OWON HDS200 Dual Channel Series

Rate this product if you’ve used it
Contents

OWON HDS200 Dual Channel Series Handheld Oscilloscope User Manual

Models: HDS242, HDS272, HDS2102, HDS2202S

For product support, visit: www.owon.com.hk/download

1. Introduction

This manual provides comprehensive instructions for the OWON HDS200 Dual Channel Series Handheld Oscilloscope, which integrates an oscilloscope, multimeter, and waveform generator into a single portable device. It covers safety information, setup procedures, operational guidelines for each function, maintenance, troubleshooting, and detailed technical specifications. Please read this manual thoroughly before using the product to ensure proper and safe operation.

Play video
Watch video online

Video: Demonstrating the waveform generator and oscilloscope functions, including frequency adjustment and waveform observation.

2. Safety Information

Before using this product, please read the safety information in advance to prevent personal injury and damage to the product or any other connected devices. This product can only be used within the specified range.

2.1. Safety Terms and Symbols

The following terms and symbols may appear in this manual and on the product:

  • Warning: Indicates conditions or practices that could result in injury or loss of life.
  • Caution: Indicates conditions or practices that could result in damage to this product or other property.
  • Danger (on product): Indicates an immediate hazard or injury possibility.
  • Warning (on product): Indicates a possible hazard or injury.
  • Caution (on product): Indicates potential damage to the instrument or other property.

Safety Symbols on the Product:

  • Hazardous Voltage symbol Hazardous Voltage
  • Refer to Manual symbol Refer to Manual
  • Protective Earth Terminal symbol Protective Earth Terminal
  • Chassis Ground symbol Chassis Ground
  • Test Ground symbol Test Ground

2.2. Safety Requirements

  • Only a qualified person should perform internal maintenance.
  • Check all Terminal Ratings to avoid fire or shock hazard.
  • Do not operate without covers or panels removed.
  • Avoid exposed circuits. Be careful when working on exposed circuitry.
  • Do not operate if any damage is suspected.
  • Do not operate in damp conditions or an explosive atmosphere.
  • Keep product surfaces clean and dry.
  • Using the equipment not in accordance with the manufacturer's specified method may damage the protection provided.
  • To prevent electric shock or fire, use a suitable power adapter dedicated to this product.
  • The two channels of the oscilloscope are non-isolated. Use a common reference when measuring.
  • If the input port is connected to a voltage higher than 42V (30Vrms) or a circuit with a peak value of more than 4800 VA, use only voltage probes, test wires, and adapters with proper insulation.
  • Before use, check all probes and accessories for mechanical damage.
  • Remove all unused multimeter test probes, oscilloscope probes, and accessories.
  • Do not use an input voltage higher than the rated value.
  • Do not touch the bare metal BNC or banana plug.
  • Do not insert metal objects into the connector.
  • Do not input any signal, voltage, or current into the GEN Out port of the signal generator.

2.2.1. Grounding Schematics

The following diagrams illustrate the internal ground wire connections for the oscilloscope:

Schematic diagram of the internal ground wire connection of the oscilloscope
Figure 2-1: Schematic diagram of the internal ground wire connection of the oscilloscope.
Schematic diagram of internal ground connection when oscilloscope is connected with computer through the port
Figure 2-2: Schematic diagram of internal ground connection when oscilloscope is connected with computer through the port.

When the oscilloscope is AC powered by an adapter or connected to an AC powered computer through the port, it is not allowed to measure the primary power supply of the power grid.

3. Setup

3.1. General Inspection

Upon receiving your new oscilloscope, perform the following checks:

  1. Check for transportation damage: Inspect the packaging carton and foam for damage. If damage is found, retain the packaging until electrical and mechanical property tests are completed.
  2. Check accessories: Refer to the "List of Accessories" in Appendix A to verify all items are present and undamaged. Contact your distributor if any items are missing or damaged.
  3. Check the complete instrument: Inspect the instrument for any physical damage or functional issues. If the instrument does not work normally or fails performance tests, contact your distributor.

3.2. Front Panel and Keys Overview

Front Panel of the Oscilloscope with numbered labels
Figure 3-1: Front Panel of the Oscilloscope
  • CH1 and CH2 input connectors.
  • Waveform generator output connector (optional).
  • Display area.
  • F1-F4 keys: Multi-function keys for menu selection.
  • HOR key: Adjusts horizontal time base and signal displacement.
  • Return key: Returns to the previous menu or closes the menu.
  • Measurement menu key (oscilloscope) or range key (multimeter).
  • Zoom or move key: Directional keys for waveform movement, time base changes, voltage cursor movement, and trigger level changes.
  • Automatic setting key (oscilloscope) or automatic range key (multimeter).
  • Stop / run key (oscilloscope) or value hold key (multimeter) or turn on/off signal output (waveform generator).
  • Input end of the multimeter.
  • Trigger menu key (oscilloscope) or relative value key (multimeter).
  • Power switch key.
  • Enter the save settings key.
  • Enter the system settings key.
  • Switch key for working state of oscilloscope and multimeter.
  • CH1 / CH2 - channel switch key.

3.3. Side Panel Overview

Side Panel of the Oscilloscope with numbered labels
Figure 3-2: Side Panel of the Oscilloscope
  • Probe compensation: 3.3V/1kHz square wave signal output.
  • Charging or USB communication interface.
  • Bracket.

3.4. Functional Check

To quickly verify the instrument is working properly:

  1. Press the power switch key (13 on Figure 3-1). The instrument will perform self-checks and display the startup screen.
  2. Set the oscilloscope probe switch to 10X and connect it to the CH1 channel. Align the slot on the probe with the plug on the BNC connector, insert, and turn to tighten. Connect the probe tip and ground clamp to the probe compensator's signal output and reference ground, respectively.
  3. Press the "Auto" key on the front panel. Within a few seconds, a square wave display (1kHz/3.3Vpp) should appear.
Oscilloscope display showing an automatically set square wave
Figure 3-3: Automatic Setting Display

Repeat steps 2 and 3 for the CH2 channel.

3.5. Probe Compensation

When connecting a probe to an input channel for the first time, adjust the probe to match the input channel. Uncompensated or improperly compensated probes can lead to measurement errors.

  1. Set the probe menu attenuation coefficient to 10X and ensure the probe switch is also set to 10X. Connect the probe tip to the signal output connector of the probe compensator and the reference wire clamp to the ground wire connector. Press the Auto key.
  2. Check the displayed waveform and adjust the probe until the compensation is correct.
Waveforms showing overcompensation, correct compensation, and under-compensation
Figure 3-4: Display Waveform of Probe Compensation
Diagram showing how to adjust a probe
Figure 3-5: Probe Adjustment

Repeat the steps if necessary.

3.6. Probe Attenuation Coefficient Setting

Probes have various attenuation coefficients that affect the vertical position factor of the oscilloscope. To change or check this setting:

  1. Press the CH1/CH2 key to switch to the desired channel.
  2. Press F3 to select the desired attenuation coefficient.
Diagram of a probe attenuation switch
Figure 3-6: Probe Attenuation Switch

Note: The default probe attenuation coefficient is 10X. Ensure the probe's physical switch matches the oscilloscope's menu setting. When the probe switch is set to 1X, the oscilloscope's bandwidth is limited to 5 MHz. For full bandwidth, set the switch to 10X.

3.7. Safe Use of Probe

The safety ring on the probe body protects against electric shocks.

Diagram of a probe showing the finger safety ring
Figure 3-7: Finger Safety Ring of the Probe

Warning: Keep fingers behind the safety ring. Do not touch the metal part of the probe head when connected to a voltage source.

4. Operating Instructions

4.1. Oscilloscope Operation

The oscilloscope function allows for viewing and analyzing electrical signals. Before making any measurements, connect the probe to the instrument and the ground terminal to the ground.

4.1.1. User Interface

Oscilloscope display showing various interface elements
Figure 4-1: Oscilloscope User Interface

Key elements of the user interface include:

  • Trigger Status: Indicates Auto (automatic mode), Trig (trigger detected), Ready (pre-trigger data obtained), Scan (continuous collection), or Stop (data collection stopped).
  • Run/Stop: Controls waveform acquisition.
  • Time Base Display: Shows the horizontal time scale.
  • Trigger Horizontal Position Pointer: Indicates the trigger point on the horizontal axis.
  • Trigger Position within Storage Depth Pointer: Indicates the trigger position relative to the stored data.
  • Trigger Horizontal Displacement Value: Displays the current horizontal displacement and waveform window position in memory.
  • USB Disk Connection Indicator: Shows if a USB disk is connected.
  • Battery/External Power Indicator: Shows power status.
  • Channel 1/2 Waveform: Displays the signal for each channel.
  • Trigger Electrical Level Position Pointer: Indicates the trigger level for the channel.
  • Trigger Information Icon: Displays trigger channel, coupling mode, type, and electrical level.
  • Current Sampling Rate: Shows the active sampling rate.
  • Channel Information Reading: Indicates the voltage position of the channel.
  • Coupling Mode Icon: Shows DC, AC, or Ground coupling.
  • Grounding Reference Point Pointer (CH1/CH2): Indicates the zero position of the waveform.
  • Waveform Display Area: The main area where waveforms are shown.

4.1.2. Vertical System

The vertical system adjusts the vertical scale, position, and other channel settings. Each channel has independent settings.

  • Vertical Position: Press CH1/CH2 to select a channel, then use the Up/Down direction keys to move the waveform vertically. Press Up and Down simultaneously to center.
  • Vertical Volt/Div Setting: The range is 10mV/div to 10V/div (1X probe) or 100mV/div to 100V/div (10X probe), etc. Press CH1/CH2, then use Left/Right direction keys to change the volts/div setting.
Vertical System Setting Menu
Function Menu Setting Description
Switch ON / OFF Turn on/off waveform display.
Coupling DC / AC / GND DC: Uses AC and DC components. AC: Blocks DC component. GND: Disconnects input signal.
Probe 1X, 10X, 100X, 1000X, 10000X Select based on probe attenuation factor for accurate reading.
Bandwidth 20MHz / Full band Limits bandwidth to 20MHz to reduce noise or uses full bandwidth.

4.1.3. Horizontal System

The horizontal system controls the time base and horizontal trigger position.

  • Press the HOR key to enter the horizontal system setting menu.
  • Use the Left/Right direction keys to change the horizontal scale (time base) and horizontal trigger position.
  • Press Left and Right simultaneously to center the horizontal position.
Horizontal System Setting Menu
Function Menu Setting Description
Acqu Mode Sample / Peak Detect Normal sampling or peak detection for glitches.
Length 4K / 8K Select record length.
XY Mode ON / OFF Turn XY mode on or off.
Refresh High / Low Set refresh rate.
Hor center - Set horizontal trigger position to screen center.
Counter ON / OFF Turn on/off counter.

4.1.4. Measuring System

The oscilloscope offers automatic and cursor-based measurements.

  • Automatic Measurement: Press the Measure Range key and F1 to perform automatic measurements. Up to 9 types of measurements can be displayed.
Oscilloscope display showing automatic voltage measurements
Figure 4-2: Automatic Voltage Measurements (Mean, PK-PK, Vamp, Min)
Oscilloscope display showing automatic time measurements
Figure 4-3: Automatic Time Measurements (Rise Time, Fall Time, +Width, -Width)
Automatic Measurement Function Menu
Function Menu Setting Description
ON / OFF - Turn automatic measurement on or off.
Source CH1 / CH2 Set the source channel.
Add / Del Freq (F), Period (T), Amp (Va), Min (Mi), Max (Ma), PK-PK (Vpp), Mean(V), RMS (RMS), Rise Time (RT), Fall Time (FT), +PulseWidth (PW), -PulseWidth (NW) Add or delete measurement types.
  • Cursor Measurement: Press the Measure Range key and F2 to perform cursor measurements.
Cursor Measurement Menu
Function Menu Setting Description
Type CH1 Voltage, CH2 Voltage, Time, None Selects voltage or time cursors for CH1/CH2, or turns off cursors.
A - Moves cursor line A (voltage or time) using direction keys.
B - Moves cursor line B (voltage or time) using direction keys.
AB - Links A and B; moves both cursors simultaneously.

4.1.5. Trigger System

The trigger system determines when the oscilloscope starts acquiring data to display waveforms, ensuring a stable display.

  • The trigger mode is edge trigger, triggering on rising or falling edges of the input signal.
  • To enter the edge trigger menu, press the Trig/Δ key.
  • The trigger setting information (e.g., CH1:DCF-20.0mV) is displayed at the bottom right corner of the screen.
Trigger System Setting Menu
Function Menu Setting Description
Source CH1 / CH2 Set Channel 1 or 2 as the source trigger signal.
Coupling AC / DC AC: Prevents DC component. DC: Allows all components.
Type Auto, Normal, Single Auto: Acquires without trigger. Normal: Acquires when trigger conditions met. Single: Samples and stops when trigger detected.
Slope Rising edge / Falling edge Trigger on the rising or falling edge of the signal.
Trigger center - Sets trigger position to the middle of the waveform.
Force - Forcibly generates a trigger signal (Normal and Single modes).

Trigger Electrical Level: The amplitude level the signal must cross. Press Trig/Δ and use direction keys to adjust.

4.1.6. Save Settings

Press the Save key to access the save function menu. You can store oscilloscope settings, reference waveforms, and files.

  • Configure: Save and recall instrument settings.
  • Reference Waveform: Compare actual waveforms with stored reference waveforms.
  • File: Save waveforms or images. Files can be read via USB (MSC mode) or PC software (HID mode).

4.1.7. System Settings

Press the System key to enter the system function menu.

  • Display: Adjust brightness, backlight time, and menu display time.
  • System: Set language, automatic shutdown time, view instrument model/serial number/version, and upgrade firmware.
  • Default Setting: Press System, then F3 to restore default settings.
  • USB Connection: Press System, then F4 to enter the next page. Select F1 for HID (Human Interface Device) to control via PC, or MSC (Mass Storage Class) to access files.
  • Factory Settings: Press System, then F4, then F2 to restore factory settings.
  • Automatic Calibration: Disconnect all probes, press System, then F4, then F3 to perform auto-calibration for optimal measurement accuracy.

4.2. Multimeter Operation

The multimeter function provides various electrical measurements.

4.2.1. Multimeter Interface

Multimeter display showing various interface elements
Figure 4-4: Multimeter Interface

The multimeter uses four 4-mm safety banana plug input ends: A, mA, COM, and VΩ⤳C.

  • Measurement type indication (DCV, ACV, DCA, ACA, Resist, Diode, Cont, Cap).
  • Range indication (Manual or Auto).
  • Current measurement range.
  • USB cable inserted indicator.
  • Battery power indication.
  • "Hold" function to keep current reading.
  • Measurement value and unit.
  • Display of switching resistance, buzzer, diode, and capacitance measurement function.
  • Selected range V or mV (voltage); A or mA (current).
  • AC or DC current measurement selection.
  • AC or DC voltage measurement selection.
  • Relative value measurement function (for DC current, DC voltage, resistance).

4.3. Waveform Generator Operation (Optional)

The instrument can provide 4 basic waveforms (sine, square, ramp, pulse) and 8 arbitrary waveforms.

4.3.1. Connect the Output

Press the Mode button to switch to the waveform generator interface. Ensure "ON" is displayed in the upper left corner; if not, press the Run/Stop key. Connect a BNC cable to the GEN Out port on the top of the oscilloscope.

Diagram of the generator output ports
Figure 4-5: Generator Output Ports

Warning: Do not input any signal (voltage, current, etc.) into the BNC output ports, as this may damage the device.

To observe the output, connect the other end of the BNC cable to the oscilloscope's signal input connector.

4.3.2. Set the Waveform

  1. Press the Mode button to switch to the waveform generator interface.
  2. Press F1 to select the desired waveform (e.g., Sine, Square, Ramp, Pulse, Arbitrary). The corresponding setting menu will appear.
  3. Use F2-F4 and the direction keys to set waveform parameters (Frequency/Period, Amplitude/High level, Offset/Low level, etc.).

Note: When setting parameters, long-press a direction key to accelerate value changes.

4.3.3. Set the Load

  1. Press the System key.
  2. Press F4 to go to the next menu page.
  3. Press F3 to switch the load setting (High Z / *Ω). The default is 50Ω.

Use the direction keys to change the load value. The load range is 1Ω - 10KΩ.

4.3.4. Built-in Waveform List (Arbitrary Waveform)

The system has 8 built-in arbitrary waveforms. To select one:

  1. Press F3 or F4 to switch to the Type parameter.
  2. Use F2 or the direction keys to select the desired built-in waveform.
Built-in Waveform List
Name Explanation
AmpALT Attenuation oscillation curve
AttALT Gain oscillation curve
StairDn Stair-down waveform
StairUD Stair-up and stair-down waveform
StairUp Stair-up waveform
Besselj Besselj function
Bessely Bessely function
Sinc Sinc function

4.4. Communication with PC

The oscilloscope supports communication with a PC via USB for data storage, analysis, display, and remote control. Download the Oscilloscope software from the official website and follow the installation prompts.

  1. Connection: Use a USB data cable to connect the USB Device port on the oscilloscope's right panel to a PC's USB port.
  2. USB Port Settings: On the oscilloscope, go to System settings (Press System, then F4). Switch the USB protocol type to HID (Human Interface Device) for PC control. For file access, select MSC (Mass Storage Class).
  3. Software Operation: Run the Oscilloscope software on your PC. The connection information in the bottom right corner of the software will turn green upon successful connection.
Screenshot of oscilloscope software connected to PC
Figure 4-6: Connect with PC through USB port

5. Maintenance

5.1. General Maintenance

  • Do not store the instrument in direct sunlight for extended periods.
  • Caution: Do not allow spray, liquid, or solvent to touch the instrument or probes.
  • Cleaning: Regularly clean the external surfaces. Wipe floating dust with a soft cloth. For the LCD, be careful not to scratch the screen. For deeper cleaning, disconnect power and wipe with a damp, non-dripping soft cloth and mild detergent. Avoid abrasive chemical cleaning agents.

Warning: Ensure the instrument is dry before re-energizing to prevent electrical short circuits or injury from moisture.

5.2. Charging and Replacement of Battery

The device uses a lithium battery. For long-term storage, the battery may self-discharge. Pre-charge the device for 0.5 to 1 hour before turning it on if it has been stored for a long time. Charge regularly to prevent over-discharge.

5.2.1. Battery Charging

The lithium battery may not be fully charged upon delivery. Refer to the table below for charging and discharging times:

Battery Charging and Discharging Times
Model Charging Discharging
For models <100MHz ≥4.5h ≥4h
For models ≥100MHz ≥4h ≥3h

Battery indicator symbols:

  • Power-on charging status symbol Power-on charging status
  • Battery power supply symbol Battery power supply
  • Low battery symbol Only about five minutes of use time left. Charge immediately.

Charging Method: Connect the oscilloscope to a power socket using the provided USB data cable and power adapter. Alternatively, charge via a computer or other equipment's USB interface, ensuring the power supply's load capacity is sufficient.

Note: Do not exceed the allowable ambient temperature for charging to prevent battery overheating.

5.2.2. Replacement of Lithium Battery

Battery replacement should generally not be necessary. If required, it must be performed by qualified personnel using only lithium batteries of the same specification.

6. Troubleshooting

  • Oscilloscope cannot be turned on: The battery may be exhausted. Charge for 15 minutes before attempting to turn on. If the issue persists, contact support.
  • Oscilloscope turns off after a few seconds of startup: The battery is exhausted. Check the battery symbol and charge the device.
  • After switching to multimeter, measurement type displays as 'E': The measurement type may not be selected. Press F4 to select the type. If 'E' persists, restart the oscilloscope.
  • Measured voltage amplitude is 10 times larger or smaller than actual: Check if the channel attenuation coefficient matches the actual probe error ratio.
  • Waveform displays but cannot be stabilized: Check if the trigger mode menu's source matches the signal channel. Ensure the trigger electrical level is within the waveform range.
  • Nothing displayed after pressing RUN/STOP: Check if the trigger mode is normal or single, and if the trigger electrical level is out of range. Center the trigger electrical level or set trigger mode to automatic. Use Auto to complete settings.
  • Display speed becomes slower when average value sampling is set or duration is longer: This is normal behavior.

7. Technical Specifications

All technical specifications apply to probes with 10X attenuation. The oscilloscope must operate continuously for over 30 minutes at the specified temperature. If the temperature changes by more than 5°C, perform automatic correction.

7.1. Oscilloscope Specifications

Oscilloscope specifications table part 1
Figure 7-1: Oscilloscope Specifications (Part 1)
Oscilloscope specifications table part 2
Figure 7-2: Oscilloscope Specifications (Part 2)

7.2. Multimeter Specifications

Multimeter specifications table part 1
Figure 7-3: Multimeter Specifications (Part 1)
Multimeter specifications table part 2
Figure 7-4: Multimeter Specifications (Part 2)

7.3. Arbitrary Waveform Generator Specifications (Optional)

Waveform Generator specifications table
Figure 7-5: Waveform Generator Specifications

7.4. General Technical Specifications

General Technical Specifications
Characteristic Description
Display type 3.5-inch color LCD display
Display resolution 320 horizontal × 240 vertical pixels
Display color 65536 colors
Display Contrast Adjustable
Power supply 100 - 240 VACRMS, 50/60 Hz, CAT II; DC INPUT: 5VDC, 2A
Power consumption <5 W (HDS242(S), HDS272(S), HDS2102(S)); ≤6 W (HDS2202(S))
Battery 2200mAh*2 (3.7V, 18650) for <100MHz models; 2600mAh*2 (3.7V, 18650) for ≥100MHz models
Working temperature 0°C - 40°C
Storage temperature -20°C - +60°C
Relative humidity ≤90%
Operating height 3,000 meters
Non-operating height 15,000 meters
Cooling method Natural cooling
Dimensions 198 mm (length) x 96mm (height) x 38 mm (width)
Weight About 0.6 kg (main unit, without battery)
Calibration interval One year (recommended)

8. User Tips

  • The device is generally precise and easy to control, with the oscilloscope function working great.
  • Be aware that multimeter measurements (e.g., capacitor, voltage) may take 4-10 seconds to display results. This is normal behavior for the device.
  • The product is considered a professional-grade device by many users.

9. Warranty

The product is warranted to be free from defects in materials and workmanship for a period of 3 years from the date of purchase by the original purchaser. Accessories (probes, adapter) have a 12-month warranty. This warranty is non-transferable.

If the product is defective during the warranty period, the company will repair or replace it without charge. Replaced parts become the property of the company.

This warranty does not cover defects or damage caused by improper use, inadequate maintenance, unauthorized repairs, or connection to incompatible equipment. It also does not cover products modified or integrated with other products if such modifications increase servicing difficulty.

For service under warranty, notify the company before the warranty period expires. The customer is responsible for packaging and shipping the defective product to the designated service center, along with proof of purchase.

Except for the after-sales services provided, no other guarantees for maintenance, express or implied, including marketability and special-purpose acceptability, are offered. The company is not responsible for any indirect, special, or consequential damages.

10. Support & Contact Information

For product support, including user manuals and software downloads, visit the official OWON website: www.owon.com.hk/download

For further assistance, you may contact Fujian LILLIPUT Optoelectronics Technology Co., Ltd.:

  • Address: No. 19, Heming Road, Lantian Industrial Zone, Zhangzhou 363005 P.R. China
  • Tel: +86-596-2130430
  • Fax: +86-596-2109272
  • Web: www.owon.com.cn
  • E-mail: info@owon.com.cn

Please contact the nearest Sales and Service Offices for services.