MASTECH MS8301A/B/D

Contents

Mastech MS8301 Series Digital Multimeter User Manual

Models: MS8301A, MS8301B, MS8301D

Brand: MASTECH

1. Introduction

This manual provides detailed instructions for the safe and effective operation of the Mastech MS8301 series digital multimeters (MS8301A, MS8301B, MS8301D). These devices are designed for measuring AC/DC voltage, AC/DC current, resistance, diode, and continuity. Specific models offer additional features such as temperature measurement, non-contact voltage (NCV) detection, and auto-ranging. Please read this manual thoroughly before use and keep it for future reference.

2. Safety Information

Always observe the following safety precautions to prevent electric shock or personal injury, and to avoid damage to the meter or the equipment under test.

  • Do not apply more than the rated voltage, as marked on the meter, between the terminals or between any terminal and earth ground.
  • Use extreme caution when working with voltages above 30V AC RMS, 42V peak, or 60V DC. These voltages pose a shock hazard.
  • Before making current measurements, ensure the meter's fuses are intact and the test leads are connected to the correct input jacks.
  • Always disconnect the test leads from the circuit before changing functions or ranges.
  • Do not use the meter if it appears damaged or if the case is open.
  • Ensure the battery cover is securely closed before operation.
  • Adhere to the specified measurement category (CAT rating) for the meter.

3. Product Overview

The Mastech MS8301 series multimeters are compact and robust, featuring a clear digital display and a rotary switch for function selection.

Mastech MS8301A Digital Multimeter front view

Figure 3.1: Mastech MS8301A Digital Multimeter

Mastech MS8301B Digital Multimeter front view

Figure 3.2: Mastech MS8301B Digital Multimeter

Mastech MS8301D Digital Multimeter front view

Figure 3.3: Mastech MS8301D Digital Multimeter

3.1. Key Components

  • Digital Display: Shows measurement readings.
  • Rotary Switch: Selects the desired measurement function (e.g., V~, V-, A~, A-, Ω, Diode, Continuity, Temp, NCV).
  • Function Buttons: (Vary by model) Include FUNC (Function), HOLD (Data Hold), MAX/MIN, NCV (Non-Contact Voltage).
  • Input Jacks:
    • COM: Common terminal for the black test lead.
    • VΩHz%: Input for voltage, resistance, frequency, and duty cycle measurements (red test lead).
    • mAµA: Input for small current measurements (red test lead).
    • 10A: Input for large current measurements (red test lead).

4. Specifications

Detailed specifications for each model are provided below. Please refer to the specific model you own.

4.1. MS8301A Specifications

MS8301A Features and Specifications

Figure 4.1: MS8301A Features and Specifications Overview

MS8301A Detailed Specifications Table

Figure 4.2: MS8301A Detailed Specifications Table

MS8301A Key Specifications
FeatureValue
Display Counts2000
Data HoldYes
Battery Tester1.5V/9V
AC Voltage600V
DC Voltage600V
AC Current10A
Resistance2MΩ
Continuity Test<30Ω

4.2. MS8301B Specifications

MS8301B Features and Specifications

Figure 4.3: MS8301B Features and Specifications Overview

MS8301B Detailed Specifications Table

Figure 4.4: MS8301B Detailed Specifications Table

MS8301B Key Specifications
FeatureValue
Display Counts4000
Back LightYes
Non-contact Voltage DetectorYes
AC Voltage600V
DC Voltage600V
AC Current10A
DC Current10A
Resistance40MΩ
Capacitance100µF
Frequency10kHz
Continuity Test<50Ω
Temperature-4°F ~ 1832°F

4.3. MS8301D Specifications

MS8301D Features and Specifications

Figure 4.5: MS8301D Features and Specifications Overview

MS8301D Detailed Specifications Table

Figure 4.6: MS8301D Detailed Specifications Table

MS8301D Key Specifications
FeatureValue
Display Counts6000
Auto RangingYes
Auto Power OffYes
Dual DisplayYes
Continuity BuzzerYes
Data HoldYes
MAX/MINYes
Display BacklightYes
Low Battery DisplayYes
Non-Contact Voltage DetectorYes
AC Voltage600V
DC Voltage600V
Resistance10MΩ
Frequency3kHz
Continuity Test<40Ω

5. Setup

5.1. Battery Installation

The MS8301A and MS8301D models require 2x1.5V AAA batteries. The MS8301B model requires 3x1.5V AAA batteries. To install or replace batteries:

  1. Turn off the multimeter and disconnect all test leads.
  2. Locate the battery compartment on the back of the meter.
  3. Unscrew the retaining screw(s) and remove the battery cover.
  4. Insert the new batteries, observing the correct polarity (+ and -).
  5. Replace the battery cover and secure it with the screw(s).

5.2. Connecting Test Leads

Always connect the black test lead to the 'COM' (Common) jack. Connect the red test lead to the appropriate input jack based on the measurement you intend to make:

  • For Voltage, Resistance, Frequency, and Duty Cycle: Connect the red lead to the 'VΩHz%' jack.
  • For Current (small): Connect the red lead to the 'mAµA' jack.
  • For Current (large, up to 10A): Connect the red lead to the '10A' jack.

6. Operating Instructions

The MS8301D model features a 'Smart Function' for automatic detection of measurement type. For other models, manually select the function using the rotary switch.

Play video
Watch video online

Video 6.1: Demonstration of Mastech MS8301D Smart Digital Multimeter features, including automatic detection and NCV measurement.

6.1. Automatic Detection (MS8301D)

For the MS8301D, turn the rotary switch to the 'SMART' position. The meter will automatically detect and measure AC/DC voltage, resistance, or continuity when the test leads are applied to the circuit.

Automatic Detection of AC, DC, Resistance, and Continuity

Figure 6.1: Automatic detection of various electrical parameters by the multimeter.

6.2. Voltage Measurement (AC/DC)

  1. Connect the black test lead to the 'COM' jack and the red test lead to the 'VΩHz%' jack.
  2. Select the appropriate AC Voltage (V~) or DC Voltage (V-) range using the rotary switch. For MS8301D, select 'SMART' or 'V'.
  3. Connect the test probes across the component or circuit to be measured.
  4. Read the voltage value on the display.

6.3. Current Measurement (AC/DC)

  1. Connect the black test lead to the 'COM' jack.
  2. For currents up to 200mA, connect the red test lead to the 'mAµA' jack. For currents up to 10A, connect the red test lead to the '10A' jack.
  3. Select the appropriate AC Current (A~) or DC Current (A-) range using the rotary switch.
  4. Open the circuit where current is to be measured and connect the meter in series with the load.
  5. Read the current value on the display.

6.4. Resistance Measurement

  1. Connect the black test lead to the 'COM' jack and the red test lead to the 'VΩHz%' jack.
  2. Select the 'Ω' (Resistance) function using the rotary switch. For MS8301D, select 'SMART' or 'Ω'.
  3. Ensure the circuit is de-energized before measuring resistance.
  4. Connect the test probes across the component to be measured.
  5. Read the resistance value on the display.

6.5. Continuity Test

  1. Connect the black test lead to the 'COM' jack and the red test lead to the 'VΩHz%' jack.
  2. Select the continuity function (often indicated by a speaker icon or combined with diode test) using the rotary switch. For MS8301D, select 'SMART' or the continuity symbol.
  3. Connect the test probes across the circuit or component.
  4. If the resistance is below a certain threshold (e.g., <30Ω for MS8301A, <50Ω for MS8301B, <40Ω for MS8301D), the buzzer will sound, indicating continuity.

6.6. Diode Test

  1. Connect the black test lead to the 'COM' jack and the red test lead to the 'VΩHz%' jack.
  2. Select the diode test function (often indicated by a diode symbol) using the rotary switch.
  3. Connect the red test probe to the anode and the black test probe to the cathode of the diode.
  4. The display will show the forward voltage drop. Reverse the probes; the display should show 'OL' (Open Loop) for a good diode.

6.7. Temperature Measurement (MS8301B)

  1. Connect the K-type thermocouple to the appropriate input jacks (usually VΩHz% and COM, or dedicated temperature jacks if available).
  2. Select the temperature function (often indicated by °C/°F) using the rotary switch.
  3. Place the thermocouple probe on or near the object whose temperature is to be measured.
  4. Read the temperature value on the display.

6.8. Non-Contact Voltage (NCV) Measurement (MS8301B, MS8301D)

  1. Select the 'NCV' function using the rotary switch or dedicated button.
  2. Hold the top part of the meter near the live conductor.
  3. The meter will indicate the presence of AC voltage through an audible beep and/or visual indicator (e.g., LED).
NCV Measurement in an electrical panel

Figure 6.2: Performing Non-Contact Voltage (NCV) measurement.

6.9. Data Hold Function

Press the 'HOLD' button to freeze the current reading on the display. Press it again to release the hold and resume live measurements.

6.10. MAX/MIN Function (MS8301D)

Press the 'MAX/MIN' button to record the maximum and minimum values during a measurement session. Press again to cycle through MAX, MIN, and current readings.

6.11. Duty Cycle Measurement (MS8301D)

Select the 'Hz%' function. The meter can measure the duty cycle of a signal, which is the ratio of the pulse width to the period of a square wave or pulse train.

7. Maintenance

7.1. Cleaning

Wipe the meter with a damp cloth and a mild detergent. Do not use abrasives or solvents. Keep the display clean and dry.

7.2. Battery Replacement

When the battery indicator appears on the display, replace the batteries as described in Section 5.1. Prompt battery replacement ensures accurate readings.

7.3. Fuse Replacement

If the meter fails to measure current, the fuse may be blown. Refer to the specifications for the correct fuse type and rating for your model. To replace a fuse:

  1. Turn off the meter and disconnect all test leads.
  2. Open the battery compartment/rear case (may require additional screws beyond battery cover).
  3. Carefully remove the old fuse and replace it with a new one of the identical type and rating.
  4. Reassemble the meter, ensuring all screws are tightened.

8. Troubleshooting

ProblemPossible CauseSolution
No display or faint displayLow battery or dead batteryReplace batteries (refer to Section 7.2)
Incorrect readingsIncorrect function/range selected; poor test lead connection; external interferenceVerify function/range; ensure secure connections; move away from strong electromagnetic fields
Cannot measure currentBlown fuse; incorrect input jackCheck/replace fuse (refer to Section 7.3); ensure red lead is in 'mAµA' or '10A' jack
Continuity buzzer not soundingCircuit resistance too high; function not selectedEnsure circuit is closed and resistance is below threshold; select continuity function
Meter shows 'OL' (Open Loop)Measurement exceeds range; open circuit; no connectionSelect higher range (if manual ranging); check circuit for breaks; ensure good probe contact

9. User Tips

  • Always start with the highest range: When measuring an unknown voltage or current, always start with the highest range setting and work your way down to avoid overloading the meter.
  • Check test leads: Before any critical measurement, perform a quick continuity test on your test leads to ensure they are not broken or faulty.
  • Power off circuits: For resistance and continuity measurements, always ensure the circuit under test is de-energized to prevent damage to the meter and ensure accurate readings.
  • Use Data Hold: For measurements in hard-to-reach or poorly lit areas, use the 'HOLD' function to capture the reading for later review.
  • Understand CAT ratings: Be aware of the CAT rating of your multimeter and the circuit you are testing. CAT III 600V indicates suitability for distribution level circuits.

10. Certifications

The Mastech MS8301 series multimeters are certified to meet various international standards, ensuring quality and safety:

  • CE (Conformité Européenne)
  • FCC (Federal Communications Commission)
  • RoHS (Restriction of Hazardous Substances)
  • UL (Underwriters Laboratories)
  • WEEE (Waste Electrical and Electronic Equipment)

11. Packaging Contents

Upon opening the product package, please verify that all items are present and in good condition:

  • Digital Multimeter (MS8301A/B/D, depending on model purchased)
  • Test Leads
  • Battery (2x1.5V AAA for MS8301A/D, 3x1.5V AAA for MS8301B)
  • K-type Thermocouple (for MS8301B only)
  • Calibration Certificate
  • Quick Start Guide
  • Color Box (L145xW60xH188mm)
  • Carton Box (40x32x31.5cm, containing 20 PCS)
Documents and downloads

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.