1. Introduction
The Hantek HT360 series portable hardness tester is a sophisticated device designed for accurate measurement of metal hardness using the Richter hardness principle. It is capable of testing a wide range of metal materials, including steel, cast steel, alloy tool steel, stainless steel, gray cast iron, ductile iron, cast aluminum alloy, copper-zinc alloy, copper-tin alloy, pure copper, and forged steel.
This instrument facilitates easy conversion between seven common hardness values: HL (Leeb), HB (Brinell), HRB (Rockwell B), HRC (Rockwell C), HRA (Rockwell A), HV (Vickers), and HS (Shore). Featuring a streamlined ergonomic design, a standard bracket, and a clear 2.8-inch high-definition color display, the HT360 provides rich and intuitive interface information. It includes a large-capacity memory for up to 500 data sets and supports optional Bluetooth printing for convenient result output. Furthermore, it offers Windows program-controlled connectivity via USB Type-C for seamless data storage, export, and printing on a computer.

2. Package Contents
Please check the contents of your package upon receipt. The specific items may vary slightly depending on the model (HT360 I, HT360 II, or HT360C) purchased.
2.1 HT360 I Package List

- Hantek HT360 Hardness Tester Unit
- Soft Carrying Case
- USB Type-C Cable
- Power Adapter
- D-type Impact Device
- Cleaning Brush
- Certification Card
2.2 HT360 II Package List

- Hantek HT360 Hardness Tester Unit
- Calibration Block
- USB Type-C Cable
- Power Adapter
- D-type Impact Device
- Cleaning Brush
- Certification Card
- Hard Carrying Case
2.3 HT360C Package List

- Hantek HT360 Hardness Tester Unit
- Portable Printer
- Calibration Block
- USB Type-C Cable
- Power Adapter
- D-type Impact Device
- Cleaning Brush
- Certification Card
- Hard Carrying Case
3. Key Features
The Hantek HT360 series offers a range of features designed for precision and ease of use:
- Richter Hardness Measurement: Utilizes the Richter hardness principle for accurate metal hardness testing.
- Wide Material Compatibility: Capable of measuring hardness for steel, cast steel, alloy tool steel, stainless steel, gray cast iron, ductile iron, cast aluminum alloy, copper-zinc alloy, copper-tin alloy, pure copper, and forged steel.
- Multi-Directional Testing: Supports 5 different impact directions: vertical down, oblique down, horizontal, oblique up, and vertical up.
- Seven Hardness Scales: Provides conversion and display across HL, HB, HRB, HRC, HRA, HV, and HS scales.
- Versatile Impact Devices: Comes standard with a D-type impact device and supports other irregular impact devices like DC, DL, D+15, and G.
- Multiple Impact Tests: Allows for automatic calculation of average values from multiple shocks, with user-definable shock limits up to 32.
- Data Storage: Large internal memory capable of storing up to 500 groups of measurement data, including value, average, date, impact direction, number of times, material, and hardness system.
- Connectivity: Features a USB Type-C interface for easy connection to a PC, enabling data storage, export, and printing via Windows software.
- Display: Equipped with a 2.8-inch high-definition color display for clear and intuitive information presentation.
- Power Management: Powered by a rechargeable lithium battery, offering over 80 hours of continuous operation. Includes adjustable backlight time and automatic power-saving shutdown functions.
- Multi-Language Support: Easy to switch between English and Chinese display languages.
- Optional Bluetooth Printing (HT360C): The HT360C model supports an optional Bluetooth printer for immediate hard copy results.

4. Setup Guide
4.1 Charging the Device
- Connect the provided USB Type-C cable to the HT360 tester's charging port.
- Plug the other end of the USB cable into the power adapter, and then into a standard wall outlet.
- The battery indicator on the display will show charging status. Ensure the device is fully charged before first use for optimal battery life.
4.2 Connecting the Impact Device
- Locate the impact device connection port on the top of the HT360 tester (refer to Figure 6).
- Align the connector of the D-type impact device with the port and gently push it in until it clicks securely into place.
- Ensure the connection is firm to prevent measurement inaccuracies.

4.3 Powering On/Off
- To power on, press and hold the power button (usually located on the front panel) until the display illuminates.
- To power off, press and hold the power button again until the display shuts down.
4.4 Initial Settings
- Upon first power-on, you may be prompted to select a language. Use the navigation buttons to select your preferred language (English or Chinese) and confirm.
- If your model includes a calibration block (HT360 II, HT360C), it is recommended to perform an initial calibration to ensure accuracy. Refer to Section 5.5 for calibration instructions.
5. Operating Instructions
5.1 Device Overview

The device features a clear display and an intuitive button layout for easy operation. Key buttons include function keys (F1-F4), navigation arrows, Scale, Direction, Material, Times, Esc, Save, and Print (for HT360C).
5.2 Selecting Material and Hardness Scale
- From the main measurement screen, press the 'Material' button to enter the material selection menu.
- Use the up/down arrow keys to navigate through the list of materials (e.g., Cast Steel, Stainless Steel, Copper-Zinc Alloy). Press 'OK' or 'Enter' to confirm your selection.
- Press the 'Scale' button to select the desired hardness scale (e.g., HL, HB, HRC). Use the arrow keys to choose and confirm.
5.3 Selecting Impact Direction
- Press the 'Direction' button to cycle through the available impact directions (vertical down, oblique down, horizontal, oblique up, vertical up).
- Select the direction that matches your measurement setup.
5.4 Performing Measurements
- Ensure the impact device is properly connected and the material and scale settings are correct.
- Place the impact device firmly and perpendicularly on the surface of the test piece.
- Press the trigger button on the impact device. The impact body will strike the surface and rebound.
- The hardness value will be displayed on the screen.
- Repeat the measurement multiple times (as per your preset 'Times' setting) to obtain an average value.
5.5 Calibration
For models HT360 II and HT360C, a calibration block is included. Regular calibration ensures the accuracy of your measurements.
- Ensure the device is powered on and the impact device is connected.
- Select the appropriate material and hardness scale for the calibration block.
- Perform several measurements on the calibration block as described in Section 5.4.
- Compare the displayed average value with the known hardness value of the calibration block.
- If there is a significant deviation, consult the device's menu for calibration adjustment options or contact support.
5.6 Viewing and Saving Data
- After performing measurements, the device automatically stores the data.
- To view saved data, navigate to the 'Save' or 'Memory' section using the menu buttons.
- Use the arrow keys to browse through stored measurement groups.
5.7 Connecting to PC for Data Transfer
- Connect the HT360 tester to your computer using the provided USB Type-C cable.
- The device should be recognized as a removable storage device or require specific Hantek software (check Hantek's official website for software downloads).
- Follow the software instructions to transfer, analyze, and print measurement data.
5.8 Using the Optional Bluetooth Printer (HT360C)
- Ensure the Bluetooth printer is charged and powered on.
- On the HT360C, navigate to the 'Print' menu or press the dedicated 'Print' button.
- Follow the on-screen prompts to pair with the Bluetooth printer.
- Once paired, you can print current measurement results or stored data directly from the device.
6. Maintenance
6.1 Cleaning the Impact Device
Regular cleaning of the impact device is crucial for maintaining measurement accuracy.
- Disconnect the impact device from the main unit.
- Use the provided cleaning brush to remove any debris or dust from the impact body and guide tube.
- Avoid using harsh chemicals or abrasive materials that could damage the device.
6.2 Battery Care
- Recharge the device when the battery indicator is low.
- Avoid completely draining the battery frequently to prolong its lifespan.
- If storing the device for an extended period, ensure it is partially charged (around 50%) and recharge it every few months.
6.3 General Storage
- Store the HT360 tester and its accessories in a clean, dry environment, preferably in its carrying case, away from direct sunlight, extreme temperatures, and high humidity.
- Keep the device away from strong magnetic fields.
7. Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| Device does not power on | Low battery; Device malfunction | Charge the device; Contact customer support |
| Inaccurate readings | Improper calibration; Dirty impact device; Incorrect material/scale selection; Unstable contact with test piece | Perform calibration; Clean impact device; Verify settings; Ensure firm and perpendicular contact |
| Impact device not triggering | Loose connection; Device malfunction | Ensure impact device is securely connected; Contact customer support |
| Cannot connect to PC | Faulty USB cable; Driver issues; Software not installed | Try a different USB cable; Install necessary drivers/software from Hantek website |
| Bluetooth printer not connecting (HT360C) | Printer off/low battery; Not paired; Out of range | Ensure printer is on and charged; Re-pair the devices; Move printer closer to the tester |
8. Specifications
| Parameter | Value |
|---|---|
| Display | 2.8-inch Color Display |
| Hardness Scale | HL, HB, HRB, HRC, HRA, HV, HS |
| Measuring Range (HLD) | (170-960)HLD |
| Measuring Range (HRC) | (17.9-69.5)HRC |
| Measuring Range (HRB) | (13.5-101.7)HRB |
| Measuring Range (HB) | (19-683)HB |
| Measuring Range (HV) | (80-1042)HV |
| Measuring Range (HS) | (30.6-102.2)HS |
| Measuring Range (HRA) | (59.1-85.4)HRA |
| Measuring Materials | Steel and cast steel, alloy tool steel, stainless steel, gray cast iron, ductile iron, cast aluminum alloy, copper-zinc alloy, copper-tin alloy, pure copper, forged steel |
| Impact Device | D-type (standard), supports DC, DL, D+15, G |
| Accuracy | ±6HLD (when HLD=760) |
| Repeatability | ±6HLD (when HLD=760) |
| Measurement Direction | 360° (vertical down, oblique down, horizontal, oblique up, vertical up) |
| Data Storage | 500 groups (impact times 32-1) |
| Interface | USB Type-C |
| Power Supply | Rechargeable Lithium Battery |
| Continuous Working Time | No less than 80 hours |
| Operating Temperature | -10°C to +50°C |
| Storage Temperature | -30°C to +60°C |
| Dimensions | 150mm x 76mm x 38mm |
| Weight | 200g (main unit) |
9. User Tips
- Stable Contact: Always ensure the impact device is held firmly and perpendicularly against the test surface to achieve accurate and repeatable results.
- Surface Preparation: For best results, the test surface should be clean, smooth, and free of scale, oil, or other contaminants.
- Multiple Readings: Take multiple readings at different points on the test piece and average them to minimize error and ensure a representative hardness value.
- Temperature Considerations: Ensure the test piece and the hardness tester are at a stable temperature, ideally within the operating temperature range, for consistent measurements.
10. Warranty and Support
For warranty information, technical support, or service inquiries, please refer to the documentation provided with your purchase or visit the official Hantek website. Keep your purchase receipt as proof of purchase for warranty claims.