Contents

MESTEK IR05A/IR02C Non-contact Infrared Thermometer User Manual

Models: IR05A (High Temperature), IR02C (Standard Temperature)

1. Introduction

This manual provides detailed instructions for the safe and effective operation of your MESTEK IR05A or IR02C Non-contact Infrared Thermometer. These devices are designed for industrial temperature measurement, offering both non-contact infrared and contact K-type thermocouple measurement capabilities. Please read this manual thoroughly before use and keep it for future reference.

MESTEK IR05A Infrared Thermometer with K-type probe, manual, cloth bag, and packaging.

Figure 1: MESTEK IR05A Infrared Thermometer with accessories.

MESTEK IR02C Infrared Thermometer with K-type probe, manual, cloth bag, and packaging.

Figure 2: MESTEK IR02C Infrared Thermometer with accessories.

2. Safety Precautions

  • This device is not intended for measuring human or animal body temperature. Readings for living beings will be inaccurate.
  • Do not measure through transparent surfaces such as glass or plastic. These materials will reflect infrared energy, leading to inaccurate readings.
  • Do not aim the laser directly at eyes or indirectly at eyes through reflective surfaces. Laser radiation can cause eye damage.
  • If the ambient temperature changes significantly or suddenly, allow the thermometer to stabilize for at least 30 minutes before use to ensure accuracy.

3. Product Overview

3.1. Key Components

Internal components of the infrared thermometer, showing high precision laser indication, high precision sensors, aluminum alloy sensing housing, and high speed intelligent chip.

Figure 3: Internal components of the thermometer.

  • High Precision Laser Indication: For accurate targeting of the measurement area.
  • High Precision Sensors: Ensures reliable temperature readings.
  • Aluminum Alloy Sensing Housing: Durable and provides stable performance.
  • High Speed Intelligent Chip: For fast processing and response time.

3.2. Display and Controls

Diagram of the thermometer's display and controls, with numbered indicators for various functions.

Figure 4: Display and control layout.

  1. Emissivity
  2. Measurement indication
  3. Laser indication
  4. Battery undervoltage indication
  5. Data locking (HOLD)
  6. Irradiation measurement temperature value
  7. Temperature unit (°C/°F)
  8. Maximum lock (MAX)
  9. Minimum lock (MIN)
  10. Average (AVG)
  11. Temperature difference value (DIFF)
  12. Alarm
  13. High alarm (Alarm Hi)
  14. Low alarm (Alarm Low)
  15. K-type thermocouple temperature

4. Setup

4.1. Battery Installation

The thermometer requires 2x 1.5V AA batteries (not included in all packages). To install:

  1. Open the battery compartment located on the handle.
  2. Insert two AA batteries, ensuring correct polarity.
  3. Close the battery compartment securely.
Diagram showing how to open the battery compartment and insert AA batteries into the thermometer.

Figure 5: Battery installation steps.

5. Operating Instructions

5.1. Basic Measurement

To take a non-contact temperature reading, simply pull and hold the trigger. The device will display the temperature of the targeted surface. Release the trigger to hold the reading on the display.

Video 1: Demonstrates basic operation and feature switching.

5.2. High and Low Alarm Mode

The thermometer features adjustable high and low temperature alarms.

  • Setting High Alarm: Long press the "SET" button to enter the high alarm setting mode. Use the up/down arrows to adjust the desired high temperature threshold.
  • Setting Low Alarm: Gently press the "SET" button again to switch to the low alarm setting mode. Use the up/down arrows to adjust the desired low temperature threshold.
Images showing the steps to set high and low temperature alarms on the thermometer.

Figure 6: Setting high and low alarms.

5.3. Adjustable Emissivity

Emissivity (ε) is adjustable from 0.10 to 1.00. The default preset emissivity is 0.95. To adjust:

  1. Press the "SET" button until the emissivity value flashes.
  2. Use the up/down arrows to adjust the emissivity value according to the material being measured.
  3. Short press the trigger to exit setup mode.
Image showing the thermometer display with emissivity setting and a table of common material emissivities.

Figure 7: Adjustable emissivity setting and comparison table.

5.4. Temperature Unit Switching (°C/°F)

To switch between Celsius (°C) and Fahrenheit (°F):

  1. Press the "SET" button until the temperature unit flashes.
  2. Use the up/down arrows to select the desired unit.
  3. Short press the trigger to exit setup mode.
Image illustrating the process of switching between Celsius and Fahrenheit temperature units.

Figure 8: Switching temperature units.

5.5. Laser ON/OFF

To turn the laser pointer ON or OFF, short press the laser button (triangle icon) on the control panel.

5.6. MAX/MIN/AVG/DIFF Mode

The thermometer can display maximum, minimum, average, and temperature difference values. Press the "MIN/MAX" button to cycle through these modes.

Image showing the thermometer display cycling through MAX, MIN, AVG, and DIFF modes.

Figure 9: MAX/MIN/AVG/DIFF mode selection.

5.7. K-type Thermocouple Measurement (IR05A and IR02C only)

For contact temperature measurement, insert the K-type thermocouple plug into the dedicated socket on the instrument, observing the positive and negative pole markings. The probe tip should directly contact the object to be measured.

  • Exposure Temperature Range: -40°C~1600°C (-40°F~2912°F)
  • K-type Temperature Range: -10°C~500°C (14°F~932°F)
Image showing the K-type thermocouple connected to the thermometer and measuring water temperature.

Figure 10: K-type thermocouple in use.

5.8. Sleep Mode

The thermometer is designed with an energy-saving feature. It will automatically enter a sleep state after approximately 35 seconds of inactivity.

6. Measurement Methods

6.1. Distance to Spot Ratio (D:S)

The D:S ratio for this thermometer is 50:1. This means that at a distance of 50 units from the target, the measurement spot diameter will be 1 unit. For example, at 50 inches distance, the spot size is 1 inch. Ensure the target is larger than the spot size for accurate readings.

Diagram illustrating the Distance to Spot (D:S) ratio of 50:1, showing how the measurement spot size increases with distance.

Figure 11: Distance to Spot Ratio (D:S).

6.2. Water/Oil Temperature Measurement

When measuring boiling liquids, do not let smoke or steam come into contact with the thermometer's sensor. Maintain a measurement angle of approximately 45° to 60° in the temperature sensing area for optimal results.

Diagram showing correct and incorrect methods for measuring liquid temperatures, emphasizing avoiding steam and maintaining an angle.

Figure 12: Correct method for measuring liquid temperatures.

7. Specifications

FeatureSpecification (IR05A)
ModelIR05A
Temperature Range-40°C~1600°C (-40°F~2912°F)
Accuracy (IR)-40°C~800°C (±1.5%), 800°C~2000°C (±3%)
K-type Range-10°C~500°C (14°F~932°F)
K-type Accuracy±(1.5%+2°C/4°F)
DisplayVA reverse display color screen
Distance to Spot (D:S)50:1
Spectral Response8~14um
Resolving Power0.1°C/0.1°F
Emissivity0.10~1.00 (Adjustable)
Laser<1mW /620-690nm Level 2
Response Time<0.5S
Automatic ShutdownAbout 35 seconds
Usage Environment0~50°C / 10%-90%RH
Storage Temperature-20°C~60°C
Size180x120x60mm
Power Source2x1.5V AA battery
WeightApproximately 265g (excluding battery)
CertificationCE
Diagram showing the dimensions of the thermometer and a table of its technical specifications.

Figure 13: Product dimensions and detailed specifications.

8. User Tips

  • Be aware of the measurement spot size. Due to the 50:1 D:S ratio, the measurement area can be quite large at longer distances. Ensure your target is adequately covered by the spot for accurate readings.

9. Troubleshooting

9.1. Inaccurate Readings

  • Check Emissivity: Ensure the emissivity setting matches the material being measured. Refer to the emissivity table in Section 5.3.
  • Distance to Spot Ratio: Verify that the target object is larger than the measurement spot size at your chosen distance (D:S 50:1).
  • Ambient Temperature Changes: If the device was recently moved to an environment with a significantly different temperature, allow it to stabilize for 30 minutes.
  • Transparent Surfaces: Do not measure through glass or plastic.

9.2. Device Not Turning On / Display Issues

  • Battery Check: Ensure batteries are correctly installed and have sufficient charge. Replace if necessary.
  • Ambient Temperature Sensor: If the ambient temperature reading appears stuck or inaccurate, ensure the device is allowed to acclimate to the environment. If the issue persists, contact support.

10. Maintenance

  • Keep the device clean and free from dust.
  • Use clean, dry air to gently clean the lens and sensor area. Avoid abrasive materials or harsh chemicals.
  • Store the thermometer in a dry, cool place within the specified storage temperature range (-20°C~60°C).
  • Remove batteries if the device will not be used for an extended period to prevent leakage.

11. Warranty and Support

For warranty information, technical support, or service inquiries, please refer to the contact information provided with your purchase or visit the official MESTEK website. Keep your purchase receipt as proof of purchase.