Build-up on a sensor, air bubble present. 2. Sensor lifespan depleted, damaged sensor. |
9. User Tips
- Hydroponics Use: The COM-300 is suitable for hydroponic farming. Ensure regular calibration, especially for EC/TDS, to maintain accuracy for nutrient solutions.
- Accuracy Check: If you suspect inaccurate readings, always perform a calibration with a fresh, known-value solution before troubleshooting other issues.
- Sensor Care: Always keep the electrode cap on when not in use to prevent the sponge from drying out, which can affect sensor performance.
10. Warranty and Support
10.1. Limited Warranty
This HM Digital, Inc. ("the Company") product ("COM-300") is warranted to the purchaser against defective materials and workmanship for one (1) year from the date of purchase.
The detachable sensor (SP-C3) is warranted for six (6) months from the date of purchase against defective materials and workmanship.
What is covered: Repair parts and labor, or replacement at the Company's discretion. Transportation Charges for repaired or new product to be returned to the purchaser.
What is NOT covered: Transportation charges for the defective products to be sent to the Company. Any consequential damages, incidental damages, or incidental expenses, including damages to property. This includes damages from abuse or improper maintenance such as tampering, wear and tear, water damage, or any other physical damage. The COM-300 is watertight and completely submersible. However, please ensure that the battery compartment and probe gasket ring are firmly tightened before submerging in water. The warranty does not cover water damage to the COM-300 due to parts not securely closed.
10.2. Obtaining Warranty Service
To obtain warranty service, please contact HM Digital, Inc. at 800.383.2777 or email Warranty@HMDigital.com to receive further instructions. Before sending the product back, please include the following:
- Your name
- Phone number/Address
- Description of problem
- Proof of purchase, must include Date
* If a returned product does not include the above-mentioned items, the Company Reserves the right to refuse warranty service.
10.3. Implied Warranties
Any implied warranties, including implied warranties of merchantability and fitness for a particular purpose, are limited in duration to ONE YEAR from the date of purchase. Some states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you. To the extent any provision of this warranty is prohibited by federal and state law and cannot be preempted, it shall not be applicable. This warranty gives you specific legal rights, and you may also have other rights, which vary from state to state.
NOTE: Warranties are product-specific. Third-party products and products deemed by HM Digital as “accessories” are not covered under warranty. Third-party products include, but are not limited to, batteries and fittings. Accessories included, but not limited to, lanyards and product boxes.
This manual and all information contained herein copyright HM Digital, Inc. The contents of this manual cannot be reprinted without the permission of HM Digital, Inc.
442 and 442 Natural Water are registered trademarks of the Myron L Company.
For the full user manual in PDF format, please refer to the official document: HM COM-300 User Manual (PDF)