This manual provides detailed instructions for the safe and efficient operation of your HT-RT01 tester.

The HT-RT01 is designed for testing and high-speed sorting of low-resistance and large-scale lithium battery packs. It measures internal resistance (in mΩ) and voltage, crucial for identifying batteries with high internal resistance that can lead to increased power consumption, heat generation, accelerated aging, and reduced lifespan. By screening and matching cell capacity, internal resistance, and voltage, the tester helps ensure optimal battery pack performance based on the "barrel" principle, where overall performance is limited by the weakest cell.

1. Setup

1.1. Unpacking and Contents

Carefully unpack the HT-RT01 tester and verify that all standard accessories are present:

  • HT-RT01 Tester x1
  • 4-wire resistance test pen x1
  • Test fixture x1
  • USB data cable x1
  • Power supply cord x1
  • Manual x1 (this document)
HT-RT01 Standard accessories
Figure 1: HT-RT01 Standard Accessories

1.2. Power Connection

  1. Connect the power supply cord to the AC 110V/220V power input on the back of the tester.
  2. Ensure the main power switch is in the OFF position before connecting to a power outlet.
  3. Plug the power supply cord into a suitable power outlet.
Back view of HT-RT01 tester showing power input and switch
Figure 2: Rear Panel with Power Input and Main Switch

1.3. Test Probe Connection

Connect the 4-wire resistance test pen or the test fixture to the red and black measurement input ports on the front panel of the tester. Ensure correct polarity (red to positive, black to negative) for accurate readings.

2. Operating Instructions

2.1. Basic Measurement

  1. Turn on the main power switch on the back of the tester.
  2. Press the O/Enter button on the front panel to power on the device.
  3. The instrument will automatically default to the automatic gear selection mode and intelligently identify the battery/battery pack under test. Manual adjustment is not required in this mode.
  4. Connect the battery to be tested using the 4-wire test pen or test fixture. Ensure the positive and negative electrodes are correctly connected to the corresponding red and black sockets.
  5. The screen will display the internal resistance and voltage values of the battery under test.
User measuring battery pack with HT-RT01 tester
Figure 3: Measuring a Battery Pack

2.2. Automatic Polarity Discrimination

The tester automatically recognizes if the positive (red) and negative (black) clips of the Kelvin four-wire clip do not correspond to the battery's polarity. In such cases, the voltage value will be displayed as a negative number (e.g., -3.9872V), indicating an inverse connection. This feature helps prevent incorrect readings due to reversed connections.

2.3. Supported Battery Types

The HT-RT01 can measure internal resistance and voltage for a wide range of battery types, including:

  • Ternary lithium batteries
  • Lithium iron phosphate (LiFePO4) batteries
  • Lead-acid batteries
  • Lithium-ion batteries
  • Lithium polymer batteries
  • Alkaline batteries
  • Dry batteries
  • Nickel-metal hydride batteries
  • Nickel-cadmium batteries
  • Button batteries
Examples of various battery types being measured
Figure 4: Various Battery Measurement Examples

2.4. Language Switching

The tester supports switching between Chinese and English display languages.

  1. Press the Set key to enter the menu.
  2. Use the Sel key to navigate to the "Language" option.
  3. Press O/Enter to select.
  4. Use the Sel key to choose between "Chinese" and "English".
  5. Press O/Enter to confirm your selection.
HT-RT01 display showing language selection menu
Figure 5: Language Selection Menu

2.5. Connecting to a Computer

The HT-RT01 features a USB port for connecting to a computer, allowing for numerical analysis of multiple measurements.

  1. Connect the tester to your computer using the provided USB data cable.
  2. On the tester, press the Set key to enter the menu.
  3. Use the Sel key to navigate to the "Serial port setting" option.
  4. Press O/Enter to enter the serial port settings.
  5. Use the Sel key to select "ON" and press O/Enter to confirm. The serial port will now be open.
  6. Open the dedicated software on your computer (contact Heltec BMS for software and drivers).
  7. In the software, select the correct COM port for the tester and open the serial port to view real-time resistance and voltage data.
  8. Data can be exported from the software for further analysis.
HT-RT01 connected to a laptop via USB, showing software interface
Figure 6: Connecting the Tester to a Computer for Data Analysis

2.6. Calibration

The HT-RT01 supports manual calibration for both resistance and voltage. This process requires a known reference for accurate calibration.

2.6.1. Resistance Calibration

  1. Connect a standard resistance (e.g., 20mΩ) to the tester.
  2. Press the Set key, then navigate to "Cal R" (Calibration Resistance) and press O/Enter.
  3. Use the Vol-/Res+ keys to adjust the displayed value to match the standard resistance.
  4. Press O/Enter to save the calibration.

2.6.2. Voltage Calibration

  1. Connect a battery with a precisely measured voltage (reference voltage) to the tester.
  2. Press the Set key, then navigate to "Cal V" (Calibration Voltage) and press O/Enter.
  3. Use the Vol-/Res+ keys to adjust the displayed value to match the reference voltage.
  4. Press O/Enter to save the calibration.

2.7. Battery Screening

The tester allows setting specific internal resistance and voltage ranges to filter batteries in batches, which is useful for assembling battery packs. This ensures that only batteries meeting desired criteria are used, improving overall pack performance and longevity.

If a battery's parameters are within the set range, the computer software will display "PASS". If they exceed the set values, it will display "NG" (Not Good).

Video 1: Demonstration of HT-RT01 operation, including setup, calibration, battery screening, and computer connection.

3. Maintenance

  • Keep the instrument in a dry environment, away from mist or high humidity, to ensure accurate measurements and prevent damage.
  • Clean the exterior of the device with a soft, dry cloth. Do not use abrasive cleaners or solvents.
  • Regularly inspect the test probes and cables for any signs of wear or damage. Replace if necessary.
  • When measuring internal resistance, ensure the positive and negative electrodes of the battery are clean and free of dirt or oxide layers. If necessary, gently rub them with fine sandpaper to expose the metallic luster for optimal contact.

4. Troubleshooting

4.1. Inaccurate or Erratic Readings

  • Interference: Do not use a mobile phone near the instrument during measurement, as its radio frequency signals can interfere with the instrument and cause errors. Keep the instrument away from high-frequency switching power supplies.
  • Environmental Conditions: Ensure the instrument is used in a dry environment, free from mist.
  • Poor Contact: When using the test fixture or test clamp, ensure good contact with the battery terminals. Poor contact can lead to excessively high or inconsistent internal resistance readings.
  • Dirty/Oxidized Terminals: Clean battery terminals by rubbing them with fine sandpaper to expose the metallic luster.
  • Test Clip Alignment: Keep the test clips parallel to reduce the influence of eddy currents and improve test accuracy.

4.2. Negative Voltage Display

A negative voltage value (e.g., -3.9872V) indicates that the positive and negative test clips are inversely connected to the battery terminals. Correct the connection to resolve this.

4.3. No Display or Power Issues

  • Check if the power supply cord is securely connected to both the tester and the power outlet.
  • Ensure the main power switch on the back of the tester is in the ON position.
  • Verify that the power outlet is functional.

5. Specifications

ParameterSpecification
Measurement ParametersAC resistance, DC resistance
IR Precision±0.5%
Voltage Precision±0.5%
IR Measuring Range0.001mΩ - 200Ω
Voltage Measuring Range0.001V - ±100VDC
FrequencyAC 1KHz
Signal Source Current2mΩ/20mΩ gear 50mA, 200mΩ/2Ω gear 5mA, 20Ω/200Ω gear 0.5mA
Resistance Measurement Range6 gear adjustment
Voltage Measurement Range3 gear adjustment
Test Pace5 times /S
Resistance CalibrationManual calibration
Voltage CalibrationManual calibration
Power SupplyAC110V/AC220V
Supply Current50mA~100mA
RangeAutomatic Switching
Measuring InputBalance System
Measuring ProbesLCR Kelvin 4-wire clamp
LanguageSwitch language (Chinese / English)
USB PortConnecting to computer
Size (L x W x H)190*180*80mm (approx. 11 x 12 x 4 cm package dimensions)
Weight1.1kg (approx. 3.0 kg package weight)
Dimensions of the HT-RT01 tester
Figure 7: HT-RT01 Tester Dimensions

6. User Tips

  • The HT-RT01 is noted for its fast operation, making it an efficient tool for battery testing.
  • Users have found it to be a valuable tool for its price point, offering good performance for the cost.

7. Warranty and Support

7.1. Warranty

The Heltec BMS HT-RT01 comes with a one-year warranty from the date of purchase.

7.2. Technical Support

For technical assistance, software drivers, installation videos, or any other support inquiries, please contact Heltec BMS directly. Refer to the contact information provided with your purchase or on the official Heltec BMS website.

This instruction manual is designed to help you operate your HT-RT01 tester safely and effectively. Please keep it in a safe place for future reference.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.