GVDA GD120B

Contents

GVDA GD120B Smart Digital Multimeter User Manual

Intelligent True RMS Multimeter with Auto Range and NCV Function

1. Introduction

This GVDA GD120B Smart Digital Multimeter is an intelligent true RMS digital multimeter designed for precise and reliable electrical measurements. It features both intelligent (AUTO) and professional measurement functions, a full-function display with gear indication, and an analog bar for multiple displays. With a sampling speed of 3 times per second, it provides quick and accurate readings.

GVDA GD120B Smart Digital Multimeter with test leads
GVDA GD120B Smart Digital Multimeter with included test leads.

Product Features:

  • HD Colour LCD Display for clear readings.
  • Maximum display: 6000 Counts.
  • Non-contact AC voltage detection (NCV).
  • Sampling Rate: Approximately 3 times per second.
  • Over range display: "OL".
  • Low voltage display: Low battery indication.
  • Working environment: 0-40°C, relative humidity <80%.
  • Power Supply: 4x1.5V AAA Batteries (NOT included).
  • Stable performance, high precision, high reliability, and overload protection.
  • True RMS measurement capability.
  • Diode, Capacitance, Frequency, and Live Wire detection.
  • Integrated LED Flashlight.

2. Safety Information

Please read this manual carefully before using the instrument and pay special attention to all safety warnings. Failure to follow these instructions may result in electric shock, personal injury, or damage to the meter.

2.1 Safety Statement

⚠ Warning: Indicates an operation that may cause danger to users.

⚠ Caution: Indicates an operation that may cause damage to the meter or equipment.

2.2 Safety Instructions

  • The meter conforms to IEC61010-1 CAT.III 600V overvoltage safety standard and pollution level 2.
  • Operate the meter according to this manual. Otherwise, the protection function provided by the instrument may be damaged or weakened.
  • Take special care when measuring values that exceed 60VDC, 30VAC RMS, or 42V, as these voltages pose a risk of electric shock.
  • Do not measure voltage higher than the rated value between terminals or between terminals and ground.
  • Always test a known voltage first to ensure the meter is working correctly before performing critical measurements.
  • Before using, inspect the meter for any cracks or damaged plastic parts in the instrument shell. Do not use if damaged.
  • Check if the probes are cracked or damaged. Replace them with probes of the same model and electrical specification if necessary.
  • Use the meter according to the measurement category, voltage, or current rating specified in this manual.
  • Observe local and national safety regulations. Wear personal protective equipment (e.g., rubber gloves, masks, flame retardant clothing) when exposed to dangerous live conductors.
  • If the low battery symbol "L" is displayed, replace the battery promptly to prevent measurement errors.
  • Do not use the meter in environments with explosive gas, steam, or high humidity.
  • When using probes, always hold your fingers behind the probe finger guards.
  • When measuring, connect the null or ground wire first, then the live wire. When disconnecting, disconnect the live wire first, then the null or ground wire.
  • Remove the probes from the meter before opening the case or battery cover. Do not use the meter if it is disassembled or the battery cover is open.

3. Product Overview

The GVDA GD120B is designed for ease of use and durability. Below is a diagram of its main components and controls.

Diagram of GVDA GD120B Multimeter showing controls and display
Front view of the GD120B Multimeter with key components labeled.
  1. Warning Indicator: Provides visual alerts.
  2. Display: HD Color LCD screen for readings.
  3. Function Key (FUNC): Used to cycle through manual measurement modes.
  4. Current Input Jack (A): For current measurements.
  5. COM Input Jack: Common ground for all measurements.
  6. Input Jack (VΩHz%C/F): For voltage, resistance, frequency, capacitance, and temperature measurements.
  7. NCV Sensor Area: For non-contact voltage detection.
  8. Flashlight: Integrated LED light for illuminating work areas.
Dimensions of the GVDA GD120B Multimeter
Physical dimensions of the GD120B Multimeter: 145mm x 71mm x 26mm.

4. Setup

4.1 Battery Installation

The GVDA GD120B requires 4 x 1.5V AAA batteries (not included) for operation.

  1. Turn off the multimeter and remove the test probes.
  2. Locate the battery cover on the back of the device.
  3. Remove the screw fixing the battery cover and carefully lift off the cover.
  4. Insert 4 new 1.5V AAA batteries, ensuring correct polarity (+/-).
  5. Replace the battery cover and secure it with the screw.

⚠ Warning: To avoid electric shock or personal injury, replace batteries immediately when the low battery indicator appears. Do not discharge batteries by shorting or reversing polarity. Remove batteries if the meter will not be used for a long time to prevent leakage.

Battery compartment showing 4xAAA batteries
Battery compartment for 4xAAA batteries.

4.2 Probe Connection

Connect the test probes to the appropriate input jacks on the multimeter:

  • Insert the black test lead into the COM (Common) input jack.
  • For most measurements (voltage, resistance, capacitance, frequency, temperature, NCV, live wire), insert the red test lead into the VΩHz%C/F input jack.
  • For current measurements (AC/DC Current), insert the red test lead into the A input jack.

The input jack indicator lights will flash for 5 times to prompt you to insert the probe into the correct jack when the gear is changed.

5. Operating Instructions

5.1 Power On/Off

Press and hold the Power button (P) for approximately 2 seconds to turn the multimeter on or off.

5.2 Gear Selection (Mode Switching)

  • Smart (AUTO) Mode: This is the default measurement mode when the meter is powered on. In this mode, the meter automatically identifies the measurement signal for DC voltage, AC voltage, resistance, and continuity. The display will show "Auto".
  • Manual Mode: Press the FUNC button to switch to manual mode and cycle through different measurement functions (e.g., AC/DC Voltage, Resistance, Capacitance, Frequency, Diode, Continuity, Temperature).
  • To return to Smart (AUTO) mode, press the SMART button.
Video demonstrating the GVDA GD120B multimeter's various measurement functions, including auto-ranging, DC voltage, resistance, diode test, AC voltage, capacitance, temperature, and NCV detection.

5.3 Data Hold

Press the H button to hold the current reading on the display. Press it again to release.

5.4 Flashlight

Press and hold the SEL/ button for about 2 seconds to turn the LED flashlight on or off.

Multimeter with LED flashlight on
The integrated LED flashlight illuminates the work area.

5.5 Backlight

Press the H/* button to turn the display backlight on or off. Note: VA display models may not have this function.

5.6 Auto Power Off

After power on, the auto power off function is enabled by default, indicated by the "P" symbol. If no key operation occurs for about 15 minutes, the meter will automatically shut down to save battery energy. To cancel this function, press and hold the SEL/ key when turning on the meter; the "P" symbol will not be displayed.

5.7 Measurement Operations

Always ensure correct probe connection and mode selection before taking measurements.

5.7.1 Smart (AUTO) Measurement

This is the default mode. Connect probes as described in Section 4.2. The meter will automatically identify and measure DC voltage, AC voltage, resistance, and continuity. The minimum measurable voltage in this mode is 0.8V.

Multimeter in Smart mode measuring AC voltage
Multimeter in Smart mode automatically detecting and measuring AC voltage.

5.7.2 AC/DC Voltage Measurement

  1. Power on the meter.
  2. Press FUNC to select the voltage gear (V).
  3. Press SEL/ to switch between AC voltage (displaying "AC") and DC voltage (displaying "DC").
  4. Insert the red probe into the VΩHz%C/F jack and the black probe into the COM jack.
  5. Contact the probes with both ends of the measured power supply (in parallel).
  6. Read the results from the display.

5.7.3 Resistance Measurement

  1. Power on the meter.
  2. Press FUNC to select the resistance gear (Ω).
  3. Insert the red probe into the VΩHz%C/F jack and the black probe into the COM jack.
  4. Contact the probes with both ends of the measured resistance (in parallel).
  5. Read the results from the display.

5.7.4 Continuity Test

  1. Power on the meter.
  2. Press FUNC to select the continuity gear ()).
  3. Insert the red probe into the VΩHz%C/F jack and the black probe into the COM jack.
  4. Contact the probes with both ends of the circuit or resistance (in parallel).
  5. If the resistance value is less than approximately 50Ω, the buzzer will sound, and the indicator light will turn on.
  6. Read the results from the display.

5.7.5 Frequency/Duty Measurement

  1. Power on the meter.
  2. Press FUNC to select the Hz% gear.
  3. Insert the red probe into the VΩHz%C/F jack and the black probe into the COM jack.
  4. Contact the probes with both ends of the measured power supply.
  5. Read the results from the display.

5.7.6 Capacitance Measurement

  1. Power on the meter.
  2. Press FUNC to select the capacitance gear (F).
  3. Insert the red probe into the VΩHz%C/F jack and the black probe into the COM jack.
  4. Contact the probes with both ends of the measured capacitance (in parallel).
  5. Read the results from the display.

5.7.7 Diode Test

  1. Power on the meter.
  2. Press FUNC to select the diode test gear (+).
  3. Insert the red probe into the VΩHz%C/F jack and the black probe into the COM jack.
  4. Contact the red probe to the anode of the diode and the black probe to the cathode.
  5. If the probe polarity is opposite to the diode polarity, the display will show "OL".
  6. Read the results from the display.

5.7.8 Temperature Measurement

A K-type thermocouple probe is included for temperature measurements.

  1. Power on the meter.
  2. Press FUNC to select the temperature gear (°C/°F).
  3. Insert the positive pole of the K-type thermocouple into the VΩHz%C/F jack and the negative pole into the COM jack.
  4. Contact the thermocouple probe with the measured object.
  5. Read the results from the display.
Multimeter performing temperature test with K-type probe
Temperature measurement using the K-type thermocouple probe.

5.7.9 Non-Contact AC Voltage Detection (NCV)

  1. Power on the meter.
  2. Press FUNC to select the NCV gear. The display will show the "NCV" symbol.
  3. Bring the NCV sensor area (top of the meter) gradually close to the conductor.
  4. When a weak electric field signal is detected, it will display "---L"; the buzzer will sound slowly, and the green light will turn on.
  5. When a strong electric field signal is detected, it will display "---H"; the buzzer will sound quickly, and the red light will turn on.
Multimeter performing NCV detection
Non-Contact Voltage (NCV) detection in progress.

5.7.10 Live Wire Detecting

  1. Power on the meter.
  2. Press FUNC to select the Live gear.
  3. Press SEL/ to display the "LIVE" symbol.
  4. Insert the red probe into the VΩHz%C/F jack and remove the black probe.
  5. Use the red probe to contact the conductor.
  6. When a weak electric field signal is detected, it will display "---L"; the buzzer will sound slowly, and the green light will turn on.
  7. When a strong electric field signal is detected, it will display "---H"; the buzzer will sound quickly, and the red light will turn on.
Multimeter detecting live wire
Live wire detection using the red test lead.

5.7.11 AC/DC Current Measurement

⚠ Caution: Do not measure current greater than 10A, otherwise the fuse will be burnt out.

  1. Power on the meter.
  2. Press FUNC to select the current gear (A). Alternatively, inserting the red probe into the A jack will automatically select the current measurement function.
  3. The display will show "DC" for DC current. Press SEL/ to switch to AC current (displaying "AC").
  4. Insert the red probe into the A jack and the black probe into the COM jack.
  5. Disconnect the measured power supply, connect the meter in series with the power supply, and then turn on the measured power supply.
  6. Read the results from the display.

6. Maintenance

6.1 Cleaning

  1. Turn off the meter power and remove the probes.
  2. Wipe the case with a damp cloth or mild detergent. Do not use abrasives or solvents.
  3. Wipe the contacts in each input socket with a clean swab soaked in alcohol.

⚠ Warning: Always keep the inside of the meter clean and dry to prevent electric shock or damage to the meter.

6.2 Battery Replacement

Refer to Section 4.1 for detailed battery replacement instructions.

6.3 Fuse Replacement

⚠ Warning: After opening the back cover of the meter, do not use the instrument for measurement to prevent electric shock or damage to the instrument.

If the fuse is burnt out, the symbol "E" will be displayed. If the current gear is selected, "FUSE" will also be displayed. Replace the fuse immediately.

  1. Turn off the meter power and remove the probes.
  2. Remove the screw fixing the back cover and carefully lift off the cover.
  3. Remove the burnt out fuse and replace it with a new one of the same specification (F10A/250V fuse).
  4. Ensure that the new fuse is installed in the safety clip and clamped tightly.
  5. Install the back cover and fix it with screws.

7. Troubleshooting

ProblemPossible CauseSolution
Meter does not power onDead batteries, incorrect battery installationReplace batteries, check polarity
Low battery indicator "L" displayedBatteries are lowReplace batteries (4x1.5V AAA)
"OL" displayedOver range measurement, open circuit (resistance/continuity), reversed diode polarityCheck measurement range, ensure circuit is closed, reverse diode probes
"FUSE" or "E" displayed during current measurementBlown fuseReplace fuse (F10A/250V)
Inaccurate readingsDamaged probes, environmental factors, meter malfunctionCheck/replace probes, ensure stable environment, contact support if persistent
No NCV/Live detectionNo voltage present, incorrect mode, sensor not close enoughVerify voltage, select correct mode, ensure close proximity to conductor

If the meter is not functioning normally or appears damaged, discontinue use immediately and refer to the safety guidelines or contact customer support.

8. Specifications

8.1 General Technical Specifications

  • Display Type: Digital Display (HD Color LCD)
  • Max Display: 6000 Counts
  • Operating Mode: Auto Range (Smart Mode)
  • Sampling Rate: Approx. 3 times/second
  • Over Range Indication: "OL"
  • Low Battery Indication: Yes
  • Input Polarity Indication: Display "-"
  • True RMS: Yes
  • Operating Temperature: 0-40°C (<80% RH, <10°C non-condensing)
  • Storage Temperature: -10-60°C (<70% RH, remove battery)
  • Safety Standard: IEC61010-1 CAT.III 600V, Pollution level 2, Altitude < 2000m
  • Power Supply: 4 x 1.5V AAA batteries (not included)
  • Dimensions: 145mm x 71mm x 26mm
  • Weight: 0.398 kg (approx. 400g package weight)
  • Certifications: CE, FCC

8.2 Accuracy Specifications (within one year after calibration, 18-28°C, <80% RH)

MeasurementRangeResolutionAccuracy
DC Voltage600mV0.1mV±(0.5% +3)
6V0.001V
60V0.01V
600V1V
AC Voltage6V0.001V±(0.8% +3)
60V0.01V
600V1V
Resistance600Ω0.1Ω±(1.0% +5)
6KΩ0.001KΩ
60KΩ0.01KΩ
600KΩ0.1KΩ
6MΩ0.001MΩ
60MΩ0.01MΩ±(1.5% +10)
DC Current600mA0.1mA±(1.2% +3)
6A0.001A
10A0.01A
AC Current600mA0.1mA±(1.2% +3)
6A0.001A
10A0.01A
Capacitance6nF0.001nF±(4.0% +5)
60nF0.01nF
600nF0.1nF
6µF0.001µF
60µF0.01µF
600µF0.1µF
6mF0.001mF±(5.0% +5)
60mF0.01mF
Frequency6Hz0.001Hz±(1.0% +3)
60Hz0.01Hz
600Hz0.1Hz
6KHz0.001KHz
60KHz0.01KHz
600KHz0.1KHz
6MHz0.001MHz
10MHz0.01MHz
Duty Cycle1.0-99.0%0.1%
Temperature-40°C ~ 0°C1°C±3°C
0°C ~ 1000°C1°C±2.0% or ±2°C
Diode TestDisplay diode voltage drop
ContinuityApprox. 50Ω, Buzzer will sound and indicator light will be on.

9. User Tips

  • For general electrical checks, start with the Smart (AUTO) mode. It simplifies common measurements like voltage and resistance by automatically detecting the signal type.
  • Utilize the NCV (Non-Contact Voltage) feature for a quick and safe preliminary check for voltage presence without needing to touch live wires directly.
  • When measuring current, always connect the multimeter in series with the circuit and ensure the power is off before connecting to avoid damaging the device.
  • The integrated LED flashlight is very useful for working in dimly lit areas, improving visibility and safety.
  • Remember to remove the test leads from the multimeter before replacing batteries or fuses to prevent accidental shock or damage.
  • Keep the multimeter and its accessories in the provided cloth bag for protection and easy transport.
Multimeter in cloth bag
The multimeter fits securely in its protective cloth bag.

10. Warranty and Support

Specific warranty details for this product are not provided in the available documentation. Please refer to your purchase receipt or contact the seller/manufacturer directly for warranty information.

For additional item guides and documents, including a PDF version of the user manual, please refer to the official product page or contact GVDA customer support.

A PDF version of the user manual is available for download: User Manual (PDF)

© 2023 GVDA. All rights reserved.


Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.