1. Product Description
The NJTY 3266 Series Clamp Meters (models 3266TA and 3266TD) are designed for professional electricians, offering smart and reliable solutions for diverse electrical measurement needs. These meters feature True RMS technology for accurate measurement of complex waveforms up to a 2kHz frequency response. Their 25mm large-diameter clamp heads easily accommodate industrial-grade thick cables. An intelligent safety system provides 600V over-voltage protection and a non-contact voltage (NCV) detection function with both sound and light alarms. Data integrity is secured with dual insurance features, including an automatic shutdown memory and manual hold mode. Built to endure tough environments, the meters boast a military-grade design with IP54-rated dust and water resistance, and a lightweight 165g body for extended use.
The series includes four models (A, B, C, D), with 3266TA corresponding to models A/B and 3266TD to models C/D, each offering specific functionalities as detailed in the specifications section.
2. Safety Precautions
All NJTY NCV series clamp meters are designed to meet international safety standards, including IEC61010-1, IEC61010-031:2002 & IEC61010-2-032 Measurement Category (CAT.) II 600V.
2-1: Safety Specifications
The device is rated for Measurement Category II 600V.
2-2: Precautions
- Do not use the instrument on a circuit where the voltage exceeds AC 600V.
- Do not use the instrument in the presence of flammable gases.
- Do not use the instrument above 2000m elevation.
- The working condition of the instrument is: less than 2000m high.
- Operating temperature range is between -10℃ and 50℃.
3. Symbols on Panel
Familiarize yourself with the various symbols and controls on the clamp meter's panel for proper operation.

- Current Clamp: Used for non-contact AC current measurement.
- NCV Indicator (Red LED): Flashes and buzzes when non-contact voltage is detected.
- Rotary Switch: Used to select the measurement function and range.
- HOLD/BL Button: Press to hold the current reading on the display. Press again to release. For 3266TD, this button also controls the backlight.
- FUNC Button: Used to switch between multiple functions within a single rotary switch position (e.g., ACV/Frequency, Diode/Continuity).
- MAX/REL Button: For 3266TA (models A/B), this is the MAX button to hold the maximum value. For 3266TD (models C/D), this is the REL button for relative value measurement.
- Trigger: Used to open and close the current clamp jaws.
- LCD Screen: Displays measurement readings, symbols, and indicators.
- V/Ω/Hz Input Socket: Positive input terminal for voltage, resistance, diode, temperature, capacitance, and frequency measurements.
- COM Input Terminal: Common (negative) input terminal for all measurements.
- NCV Induction Zone: Area where non-contact voltage detection occurs.
4. Specifications
The NJTY 3266 series offers two main models, 3266TA and 3266TD, with varying capabilities. Below are the detailed specifications for each.
4-1: 3266TA (A/B Model) Specifications
| Function | Range | Max. Resolution | Accuracy |
|---|---|---|---|
| AC Current (ACA) | 2A, 20A, 200A/600A | 1mA, 0.1A | ±(3%rdg+5dgt), ±(2%rdg+5dgt) |
| AC Voltage (ACV) | 2V/20V/200V/600V | 1mV | ±(1.2%rdg+5dgt) |
| DC Voltage (DCV) | 200mV/2V/20V/200V/600V | 0.1mV | ±(0.8%rdg+5dgt) |
| Resistance | 200Ω/2KΩ/20KΩ/200KΩ/2MΩ/20MΩ | 0.1Ω | ±(1.2%rdg+5dgt) |
| Diode | Open voltage: 1.999V | ||
| Continuity | Buzzer sounds if resistance < 50Ω |
Note: Temperature measurement is only available for B model. The accuracy of current is guaranteed when the current is more than 1A. The input resistance impedance of voltage is 10MΩ.
4-2: 3266TD (C/D Model) Specifications
| Function | Range | Max. Resolution | Accuracy |
|---|---|---|---|
| AC Current (ACA) | 40A/600A | 10mA | ±(2.5%rdg+5dgt) |
| AC Voltage (ACV) | 4V/40V/400V/600V | 1mV | ±(1.2%rdg+5dgt) |
| DC Voltage (DCV) | 400mV/4V/40V/400V/600V | 0.1mV | ±(0.8%rdg+5dgt) |
| Frequency | 4Hz/40Hz/400Hz/4KHz/40KHz/400KHz/4MHz/10MHz | 0.001Hz | ±(0.5%rdg+5dgt) |
| Capacitance | 4nF/40nF/400nF/4uF/40uF/400uF/4mF | 1pF | ±(3%rdg+5dgt) |
| Temperature | -20℃ to 1000℃ (-4℉ to 1832℉) | 1℃ (1℉) | ±(3%rdg+5dgt) |
| Resistance | 400Ω/4KΩ/40KΩ/400KΩ/4MΩ/40MΩ | 0.1Ω | ±(1.2%rdg+5dgt) |
| Diode | Open voltage: 1.999V | ||
| Continuity | Buzzer sounds if resistance < 50Ω |
Note: Temperature and capacitance measurements are only available for D model. The accuracy of current is guaranteed when the current is more than 1A. The input resistance impedance of voltage is 10MΩ.
4-3: Other Features
- Operating Temperature: 0℃ to 40℃ & <75%RH humidity.
- Storage Temperature: -10℃ to 50℃ & <75%RH humidity.
- Auto Power-OFF: The instrument will automatically turn off after 15 minutes of no operation to conserve energy. Fluctuations in the device gear will wake up the instrument again.
- Display: LCD high-definition backlit display for clear readings in various lighting conditions.
- Flashlight: Built-in flashlight for working in dark environments.
- Jaw Design: 25mm extra-large jaw for easy handling of thick cables.
5. Operating Instructions
5-1: AC Current Measurement
For 3266TA (A/B models): Two ranges: ~200/600A and ~2/20A.
For 3266TD (C/D models): One range: ~40A/600A.
- Set the Rotary Switch to the desired AC Current range.
- Press the trigger to open the transformer jaws and clamp them onto a single conductor under test.
- Read the measurement value on the LCD display.
Notes:
- Do not clamp two conductors with opposite current directions simultaneously, as the currents may cancel each other out, leading to incorrect readings.
- Keep your fingers under the arc-shaped part of the clamp to ensure safety during measurement.

5-2: AC and DC Voltage Measurement
The maximum input voltage is 600V.
- Switch the Rotary Function to the desired ACV (V~) or DCV (V-) position.
- Insert the red test lead into the V/Ω/Hz terminal and the black test lead into the COM terminal.
- For DC Voltage measurement, connect the red and black test leads to the positive (+) and negative (-) sides of the circuit under test, respectively.
- For AC Voltage measurement, connect the test leads across the circuit.
- For 3266TD (C/D models), pressing the 'FUNC' button in ACV mode can measure frequency (about 10Hz to 30kHz).
5-3: Resistance Measurement
- Set the Rotary Function Switch to the Ω position. For 3266TD (C/D models), use the 'FUNC' key to switch until 'Ω' is displayed. (3266TA/A/B models do not need this extra step).
- Insert the red test lead into the V/Ω/Hz terminal and the black test lead into the COM terminal.
- Read the measurement result directly from the LCD display.
Note:
- When measuring small resistances (considering test lead contact resistance of about 0.2-0.3Ω), connect the two leads directly and record this resistance first. Then, subtract this recorded value from subsequent resistance readings.
- When checking in-circuit resistance, ensure the circuit under test has all power removed and all capacitors are fully discharged.
5-4: Diode & Continuity Check
- Set the Rotary Function Switch to the position indicated by the diode and continuity symbols (e.g., '→|' and '•)))').
- Press the 'FUNC' key to switch between Diode ('→|') and Continuity ('•)))') modes.
- In Continuity mode, insert the test leads across the conductor under test. The buzzer will sound if the resistance is 50Ω or less.
- For Diode test, insert the red test lead into the V/Ω/Hz terminal and the black test lead into the COM terminal. Connect the red and black test leads to the positive and negative terminals of the diode, respectively. The display will show the forward voltage drop. If the test leads are connected with reverse polarity and 'OL' is displayed, the diode is functioning correctly.
5-5: Frequency Measurement (3266TD only)
When the input voltage is more than 30V, use the 'Hz/V' function to measure frequency. Frequency measurement range is 0.001Hz to 10MHz (Sensitivity: 0.2Vpp-8Vpp).
- Connect the output wires of the frequency power source to the 'COM' and 'V/Ω/Hz' terminals. If direct connection is not suitable, use test leads.
- The reading on the LCD screen is the measured frequency.
5-6: Capacitance Measurement (3266TD only)
- Set the Rotary Function Switch to the capacitance position (e.g., 'F') and then press the 'FUNC' key to select the Capacitance Measurement Mode.
- Insert the test leads across the capacitor and read the result on the LCD display.
Note:
- When testing 'pF' capacitance, use the shortest possible test leads to reduce the impact on readings.
- To avoid damage to the meter or equipment, disconnect power and discharge all high-voltage capacitors before measuring capacitance. If capacitance is larger than 10uF, it may take more than 30 seconds to get a stable reading.
5-7: Temperature Measurement (3266TD only)
This function allows you to measure temperature. When switching to this range, the ambient temperature is displayed. After inserting the K-type Temperature Probe, the temperature of objects can be measured.

5-8: NCV Function (Non-Contact Voltage Detection)
The NCV function detects the presence of AC voltage without physical contact. When an electric field exceeding a certain threshold is detected by the inductive sensor in the jaws, the red LED on the upper right area of the panel will flash rapidly, and an alarm buzzer will sound. This indicates the presence of voltage in an electrical circuit or equipment without touching it.
- Set the Rotary Function Switch to the 'NCV' position.
- As the clamp jaws approach a voltage source (power strip or strong electric field), the NCV LED will flash and the buzzer will sound. The closer the meter is to the electric field, the faster the flashing and buzzing frequency. The display will show more '-' symbols (e.g., '----') for stronger fields and fewer for weaker fields.
- This position can also be used to identify live wires. Randomly insert black or red test leads; live wires will cause continuous flashing and buzzing.


6. Maintenance
- Battery Replacement: When the low battery symbol is displayed on the screen, it is necessary to remove the back cover and immediately replace the batteries with two new 1.5V AAA batteries of the same type.
- Cleaning: Do not use abrasives or solvents on the meter. Clean it using only a damp cloth and mild detergent.
- Storage: When not in use for a long period of time, remove the batteries to prevent leakage and corrosion. Avoid storing the instrument in high temperature and high humidity environments.
7. Accessories
The following accessories are typically included with your NJTY Digital Clamp Meter:
- Test Leads: 1 set
- Battery: 1.5V AAA Type x 2pcs (not included in some packages)
- User's Manual: 1pcs
- Carrying Case: 1pcs
- K-Type Temperature Probe: 1 set (Only for 3266TD/B & D models)
8. User Tips
- Battery Contact: If you experience issues with the device not powering on or intermittent operation, check the battery contacts. Some AAA batteries may have slight variations in size, leading to contact issues. Ensure batteries are firmly seated.
- Portability: The compact size and lightweight design make this clamp meter highly portable. Utilize the included carrying case for easy transport and protection, especially for on-the-go troubleshooting.
- Backlight: For measurements in dimly lit areas, remember to activate the backlight function (if available on your model) for clear visibility of readings.
9. Troubleshooting
This section addresses common issues you might encounter with your NJTY Digital Clamp Meter.
Common Questions & Solutions
- Q: Does the clamp meter measure direct current (DC) or only alternating current (AC)?
A: The NJTY 3266TA/3266TD models are primarily designed for AC current measurement via the clamp. However, they can measure both AC and DC voltage using the test leads. Please refer to the specifications for detailed ranges. - Q: The display shows 'OL' (Overload) during measurement. What does this mean?
A: 'OL' indicates that the measured value exceeds the selected range of the instrument. Try switching to a higher range or ensure the input is within the meter's maximum capabilities. - Q: The meter does not power on or turns off unexpectedly.
A: Check the batteries. Ensure they are correctly installed and have sufficient charge. Replace them if necessary. The meter also has an auto power-off feature after 15 minutes of inactivity; rotate the function switch to wake it up. - Q: Readings are unstable or inaccurate.
A: Ensure test leads are properly connected and not damaged. Verify that the correct function and range are selected for the measurement. For resistance measurements, ensure the circuit is de-energized and capacitors are discharged. Avoid strong electromagnetic interference.
10. Warranty and Support
For warranty information and technical support, please contact your retailer or the official NJTY customer service. Keep your purchase receipt as proof of purchase. This manual contains general instructions; for specific issues not covered, professional assistance is recommended.