ELECTESLA ES-PC300

Contents

ELECTESLA ES-PC300 LCD Pressure Transmitter User Manual

Model: ES-PC300 | Brand: ELECTESLA

1. Introduction

The ELECTESLA ES-PC300 LCD Pressure Transmitter is designed for precise measurement of pressure in various media including water, gas, oil, and other liquids. It utilizes an imported diffused silicon sensor, high-performance circuit processor, and temperature compensation technology to convert pressure into standard electrical signals (4-20mA, 0-5V, 0-10V, RS485). These signals can then be used by secondary instruments for measurement, indication, and process regulation. The device features an LCD display for clear readings and unit adjustment.

ELECTESLA ES-PC300 LCD Pressure Transmitter overview
Figure 1: ELECTESLA ES-PC300 LCD Pressure Transmitter with key specifications.

2. Safety Precautions

  • This product uses a diffused silicon oil-filled core. Improper use may cause explosion accidents. Measuring oxygen is strictly prohibited.
  • Do not measure media that are incompatible with the material in contact with the transmitter.
  • No modifications or changes are allowed on the device.
  • When measuring steam or other high-temperature media, ensure the medium temperature does not exceed the transmitter's operating temperature. Install a cooling device if necessary.
  • This product is weak current equipment; wire it separately from strong current cables.
  • Ensure the power supply voltage meets the transmitter's requirements and the pressure source's high pressure is within the transmitter's range.
  • During pressure measurement, increase and release pressure slowly to avoid sudden high pressure or product damage.
  • This product is not explosion-proof. Do not use it in explosion-proof areas.

3. Specifications

The ES-PC300 pressure transmitter is designed with robust features for reliable performance.

Exploded view of pressure transmitter components
Figure 2: Internal components and features of the pressure transmitter.

3.1. Product Features

  • Sensor: German imported diffused silicon sensor for fast response, accurate, and stable measurement.
  • Design: Adopts diaphragm isolation technology, integrated core, and wide voltage power supply.
  • Accuracy: 0.5% F.S. (default), 0.2% F.S.
  • Output: 4-20mA, 0-5V, 0-10V, RS485.
  • Power Supply: 12V/24VDC, 12V-24VDC, 5VDC (customizable).
  • Thread Size: M20*1.5, G1/4, G1/2, G3/8, NPT1/2, NPT1/4 (customizable).
  • Overload Capacity: 150% F.S.
  • Response Time: ≤10ms.
  • Protection Grade: IP65 (Waterproof and dustproof).
  • Stability: ±0.2% F.S./year.
  • Temperature Drift: 0.05% F.S./°C (-10°C to 65°C).
  • Housing: 304 stainless steel (customizable SS316).
  • Connector: Hirschmann Connectors.
  • Display: LCD display with units in Kpa, Mpa, bar, mbar, psi, $KG/Fcm^2$, mm, m.
  • Anti-interference: Digital circuit board with strong anti-interference capability.
  • Pressure Shock Resistance: 3mm pilot hole for 1.5 times instant pressure shock resistance.

3.2. Technical Parameters Table

Product Technical Parameters Table
Figure 3: Detailed technical parameters of the pressure transmitter.

3.3. Pressure Range and Thread Size Selection

Pressure Range and Thread Size Tables
Figure 4: Tables for selecting pressure range and thread size.

4. Installation

4.1. General Installation Guidelines

  • Installation requires no pressure and no power supply.
  • Install in locations where temperature gradient changes are minimal.
  • For gases, if the measuring range is less than 0.03Mpa, mount vertically to ensure accuracy. Other media can be mounted at any angle.
  • In hydraulic systems, mount the pressure connections upwards for easy gas venting.
  • If the transmitter is installed with the pressure connection facing upwards or sideways, ensure no liquids flow outside the transmitter housing to prevent moisture and dirt from blocking atmospheric outlets near the electrical connection.
  • Keep threaded connection edges free of dust and dirt.
  • It is recommended to install a pressure shut-off valve in front of the measurement point for easy access and removal. This also helps reduce pressure shock during initial installation.
  • For harsh environments (e.g., lightning strikes, over-voltage), protect the transmitter with lightning and over-voltage protection at the distribution box or between the power supply and the transmitter.
  • Select appropriate sealing fittings and sealing positions based on the measuring medium and input pressure. Ensure thread length is greater than the bottom thread depth for an effective end gasket seal, and less than the base thread depth for an effective root gasket seal.

4.2. Product Dimensions

Product size diagram
Figure 5: Product dimensions. Note: For other requests, contact customer service.

5. Wiring Diagrams

Proper wiring is crucial for the correct operation of the pressure transmitter. Refer to the diagrams below for electrical connections.

Detailed wiring diagram for 2-wire current output and 3-wire voltage output
Figure 6: Electrical connection diagrams for 2-wire current output and 3-wire voltage output.

5.1. No Display Version Wiring (Current Output)

Wiring diagram for no display version with 4-20mA output
Figure 7: Wiring diagram for the ES-PC300 Pressure Transmitter without display, showing 4-20mA current output.

6. Operating Instructions

The ES-PC300 features an LCD display for easy monitoring and configuration. The display supports multiple units and can be freely adjusted.

6.1. LCD Display Features

  • Two-wire passive working mode, 4-bit LCD display.
  • Supports zero clearing, small signal removal, filter settings, and other functions for ease of use.
  • Pressure, temperature, liquid level, and 12 other types of units can be selected.
  • Equipped with over-current, over-voltage, short-circuit, and reverse connection protection. The 24VDC power supply will not be damaged even if the load is short-circuited.
  • Strengthened anti-interference design for harsh industrial environments with electromagnetic interference.
Digital Circuit Board with anti-interference capability
Figure 8: Digital Circuit Board with strong anti-interference capability.

6.2. Working Principle

The pressure sensor operates on the principle of a Wheatstone bridge diffused on a single crystal silicon wafer. The measured medium (gas or liquid) applies pressure, causing a change in the resistance of the bridge wall (piezoresistive effect). This generates a differential voltage signal, which is then converted into a standard analog or digital signal corresponding to the pressure range via a dedicated amplifier.

7. Maintenance

  • Before cleaning, ensure the pressure transmitter is relieved of pressure and disconnected from power.
  • Do not touch the diaphragm with objects to avoid damage.
  • Wipe the surface of the transmitter with a soft, damp cloth.
  • The product is enclosed with a silicone gel process inside, making it waterproof and moisture-proof to ensure a long service life of components.
Silicone gel infusion process for waterproofing
Figure 9: Standardized production process with silicone gel infusion.

8. Troubleshooting

  • Sudden Pressure Shock: Cavitation, liquid hammer, and peak pressure can occur in filling systems, especially when quickly closing valves or starting/closing pumps. The sensor is equipped with a pulse buffer to mitigate these issues. Note that after installing a pulse buffer, the frequency of the dielectric pulse may decrease below 30Hz.
  • Inaccurate Readings: Ensure proper vertical mounting for gas measurements below 0.03Mpa. Verify that the measuring medium is compatible with the transmitter's contact material.
  • No WiFi Connectivity: This pressure transmitter does not support WiFi connectivity. It provides reliable signal output via 4-20mA, 0-5V, 0-10V, or RS485.
  • Power/Signal Issues: Check all wiring connections against the provided diagrams. Ensure the power supply voltage is correct and stable.
Diagram of 3mm pilot hole for pressure shock resistance
Figure 10: International Damping Design with 3mm pilot hole for instant pressure shock resistance.

9. User Tips

  • Users have reported this product to be very effective and reliable for its intended purpose.
  • The ES-PC300 with LCD display is confirmed to support specific configurations like 0-250 mbar range with 4-20mA output, making it versatile for various applications.

10. Warranty and Support

For any questions, technical guidance, or support, please contact the manufacturer directly. Technical support is provided for the life of the product.

Contact Information:

  • WhatsApp: +8619356149608
  • Email: whitney@jinqu-electric.com

A PDF version of the user manual is available for download: User Manual (PDF)

11. Product Video Demonstration

Watch this video for a visual overview and demonstration of the pressure transmitter.

Play video
Watch video online
Video: Product demonstration of the Herssman Pressure Transmitter with LCD Display.
Documents and downloads

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.