1. Introduction
The Yieryi PH-W218 is an advanced 8-in-1 WiFi Tuya Smart Water Quality Tester designed for continuous online monitoring of various water parameters. This device provides real-time data for pH, ORP, TDS, EC, Salt, S.G (Specific Gravity), Temperature, and CF (Conductivity Factor). It is suitable for a wide range of applications including aquariums, mariculture, drinking water analysis, hydroponics, laboratories, spas, and swimming pools. The tester integrates with the Tuya Smart APP for convenient remote monitoring, data logging, and alert notifications.
2. Package Contents
Please verify that all items are present and in good condition upon unpacking:
- 1 x Main Monitor Unit
- 1 x PH Probe
- 1 x ORP Probe
- 1 x PPM/EC/Salt/S.G./CF Probe
- 1 x Temperature Probe
- 1 x Power Adapter (EU/US/UK/AU plug type, depending on region)
- 1 x English User Manual
- 2 x Suction Cups
- 1 x Clip
- 1 x 4.0 PH Calibration Powder
- 1 x 6.86 PH Calibration Powder
- 1 x 9.18 PH Calibration Powder
- 1 x ORP Calibration Powder
- 1 x 1413us/cm/25ml Calibration Solution
- 1 x 12.88ms/cm/25ml Calibration Solution

3. Setup
3.1 Physical Installation
The PH-W218 features a wall-mounted design for easy installation. Use the provided suction cups and clip to securely mount the main monitor unit in a suitable location. Ensure the location is dry and not exposed to water or dampness, as the tester is not waterproof.


3.2 Probe Connection
Connect each probe to its corresponding port on the main monitor unit. The probes are labeled for PH, ORP, PPM/EC/Salt/S.G./CF, and Temperature. Ensure connections are secure.

3.3 Power Connection
Connect the provided power adapter to the main monitor unit and a suitable power outlet. Important: Do not use a mobile phone USB cable to power the device; use only the supplied power adapter.
3.4 Tuya Smart APP & WiFi Connection
- Download the "Tuya Smart" APP from your device's app store (compatible with Apple and Android systems).
- Register or log in to your Tuya account.
- On the PH-W218 main unit, long press the "MODE" key (PPM/EC/SALT/S.G/CF section) to turn WiFi on/off. The WiFi indicator should blink.
- In the Tuya Smart APP, follow the on-screen instructions to add a new device. The app should discover the "WiFi smart online 8-in-1" device.
- Enter your WiFi network details (SSID and password) when prompted.
- Once connected, the device will appear in your app, and you can begin real-time monitoring.

4. Operation
4.1 Basic Measurement
Once the device is powered on and probes are connected, immerse the electrodes into the water sample you wish to test. The LCD screen on the main unit will display real-time readings for all parameters. The Tuya Smart APP will also show these readings.

4.2 Switching Temperature Units (°C/°F)
Short press the "TEMP CAL" key (PH/TEMP section) to switch the temperature display between Celsius (°C) and Fahrenheit (°F).

4.3 Switching PPM/EC/SALT/S.G/CF Display
Short press the "MODE" key (PPM/EC/SALT/S.G/CF section) to cycle through the display of PPM, EC, SALT, S.G, and CF parameters.

4.4 Real-time Monitoring via APP
The Tuya Smart APP provides a comprehensive interface for real-time monitoring. You can view current data, historical data, and set alarm thresholds for various parameters. The app automatically converts language and region settings based on your phone's configuration.






4.5 Data Logging and Export
The device logs data continuously, transmitting a set of data every 15 seconds. You can view data curve graphs for daily, weekly, and monthly periods within the app for analysis.

To export data:
- In the Tuya Smart APP, navigate to the data export section.
- Click "Export" and enter your email address.
- The data will be sent to your mailbox in an Excel format. You can then download, view, and share the data.

4.6 Alarm Settings
You can set high and low alarm thresholds for pH and other parameters within the Tuya Smart APP. The app can provide SMS notifications or pop-up screen prompts when values exceed or fall below these set limits.
5. Calibration
Regular calibration ensures the accuracy of your water quality tester. Use the provided calibration powders and solutions as instructed.
5.1 PH Calibration
The PH-W218 supports 3-point pH calibration. You can choose between two sets of calibration points: 6.86, 4.00, and 9.18, or 7.00, 4.00, and 10.00. The factory default calibration is 6.86, 4.00, and 9.18.
To calibrate pH:
- Prepare the pH buffer solutions (e.g., pH 4.01, pH 6.86, pH 9.18).
- Long press the "TEMP CAL" key (PH/TEMP section) to enter PH calibration mode.
- Follow the on-screen prompts or app instructions to calibrate each point by immersing the PH probe in the respective buffer solution.

5.2 ORP Calibration
ORP calibration points are 265mV and 100mV. Use the provided ORP calibration powder to prepare the solution and follow the calibration procedure as outlined in the app or detailed manual.
5.3 EC Calibration
EC calibration points are 12.88ms/cm and 1.413ms/cm. Use the provided conductivity solutions and follow the calibration procedure.
6. Specifications
| Parameter | Range | Resolution | Accuracy |
|---|---|---|---|
| PH | 0.00 - 14.00PH | 0.01PH | ± 0.03PH |
| ORP | -2000mv - +2000mv | 1mV | ± 5mv |
| EC | 0-19000us/cm, 0-199.0ms/cm | 1us/cm (<1999us/cm), 10us/cm (>1999us/cm), 0.1ms/cm | ±2% F.S |
| TDS | 0-19990ppm, 0-199.0ppt | 1ppm (<1999ppm), 10ppm (>1999ppm), 0.1ppt | ±2% F.S |
| SALT | 0-199.9ppt, 0%-20% | 0.01% | ±2% F.S |
| S.G | 0.990-1.400 | 0.001 | N/A |
| Temperature | 0-50°C (32-122°F) | 0.1°C (0.2°F) | ±1.0°C (±2°F) |
| CF | 0-1999.0 CF | 1CF | N/A |
- Operating Temperature: 0°C ~ 50°C
- Temperature Compensation: 0°C ~ 50°C
- Power Supply: DC6V
- Dimensions: 152 mm x 80 mm x 26 mm (5.98 in x 3.07 in x 0.98 in)
- Weight: 245g
- Connectivity: WiFi (Tuya Smart APP)
- Display Type: DIGITAL with Backlight
- Battery Included: No
- Certification: CE

7. Maintenance
- Probe Care: Always keep the probes clean. Rinse them with distilled water after each use to prevent contamination.
- Storage: Store the probes in their protective caps, ideally with a small amount of storage solution (for pH and ORP probes) to keep them hydrated.
- Cleaning: Wipe the main unit with a soft, dry cloth. Do not use abrasive cleaners or immerse the unit in water.
- Calibration: Calibrate the device regularly, especially if you notice inconsistent readings or after prolonged storage, to maintain accuracy.
8. Troubleshooting
- Device not powering on: Ensure the power adapter is correctly connected to the unit and a working power outlet. Verify you are using the supplied power adapter, not a mobile phone USB cable.
- WiFi connection issues:
- Ensure the device's WiFi is enabled (long press "MODE" key).
- Check your WiFi network name and password in the Tuya Smart APP.
- Make sure the device is within range of your WiFi router.
- Restart the device and try connecting again.
- Inaccurate readings:
- Clean the probes thoroughly.
- Perform a full calibration for the affected parameter.
- Ensure probes are fully immersed in the sample.
- Check for any damage to the probes or their cables.
- App not displaying data:
- Verify the device is successfully connected to WiFi and the Tuya Smart APP.
- Check your internet connection on your smartphone.
- Restart the app or the device.
9. User Tips
- For best results, allow the probes to stabilize in the water sample for a few minutes before taking a final reading.
- When setting alarm thresholds in the app, consider the specific needs of your application (e.g., fish tank, hydroponics) to prevent false alarms or missed critical changes.
- Utilize the data logging and export features to track trends over time, which can be invaluable for maintaining stable water conditions.
- Always use fresh calibration solutions for accurate calibration. Do not reuse old solutions.
10. Warranty and Support
For warranty information, technical support, or service inquiries, please contact the seller or manufacturer directly. Keep your purchase receipt as proof of purchase.