FeelTech FY2300 Series Dual-Channel Function/Arbitrary Waveform Generator Instruction Manual

Model: FY2300

1. Introduction

The FeelTech FY2300 series Dual Channel Function/Arbitrary Waveform Generator is a versatile instrument designed for generating various types of signals. It integrates a Function Signal Generator, Arbitrary Waveform Generator, Pulse Signal Generator, Noise Signal Generator, Counter, and Frequency Meter. Utilizing large-scale CMOS integrated circuits and a high-speed MCU microprocessor, the internal circuit employs an SPXO as the basic standard for signal stability. Surface mounting technology enhances interference immunity and extends operational lifespan.

This instrument features two independent channels capable of outputting 31 preset DDS signals, including Sine, Square, Triangle (Ramp), Rise Sawtooth, Fall Sawtooth, Lorenz Pulse, Multitone, Noise, Cardiogram, Trapezoidal Pulse, Sinc Pulse, Narrow Pulse, Gauss White Noise, AM, FM, and Step waveforms, as well as user-defined waveforms. It can also output TTL electric level signals.

The user-friendly interface includes a 2.4-inch TFT Color LCD with 320×240 resolution, displaying all parameters for both channels and current button functions. Shortcut keys streamline complex operations, allowing users to become proficient quickly.

The FY2300 series is an ideal tool for electronic engineers, laboratories, industrial production lines, universities, and R&D institutes, offering significant advantages in signal generation, waveform sweeping, and parameter measurement.

2. Package Contents

Upon opening the package, please verify that all items listed below are present and in good condition.

Video: Unboxing the FeelTech FY2300 Function Generator and its accessories.

  • 1 x FeelTech FY2300 DDS Signal Generator (12M or 20M variant)
  • 2 x Signal Output Cables
  • 1 x USB Cable
  • 1 x DC 5V Power Adapter (US/UK/EU/AU Optional)
  • 1 x Quality Certificate (may vary)
  • 1 x Software CD (may vary)
FeelTech FY2300 package contents including generator, cables, power adapter, and CD

Image: Complete package contents of the FeelTech FY2300.

3. Setup

3.1 Device Overview

Familiarize yourself with the main components and ports of the FeelTech FY2300.

Labeled diagram of FeelTech FY2300 front, rear, and side panels

Image: Labeled diagram of the FeelTech FY2300's interfaces.

FeelTech FY2300 Component Descriptions
NO.Description
1Function button area
2Color LCD display
3Menu
4Arrow keys
5Channel control switch
6Adjustment knob/OK key
7DC 5V power supply interface
8USB communication interface
9Power Switch
10CH2 output port
11CH1 output port
12External measurement port
13TTL input/output port

3.2 Power Connection

  1. Connect the provided DC 5V power adapter to the DC 5V power supply interface (7) on the device.
  2. Plug the power adapter into a suitable electrical outlet.
  3. Ensure the power switch (9) is in the OFF position before connecting power.
  4. Turn the power switch (9) to the ON position to power on the device.
Side view showing DC5V power and USB ports

Image: Side panel showing DC 5V power input and USB communication interface.

3.3 Signal Output Connection

  • Connect signal output cables to CH1 output port (11) and/or CH2 output port (10) as needed.
  • Connect the other end of the signal output cables to your testing equipment (e.g., oscilloscope, circuit).

3.4 PC Connection

  • To use PC software for control or arbitrary waveform generation, connect the USB cable to the USB communication interface (8) on the device and to your computer.
  • Install the necessary drivers and software from the provided CD or by contacting support.

4. Operating Instructions

4.1 User Interface Overview

The device features a 2.4-inch TFT Color LCD (2) and a function button area (1) for navigation and parameter adjustment. The adjustment knob/OK key (6) and arrow keys (4) are used to select and modify settings.

Front view of FeelTech FY2300 showing LCD and control buttons

Image: Front panel of the FeelTech FY2300 with its display and controls.

4.2 Basic Waveform Generation

The FY2300 can generate 31 preset DDS signals. Use the 'WAVE' button to cycle through available waveform types. Adjust frequency, amplitude, offset, and duty cycle using the corresponding buttons and the adjustment knob.

Screenshots of FeelTech FY2300 display showing various function interfaces

Image: Display interfaces for CH1, CH2, Burst, Sweep, Measure, and System functions.

  • Sine Wave: Smooth, continuous oscillation.
  • Square Wave: Ideal for digital circuit testing.
  • Triangle Wave (Ramp): Linear rise and fall.
  • Sawtooth Wave: Linear rise with a sharp fall, or vice-versa.
  • Pulse Wave: Adjustable pulse width and duty cycle.
  • TTL/CMOS Wave: Digital logic level signals.

4.3 Arbitrary Waveform Generation

The device supports user-defined arbitrary waveforms. These can be created and loaded via the PC software.

  1. Connect the FY2300 to your PC via the USB cable.
  2. Launch the PC software.
  3. Use the waveform editor in the software to draw or import your desired arbitrary waveform.
  4. Click "Send Waveform Data" to transfer the waveform to the generator.
PC software interface for arbitrary waveform generation and oscilloscope display

Image: PC software interface for arbitrary waveform creation and generator display example.

4.4 Sweep Function

The sweep function allows the output frequency, amplitude, offset, or duty cycle to change over a defined range and time. Access this function via the 'SWEEP' button.

  • Sweep Type: Select Linear or Logarithm sweep.
  • Sweep Objects: Choose to sweep Frequency, Amplitude, Offset, or Duty Cycle.
  • Sweep Time: Set the duration for each step (0.01S to 999.99S).
  • Setting Range: Define the starting and finishing positions for the sweep parameters.
Oscilloscope display showing various sweep waveforms generated by FY2300

Image: Examples of sweep waveforms displayed on an oscilloscope.

4.5 Measurement Function (Frequency Meter & Counter)

The FY2300 includes a built-in frequency meter and counter. Connect the signal to be measured to the External Measurement Port (12).

  • Frequency Meter: Measures frequencies from 0.01Hz to 100MHz.
  • Counter: Counts events up to 4,294,967,295.
  • Coupling Mode: Select DC or AC coupling.
  • Gate Time: Adjustable in 1S, 10S, 100S increments for frequency measurement sensitivity.

4.6 Spooky2 Compatibility

The FeelTech FY2300 signal generator is compatible for use with Spooky2 software. For detailed instructions on integrating the FY2300 with Spooky2, please refer to the Spooky2 documentation or contact FeelTech support for specific guidance.

5. Maintenance

  • Keep the device clean and free from dust. Use a soft, dry cloth for cleaning.
  • Avoid exposing the device to extreme temperatures, humidity, or direct sunlight.
  • Do not attempt to open the casing or perform repairs yourself, as this will void the warranty and may cause damage. Refer to qualified service personnel for any repairs.
  • Store the device and its accessories in a safe, dry place when not in use.

6. Troubleshooting

  • Device does not power on:
    • Ensure the power adapter is securely connected to the device and a working power outlet.
    • Verify the power switch is in the ON position.
    • Check if the power adapter is functioning correctly.
  • No signal output:
    • Check if the signal output cables are correctly connected to the CH1/CH2 ports and the receiving equipment.
    • Verify that the output channels are enabled and parameters (frequency, amplitude) are set appropriately on the device display.
    • Ensure the receiving equipment (e.g., oscilloscope) is correctly configured to display the signal.
  • PC software not connecting:
    • Ensure the USB cable is securely connected to both the device and the PC.
    • Verify that the correct USB drivers are installed on your computer.
    • Try restarting both the device and the PC software.
  • Inaccurate measurements:
    • Check the coupling mode (AC/DC) and gate time settings in the measurement function.
    • Ensure the input signal is within the specified voltage range (0.7Vpp~20Vpp).
    • Verify proper connection to the External Measurement Port (12).

7. Specifications

7.1 Frequency Range

Waveform TypeFY2300-02MFY2300-06MHzFY2300-12MHzFY2300-20MHz
Sinewave0~2MHz0~6MHz0~12MHz0~20MHz
Squarewave0~2MHz0~6MHz0~6MHz0~6MHz
Trianglewave0~2MHz0~6MHz0~6MHz0~6MHz
Sawtooth wave0~2MHz0~6MHz0~6MHz0~6MHz
Pulsewave0~2MHz0~6MHz0~6MHz0~6MHz
TTL/CMOS Wave0~2MHz0~6MHz0~6MHz0~6MHz
Arbitrary Waveform0~2MHz0~6MHz0~6MHz0~6MHz
Resolution1µHz (0.000001Hz)
Accuracy± 5×10-6
Stability±1×10-6/ 3 Hours
Impedance50Ω±10% (Typical)
Phase Range0~359°
Phase Resolution1°

7.2 Waveform Characteristics

CharacteristicValue
Waveform TypesSine, Square, Triangle, Sawtooth, Pulse, Ramp, TTL, CMOS, Multitone, Noise, Cardiogram, Sinc Pulse, Narrow Pulse, Gauss White Noise, AM, FM, Step, Arbitrary Waveform, etc.
Waveform Length2048 Points
Sampling Rate200MSa/s
Vertical Resolution12 Bits
SinewaveHarmonic Suppression≥45dBc (<1MHz); ≥40dBc (1MHz~20MHz);
Total Harmonic Distortion<0.8% (20Hz~20kHz, 0dBm)
SquarewaveRise/Fall Time≤20ns
Overshoot≤7.5%
Duty Cycle0.1%~99.9%
SawtoothwaveLinearity≥98% (0.01Hz~10kHz)

7.3 Output Characteristics

CharacteristicValue
Amplitude (50Ω)10mVpp~20Vpp (Freq. ≤ 15MHz); 10mVpp~15Vpp (Freq. > 15MHz)
Amplitude Resolution1mV (Attenuate 20dB)
Amplitude Stability±0.5%/ 5 Hours
Amplitude Flatness±5% (<10MHz); ±10% (>10MHz);
ProtectionAll channels can work more than 60 seconds when the load is short-circuited.

7.4 DC Offset

CharacteristicValue
Offset Range±10V
Offset Resolution0.01V (0.001V With Attenuation)

7.5 TTL Output

CharacteristicValue
Electrical Level Amplitude>3Vpp
Fan-out>20 TTL Load
Rise/Fall Time≤20ns

7.6 CMOS Output

CharacteristicValue
Low Electric Level<0.3V
High Electric Level1V~10V
Rise/Fall Time≤20ns

7.7 External Measurement

CharacteristicValue
Frequency MeterRange0.01Hz~100MHz
SensitivityGate Time 3 grades adjustable (1S, 10S, 100S)
CounterRange0-4294967295
Coupling ModeDC, AC
Input Voltage Range0.7Vpp~20Vpp
Period MeasurementResolution 5nS, Max. Limit 20S
Pulse Width MeasurementResolution 5nS, Max. Limit 10S

7.8 Sweep Function Specifications

CharacteristicValue
Sweep TypeLinear Sweep or Logarithm Sweep
Sweep ObjectsFrequency, Amplitude, Offset and Duty Cycle
Sweep Time0.01S~999.99S/Step
Setting RangeStarting position and Finishing position can be set arbitrarily.
Sweep RangeDetermined by the sweep parameter settings

7.9 General Specifications

CharacteristicValue
Display Type2.4 inch TFT Color Display
Save & Load Amount20
Save & Load Position01 to 20 (01 for default value)
Interface TypeUSB to Serial interface
Communicating Speed9600bps
ProtocolCommand line mode, protocol complete open.
Power Supply Voltage RangeDC5V ±0.5V
BuzzerCan be turned on/off by setting.
Environment Temperature0~40℃
Environment Humidity<80%
Size (L * W * H)140 * 95 * 54.2mm / 5.51 * 3.74 * 2.13in
Weight419g / 14.8oz
Package Size (L * W * H)22.5 * 15 * 7.5cm / 8.86 * 5.9 * 2.95in
Package Weight700g / 24.7oz

8. Warranty and Support

For technical support, English manual, PC software, or operational guidance, please contact the seller directly via email or through the AliExpress messaging system. The product is covered by a free returns policy within 90 days. Additionally, services include fast delivery with potential coupon codes if delayed, and refunds for lost or damaged packages.

9. User Tips

  • When generating arbitrary waveforms, ensure the PC software is properly installed and the device is connected via USB for seamless data transfer.
  • For precise frequency measurements, select an appropriate gate time to balance accuracy and measurement speed.
  • If using with Spooky2 software, consult the Spooky2 community forums or documentation for optimal setup and usage configurations.