1. Setup
1.1. Unpacking and Contents
Carefully unpack the HT-RT01 tester and verify that all standard accessories are present:
- HT-RT01 Tester x1
- 4-wire resistance test pen x1
- Test fixture x1
- USB data cable x1
- Power supply cord x1
- Manual x1 (this document)

1.2. Power Connection
- Connect the power supply cord to the AC 110V/220V power input on the back of the tester.
- Ensure the main power switch is in the OFF position before connecting to a power outlet.
- Plug the power supply cord into a suitable power outlet.

1.3. Test Probe Connection
Connect the 4-wire resistance test pen or the test fixture to the red and black measurement input ports on the front panel of the tester. Ensure correct polarity (red to positive, black to negative) for accurate readings.
2. Operating Instructions
2.1. Basic Measurement
- Turn on the main power switch on the back of the tester.
- Press the O/Enter button on the front panel to power on the device.
- The instrument will automatically default to the automatic gear selection mode and intelligently identify the battery/battery pack under test. Manual adjustment is not required in this mode.
- Connect the battery to be tested using the 4-wire test pen or test fixture. Ensure the positive and negative electrodes are correctly connected to the corresponding red and black sockets.
- The screen will display the internal resistance and voltage values of the battery under test.

2.2. Automatic Polarity Discrimination
The tester automatically recognizes if the positive (red) and negative (black) clips of the Kelvin four-wire clip do not correspond to the battery's polarity. In such cases, the voltage value will be displayed as a negative number (e.g., -3.9872V), indicating an inverse connection. This feature helps prevent incorrect readings due to reversed connections.
2.3. Supported Battery Types
The HT-RT01 can measure internal resistance and voltage for a wide range of battery types, including:
- Ternary lithium batteries
- Lithium iron phosphate (LiFePO4) batteries
- Lead-acid batteries
- Lithium-ion batteries
- Lithium polymer batteries
- Alkaline batteries
- Dry batteries
- Nickel-metal hydride batteries
- Nickel-cadmium batteries
- Button batteries

2.4. Language Switching
The tester supports switching between Chinese and English display languages.
- Press the Set key to enter the menu.
- Use the Sel key to navigate to the "Language" option.
- Press O/Enter to select.
- Use the Sel key to choose between "Chinese" and "English".
- Press O/Enter to confirm your selection.

2.5. Connecting to a Computer
The HT-RT01 features a USB port for connecting to a computer, allowing for numerical analysis of multiple measurements.
- Connect the tester to your computer using the provided USB data cable.
- On the tester, press the Set key to enter the menu.
- Use the Sel key to navigate to the "Serial port setting" option.
- Press O/Enter to enter the serial port settings.
- Use the Sel key to select "ON" and press O/Enter to confirm. The serial port will now be open.
- Open the dedicated software on your computer (contact Heltec BMS for software and drivers).
- In the software, select the correct COM port for the tester and open the serial port to view real-time resistance and voltage data.
- Data can be exported from the software for further analysis.

2.6. Calibration
The HT-RT01 supports manual calibration for both resistance and voltage. This process requires a known reference for accurate calibration.
2.6.1. Resistance Calibration
- Connect a standard resistance (e.g., 20mΩ) to the tester.
- Press the Set key, then navigate to "Cal R" (Calibration Resistance) and press O/Enter.
- Use the Vol-/Res+ keys to adjust the displayed value to match the standard resistance.
- Press O/Enter to save the calibration.
2.6.2. Voltage Calibration
- Connect a battery with a precisely measured voltage (reference voltage) to the tester.
- Press the Set key, then navigate to "Cal V" (Calibration Voltage) and press O/Enter.
- Use the Vol-/Res+ keys to adjust the displayed value to match the reference voltage.
- Press O/Enter to save the calibration.
2.7. Battery Screening
The tester allows setting specific internal resistance and voltage ranges to filter batteries in batches, which is useful for assembling battery packs. This ensures that only batteries meeting desired criteria are used, improving overall pack performance and longevity.
If a battery's parameters are within the set range, the computer software will display "PASS". If they exceed the set values, it will display "NG" (Not Good).
3. Maintenance
- Keep the instrument in a dry environment, away from mist or high humidity, to ensure accurate measurements and prevent damage.
- Clean the exterior of the device with a soft, dry cloth. Do not use abrasive cleaners or solvents.
- Regularly inspect the test probes and cables for any signs of wear or damage. Replace if necessary.
- When measuring internal resistance, ensure the positive and negative electrodes of the battery are clean and free of dirt or oxide layers. If necessary, gently rub them with fine sandpaper to expose the metallic luster for optimal contact.
4. Troubleshooting
4.1. Inaccurate or Erratic Readings
- Interference: Do not use a mobile phone near the instrument during measurement, as its radio frequency signals can interfere with the instrument and cause errors. Keep the instrument away from high-frequency switching power supplies.
- Environmental Conditions: Ensure the instrument is used in a dry environment, free from mist.
- Poor Contact: When using the test fixture or test clamp, ensure good contact with the battery terminals. Poor contact can lead to excessively high or inconsistent internal resistance readings.
- Dirty/Oxidized Terminals: Clean battery terminals by rubbing them with fine sandpaper to expose the metallic luster.
- Test Clip Alignment: Keep the test clips parallel to reduce the influence of eddy currents and improve test accuracy.
4.2. Negative Voltage Display
A negative voltage value (e.g., -3.9872V) indicates that the positive and negative test clips are inversely connected to the battery terminals. Correct the connection to resolve this.
4.3. No Display or Power Issues
- Check if the power supply cord is securely connected to both the tester and the power outlet.
- Ensure the main power switch on the back of the tester is in the ON position.
- Verify that the power outlet is functional.
5. Specifications
| Parameter | Specification |
|---|---|
| Measurement Parameters | AC resistance, DC resistance |
| IR Precision | ±0.5% |
| Voltage Precision | ±0.5% |
| IR Measuring Range | 0.001mΩ - 200Ω |
| Voltage Measuring Range | 0.001V - ±100VDC |
| Frequency | AC 1KHz |
| Signal Source Current | 2mΩ/20mΩ gear 50mA, 200mΩ/2Ω gear 5mA, 20Ω/200Ω gear 0.5mA |
| Resistance Measurement Range | 6 gear adjustment |
| Voltage Measurement Range | 3 gear adjustment |
| Test Pace | 5 times /S |
| Resistance Calibration | Manual calibration |
| Voltage Calibration | Manual calibration |
| Power Supply | AC110V/AC220V |
| Supply Current | 50mA~100mA |
| Range | Automatic Switching |
| Measuring Input | Balance System |
| Measuring Probes | LCR Kelvin 4-wire clamp |
| Language | Switch language (Chinese / English) |
| USB Port | Connecting to computer |
| Size (L x W x H) | 190*180*80mm (approx. 11 x 12 x 4 cm package dimensions) |
| Weight | 1.1kg (approx. 3.0 kg package weight) |

6. User Tips
- The HT-RT01 is noted for its fast operation, making it an efficient tool for battery testing.
- Users have found it to be a valuable tool for its price point, offering good performance for the cost.
7. Warranty and Support
7.1. Warranty
The Heltec BMS HT-RT01 comes with a one-year warranty from the date of purchase.
7.2. Technical Support
For technical assistance, software drivers, installation videos, or any other support inquiries, please contact Heltec BMS directly. Refer to the contact information provided with your purchase or on the official Heltec BMS website.
This instruction manual is designed to help you operate your HT-RT01 tester safely and effectively. Please keep it in a safe place for future reference.