1. Introduction

The BSIDE O9 is a versatile, high-performance 3-in-1 handheld device that integrates the functions of a dual-channel oscilloscope, a digital multimeter, and a function signal generator. It features an advanced FPGA+MCU+ADC hardware architecture, ensuring reliable and accurate measurements.

Equipped with a 3.98-inch 480*320px resolution full-view high-definition TFT display, the O9 provides clear and detailed readings. Its robust design includes a built-in high-voltage protection module, supporting peak voltage measurements up to ±400V, and intelligent burn prevention for enhanced safety.

The device is powered by a built-in 5000mAh rechargeable lithium battery, offering approximately 10 hours of continuous operation in oscilloscope mode and up to 24 hours in multimeter mode, ensuring long-lasting performance for various tasks.

The BSIDE O9 is designed for both electrical professionals and DIY enthusiasts, providing a comprehensive solution for testing, troubleshooting, and signal generation needs.

2. Package Contents

Please check the package carefully to ensure all items are present and undamaged:

  • 1 x BSIDE O9 Handheld Oscilloscope Multimeter
  • 1 x EVA Carrying Case
  • 2 x Oscilloscope Probes
  • 1 x Oscilloscope Accessory Kit
  • 1 x Multimeter Test Leads (pen)
  • 1 x Crocodile Clip
  • 1 x Packaging Box
  • 1 x USB Charging Cable
  • 1 x Instruction Manual
  • 1 x Warranty Card
  • 1 x K-Type Thermocouple
BSIDE O9 with all included accessories
Figure 2.1: BSIDE O9 device and all included accessories.

3. Product Overview

The BSIDE O9 features a user-friendly interface and robust construction. Key components include:

  • Display Screen: A 3.98-inch TFT full-view color screen (480*320 pixels) for clear data visualization.
  • Input Ports: Dedicated CH1 and CH2 inputs for oscilloscope functions, and multimeter measurement input ports.
  • Output Port: A signal generator output port (OSC Port).
  • USB Interface: For charging and data export.
  • Button Board: Intuitive controls including F1-F4 function keys, CH1/2 switch, MENU, x1/x10 attenuation, AC/DC mode, AUTO PEAK, SAVE, and power button.
  • Protective Sheath: Full body silicone sheath provides anti-burn and anti-fall protection.
Main interface description for Oscilloscope and Multimeter
Figure 3.1: Main interface description for Oscilloscope and Multimeter modes.
Internal view of dual channel signal source and physical features
Figure 3.2: Detailed view of the device's physical features, including dual channel inputs.

4. Initial Setup and Charging

4.1 Charging the Device

The BSIDE O9 comes with a built-in 5000mAh lithium battery. Before first use, or when the low battery alarm indicates, charge the device using the provided USB charging cable.

  1. Connect the small end of the USB charging cable to the USB interface on the BSIDE O9.
  2. Connect the larger end of the USB cable to a standard USB power adapter (not included) or a computer's USB port.
  3. The charging indicator on the device will show the charging status. Ensure the device is fully charged for optimal performance.
BSIDE O9 charging via USB
Figure 4.1: Charging the BSIDE O9 via USB cable.

4.2 Powering On/Off

To power on the device, press and hold the power button until the screen illuminates. To power off, press and hold the power button again until the device shuts down.

5. Operating Modes

The BSIDE O9 offers three primary operating modes: Oscilloscope, Multimeter, and Signal Generator. Use the 'MODE' button to switch between these functions.

Dual mode display, switching between oscilloscope and multimeter
Figure 5.1: The device supports dual mode display, allowing easy switching between oscilloscope and multimeter functions.

5.1 Oscilloscope Function

The oscilloscope mode provides dual-channel waveform display with a 50MHz bandwidth and a 250MSa/s real-time sampling rate. It features an automatic range waveform function for ease of use.

  1. Connect the oscilloscope probes to the CH1 and/or CH2 input ports.
  2. Connect the probe tips to the circuit or signal source you wish to measure.
  3. Press the 'MODE' button to select the oscilloscope interface.
  4. Use the 'CH1/2' button to switch between channels or view both simultaneously.
  5. Utilize the 'MENU' button to access settings for trigger, time base, vertical sensitivity, and other parameters.
  6. The 'AUTO PEAK' button can automatically adjust the display for optimal waveform viewing.
  7. Use the navigation buttons (up, down, left, right) to adjust waveform position and cursor measurements.
Dual channel display of the oscilloscope, showing waveforms
Figure 5.2: Dual channel oscilloscope display showing signal waveforms.

5.2 Multimeter Function

The digital multimeter mode offers 20000 counts with True RMS measurement capabilities. It supports a wide range of electrical measurements:

  • AC/DC Voltage
  • AC/DC Current (including flexible current clamp support for up to 2000A)
  • Resistance
  • Capacitance
  • Temperature (using K-Type Thermocouple)
  • Diode Test
  • Continuity Test
  1. Connect the multimeter test leads to the appropriate input ports (e.g., VΩmA for voltage/resistance/current, COM for common ground).
  2. Press the 'MODE' button to select the multimeter interface.
  3. Select the desired measurement function (e.g., AC V, DC V, Ω, A) using the function buttons or 'MENU' if applicable.
  4. Place the test leads across the component or circuit to take a reading.
Measuring AC voltage with the BSIDE O9 multimeter function
Figure 5.3: Measuring AC voltage using the multimeter function.

5.3 Signal Generator Function

The built-in DDS function signal generator can output 7 types of waveforms with adjustable frequency, amplitude, and duty cycle. The maximum output frequency for all signals is 2MHz, with a step size of 1Hz.

  • Available Waveforms: Sine wave, Half wave, Square wave, Direct current, Full wave, Triangle wave, Sawtooth wave.
  1. Press the 'MODE' button to navigate to the signal generator function (if available as a distinct mode, otherwise it might be integrated into oscilloscope output).
  2. Connect a suitable cable from the 'OSC Port' output to the device or circuit requiring the signal.
  3. Use the function buttons or menu options to select the desired waveform type.
  4. Adjust the output frequency, amplitude, and duty cycle as needed.
Function Introduction of Signal Generator, showing 7 waveform types
Figure 5.4: Overview of the 7 waveform types available from the signal generator.

6. Data Management

6.1 Waveform Saving

The BSIDE O9 allows for one-click waveform screenshot saving. The device supports storing up to 36 records.

  1. While in oscilloscope mode, press the 'SAVE' button to capture the current waveform display.
  2. The saved image can be viewed directly on the oscilloscope.
Interface saving view with storage
Figure 6.1: Waveform saving interface and storage capability.

6.2 Data Export

Saved waveforms and other data can be exported to a computer for further analysis.

  1. Connect the BSIDE O9 to a computer using the provided USB charging cable.
  2. The device should appear as a removable storage device on your computer.
  3. Access the stored data files (e.g., waveform screenshots) and copy them to your computer.
Waveform saving and viewing on a laptop, connected to the BSIDE O9
Figure 6.2: Exporting saved waveforms to a computer for secondary analysis.

7. Specifications

7.1 General Specifications

  • Model Number: O9
  • Brand Name: BSIDE
  • DIY Supplies: Electrical
  • Certification: CE, FCC, RoHS
  • Origin: Mainland China
  • Display Size: 3.0 - 4.9 Inches (3.98-inch TFT full-view color screen)
  • Display Resolution: 480*320 pixels
  • Power: 5000mAh (Built-in Lithium Battery)
  • Size: 216.6*84.5*36.0mm
  • Weight: 384g
  • Battery Included: Yes
  • IS Battery Demountable: No

7.2 Oscilloscope Parameters

Table 7.1: Oscilloscope Functional Parameters
CharacteristicExplanation
BandwidthDual channel real-time sampling, 50MHz
Sampling MethodReal-time sampling
Real Time Sampling Rate250MSa/s
Number of Channels2 (Dual channel DC and AC)
Input CouplingDC, AC
Input Impedance1MΩ, @16pF
Probe AttenuationX1, X10
Maximum Input VoltageX1 range <150V, X10 range <300V (DC+AC Peak)
Sampling Rate Interpolation(sinx)x
Scanning Speed Range100ns/div - 20s/div
Time Base Accuracy20ppm
Record LengthHighest 128KByte
Vertical Sensitivity20mV/div - 10V/div
Displacement Range4 grids (positive and negative)
Low Frequency ResponseGreater than 10Hz
Rise TimeLess than 10ns
DC Gain Accuracy±3%
Automatic MeasurementPeriod, Frequency, Peak to Peak value, Maximum value, Minimum value, Duty cycle, Root Mean Square
Trigger MethodAutomatic, Normal, Single Line
Trigger EdgeRising edge, Falling edge
Display ModeYT, Xy, Rolling
Afterglow TimeMin, 500ms, 1S, 10S, Infinite
Waveform Output TypeSine wave, Square wave, Triangular wave, Half wave, Full wave, Sawtooth wave, DC
Frequency Range0Hz-2MHz
Output Channel1
Output Amplitude0.1V-3.0V
Minimum Resolution1Hz
Square Wave Duty Cycle1%-99%
Table of Oscilloscope functional parameters
Figure 7.1: Detailed Oscilloscope Functional Parameters.

7.3 Multimeter Parameters

Table 7.2: Multimeter Function Parameters
FunctionRangeResolving PowerAccuracy
DC Voltage (V)1000V0.001V±(0.05%+3)
DC Voltage (mV)200mV0.01mV±(0.1%+3)
AC Voltage (V)750V0.001V±(0.05%+3)
AC Voltage (mV)200mV0.01mV±(0.1%+3)
DC Current (mA)200mA0.01mA±(0.1%+3)
AC Current (mA)200mA0.01mA±(0.1%+3)
AC Current (A) (Flexible coil)2000A0.1A±(0.1%+3)
Resistance200MΩ0.001MΩ±(0.5%+3)
Capacitance99.999mF0.001mF±(4.0%+10)
Temperature-20~1000°C / -4~1832°F1°C/°F±(2.0%+5)
DiodeYesOn/OffYes
True Effective ValueYesScreen backlightYes
Low Battery WarningYesAutomatic shutdownYes
Battery Type5000mAh Lithium battery
Table of Multimeter function parameters
Figure 7.2: Detailed Multimeter Function Parameters.

8. Safety Information

The BSIDE O9 is designed with user safety in mind, incorporating several protective features:

  • Intelligent Burn Prevention: This feature helps prevent damage to the meter from overload during high-precision measurements, making it safer for both experienced users and beginners.
  • High-Voltage Protection Module: Built-in protection supports maximum peak voltage measurements of ±400V.
  • Silicone Sheath: The full body silicone sheath provides physical protection against daily bumps and falls.

Always adhere to standard electrical safety practices when using the device. Do not attempt to measure voltages or currents beyond the specified limits. Ensure test leads are in good condition before use.

Intelligent burn prevention feature
Figure 8.1: Intelligent burn prevention for enhanced safety.

9. User Tips

  • Flexible Current Clamp: For measuring high AC currents (up to 2000A), the BSIDE O9 is compatible with the optional BSIDE TP01 Flexible Current Clamp. This accessory is sold separately and can significantly extend the current measurement capabilities of your device.
  • Data Review: Regularly export saved waveforms to your computer for detailed analysis and record-keeping. This is especially useful for tracking changes over time or documenting specific events.
  • Battery Life: To maximize battery life, ensure the device is fully charged before extended use. The automatic shutdown feature helps conserve power when the device is idle.

10. Troubleshooting

If you encounter issues with your BSIDE O9, please refer to the following common problems and solutions:

  • Device does not power on: Ensure the battery is sufficiently charged. Connect the device to a power source using the USB cable and allow it to charge for at least 30 minutes before attempting to power on again.
  • Inaccurate readings: Check that the test leads or probes are correctly connected and not damaged. Verify that the correct measurement mode is selected for the task. For oscilloscope measurements, ensure proper probe attenuation (x1/x10) is set.
  • Screen is blank or frozen: Try restarting the device by holding down the power button until it shuts off, then power it back on. If the issue persists, ensure the battery is charged.
  • Cannot save or export data: Check the available storage on the device. Ensure the USB connection to the computer is stable and recognized. Try using a different USB port or cable.
  • Flexible Current Clamp not working: Ensure the clamp is correctly connected to the device and the appropriate current measurement mode is selected. Verify the clamp itself is not damaged.

If these steps do not resolve your issue, please contact customer support.

11. Maintenance

11.1 Cleaning

To maintain the accuracy and longevity of your BSIDE O9:

  • Wipe the device with a soft, damp cloth. Do not use abrasive cleaners or solvents.
  • Ensure all ports are free of dust and debris.
  • Keep the display screen clean and free from scratches.

11.2 Storage

When not in use, store the device in its EVA carrying case in a cool, dry place, away from direct sunlight and extreme temperatures. Remove test leads and probes before storing.

12. Warranty and Support

The BSIDE O9 comes with a 12-month warranty from the date of purchase, covering manufacturing defects. Please refer to the included warranty card for detailed terms and conditions.

For technical support, troubleshooting assistance, or warranty claims, please contact the BSIDE Official Store customer service. Keep your purchase receipt and warranty card handy when contacting support.

BSIDE quality service banner with warranty information
Figure 12.1: BSIDE Quality Service highlights a 12-month warranty.

13. Optional Accessories

To expand the capabilities of your BSIDE O9, consider the optional BSIDE TP01 Flexible Current Clamp. This accessory is designed for measuring high AC currents up to 2000A and is suitable for use with the O9 model.

You can acquire the BSIDE TP01 Flexible Current Clamp from the official seller's store. For more information or to purchase, please visit: https://www.aliexpress.com/item/1005008227777921.html

BSIDE TP01 Flexible Current Clamp accessory
Figure 13.1: BSIDE TP01 Flexible Current Clamp, an optional accessory for high current measurements.

Documents and downloads

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.