1. Introduction
The PEAKMETER PM2128 is a multifunctional digital clamp meter designed for high-precision measurements in various electrical applications. It features True RMS measurement, AC/DC current up to 1000A, AC/DC voltage up to 1000V, VFD frequency measurement, and a 6000-count digital display with backlight. This instrument is an indispensable assistant for your measurement tasks, offering fast response and accuracy.

The meter features a double-layer injection molded, double-coated handle for a comfortable and non-slip grip, ensuring safe and easy handling during operation. Its high-speed encapsulated chip provides fast response and accurate readings for up to 6000 counts.


2. Safety Information
WARNING: Pay special attention to the use of this instrument, as improper use may cause electric shock or damage to the instrument. During use, observe usual safety regulations and safety measures regulated in this manual.
To make full use of instrument functions and guarantee safe operation, please carefully read and observe the use methods in this manual.
This instrument complies with safety requirements for electronic measuring instruments: EN-61010-1, EN-61010-2-030, and EN-61010-2-032, level II pollution, and over-voltage standard CAT III 1000V, CAT IV 600V.
2.1 Preparation
- When using this instrument, users must observe standard safety rules:
- General electric shock prevention.
- Prevention of misuse of instrument.
- After receiving the instrument, check whether it is damaged during transportation.
- After storing and shipping under adverse conditions, check whether the instrument has been damaged.
- Test leads must be in good condition. Before use, check whether the insulation of the leads is damaged, and whether metal wire is exposed.
2.2 Safety Symbols
| Symbol | Description |
|---|---|
| ⚠ | Note (refer to use manual for important safety information) |
| ⚡ | Able to be used on dangerous electrified conductors. |
| □ | Dual-insulation protection (Category II) |
| CAT III CAT IV | Over-voltage according to IEC-61010-1 standard (installation), level III, IV pollution degree 2 refers to protection level of pulse withstand voltage provided. |
| CE | Comply with EU standard. |
| ⏚ | Grounded |
3. Product Description
3.1 Part Names

- Current clamp: Used in current measurement
- Rotary switch
- Function key
- Liquid crystal display (LCD)
- Input socket
- Function key
- Trigger
- Non-contact voltage detecting & inducing area
3.2 Rotary Switch Description
| Symbol | Function |
|---|---|
| A≈ | AC & DC current measurement |
| TEMP | Temperature measurement |
| Ω | Capacitor, on-off and resistance measurement |
| VFDV | AC/DC voltage measurement |
| LoZ V | Low-impedance voltage measurement |
| µA | DC µA current measurement |
| OFF | Instrument OFF |
3.3 LCD Display

| Indicator | Description |
|---|---|
| AC & DC | AC & DC measurement mode |
| •)) | Connection/Disconnection indication |
| AUTO | Automatic range mode |
| ⏱ | Automatic shutdown indication |
| ⏸ | Low battery |
| H | Reading hold state |
| V,A, µA | Volt (voltage), ampere (current) |
| Ω, KΩ, MΩ | Ohm, kilohm and megohm (resistance) |
| Hz, kHz | Hertz, kilohertz |
| VFD | Variable frequency measurement |
| µF | Microfarad |
| °C°F | Centigrade and Fahrenheit |
| LowZ | Low-impedance measurement mode |
| % | Duty ratio measurement |
| REL | Relative measurement |
4. Specifications
4.1 Overview
- Select measurement function and range automatically.
- Overload protection throughout the range.
- Max. voltage between measurement terminal and ground: 1000V DC or 1000V AC.
- Operating height: Max. 2000m.
- Display: LCD.
- Max. display value: 6000 digit.
- Polarity indication: Automatic indication, and '-' indicates negative.
- Over range display: 'OL' or '-OL'.
- Sampling time: About 3 times/second, bargraph 10 times/second.
- Unit display: Function and electricity quantity unit display.
- Automatic shutdown time: 10 minutes.
- Power supply: 1.5V AAA battery × 3.
- Battery under-voltage indication: LCD display symbol ⏸.
- Temperature coefficient: Less than 0.1×accuracy/°C.
- Operating temperature: 18°C ~ 28°C.
- Storage temperature: -10°C ~ 50°C.
4.2 Technical Indexes
The instrument specifies one year as a cycle, and shall be re-calibrated under 18°C ~ 28°C, with relative humidity less than 75%.
AC/DC Clamp Meter Specifications


Detailed Technical Indexes
| Measurement | Range | Resolution | Accuracy |
|---|---|---|---|
| AC Current | 60A | 0.01A | ± (2.5% reading + 8 digits) |
| 600A | 0.1A | ||
| 1000A | 1A | ||
| Min. input current: 0.1A AC. Max. input current: 1000A AC. Frequency range: 45 ~ 65Hz. | |||
| DC Current | 60A | 0.01A | ± (3% reading + 10 digits) |
| 600A | 0.1A | ||
| 1000A | 1A | ||
| Min. input current: 0.01A DC. Max. input current: 1000A DC. | |||
| DC Voltage | 6V | 0.001V | ± (0.5% reading + 5 digits) |
| 60V | 0.01V | ||
| 600V | 0.1V | ||
| 1000V | 1V | ||
| Min. input voltage: 0.001V DC. Max. input voltage: 1000V DC. | |||
| AC Voltage | 6V | 0.001V | ± (0.8% reading + 5 digits) |
| 60V | 0.01V | ||
| 600V | 0.1V | ± (1.0% reading + 5 digits) | |
| 1000V | 1V | ||
| Min. input voltage: 0.001V AC. Max. input voltage: 1000V AC (valid value). Frequency range: 45 ~ 1000Hz. | |||
| Frequency (Clamp) | 60.0Hz | 0.1Hz | ± (1.0% reading + 5 digits) |
| 1000Hz | 1Hz | ||
| Measurement range: 40Hz ~ 1000Hz. Input signal range: ≥ 2A AC current (valid value). | |||
| Frequency (Voltage) | 60.0Hz | 0.1Hz | ± (1.0% reading + 5 digits) |
| 1000Hz | 1Hz | ||
| Measurement range: 40Hz ~ 1000Hz. Input signal range: ≥ 0.8V AC voltage (valid value). | |||
| Resistance | 6KΩ | 0.001KΩ | ± (0.8% reading + 3 digits) |
| 60KΩ | 0.01KΩ | ||
| 600KΩ | 0.1KΩ | ||
| 6MΩ | 0.001MΩ | ||
| 60MΩ | 0.01MΩ | ||
| Overload protection: 250V DC or AC (valid value). | |||
| Line On-off Test | •)) | 1Ω | If resistance is less than 30Ω, buzzer will sound continuously. |
| Overload protection: 250V DC or AC (valid value). | |||
| Capacitance | 60.00nF | 0.01nF | ± (4% reading + 3 digits) |
| 600.0nF | 0.1nF | ||
| 6.000µF | 1nF | ||
| 60.00µF | 10nF | ||
| 600.0µF | 100nF | ||
| 6.000mF | 1µF | ||
| 60.00mF | 10µF | ||
| Overload protection: 250V DC or AC (valid value). | |||
| Temperature | -20~1000°C | 1°C/2°F | ± (1% reading + 2 digits) |
| -4~1832°F | |||
| Overload protection: 250V DC or AC (valid value). | |||
| µA Current | 200.0µA | 0.1µA | ± (0.8% reading + 3 digits) |
| Overload protection: 250V DC or AC (valid value). | |||
4.3 Dimensions

5. Setup
5.1 Battery Installation
WARNING: Before opening the battery cover, remove the test leads from the circuit being measured to avoid electric shock.
- When the low battery symbol (⏸) appears on the instrument, it indicates that the battery needs to be changed.
- Unfasten the bolts on the battery cover and remove the cover.
- Replace the old batteries with 3 new 1.5V AAA batteries.
- Place the battery cover back and fasten the bolts.

5.2 Test Lead Connection
For voltage, resistance, capacitance, and µA current measurements, connect the test leads to the input sockets. Ensure the red lead is connected to the positive terminal and the black lead to the common (COM) terminal. For current clamp measurements, use the clamp jaws.
6. Operating Instructions
6.1 Reading Hold (H)
During measurement, if you need to hold the displayed reading, press the 'H' button. The value on the display will be locked. Press the 'H' button again to cancel the reading hold.
6.2 Backlight/Light
If the environment is too dark to read the display, press and hold the backlight button (often integrated with 'H' or a dedicated light button) for more than 2 seconds to turn on the backlight. It will automatically turn off after approximately 1 minute. To manually turn off the backlight, press and hold the button again for more than 2 seconds.

6.3 Automatic Shutdown
If there is no operation within 10 minutes after starting the instrument, it will enter standby state and shut down to save energy. To wake the instrument from automatic shutdown, press any key.
6.4 AC & DC Current Measurement
Turn the rotary switch to the AC current range (A≈). Hold the trigger to open the clamp jaws, and clip one lead of the line being measured within the clamp. The instrument will display the measured current. Press the FUNC key to select between AC or DC current measurement. Press the RAN key to select the range manually. During AC current measurement, press the Hz% key to shift the display between frequency and duty ratio of the measured current. Long press the FUNC key to start the VFD measurement function.

6.5 AC & DC Voltage Measurement
Turn the rotary switch to the AC & DC voltage range (VFDV). Press the FUNC key to shift between AC and DC voltage measurement modes. Connect the test leads to the signal being measured, with the red lead connected to the positive and the black lead to the negative. During AC voltage measurement, press the Hz% key to shift the display between frequency and duty ratio of the measured voltage. Long press the FUNC key to start the VFD measurement function.


6.6 Low-Impedance Measurement (LoZ V)
Turn the rotary switch to LoZ. Press the FUNC key to shift between AC and DC voltage measurement modes. Connect the test leads to the signal being measured. Note: In low-impedance measurement mode, the longest measurement time shall not be larger than 1 minute.
6.7 On-off/Resistance/Capacitor Measurement (Ω)
Turn the rotary switch to the resistance and capacitor range (Ω). Press the FUNC key to shift between on-off, resistance, and capacitor functions. When test leads are connected to the resistance being measured, the instrument will display the measured resistance. If the measured value is less than 30Ω, the instrument buzzer will alarm.



6.8 Temperature Measurement (TEMP)
Turn the rotary switch to TEMP. Insert the thermocouple probe into the input sockets, with the positive of the probe connected to the red input terminal. The primary display will show the temperature in Centigrade. Press the FUNC key to display the temperature in Fahrenheit.
6.9 µA Current Measurement (µA)
Turn the rotary switch to µA. Connect the test leads in series with the signal being measured. The primary display will show the measured current.
6.10 NCV (Non-contact Voltage Detection)
Turn the rotary switch to NCV. The instrument will display "EF". Bring the NCV detector close to the lead being measured. The instrument can detect if the lead being measured has >90V AC voltage. When the instrument detects AC voltage, the buzzer will alarm and the NCV alarm light will flash.
Note: Even if there is no alarm indication, voltage may still exist. Do not solely depend on the non-contact voltage detector to judge whether there is voltage in the lead. Detection operation may be impacted by factors like different socket designs and insulation thickness types etc.
7. Maintenance
7.1 General Maintenance
- Do not try to open the bottom case to adjust or repair the instrument. Such operations should only be conducted by technicians who fully understand the instrument and the dangers of electric shock.
- Before opening the instrument bottom case or battery cover, remove the test leads from the wire being measured.
- Use a wet cloth and gentle detergent to clean the instrument. Do not use any abrasive or solvent.
- Power off the instrument when not in use, and rotate the range switch to the OFF position.
- If the instrument is not used for a long time, take out the battery to avoid any damage to the instrument from battery leakage.
7.2 Test Lead Replacement
WARNING: When changing the test leads, it is required to replace them with identical leads or leads of the same level. The leads must be in good condition, and the level of the leads is: 1000V 10A.
If the insulation layer of the test lead is damaged, or the metal wire of the lead is exposed, it is required to change the lead immediately.
8. Troubleshooting
This section provides general troubleshooting tips for common issues. For more complex problems, consult a qualified technician.
- No Display / Instrument Not Turning On:
- Check if the rotary switch is in the OFF position.
- Verify battery installation and ensure batteries are not depleted. Replace if necessary (refer to Section 5.1).
- If the instrument was in auto-shutdown, press any key to wake it up.
- Inaccurate Readings:
- Ensure test leads are properly connected and not damaged.
- Verify the correct measurement function and range are selected.
- Check for external interference or strong magnetic fields.
- The instrument may require re-calibration if readings are consistently off (refer to Section 4.2).
- Buzzer Not Sounding for Continuity:
- Ensure the rotary switch is set to the continuity/resistance mode.
- Confirm the resistance being measured is below 30Ω.
- NCV Not Detecting Voltage:
- Ensure the rotary switch is set to NCV.
- Remember that detection can be affected by socket design and insulation thickness. Do not rely solely on NCV for safety verification.
9. User Tips
- Durability: Users have noted the robust construction of PEAKMETER products. Handle with care, but expect good durability for regular use.
- Accessories: The product typically comes with test leads, a temperature probe, a carrying case, and a user manual. Ensure all components are present upon receipt.
- Battery Life: To maximize battery life, always turn the rotary switch to the OFF position when the meter is not in use. For long-term storage, remove the batteries.
10. Accessories
The following accessories are included with your PEAKMETER PM2128 Digital Clamp Meter:
- Test Leads (Level: 1000V 10A) - 1 set
- User Manual (PDF available online: View PDF) - 1 copy
- 1.5V AAA Battery - 3 pieces (for initial use, may not be included in all packages due to shipping restrictions)
- Cloth Bag / Carrying Case - 1 piece
- K-Type Thermocouple Probe - 1 piece

11. Warranty and Support
For warranty information and technical support, please refer to the official PEAKMETER website or contact your local distributor. Keep your purchase receipt as proof of purchase for any warranty claims.