1. Introduction
The JXBS-3001-SCY-PT-NPK is a professional handheld quick test platform designed for comprehensive soil analysis. This all-in-one sensor system, developed by Jingxuntong, allows for simultaneous monitoring of multiple soil parameters including pH, NPK (Nitrogen, Phosphorus, Potassium), Electrical Conductivity (EC), Temperature, and Moisture. Utilizing the international new generation TDR measurement method, it offers high detection accuracy within 3%. The robust special metal probe is engineered for extended service life, making it an ideal tool for agricultural and environmental monitoring applications.

Figure 1: JXBS-3001-SCY-PT-NPK Handheld Soil Analyzer with Sensor
2. Product Overview
2.1 Components
The JXBS-3001-SCY-PT-NPK system includes the following main components:
- Handheld Rapid Test Platform: The main unit for display, control, and data processing.
- Soil Sensor: The probe unit with five pins for inserting into the soil to collect data.
- Charging Power Supply: For recharging the handheld unit's internal battery.
- Suitcase: A protective case for storage and transport of the device.

Figure 2: Main components of the JXBS-3001-SCY-PT-NPK system
2.2 Handheld Terminal Features
The handheld terminal is equipped with a clear display, a membrane keyboard for input, and a D-type data interface connector (DB9) which also serves as the charging port.

Figure 3: Handheld Terminal and Sensor with labeled parts
2.3 Key Capabilities
- Multi-parameter Measurement: Measures pH, NPK, EC, Temperature, and Humidity.
- RFID Identification: For efficient data management.
- Data Local Viewing: Review measurements directly on the device.
- 4G Connected: Enables cloud platform viewing for remote data access and analysis.

Figure 4: Handheld display with key features
3. Specifications
| Parameter | Value |
|---|---|
| Model No. | JXBS-3001-SCY-PT |
| DIY Supplies | Electrical |
| Origin | Mainland China |
| Certification | CE |
| PH Measurement Range | 3-9ph |
| Output Signal | RS485 output (Modbus Protocol) |
| Response Speed | ≤ 15s |
| DC Power Supply | 12-24 VDC |
| Measurement Accuracy | ± 0.3ph |
| Long Term Stability | ≤ 5% year |
| Power Consumption | ≤ 0.15w |
| Material | ABS |
| Handheld Terminal Dimension | 125mm (H) x 65mm (W) x 45mm (D) |
| Soil Sensor Dimension | 72mm (H) x 45mm (W) x 15mm (Thickness), 70mm (Probe Length) |

Figure 5: Dimensions of Handheld Terminal and Soil Sensor
4. Setup
- Unpack: Carefully remove all components from the suitcase. Verify that the handheld terminal, soil sensor, and charging power supply are present.
- Charge the Device: Connect the charging power supply to the D-type data interface connector (DB9) on the handheld terminal. Allow the device to fully charge before first use. The device is battery-powered.
- Connect Sensor: Plug the soil sensor cable into the D-type data interface connector on the handheld terminal. Ensure a secure connection.

Figure 6: Sensor connection and battery power indication
5. Operating Instructions
- Power On: Press the "Switch" button on the handheld terminal to power on the device. The display will illuminate.
- Menu Navigation: Use the arrow keys and "Menu" button on the membrane keyboard to navigate through the device's functions and settings.
- Prepare for Measurement:
- Ensure the soil is moist enough for accurate readings. If the soil is too dry, readings may be inaccurate.
- Clear any debris or large particles from the measurement area.
- Insert Sensor: Gently insert the five-pin soil sensor into the soil at the desired depth. Ensure all five pins are fully submerged and make good contact with the soil. Avoid forcing the sensor, which could damage the probes.
- Read Data: The device will display real-time measurement results for pH, NPK, EC, Temperature, and Moisture within a few seconds.
- Data Recording:
- Press "Start recording" to begin logging data.
- Press "Stop recording" to end the current recording session.
- Use the "Storage" button to save recorded data.
- Data Viewing:
- Local Viewing: Access stored data directly on the handheld terminal.
- Cloud Platform Viewing: If the device is 4G connected, data can be uploaded and viewed via a cloud platform for remote analysis. Refer to the cloud platform's specific instructions for setup and access.
- Power Off: After use, press and hold the "Switch" button to power off the device.

Figure 7: Soil sensor in use for measurement
6. Maintenance
- Cleaning the Sensor: After each use, gently clean the metal probes of the soil sensor with a soft cloth and clean water to remove any soil residue. Ensure the probes are dry before storage.
- Storage: Store the handheld terminal and sensor in the provided suitcase in a dry, cool place away from direct sunlight and extreme temperatures.
- Battery Care: Recharge the device regularly, even if not in frequent use, to maintain battery health. Avoid fully discharging the battery for extended periods.
- Probe Protection: The special metal probes are designed for durability, but avoid bending or applying excessive force to prevent damage.
7. Troubleshooting
- Q: Device does not power on.
- A: Ensure the device is fully charged. Connect it to the charging power supply and allow sufficient time for charging.
- Q: Inaccurate or inconsistent readings.
- A:
- Ensure the sensor probes are clean and free of debris.
- Verify that all five pins of the sensor are fully inserted into the soil and making good contact.
- Check soil moisture levels; very dry soil can lead to inaccurate readings. Moisten the soil if necessary.
- Ensure the sensor cable is securely connected to the handheld terminal.
- Q: Data not saving or uploading.
- A:
- Check the device's internal storage capacity.
- For cloud platform upload, ensure the device has a stable 4G connection.
- Refer to the cloud platform's specific instructions for troubleshooting connectivity issues.
8. User Tips
- For best results, take multiple readings in different spots within your test area and average them.
- Always clean the sensor probes immediately after use to prevent corrosion and ensure longevity.
- Keep the device's firmware updated if updates become available (check manufacturer's website).
- When inserting the sensor, avoid rocky or overly compacted soil to prevent damage to the probes.
9. Warranty and Support
The JXBS-3001-SCY-PT-NPK Soil Analyzer comes with a one-year warranty from the date on the Bill of Lading (B/L). In case of quality problems, the manufacturer promises to return goods or funds.
For technical support, warranty claims, or any inquiries, please contact:
- Email: 2193219721@qq.com
- WhatsApp: +86 18175132331
OEM services, including custom packaging and LOGO printing, are available upon request.