1. Introduction
The Meterk YY-1010 is a versatile 10-in-1 digital water quality meter designed for professional measurement of various parameters including pH, Electrical Conductivity (EC), Total Dissolved Solids (TDS), Oxidation-Reduction Potential (ORP), Hydrogen (H2), Resistivity (MΩ), Fertility, Salinity, Specific Gravity (S.G), and Temperature. Featuring a high-definition large screen with backlight, fast measurement capabilities, and automatic temperature compensation, this device ensures accurate and easy-to-read results. It is suitable for a wide range of applications such as household drinking water, food testing, swimming pools, laboratories, aquariums, and hydroponics.

2. Product Overview and Components
Familiarize yourself with the components of your YY-1010 water quality meter before operation.

- Battery Compartment: Houses 3 x 1.5V (LR44) button batteries.
- Power Button: Used to turn the device on and off.
- Mode Key / CAL Key: Switches between measurement modes and initiates calibration.
- Screen Display: High-definition LCD with backlight for clear readings.
- HOLD / TEMP Key: Freezes the current reading or switches between Celsius/Fahrenheit.
- Protective Cap: Covers and protects the probe when not in use.
- Replaceable Probe: The electrode head can be replaced for extended product life.
- High Precision Probe: Designed for accurate measurements and reduced interference.
3. Setup and Initial Calibration
3.1 Battery Installation
- Unscrew the top cap of the meter to access the battery compartment.
- Insert 3 x 1.5V LR44 button batteries, ensuring correct polarity.
- Securely screw the top cap back on.
3.2 Initial Calibration
The device comes with calibration powder. Calibration is crucial for accurate readings, especially for pH. Refer to the specific instructions provided with the calibration powder for detailed steps. Generally, pH calibration involves preparing buffer solutions (e.g., pH 4.00, pH 6.86, pH 9.18) and following the meter's calibration mode.
4. Operating Instructions
4.1 Power On/Off
- Press the Power Button to turn the meter on.
- Press and hold the Power Button to turn the meter off.
- The meter features an automatic shut-off after 5 minutes of inactivity to conserve battery life.
4.2 Mode Switching
The YY-1010 can measure 10 different parameters. Press the Mode Key / CAL Key to cycle through the measurement modes: PH, EC, TDS, ORP, H2, MΩ, Fertility, SALT, S.G, and Temperature.

4.3 Taking Measurements
- Ensure the protective cap is removed from the probe.
- Immerse the probe into the sample solution, ensuring the electrode is fully submerged.
- Gently stir the meter to remove any air bubbles and allow the reading to stabilize.
- Read the value displayed on the screen.
4.4 HOLD / TEMP Function
- Press the HOLD / TEMP Key briefly to freeze the current reading on the display. Press again to release.
- Press and hold the HOLD / TEMP Key to switch the temperature unit between Celsius (°C) and Fahrenheit (°F).
4.5 Automatic Temperature Compensation (ATC)
The meter features automatic temperature compensation (0-60°C / 32-140°F) to ensure accurate readings regardless of the sample's temperature. This eliminates measurement bias caused by temperature fluctuations.

5. Maintenance
5.1 Electrode Care and Cleaning
Proper care of the electrode is vital for maintaining accuracy and extending the life of your meter.

- Always keep the electrode clean.
- After each use, wash the electrode with distilled water or purified water.
- During use, avoid colliding or rubbing the electrode to prevent damage.
- Store the probe with the protective cap on, ideally with a few drops of storage solution or distilled water to keep it moist.
5.2 Probe Replacement
The electrode head is replaceable, allowing for long-term use of the product. If the probe becomes damaged or its performance degrades significantly, it can be replaced with a new one. Consult the manufacturer for compatible replacement probes.
5.3 Battery Replacement
When the battery indicator on the display shows low power, replace the 3 x 1.5V LR44 button batteries as described in Section 3.1.
6. Troubleshooting
- Inaccurate Readings: Ensure the probe is clean and properly calibrated. Check battery levels.
- No Display/Power: Verify batteries are correctly installed and charged. Replace if necessary.
- Slow Response: Clean the electrode thoroughly. If the issue persists, the probe might need replacement.
- Error Messages: If an error code appears, try recalibrating the device. If the error persists, contact customer support.
7. Specifications
| Parameter | Range | Resolution | Accuracy |
|---|---|---|---|
| pH | 0.00-14.00pH | 0.01pH | ±0.05pH |
| TDS | 0-10000ppm, 10.1-200.0ppt | 0.1ppm; 0.1ppt | ±2% F.S |
| EC | 0-10000µS/cm, 10.1-200.0mS/cm | 1µS/cm; 0.1mS/cm | ±2% F.S |
| Salinity | 0-10000ppm, 10.1-200.0ppt, 0.01%-25.0% | 1ppm, 0.1ppt, 0.01% | 0.01%-5.00% (±0.1%), 5.1-25.00% (±1%), ±2% F.S |
| ORP | ±999mV | 1mV | ±2mV |
| H2 | 0-2400ppb, 0-2.400ppm | 1ppb, 0.001ppm | ±10ppb, ±0.01ppm |
| Temperature | 0-60℃ / 32-140℉ | 1℃ / ℉ | ±0.5℃ |
| S.G | 1.000-1.222 | N/A | N/A |
| MΩ | 50Ω-20MΩ | N/A | N/A |
| Fertile | 0-3 | N/A | N/A |
- Waterproof: IP67
- Operation Environment: 0-60℃ (32-140℉); RH 100%
- TDS Factor: 0.5
- Automatic Shut-off: 5 minutes
- Battery: 3 * 1.5V (LR44)
- Item Size: 185 * 40 * 22mm / 7.28 * 1.57 * 0.86in
- Item Weight: 90g / 3.17oz
8. User Tips
- For best accuracy, always perform calibration regularly, especially for pH measurements, and after long periods of inactivity.
- When measuring, ensure the probe is fully submerged and allow a few moments for the reading to stabilize before recording.
- Rinse the probe with distilled water before and after each use to prevent cross-contamination and residue buildup.
- Keep the protective cap on the probe when not in use to prevent it from drying out or getting damaged.
- If measuring different solutions consecutively, always rinse the probe thoroughly between samples.
9. Warranty and Support
This product comes with a 1-year warranty from the date of purchase. If your purchase does not meet merchantable quality, fitness for purpose, or match the description, please contact the seller for assistance. For any issues or support inquiries, please refer to the contact information provided by your retailer or the manufacturer.