vilnason AK-CS-01

Contents

Vilnason Physiotherapy Therapeutic Ultrasound Device

Model: AK-CS-01 | Brand: Vilnason

1. Introduction

This manual provides comprehensive instructions for the safe and effective use of your new Vilnason Physiotherapy Therapeutic Ultrasound Device. This device is designed to deliver therapeutic ultrasound for body pain relief, muscle soreness, and to support tissue regeneration and wound healing. Please read this manual thoroughly before operation and retain it for future reference.

Therapeutic ultrasound utilizes high-frequency sound waves to produce deep tissue heat, which can help reduce inflammation, increase blood flow, and alleviate pain, stiffness, and muscle spasms. It is a non-invasive method for physical therapy.

2. Safety Information

  • Always use the device with the provided power adapter and ensure it matches your local voltage (100-240V).
  • Do not attempt to disassemble or repair the device. High voltage is present, and personal dismantling is not allowed.
  • Keep the device away from water or excessive moisture.
  • Ensure the device is placed on a stable, flat surface during use.
  • Only use conductive ultrasound gel with the probe to ensure proper transmission of ultrasound waves and prevent skin irritation.
  • Consult a healthcare professional before using this device, especially if you have pre-existing medical conditions, are pregnant, or have implanted electronic devices.
  • Do not use on open wounds, sensitive skin areas, or over areas with acute inflammation without medical advice.

3. Product Overview

3.1 Components

The Vilnason Physiotherapy Therapeutic Ultrasound Device consists of the main unit and an ultrasound probe. The main unit features a large backlight LCD display and touch control buttons.

Vilnason Therapeutic Ultrasound Device main unit with probe

Figure 1: Main unit of the Ultrasound Therapy Device

3.2 Control Panel

The control panel features a large backlight LCD display and touch buttons for setting time and intensity.

Control panel with LCD display and touch buttons for time and intensity adjustment

Figure 2: Control Panel Buttons

  • Time Button: Adjusts the treatment duration.
  • '+' Button: Increases intensity or time.
  • '-' Button: Decreases intensity or time.
  • ON/OFF Button: Powers the device on or off.

3.3 Back Panel Connections

The back panel includes the power switch, power connection port, and the function interface for the ultrasound probe.

Back panel showing power switch, power input, and function interface

Figure 3: Back Panel Connections

4. Setup

  1. Unpack: Carefully remove all components from the packaging. Verify that all items listed in the product list are present (see Section 8.1).
  2. Placement: Place the main unit on a stable, flat, and dry surface.
  3. Connect Probe: Connect the ultrasound probe cable to the 'Function interface' port on the back of the main unit.
  4. Connect Power: Connect the power cable to the power input port on the back of the main unit, then plug the other end into a suitable electrical outlet (100-240V AC, 50/60Hz).
  5. Power On: Flip the 'Power' switch on the back of the unit to the ON position. The LCD display will illuminate.

Video 1: Device setup and basic operation overview.

5. Operating Instructions

  1. Apply Gel: Apply a generous amount of conductive ultrasound gel to the area of skin to be treated. Ensure the entire treatment area is covered.
  2. Select Time: Press the 'Time' button on the control panel to cycle through the available treatment durations: 5, 10, 15, 20, 25, or 30 minutes. The default time is 10 minutes.
  3. Adjust Intensity: Use the '+' and '-' buttons to adjust the ultrasound intensity. There are 5 intensity levels: 0.8, 1.2, 1.6, 2.0, and 2.5 (corresponding to effective ultrasound output of 1.1, 2.5, 3.3, 4.1, 5 W/cm² respectively). Start with a lower intensity and gradually increase as comfortable.
  4. Start Treatment: Press the 'ON/OFF' button to begin the ultrasound therapy. The device will start emitting ultrasound waves.
  5. Apply Probe: Gently place the ultrasound probe head flat against the skin in the gel-covered treatment area. Move the probe in slow, circular motions over the area. Do not hold the probe stationary in one spot.
  6. Monitor: The device will automatically shut off after the selected time has elapsed.
  7. End Treatment: After the session, wipe off any excess gel from your skin and the probe head. Turn off the device by pressing the 'ON/OFF' button and then the 'Power' switch on the back.

5.1 Applicable Areas and Efficacy

The device is suitable for various body parts experiencing pain or requiring tissue support, including the knee, shoulder, waist, hands, and neck. It can help with conditions such as shoulder pain, cervical spondylosis, lumbar pain, soft tissue injury, calf strain, and leg pain.

Diagram showing various pain points on the body where the device can be applied

Figure 4: Applicable Pain Points

Diagrams illustrating massage and analgesia, metabolism promotion, muscle relaxation, and congestion absorption

Figure 5: Efficacy of Ultrasound Therapy

6. Maintenance

  • Cleaning: After each use, wipe the ultrasound probe head and the main unit with a soft, damp cloth. Do not use abrasive cleaners or immerse the device in water.
  • Storage: Store the device in a cool, dry place, away from direct sunlight and extreme temperatures. Keep it in its original packaging or a protective case when not in use.
  • Fuses: The device includes two spare fuses. If the device fails to power on, check and replace the fuses as needed. Refer to the product list for fuse location.

7. Troubleshooting

ProblemPossible CauseSolution
Device does not power onPower cable not connected; Power switch off; Blown fuseEnsure power cable is securely connected. Turn on the power switch. Check and replace fuses if necessary.
No ultrasound sensation/effectInsufficient gel; Probe not in full contact; Intensity too low; Probe not connected properlyApply more conductive gel. Ensure full contact of the probe with skin. Increase intensity. Reconnect the probe cable.
Device overheatsProlonged use; Blocked ventilationAllow the device to cool down. Ensure ventilation holes on the bottom are not blocked.

8. Specifications

ParameterValueParameterValue
Product NameUltrasound equipmentUltrasound Frequency1.05MHz ±10%
ModelAK-CS-01Power Input Rating25W
Rated VoltageAC100-240VDisplay ScreenLCD screen
Frequency50/60HzUnit Weight2.16kg
Product Size310*230*100mmTime Indication5-30 minutes
Rated Power35WPenetrating Depth7cm
Temperature of Probe<60℃MaterialABS
Relative Humidity20%-80%Ambient Temperature5-40℃
Atmospheric Pressure80-106kPaBeam TypeCollimated
Power (Probe)≤ 5WBNR6
ERA9cm²Pulse Duration2 / 4 / 6 / 8 / 10ms

8.1 Product List (What's Included)

Contents of the product package: main unit, ultrasound probe, power cable, user manual, and two fuses

Figure 6: Product List

  • Ultrasound Therapy Main Unit
  • Ultrasound Probe
  • Power Cable
  • User Manual
  • Fuses (x2)
  • Note: Conductive gel is not included and must be purchased separately.

9. User Tips

  • Gel is Essential: Always use a sufficient amount of conductive ultrasound gel. Without it, the ultrasound waves cannot effectively penetrate the skin, and you may not feel any effect.
  • Start Low, Go Slow: When first using the device or treating a new area, begin with the lowest intensity setting and gradually increase it to a comfortable level.
  • Keep Moving: Ensure the probe is constantly moving in slow, circular motions over the treatment area. Do not keep it stationary to avoid localized heat buildup.
  • Voltage Compatibility: The device supports 100-240V AC, 50/60Hz, making it compatible with various international power standards, including the USA. Ensure you have the correct plug type for your region.

10. Frequently Asked Questions (FAQ)

Q: Does this device have multiple frequencies like 1 MHz and 3 MHz?
A: The specifications indicate a frequency of 1.05MHz ±10%. It does not operate at 3 MHz.
Q: Can ultrasound be delivered in pulses?
A: Yes, the device supports pulse durations of 2, 4, 6, 8, and 10 milliseconds, allowing for pulsed ultrasound delivery.
Q: Is conductive gel included with the device?
A: No, conductive gel is not included in the packing box and needs to be purchased separately.
Q: Can this device be used for conditions like Erectile Dysfunction (ED)?
A: This device is designed for general physiotherapy applications such as relieving muscle soreness, pain, and supporting tissue regeneration in areas like shoulders, back, and limbs. For specific medical conditions, please consult a healthcare professional.

11. Warranty and Support

For warranty information, technical support, or service inquiries, please refer to the contact details provided by your retailer or the manufacturer's official website. Keep your purchase receipt as proof of purchase.