1. Introduction
This series of multimeters represents the latest high-sensitivity pointer-type multimeters, crafted from rare earth magnetic materials and encased in soft rubber sleeves for enhanced shockproof and drop-resistant properties. Designed for a wide range of applications, it can accurately measure DC voltage and current, AC voltage and current, resistance, capacitance, frequency, diode forward voltage drop, Zener tube parameters, transistor hFE parameters, circuit continuity, battery power, infrared remote control emission signals, and facilitate fire line discrimination.
Image: Multi-function analog multimeter highlighting AC/DC voltage, DC current, diode, continuity, multiple protection, and optional carrying case.
2. Safety Information
When operating this instrument, users must adhere to all standard safety procedures, specifically focusing on preventing electric shock and avoiding misuse of the instrument. For your safety, always use the provided test leads and ensure they are in good condition before use.
2.1. Precautions for Use
- Avoid using the meter near equipment with significant electromagnetic interference, as this may cause unstable readings and large errors.
- Do not use the meter or test leads if they are damaged, to prevent electric shock.
- Improper use may impair the meter's safety functions.
- When working near exposed conductors or busbars, avoid touching the wires and ensure safety during measurement.
- Do not use this instrument in environments with explosive gas, vapor, or dust.
- Always use the correct input terminal, function, and range for measurements.
- Ensure the input value does not exceed the specified input limit for each range to prevent instrument damage.
- Do not touch unused inputs when the meter is connected to the circuit under test.
- Exercise caution to avoid electric shock when the measured voltage exceeds 60VDC or 30VAC.
- When using test leads, keep your fingers behind the protective ring.
- Before changing the measurement range, ensure the test leads are disconnected from the circuit under test.
- Before performing resistance, diode, capacitance, or continuity measurements, disconnect the power supply to the circuit under test and discharge all high-voltage capacitors.
- Do not measure resistance, capacitance, or perform continuity tests on live circuits.
3. Panel Indication
Familiarize yourself with the components of your NJTY 88C multimeter:
Image: Detailed view of the multimeter's front panel with labeled components.
- Dial (Analog Display)
- Pointer Zero Adjustment
- hFE Test Socket
- Position Rotary Switch (Function/Range Selector)
- VΩmA Input Socket
- COM Input Socket (Common Terminal)
- 10A Input Socket
- 2500V Input Socket
- Resistance Zeroing Knob
- Infrared Receiving Window
- Infrared and Live Wire Identification Indicator Light
- Instrument Hanging Loop with Magnet (on the back)
- Clear Scale Dial without Ghosting
- Soft Rubber Sheath
- Infrared Emitter
- Transistor Socket
- Function Knob with Strong Detents and No Play
4. Setup
4.1. Battery Installation
The instrument requires two 1.5V AAA batteries for power supply. Ensure batteries are correctly installed in the battery compartment.
- Turn off the instrument power supply (if applicable for N model).
- Pull out all test leads from the input jacks.
- Loosen the screw securing the battery cover with a screwdriver.
- Remove the battery cover.
- Insert two new 1.5V AAA batteries, observing correct polarity.
- Replace the battery cover and tighten the screws.
Image: The multimeter package includes the manual, test leads, and two spare fuses (installed in the battery compartment).
5. Operating Instructions
Before use, check that the pointer aligns with the mechanical zero position. If not, adjust the "pointer to zero" knob on the cover until it aligns.
5.1. DC Voltage Measurement (DCV)
- Insert the black test lead into the "COM" jack and the red test lead into the "VΩmA" jack. For voltages greater than 1000V, insert the red test lead into the "2500V" jack.
- Set the rotary switch to the desired DCV range (e.g., 1000V for high voltage).
- Connect the test leads to the load or signal source under test.
Note:
- If the voltage range is unknown, start with the highest range and decrease step-by-step.
- If the pointer exceeds the maximum scale, the voltage has exceeded the range; adjust to a higher gear.
- Always exercise caution when measuring high voltage to avoid electric shock.
Image: Example of DC voltage testing using the multimeter.
5.2. AC Voltage Measurement (ACV)
- Insert the black test lead into the "COM" jack and the red test lead into the "VΩmA" jack. For voltages greater than 1000V, insert the red test lead into the "2500V" jack.
- Set the rotary switch to the desired ACV range.
- Connect the test leads to the load or signal source under test.
Note:
- Follow the same precautions as DC voltage measurement regarding unknown ranges, over-range indication, and high voltage safety.
Image: Example of AC voltage measurement using the multimeter.
5.3. Direct Current Measurement (DCA)
- Insert the black test lead into the "COM" jack and the red test lead into the "VΩmA" jack. For currents greater than 500mA, insert the red test lead into the "10A" jack.
- Set the rotary switch to the desired DCA range (e.g., 500mA or 10A).
- Connect the test leads in series with the circuit under test.
Note:
- If the current range is unknown, start with the highest range and decrease step-by-step.
- If the pointer momentarily exceeds the maximum scale, the current has exceeded the range; adjust to a higher gear.
Image: Internal circuit board with metal shielding cover, double-sided gold-plated fiber, diode dual-phase clamping circuit, self-recovery fuse design, neatly arranged wiring, and firmly soldered joints.
5.4. Resistance Measurement (Ω)
- Insert the black test lead into the "COM" jack and the red test lead into the "VΩmA" jack.
- Set the rotary switch to the desired Ω range.
- Connect the test leads to the resistance to be tested.
Note:
- Before measuring resistance in a circuit, ensure the power supply is cut off and any capacitors are discharged.
- The reading on the dial is the first scale line Ω.
Image: Example of resistance measurement using the multimeter.
5.5. Capacitance Measurement (C) (N model only)
- Set the rotary switch to the "C" range.
- Insert the black test lead into the "COM" jack and the red test lead into the "VΩmA" jack.
- Connect the measured capacitor to both ends of the test leads (red lead is positive). Observe polarity if necessary.
Note:
- Completely discharge the capacitor before each measurement to avoid errors.
- Connect polarized capacitors with correct polarity to prevent errors and loss resistance increase.
Image: Example of capacitance measurement using the multimeter.
5.6. Frequency Measurement (Hz) (N model only)
- Insert the black test lead into the "COM" jack and the red test lead into the "VΩmA" jack.
- Set the rotary switch to the "Hz" range.
- Connect the test leads to the frequency source (VPP>1V).
5.7. Transistor hFE Parameter Measurement
- Set the rotary switch to "hFE" (resistance × 10 gears).
- Determine if the transistor is PNP or NPN.
- Insert the E, B, and C pins of the transistor into the corresponding test sockets on the panel. The pointer deflection indicates the transistor's DC amplification factor hFE.
Note:
- If the pointer deflection is greater than 1000, check for incorrect pin insertion or a damaged transistor.
5.8. Diode Measurement
- Insert the black test lead into the "COM" jack and the red test lead into the "VΩmA" jack (black lead is positive).
- Set the rotary switch to the "→|" range. The pointer will deflect to the right full-scale position.
- Connect the test leads to the diode under test. The forward conduction pointer indicates the conduction voltage.
Image: Example of diode functionality test showing one-way conduction.
5.9. LED/Voltage Regulator Tube Measurement (L model only)
This function measures light-emitting tubes within 10V.
- Set the rotary switch to the "→|" position. The pointer will deviate to the right near full scale.
- The red test lead is the output voltage "+". Forward measurement indicates the forward voltage drop of the PN junction.
- It can also measure various types of LED light-emitting diodes, lighting them up while indicating the forward voltage drop.
- Reverse measurement provides the steady voltage value of the Zener tube (when reversed). Refer to the special scale for Zener tube readings.
Image: Example of light-emitting diode (LED) measurement.
5.10. Buzzer Continuity Test (•))))
- Insert the black test lead into the "COM" jack and the red test lead into the "VΩmA" jack.
- Set the rotary switch to the "•))))" range.
- Connect the test leads to both ends of the circuit to be checked.
- If the resistance is less than 30Ω (A/B/C/L type) or 15Ω (N type), the buzzer will beep.
Note:
- Always check continuity on a de-energized circuit to avoid false readings from load signals.
Image: Example of continuity buzzer measurement.
5.11. Infrared Remote Control Emission Signal Detection
This function helps determine if an infrared remote control transmitter is working correctly.
- Adjust the rotary switch to the "√" position.
- Align the transmitter head of the remote control vertically with the multimeter's infrared receiver (deviation not greater than ±15°).
- Press the remote control launch button. If the red light-emitting tube on the multimeter flashes, the transmitter is working normally.
- Move the transmitter within 1-30cm to assess its output power status.
Note:
- Keep the multimeter's receiver away from strong light sources when only detecting infrared signals, as strong light can activate the indicator.
- The remote control's transmitting head must be perpendicular to the instrument's infrared receiver (max deviation ±15°).
Image: Example of infrared remote control detection, showing infrared receiving and emitting lamps.
5.12. Live Wire Identification (LIVE) (B model only)
- Insert the red test lead into the "VΩmA" jack and the black test lead into the "COM" jack.
- Connect the red test lead to the live wire. The red indicator light will illuminate.
5.13. Battery Power Detection (BATT)
- Insert the black test lead into the "COM" jack and the red test lead into the "VΩmA" jack.
- Set the rotary switch to the "BATT" position.
- Connect the test leads to both ends of the battery, observing correct polarity.
- Observe the scale corresponding to BATT on the dial. Red indicates a low or damaged battery, while green indicates an intact battery.
Image: Example of battery level measurement, showing load measurement and static electricity prevention.
5.14. Standard Resistance Box Application (A/B/C/L only)
In specific cases, the meter can function as a standard resistance using the DC voltage or current range. For example, in the DC voltage range, the 2.5V range is equivalent to a 50k standard resistance (2.5V × 20k/V = 50k). Similarly, in the DC current range, the 5mA range is equivalent to a 100Ω standard resistance (0.5 + 0.005A = 100Ω). Other ranges can be deduced from technical specifications.
Note:
- Avoid overloading when using this function to prevent damage.
Image: Multimeter shown with a DC resistance box.
6. Maintenance
6.1. General Maintenance
- Before opening the instrument case or removing the battery cover, always pull out the test leads first.
- Only use specified replacement parts when servicing the meter.
- Before powering on, ensure all relevant power supplies are cut off and static electricity is discharged to prevent damage to components.
- Calibration and maintenance should only be performed by qualified professionals.
- Be aware that some internal capacitors may retain dangerous voltages even after the meter is powered off.
- If any abnormality is observed, remove the meter from service immediately and send it for repair. Ensure it passes inspection before reuse.
- For long-term storage, remove the batteries and store the meter in a place with moderate temperature and humidity.
- To avoid electric shock or damage, do not get the inside of the meter wet when opening the case or battery compartment.
- Regularly clean the meter case with a damp cloth and a small amount of detergent. Do not use abrasives or chemical solvents.
- Dirt or moisture in input sockets can cause oxidation of metal surfaces, affecting readings.
6.2. Input Protection Measures
- The maximum input voltage for voltage measurement is 1000V AC/DC (VΩmA jack) / 2500V (2500V jack).
- For frequency, resistance, continuity, and diode measurements, the meter can withstand no more than 250V AC or equivalent RMS voltage.
- For μA and mA current measurements, the meter is protected by a fuse or a self-recovery fuse of the same specification. If a fuse blows, replace it with an identical one. If a self-recovery fuse activates, it will generate heat; wait 30 seconds before measuring again to avoid affecting accuracy.
6.3. Replacing Battery and Fuse
- Turn off the instrument power supply (N model).
- Pull out all test leads from the input jacks.
- Loosen the screw securing the battery cover with a screwdriver.
- Remove the battery cover.
- Remove old batteries or damaged fuses.
- Replace with new 2×1.5V AAA batteries or new fuses.
- Replace the battery cover and tighten the screws.
7. Troubleshooting
- Unstable Readings / Large Errors: This may occur if the meter is used near equipment with strong electromagnetic interference. Move to a location with less interference.
- Pointer Exceeds Max Scale: If the pointer goes beyond the maximum scale during voltage or current measurement, the measured value has exceeded the current range. Adjust the rotary switch to a higher range.
- Resistance Scale Adjustment Less Than Zero: This indicates the battery in the meter is exhausted. Replace the batteries immediately to ensure measurement accuracy and prevent battery leakage.
- Buzzer Sounds on De-energized Circuit: Ensure the circuit under test is completely de-energized. Any residual load signal can cause the buzzer to sound, leading to incorrect judgment.
- Capacitance Measurement Errors: Ensure capacitors are fully discharged before measurement. Incorrect polarity for polarized capacitors can also lead to errors.
- Infrared Remote Control Detection Issues: Ensure the remote control's transmitting head is perpendicular to the multimeter's infrared receiver (max ±15° deviation). Avoid strong ambient light, which can cause false positives.
8. Specifications
8.1. Functional Parameters (Model 88C)
Image: Table detailing the functional parameters of the 88C model.
| Parameter | Value |
|---|---|
| Model | 88C |
| DC Current | 50uA-0.5mA-5mA-50mA-500mA-10A (±5%) |
| AC Voltage | 0-10V-50V-250V-500V-1000V-2500V (±5%) |
| DC Voltage | 0.25V-1V-2.5V-10V-50V-250V-500V-1000V-2500V (±5%) |
| Resistance | R×1 R×10 R×100 R×1K R×10K (±10%) |
| Standard Resistor Box | 0.025/0.5/5/50/500/5K/20K/50K/200K/1M/2.25M/4.5M/9M/22.5M |
| Continuity Buzzer | ✓ |
| Transistor | ✓ |
| Battery Detection | 1.2V/1.5V/2V/3V/3.6V |
| Circuit Board | Double-sided gold-plated PCB |
| Header | Internal magnet |
| Power Supply | AAA*2 batteries (not included) |
| Body Weight | Approx. 500g (excluding battery) |
| Body Dimensions | 179*118*50 (mm) |
8.2. Technical Parameters (Model N)
| Basic Function | Range | Resolution | Accuracy |
|---|---|---|---|
| DCA | 500μA/50mA/500mA | 10μA | ±2.0% |
| 10A | 0.2A | ±5% | |
| ACA | 50mA/500mA | 1mA | ±5% |
| 10A | 0.2A | ±5% | |
| DCV | 250mV/2.5V/10V/50V/250V/1000V (Input resistance 2MΩ) | ±2.0% | |
| 2500V (Input resistance 4.8MΩ) | |||
| ACV | 250mV/2.5V/10V/50V/250V/1000V (Input resistance 2MΩ) | ±5% | |
| 2500V (Input resistance 4.8MΩ) | |||
| Resistance (Ω) | R×1/R×10/R×100 (Central value 20) | ±10% | |
| R×1k/R×10k/R×100k | |||
| Capacitance(C) | ×1/×10/×100 (1000uF) | 1uF | ±5% |
| Frequency(Hz) VPP>1V | ×1/×10/×100 (10kHz) | 5Hz | ±5% |
| Buzzer | Works when lower than 15Ω |
8.3. Technical Parameters (Model A/B/C/L)
| Measurement Range | Sensitivity | Accuracy | Error Representation Method |
|---|---|---|---|
| DCA (50μA/0.5mA/5mA/50mA/500mA) | ±2.5% | Calculated as a percentage of indicated value | |
| DCA (10A) | 0.25V | ±5% | |
| DCV (0.25V/1V/2.5V/10V/50V) | 20KΩ | ±2.5% | |
| DCV (250V/500V/1000V) | ±2.5% | ||
| DCV (2500V) | 9KΩ | ±5% | |
| ACV (10V/50V/250V/500V/1000V/2500V) | ±5% | ||
| Resistance Ω (R×1/R×10/R×100/R×1k/R×10k/R×100k) | 中心值20 | ±10% | |
| BATT | 1.2V/1.5V/2V/3V/3.6V | RL=8Ω-12Ω (only models A/B/C) | |
| Buzzer | Works when lower than 30Ω | ||
| LV Detection (R×1 ×1k) | 3V | ||
| LV Detection (R×10k) | 12V | ||
| LED Detection | Voltage regulation tube within 10V, detection current: DC 5mA (only Model L) | ||
| hFE | 0~1000hFE (R×10) | ||
| Infrared Remoter Detection | Vertical angle±15°, distance 1-30cm (Red light-emitting tube indication (light up) (only B/C/L models)) | ||
| Standard Resistance Box Ω | 0.025/0.5/5/50/500/5k/20k/50k/200k/1M/2.25M/4.5M/9M/22.5M | ||
| LIVE Wire Detection | Red light-emitting tube indication (light up, 220V AC detection) (only model B) |
8.4. Comprehensive Index
- Environmental Conditions:
- Altitude: <2000m
- Working Environment Temperature and Humidity: 0~40°C (<80%RH)
- Storage Environment Temperature and Humidity: -10~60°C (<70%RH; remove the battery)
- Test or Calibration Ambient Temperature: 20℃±2℃
- Maximum Voltage: AC and DC 2500V allowed between the measuring terminal and earth.
- Fuse Protection:
- "mA" gear Fuse: 500mA instant fuse or self-recovery fuse of the same specification
- "10A" gear Fuse: 10A instant fuse
- Over-range Indication: Pointer exceeds the maximum scale of the dial.
- Power Supply: 2 x 1.5V AAA batteries.
- Dimensions: 179mm(L)×118mm(W)×50mm(H).
- Weight: Approximately 400g (without battery) for A/B type; Approximately 500g (without battery) for C/L/N models.
8.5. Accuracy Index
- Accuracy: ±(% of reading), guaranteed for one year from the date of delivery.
- Reference Conditions: Ambient temperature 18°C to 28°C, relative humidity not greater than 80%.
9. User Tips
- Always refer to this manual for specific measurement procedures and safety guidelines before operating the multimeter.
- For precise resistance measurements, ensure the circuit is completely de-energized and all capacitors are discharged.
- Keep the multimeter clean and dry to maintain optimal performance and prevent damage to the input terminals.