Yieryi YY-2023 High Accuracy Food pH Meter Instruction Manual

Model: YY-2023

1. Introduction

The Yieryi YY-2023 Food pH Meter is a high-accuracy digital device designed for precise pH and temperature measurement in various food items and liquids. It features a specialized spear tip probe, a backlit LCD display, and comes with essential accessories for calibration and maintenance. This meter is ideal for monitoring pH levels in products like meat, cheese, dough, fermented foods, beverages, and even cosmetics, ensuring optimal quality and safety.

Yieryi YY-2023 pH meter measuring fermented sausages, displaying pH 5.50 and temperature 73.2.

Figure 1: Yieryi YY-2023 pH meter in use for fermented sausages.

2. Package Contents

Please check the package carefully to ensure all items are present:

  • 1 x Upgraded Food pH Meter (YY-2023)
  • 2 x pH Calibration Liquid (pH 4.00 and pH 6.86)
  • 1 x Probe Protection Fluid (KCL Storage Solution)
  • 1 x English Instruction Manual
  • 1 x Exquisite Packaging Box
  • 1 x Drop Protection Module
Yieryi YY-2023 pH meter kit including the meter, pH calibration solutions, and probe storage solution.

Figure 2: Contents of the Yieryi YY-2023 pH Meter kit.

3. Specifications

FeatureSpecification
Model NumberYY-2023
Brand Nameyieryi
Display TypeDIGITAL (Backlit LCD)
pH Range0.00-14.00pH
Temperature Range0°C-50°C (32.0°F-122.0°F)
pH Calibration4.00/6.86 Automatic Calibration
Temperature Compensation0°C-50°C (32.0°F-122.0°F)
ElectrodeDetachable Replacement, Spear Tip
Waterproof RatingIP65
Power Supply2 x 1.5V AAA Batteries (Included)
Power SavingAutomatic shut-down after 8 minutes
Smart DeviceYes
CertificationCE
Detailed diagram of the Yieryi YY-2023 pH meter showing LCD display, C/F/4.00 button, ON/OFF/CAL button, electrode fixing nut, storage cover, 6.86 button, 2x1.5V AAA battery compartment, protective cover, temp probe, and pH probe.

Figure 3: Components and features of the YY-2023 pH Meter.

4. Setup

  1. Battery Installation: Open the battery compartment (refer to Figure 3 for location) and insert two 1.5V AAA batteries, ensuring correct polarity. Close the compartment securely.
  2. Probe Preparation: Gently remove the protective cover from the electrode by pulling it downwards.
  3. Initial Cleaning: Rinse the pH probe with distilled water and gently wipe it dry with a clean, soft cloth or tissue.
  4. Activation (if stored dry): If the meter has not been used for a long time (or if the probe appears dry), soak the electrode in KCL storage solution for at least 2 hours to activate it.

5. Operating Instructions

5.1. Calibration

For accurate readings, calibrate the pH meter regularly, especially before first use, after long periods of inactivity, or if you suspect inaccurate readings.

  1. Unscrew the caps on the bottles of pH 6.86 and pH 4.00 standard solutions.
  2. Rinse the electrode with distilled water and wipe dry. Turn on the pH Meter by pressing the 'ON/OFF/CAL' button.
  3. Calibrate pH 6.86: Immerse the electrode in the pH 6.86 buffer solution. Wait for the value to stabilize on the display. Then, press and hold the 'ON/OFF/CAL' button for 5 seconds. The screen will show "CAL" jumping. Once the value stabilizes around 6.86, press the '6.86' button. The calibration for pH 6.86 is now complete.
  4. Rinse the electrode with clean water and wipe dry.
  5. Calibrate pH 4.00: Repeat the above calibration steps for the pH 4.00 solution. Immerse the electrode in pH 4.00 buffer solution, wait for stability, then press and hold 'ON/OFF/CAL' for 5 seconds. Once stable around 4.00, press the '4.00' button.
Diagram showing the pH meter being calibrated with 6.86 and 4.00 pH buffer solutions.

Figure 4: pH Meter Calibration Process.

5.2. General Measurement

  1. Ensure the meter is calibrated (see 5.1).
  2. Remove the protective cover from the probe by pulling it downwards.
  3. Clean and dry the pH probe with distilled water and a soft cloth.
  4. Turn on the pH meter by pressing the 'ON/OFF/CAL' button.
  5. Insert the spear tip of the electrode directly into the solid, semi-solid, or liquid sample you wish to measure. Ensure the probe is fully immersed in the sample.
  6. Wait for the reading on the LCD display to stabilize. The display will show both pH and temperature.
  7. Record the stable pH and temperature readings.
  8. After measurement, rinse the electrode with distilled water and gently wipe it dry.
  9. Close the protective cover by pushing it upwards to protect the probe.
Four-step visual guide on how to use the pH tester: 1. Remove protective cover, 2. Clean and dry probe, 3. Insert into sample, 4. Close protective cover.

Figure 5: How to Use the pH Tester.

5.3. Changing Temperature Units

To switch between Celsius (°C) and Fahrenheit (°F), press the '°C/°F/4.00' button briefly during operation.

Video 1: Demonstration of the Yieryi Food pH Meter's use in various food applications like sourdough, sausages, cheese, and cosmetics.

6. Maintenance

6.1. Probe Storage and Life Extension

Proper storage is crucial for maintaining probe accuracy and extending its lifespan.

  • Keep Sponge Moist: Always add the supplied KCL storage solution to the electrode cover, ensuring the sponge inside remains moist. This prevents the loss of reference solution from the electrode and maintains its original pH potential balance, leading to more accurate measurements.
  • Electrode Activation: If the tester has not been used for a long time and the probe has dried out, soak it in KCL solution for at least 2 hours to reactivate it before use.
Instructions on how to extend probe life by adding KCL solution to keep the sponge moist and how to activate a dry electrode by soaking it in KCL solution.

Figure 6: Probe Maintenance and Life Extension.

6.2. Cleaning

After each use, rinse the electrode thoroughly with distilled water to remove any food residues or contaminants. Gently wipe dry with a soft, lint-free cloth.

7. Troubleshooting

  • Inaccurate or Unstable Readings:
    • Ensure the meter is properly calibrated (refer to Section 5.1).
    • Clean the probe thoroughly with distilled water.
    • Check if the probe's sponge is moist with KCL solution. If dry, reactivate the electrode (refer to Section 6.1).
    • Ensure the probe is fully immersed in the sample and there are no air bubbles around the sensor.
  • No Display or Meter Not Turning On:
    • Check if the batteries are correctly installed and have sufficient charge. Replace if necessary.
  • Slow Response Time:
    • The electrode might need activation. Soak it in KCL solution for 2 hours (refer to Section 6.1).
    • The electrode might be old or damaged and may need replacement.

8. User Tips

  • Dough pH: The ideal pH of well-mixed dough is typically 5.0-6.0. After fermentation, it decreases to about 3.8-4.6, depending on the degree of fermentation.
  • Meat pH: For fresh meat, the pH should generally be between 5.5 and 6.2 during storage to maintain quality.
  • Black Tea pH: The ideal pH for black tea is 2.5-3.5. Ensure the starting pH for the first fermentation is 4.5 or less.
  • Sushi Rice pH: The recommended pH for acidified sushi rice is 4.6 or lower.
  • Cheese pH: The pH of various cheeses typically ranges from 5.1-5.9.
  • Cosmetics/Soap Making: The meter can be directly inserted into lotions or soaps to check pH values, simplifying adjustments.
Yieryi pH meter measuring a cut of meat, with a table showing approximate pH values for various meat types like Abalone (6.30), Beef (5.55), Boston Bluefish (6.30), Chicken (5.90), Clams (6.55), Crab meat (6.75), Pate (5.90), Pork (5.90), Salmon (6.29), Scallops (6.00), Shrimp (6.75), Turkey meat (6.10), Cuttlefish (6.30), Eel (6.20), and cooked Lobster (7.27).

Figure 7: pH values for different meat qualities.

9. Warranty and Support

Yieryi is committed to providing high-quality products and customer service. If you encounter any issues with your YY-2023 Food pH Meter or have questions regarding its operation, please contact our customer support. We aim to provide a satisfactory response within 12 hours.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.