HM Digital PH-200, ORP-200, COM-300

Contents

HM Digital Water Quality Tester User Manual

Models: PH-200, ORP-200, COM-300

1. Overview

Thank you for purchasing your HM Digital Water Quality Tester. These advanced, microprocessor-based handheld meters provide superior accuracy for various water quality measurements. Depending on the model, they can test pH, Oxidation Reduction Potential (ORP), Electrical Conductivity (EC), Total Dissolved Solids (TDS), and temperature.

The meters are designed for a wide variety of applications, including hydroponics, aquariums, water purification, and environmental testing. They feature automatic temperature compensation (ATC) and digital calibration for ease of use and precise results.

Three HM Digital water quality testers: ORP-200, COM-300, and PH-200

Figure 1: HM Digital PH-200, ORP-200, and COM-300 Water Quality Testers

Video 1: Overview of HM Digital Water Quality Testers

2. Product Models and Features

2.1 HM PH-200 pH/TEMP METER

The PH-200 is designed to measure pH levels and temperature with high accuracy. It is ideal for various applications where pH balance is crucial.

  • Measures pH and Temperature
  • Auto-ranging calibration with digital fine tuning
  • Waterproof housing (IP-67 rating)
  • Simultaneous temperature display
  • Digital Calibration (push button)
  • Auto-off function, data-hold function and low-battery indicator
  • Large and easy-to-read LCD screen
PH-200 meter displaying pH 7.16 and 29.9°C, with a pH scale from 1 (acidic) to 14 (alkaline) showing common liquids.

Figure 2: PH-200 pH Measurement and Scale

PH-200 meter partially submerged in water, showing IP67 waterproof rating and pH/temperature ranges.

Figure 3: PH-200 IP67 Waterproof Feature

2.2 HM ORP-200 ORP/TEMP METER

The ORP-200 is specifically designed for Oxidation Reduction Potential (Redox) measurements, crucial for applications like water purification and ozone system monitoring.

  • Measures Oxidation Reduction Potential (Redox) and Temperature
  • Auto-ranging calibration with digital fine tuning
  • Waterproof housing (IP-67 rating)
  • Simultaneous temperature display
  • Digital Calibration (push button)
  • Auto-off function, data-hold function and low-battery indicator
  • Large and easy-to-read LCD screen
  • Factory Calibrated to +92 mV
ORP-200 meter partially submerged in water, showing IP67 waterproof rating and ORP reading.

Figure 4: ORP-200 IP67 Waterproof Feature

ORP-200 meter in a glass of water, surrounded by circular images depicting various applications like aquaculture, cooling towers, swimming pools, hot tubs, and water disinfection.

Figure 5: ORP-200 Applications

2.3 HM COM-300 EC/TDS/PH/TEMP METER

The COM-300 is a versatile combo meter that measures four parameters: pH, Electrical Conductivity (EC), Total Dissolved Solids (TDS), and temperature. It offers selectable modes for different measurement scales and conductivity conversion factors.

  • Measures pH, Electrical Conductivity (EC), Total Dissolved Solids (TDS), and Temperature
  • 2 different conductivity conversion factors (0.5 (NaCl) or 0.7 (442TM))
  • Selectable modes for measurements in S, mS, and ppm scales
  • Automatic Temperature Compensation (ATC)
  • Digital Calibration (push button)
  • Waterproof housing (IP-67 rating)
COM-300 meter in a glass of water, showing display modes for EC, TDS, and PH with temperature readings.

Figure 6: COM-300 Multiple Test Modes (EC, TDS, PH)

COM-300 meter in a glass of water, illustrating the relationship between water quality (low vs. high EC) and health, with a TDS value chart.

Figure 7: COM-300 Water Quality and Health

3. Setup

3.1 Housing and Screen Description (PH-200 Example)

Diagram of the ORP-200 meter showing its components: Lanyard hook, Battery case, Waterproof ring, Low battery indicator, Warning symbol, Electrode, On/Off button, TEMP/CAL button, HOLD button, Plastic ring, Protective cap.

Figure 8: Meter Components (ORP-200 shown, similar for PH-200/COM-300)

LCD Display Indicators:

  1. Low battery indicator
  2. Warning symbol
  3. pH/ORP/EC/TDS measurement
  4. Temperature measurement
  5. Celsius/Fahrenheit mode
  6. Hold mode

3.2 Battery Installation

The meters use 2 x AAA batteries (not included). When the meter displays a flashing battery symbol, your batteries are getting weak and should be replaced soon.

  1. Twist open the end cap on the top of the meter.
  2. Twist the battery cap to the open position.
  3. Remove the old batteries.
  4. Insert new batteries with the "+" side facing up.
  5. Close the battery cap and turn to the close position.
  6. Close and tighten the end cap to maintain waterproofness.

NOTE: Do not reverse the polarity of the batteries. This may short circuit the meter.

Diagram showing how to open the battery case and insert AAA batteries into the PH-200 meter.

Figure 9: Battery Installation (PH-200 shown, similar for other models)

3.3 Initial Sensor Preparation (PH-200)

The PH-200 meter's electrode is shipped with a wetted sponge of Potassium Chloride (KCl) storage solution. If the sponge is dry (solutions can evaporate), use the included Mylar pack of storage solution to saturate the probe before first use. Let it sit in the solution for a minimum of 30 minutes.

For new meters, for your first few tests, lightly swirl the meter in the water or solution you are testing 3 times and allow the meter to sit while adjusting to the reading.

If you need to test the pH of low conductivity/TDS water (below 50 µS/25 ppm), you should do so only after using the meter a few times in a pH buffer solution or higher conductivity water.

Never touch the glass sensor or reference tube with your fingers. Be sure not to hit the sensor or reference tube against a glass or beaker, to avoid cracking it.

Always make sure that the cap is on tight after usage.

4. Operation

4.1 Taking pH Measurements (PH-200 & COM-300)

  1. Remove the cap.
  2. Click the ON/OFF button. The display will activate.
  3. Dip the meter into the water sample, liquid or solution to be tested.
  4. Lightly swirl the meter to ensure the removal of trapped air bubbles or electric charges. Do not tap it against the glass.
  5. The meter will display a reading almost immediately. Keep the meter in the liquid until the reading stabilizes (approx. 5-60 seconds) for an accurate reading.
  6. To view the reading out of the liquid, click the HOLD button while the meter is in the liquid. This will freeze the reading on the screen. Clicking the HOLD button again will release it.
  7. Click the ON/OFF button to turn the meter off.
  8. Shake any excess water off the meter. For best results, rinse with distilled or deionized water after each usage and wipe with a tissue. Put the cap back on. Store the meter standing upright to ensure complete saturation of the sensor.

NOTE: Minor fluctuations are a normal function of pH in certain environmental situations.

NOTE: The PH-200 is extremely sensitive in low conductivity water (below 50 µS). It is not recommended to use this meter in water below 10 µS. If you do use the meter in such water, it is better to test in flowing water, or by swirling the meter continuously while obtaining the reading. The reading should stabilize in 60-90 seconds.

4.2 Taking ORP Measurements (ORP-200)

Follow similar steps as pH measurement. Ensure the ORP-200 is in ORP mode if it has multiple modes (COM-300). Dip the electrode into the sample and wait for the reading to stabilize.

4.3 Taking EC/TDS Measurements (COM-300)

Follow similar steps as pH measurement. Ensure the COM-300 is in the desired EC or TDS mode. Dip the electrode into the sample and wait for the reading to stabilize.

4.4 Taking Temperature Measurements (All Models)

The temperature reading is always displayed on the LCD panel during measurement mode and is shown simultaneously with pH/ORP/EC/TDS readings. It is not shown when the meter is in calibration mode. The default reading for the meter is Celsius.

  1. Remove the cap.
  2. Click the ON/OFF button. The display will activate.
  3. The temperature reading is always displayed on the LCD screen (except in calibration and reset mode), and can be used to take air or liquid temperature measurements.
  4. To measure the temperature of a water sample or solution, dip the meter into it. The temperature reading will change immediately (unless the liquid is at room temperature). For very hot or cold liquids, the reading may take slightly longer to stabilize.
  5. Click the ON/OFF button to turn the meter off.
  6. Shake any excess water off the meter and rinse with distilled or deionized water. Wipe the outside of the sensor dry with soft tissue and put the cap back on.

NOTE: Because of the sensitivity of the pH sensor and reference tube, it is not recommended to use your meter as a thermometer in very hot or very cold liquids. Do not keep the meter in very hot or cold liquids for extended periods.

4.5 Switching Temperature Modes (All Models)

The default temperature reading is Celsius. To change the temperature mode, click the TEMP/CAL button to switch from Celsius to Fahrenheit or from Fahrenheit to Celsius.

5. Calibration

5.1 Automatic pH Calibration (PH-200 & COM-300)

Your PH-200 (and COM-300 for pH) has been factory calibrated to pH 7.0. While this is suitable for many applications, it is recommended to recalibrate the meter as close as possible to the pH level that will be tested. Recalibrating prior to every test will provide superior results (though this is not required). The PH-200 should be recalibrated at least once per month. If tests are conducted between wide ranges, calibration should be done more frequently. Additionally, it should be noted that liquid pH buffer accuracy may vary. For best results use HM Digital's pH Buffers (Model: PH-BUF).

The meters feature digital auto-calibration to pH 4.0, 7.0 or 10.0.

NOTE: If calibrating to 10 after the meter was calibrated to 4 (or vice-versa), it is recommended to calibrate to 7 first, prior to calibrating to the next value. Also, if you are having trouble calibrating to 4 or 10, calibrate to 7 first, then 4 or 10.

  1. Turn the meter on by clicking the ON/OFF button. Note that the reading will fluctuate if the meter is not submerged in a liquid (it will stabilize in liquid).
  2. Make sure the meter is in pH mode.
  3. Insert the meter into a pH buffer solution of 4.0, 7.0 or 10.0. Press and hold the 'CAL' button for 5 seconds.
  4. The temperature reading will change to a flashing 'CAL' image for 10 seconds. Do not press any buttons. The meter will automatically recognize the solution it is in (or the closest to it). This number will flash and the letters of 'CAL' will flash as 'C...CA...CAL' indicating progress. Allow 5 - 60 seconds during this stage. Do not press any buttons and do not move the meter.
  5. When the meter is calibrated, 'End' will flash briefly and the screen will revert to measurement mode. Your meter is now recalibrated.
PH-200 display showing 'pH 4.0' flashing and 'CAL' indicating calibration in progress, then 'End' for completion.

Figure 10: PH-200 Calibration Process

5.2 ORP Calibration (ORP-200)

The ORP-200 meter is factory calibrated to +92 mV. It can be easily recalibrated with digital calibration using the push buttons, rather than a screwdriver. Refer to the specific instructions for ORP calibration in your full manual if needed.

5.3 EC/TDS Calibration (COM-300)

The COM-300 features digital calibration for EC and TDS. Refer to the specific instructions for EC/TDS calibration in your full manual if needed. A liquid calibration solution is required for TDS and EC measurements.

5.4 Temperature Calibration (All Models)

Only perform this if the temperature reading is incorrect.

  1. You must have a correct thermometer next to the meter. Double-check the temperature first.
  2. With the meter off and the cap off and the meter in the air, press and hold ON/OFF and TEMP/CAL for 5 seconds. The meter will enter temperature calibration mode.
  3. Adjust to the correct temperature by pressing the UP or DOWN buttons.
  4. Once correct, click the ON/OFF button. The temperature will flash and the screen will display a flashing C...CA...CAL.
  5. The screen will briefly display 'End' and then revert to measurement mode. The temperature is now calibrated.

6. Maintenance & Care

6.1 General Techniques

  • Store the meter upright (standing on the cap, so that you can read the text).
  • Do not store the meter in high temperature or direct sunlight.
  • Never touch the glass sensor! If you do touch the sensor, rinse immediately in a pH 7 buffer solution and soak in the solution for 2 hours.
  • Always immediately replace the cap when the meter is not in use. The cap contains a small wetted sponge or cotton ball to ensure the gel within the glass sensor does not dry up. A tightened cap will ensure a longer lifespan for your sensor.
  • It is normal for salt deposits to form on the outside of the cap or rim of the cap. This is from the storage solution contained within the cap. Simply wipe off with a soft tissue or rubbing alcohol.
  • Using the meter in high temperature liquid (e.g., hot coffee or spas) will shorten the lifespan of the sensor. If testing in hot liquid is essential, do not keep the meter in the hot liquid for an extended period of time.
  • After repeated usage in high TDS liquids, it is advised to rinse the sensor in distilled water to prevent residue build-up.
  • If testing two samples over a wide range (e.g., pH 4 and pH 10), make sure to rinse the sensor in distilled water or a pH 7 buffer solution between and after tests.
  • For optimum performance, it is best to use the meter at least once per month.

6.2 Sensor Care

  • Never touch the glass electrode or reference tube.
  • To clean the body, use a soft rag or towel with water and a mild soap or rubbing alcohol.
  • To clean the electrode (glass bulb), rinse in distilled water or a pH 7 buffer solution. If cleaning with a pH 7 buffer, lightly dab (touch) the sensor on a soft tissue after. Never rub the electrode with any type of material, as this may scratch the glass.
  • If the glass electrode is scratched or broken, it must be replaced.
  • There is a gel encased in the electrode. If this is dried, you can attempt to re-moisten it by keeping it in a pH 7 solution for at least two hours. If the meter still does not function properly or accuracy is not at optimum performance, the sensor may be irreparably damaged.
  • For best results, clean the sensor after each use, especially if used in very low or very high pH liquids, or liquids other than water. If testing liquid that is not clear, clean in distilled water.
  • SENSOR STORAGE: HM Digital's electrode storage solution is preferred. The meter is shipped with this solution in the cap, and a small extra bottle. Additional solution (PH-STOR) can be purchased. It is okay to rinse the sensor in distilled water, but never store it in distilled water.
Diagram showing how to clean the platinum band of the ORP-200 sensor with fine sandpaper.

Figure 11: Sensor Cleaning (ORP-200 shown)

6.3 Sensor Replacement

If your sensor has been damaged, you can purchase a new one without having to purchase a new meter. The part number is SP-P2 (for PH-200). To replace the sensor:

  1. Remove the sensor gasket ring by twisting counter-clockwise.
  2. Gently pull the sensor off the unit.
  3. Gently insert the new sensor into the unit. Be sure to align the grooves properly. Never force the sensor into the unit!
  4. Make sure the rubber gasket O-ring is properly positioned on the sensor.
  5. Screw the orange gasket ring back onto the unit by twisting it clockwise. Hand tighten.
  6. Reset to the factory default and calibrate to pH 7.0 (see section 5.1).

6.4 Sponge Care

  • Never add distilled or tap water to the sponge/cotton ball, as mildew or mold may form.
  • Rewet with HM Digital's storage solution (PH-STOR).

7. Troubleshooting

You will see corresponding message symbol(s) with the issues listed below:

Issue Symbol Error Code
The sensor has been disconnected ▲ Err
The electrode is broken, damaged or old ▲ ---
The TDS level of the water is less than 3 ppm ▲ ---
The pH level is below 0 or greater than 14 --- ---
There is internal damage --- Err
The meter cannot calibrate --- Err
The temperature is out of range --- Err

7.1 Resetting to Factory Default (Advanced Users Only)

Follow these steps only if:

  • You have replaced the sensor.
  • Your meter has been used to continuously check widely varying pH levels.
  • You find your meter is not working at optimum performance.

To reset to the default pH Settings:

  1. Turn on the meter in the air with the cap off (do not submerge in a liquid for this process).
  2. Press and hold the HOLD button for 10 seconds.
  3. 'End' will briefly appear. Release the HOLD button. Turn the meter off.
  4. Recalibrate to pH 7.0 (see section 5.1).
  5. Turn the meter off again. Your meter has been reset to the factory default settings.

8. Specifications

8.1 PH-200 pH/TEMP METER Specifications

  • pH Range: 0.0-14.0 pH
  • Temperature Range: 0.1-55°C; 32-130°F
  • Resolution: pH: 0.01 pH, Temperature: 0.1°C/°F
  • Accuracy: +/- 0.1 pH, +/- 1°C/°F
  • Temperature Compensation: Automatic (ATC) to 25°C (ATC range: 1-50°C)
  • Calibration: Push Button Auto-calibration to 4pH, 7pH, and 10pH
  • Auto Shut-Off: After 5 minutes
  • Probe: Glass sensor and glass reference tube (single junction) (Detachable)
  • Safe Operating Temperature: 0-50°C; 32-122°F (measuring in extreme temperatures is not recommended)
  • Display: LCD panel
  • Housing: Water resistance (splash proof)
  • Power source: 2 x AAA batteries
  • Dimensions: 18.5 x 3.4 x 3.4 cm (7.3 x 1.3 x 1.3 inches)
  • Weight: 127g (4.5 oz) with batteries

8.2 ORP-200 ORP/TEMP METER Specifications

  • ORP Range: -999 to +1000 mV
  • Temperature Range: 0-80°C; 32-176°F
  • Resolution: 1 mV; Temperature resolution is 0.1°C/F
  • Accuracy: 0.5% (+/-2mV)
  • Calibration: Digital calibration by push button.
  • Electrode: Replaceable glass sensor and reference tube electrodes
  • Housing: IP-67 Waterproof (submersible; floats)
  • Power source: 2pcs AAA Battery (not included)
  • Dimensions: 18.5 x 3.4 x 3.4 cm (7.3 x 1.3 x 1.3 inches)
  • Weight: 96.4 g (3.4 oz)

8.3 COM-300 EC/TDS/PH/TEMP METER Specifications

  • EC Range: 0 - 9990 µS; 0 - 9.99 mS
  • TDS Range: 0 - 8560 ppm (mg/L)
  • PH Range: 0.0-14.0 pH
  • Temperature Range: 0.1-50 °C; 32-122°F
  • Resolution: PH: 0.1 pH; EC: 0-999: 1 µS; 1000-9990: 10 µS; 0-9.99mS: 0.01mS; TDS: 0-999: 1ppm; 1000-9990: 10ppm; Temp: 0.1 °C/℉
  • Accuracy: PH: +/- 0.1%; EC/TDS: +/- 2%; Temp: +/- 1% °C/F
  • Calibration: Digital calibration by push button.
  • Probe: Detachable platinum electrodes
  • Housing: IP-67 Waterproof (submersible; floats)
  • Power source: 2pcs AAA Battery (not included)
  • Dimensions: 18.5 x 3.4 x 3.4 cm (7.3 x 1.3 x 1.3 inches)
  • Weight: 174g
  • ATC: Yes

9. User Tips

  • For accurate TDS and EC measurements, always use a liquid calibration solution.
  • Ensure the protective cap is always on tight after use to prevent the sensor from drying out and to maintain waterproofness.
  • Avoid touching the glass sensor with your fingers to prevent contamination from skin oils.
  • If testing low conductivity water, use the meter a few times in a pH buffer solution or higher conductivity water first for better stability.

10. Warranty & Support

10.1 Limited Warranty

HM Digital, Inc. warrants its products against defective materials and workmanship from the date of purchase:

  • PH-200 Meter: One (1) Year Limited Warranty
  • Detachable Sensor (SP-P2): Six (6) Month Limited Warranty

What is covered: Repair parts and labor, or replacement at the Company's discretion. Transportation charges for repaired or new product to be returned to the purchaser.

What is NOT covered: Transportation charges for defective products to be sent to the Company. Any consequential damages, incidental damages, or incidental expenses, including damages to property. This includes damages from abuse or improper maintenance such as tampering, wear and tear, water damage, or any other physical damage. The meters are watertight and completely submersible. However, please ensure that the battery compartment and probe gasket ring are firmly tightened before submerging in water. The warranty does not cover water damage due to parts not securely closed.

Implied Warranties: Any implied warranties, including implied warranties of merchantability and fitness for a particular purpose, are limited in duration to ONE YEAR from the date of purchase. Some states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you. To the extent any provision of this warranty is prohibited by federal and state law and cannot be preempted, it shall not be applicable. This warranty gives you specific legal rights, and you may also have other rights, which vary from state to state.

NOTE: Warranties are product-specific. Third-party products and products deemed by HM Digital as "accessories" are not covered under warranty. Third-party products include, but are not limited to, batteries and fittings. Accessories included, but not limited to, lanyards and product boxes.

10.2 Contact Information

If you have any problems or questions regarding your meter, please contact HM Digital, Inc:

  • Email: Info@HMDigital.com
  • Website: HMDigital.com
  • Phone: 1-800-383-2777

To obtain warranty service, please contact 800.383.2777 or email Warranty@HMDigital.com to receive further instructions. Before sending the product back to us, please include your name, description of problem, phone number/address, and proof of purchase (must include date).

© 2024 HM Digital, Inc. All rights reserved.