FeelTech FY6300 Dual Channel Function/Arbitrary Waveform Signal Generator User Manual

Model: FY6300 Series

1. Introduction

The FeelTech FY6300 series is a high-performance dual-channel function/arbitrary waveform signal generator designed for various electronic testing and development applications. This series offers significant upgrades over previous models, providing enhanced frequency range, sampling rate, and waveform resolution.

Key Differences from FY2300 Series:

  • Frequency Range: Sine wave output frequency up to 60MHz (FY6300-60M), square wave maximum output frequency up to 25MHz.
  • Sampling Rate: Upgraded to 250MSa/s for clearer high-frequency signal output.
  • DA Integrated Circuit: Utilizes a 14-bit high-speed DA integrated circuit, offering a waveform step as low as 0.3mV at 5V output.
  • Waveform Storage Depth: Each group's waveform storage depth is upgraded to 8192*14bits for high-precision signal restoration.
  • Phase Resolution: Achieves 0.01 degrees for precise control over signal delays.
  • Duty Cycle Accuracy: Duty cycle adjustment accuracy up to 0.01%.
  • Power Supply Design: Features a new power supply design to minimize noise interference with small amplitude signals (10mV small signal maintains perfect signal characteristics).

2. Package Contents

Please check the package carefully upon receipt to ensure all items are present and undamaged.

  • 1 x DDS Signal Generator (FeelTech FY6300 Series)
  • 1 x DC 5V Power Adapter (UK/AU/US/EU Optional)
  • 1 x USB Cable
  • 2 x Signal Output Cables
  • 1 x CD (Contains user manual and PC software)
FeelTech FY6300 Signal Generator and included accessories
FeelTech FY6300 Signal Generator and included accessories.
FeelTech FY6300 Signal Generator with power adapter, USB cable, and signal output cables
FeelTech FY6300 Signal Generator with power adapter, USB cable, and signal output cables.

3. Product Overview

3.1 Front Panel

Front panel of FeelTech FY6300 showing display, control buttons, and output channels
Front panel of the FY6300 series signal generator. Features a display, WAVE, MEAS, SWEEP, SYS buttons, F1-F5 function buttons, a rotary knob, and arrow keys. Output channels CH1 and CH2 are on the right.

The front panel features a 2.4-inch TFT color display, navigation buttons (WAVE, MEAS, SWEEP, SYS), five function buttons (F1-F5), a rotary encoder knob for parameter adjustment, and directional arrow keys. The CH1 and CH2 output connectors are located on the right side.

3.2 Rear and Side Panels

Side view of FeelTech FY6300 showing DC5V power input, USB port, and power switch
Side view of the FY6300 showing the DC5V power input, USB port, and ON/OFF power switch.
Rear panel of FeelTech FY6300 showing BNC connectors for CH1, CH2, EXT.IN, and TTL_IO
Rear panel of the FY6300 showing BNC connectors for CH1, CH2, EXT.IN, and TTL_IO.

The rear panel includes the DC 5V power input, a USB port for PC connectivity, and the main power ON/OFF switch. The side panel also features ventilation slots for heat dissipation.

3.3 Dimensions

Dimensions of the FeelTech FY6300 signal generator
Dimensions of the FeelTech FY6300: 140mm (Length) x 54.2mm (Height).

4. Initial Setup

Follow these steps to set up your FeelTech FY6300 signal generator:

  1. Power Connection: Connect the provided DC 5V power adapter to the DC5V input port on the rear panel of the device. Plug the adapter into a suitable power outlet.
  2. Power Switch: Ensure the power switch on the rear panel is in the 'OFF' position before connecting power.
  3. USB Connection (Optional): For PC control and arbitrary waveform generation, connect the device to your computer using the supplied USB cable.
  4. Signal Output: Connect the signal output cables to the CH1 and/or CH2 BNC connectors on the rear panel, and then to your testing equipment (e.g., oscilloscope).
  5. Power On: Flip the power switch on the rear panel to the 'ON' position. The device display should light up.
Various power adapter plug types for FeelTech FY6300
The power adapter comes with different plug types (AU, EU, UK, US) to suit regional standards.

5. Operation

The FY6300 series offers various functions accessible via the front panel buttons and rotary knob.

5.1 Basic Navigation

  • WAVE: Selects waveform types (Sine, Square, Triangle, Arbitrary, etc.).
  • MEAS: Enters measurement mode for frequency, period, pulse width, etc.
  • SWEEP: Configures sweep functions (linear/logarithmic, frequency, amplitude, offset, duty cycle).
  • SYS: Accesses system settings (language, buzzer, uplink, sync, save/load).
  • F1-F5 Buttons: Context-sensitive function buttons, their labels change based on the current screen.
  • Rotary Knob: Used to adjust parameter values. Pressing the knob acts as an 'OK' button.
  • Arrow Keys: Used for navigation and fine-tuning parameters.

5.2 Waveform Generation

Press the 'WAVE' button to select the desired waveform. Use the F1-F5 buttons and the rotary knob to adjust parameters such as frequency, amplitude, offset, and duty cycle for each channel (CH1 and CH2).

Various display interfaces of the FeelTech FY6300: CH1 Main, CH2 Main, Burst Function, Sweep Function, Measure Function, System
Examples of main interfaces for CH1, CH2, Burst, Sweep, Measure, and System functions.

5.3 Arbitrary Waveform Generation (PC Software)

The device supports arbitrary waveform generation using the included PC software. Install the software from the provided CD. You can draw custom waveforms in the software and send them to the generator via the USB connection.

Screenshot of PC software for arbitrary waveform generation and oscilloscope display
PC software interface for drawing arbitrary waveforms and a comparison with an oscilloscope display.

5.4 Video Overview

An unboxing and overview video demonstrating the FeelTech FY6300 signal generator and its accessories.

6. Frequency Calibration (FY6000 Series)

To ensure accurate frequency output, you can perform a frequency calibration. An oscilloscope is required for this procedure.

  1. Enter Calibration Mode: In the shutdown state, simultaneously press the 'Power' button and the 'CH1' channel button on the front panel.
  2. Image showing how to press Power and CH1 buttons simultaneously for calibration
    Pressing Power and CH1 buttons simultaneously.
  3. Power On: While holding the 'Power' and 'CH1' buttons, turn on the power switch on the rear panel. Wait for 2 seconds, then release both buttons. The instrument will enter the working interface.
  4. Image showing the power switch on the rear panel
    Location of the power switch on the rear panel.
  5. Navigate to System Interface: Click the 'SYS' button to enter the system interface.
  6. Access More Settings: Click the 'More' button located in the lower right corner of the screen.
  7. System interface screen with 'More' button highlighted
    System interface with 'More' button.
  8. Start OSCI Calibration: In the 'MORE' interface, click the 'OSCI' button in the lower right corner of the screen.
  9. More interface screen with 'OSCI' button highlighted
    More interface with 'OSCI' button.
  10. Adjust Frequency: Use an oscilloscope to measure the frequency of the CH1 channel output. If the measured frequency is not 10.00000KHz, turn the parameter adjustment knob (rotary encoder) counter-clockwise to decrease the displayed value until the oscilloscope reads 10kHz.
  11. OSCI interface screen for frequency adjustment
    OSCI interface for frequency adjustment.
  12. Save and Restart: Once the CH1 output frequency is precisely 10kHz, press down the parameter adjustment knob (OK button). Then, turn off the power and restart the device. The frequency calibration is now complete.

7. Technical Specifications

CategoryParameterFY6300-20MFY6300-30MFY6300-60M
FrequencySine0~20MHz0~30MHz0~60MHz
Square0~20MHz0~25MHz0~25MHz
Ramp, Triangle0~10MHz0~20MHz0~20MHz
Pulse0~10MHz0~20MHz0~20MHz
TTL/CMOS0~10MHz0~40MHz0~60MHz
Arbitrary Waveform0~10MHz0~20MHz0~20MHz
Resolution1μHz (0.000001Hz)
Accuracy± 5×10-6
Stability±1×10-6/ 3Hours
Impedance50Ω±10% (Typical)
Phase Range0~359.9°
Phase Resolution0.1°
Waveform CharacteristicsWaveformsSine, Square, Triangle(Ramp), Arbitrary, Rise Sawtooth, Fall Sawtooth, Lorenz Pulse, Multitone, Noise, Cardiogram, Rapezoidal Pulse, Sinc Pulse, Narrow Pulse, Gauss White Noise, AM, FM, Step
Waveform Length8192 points
Sampling Rate250MSa/s
Vertical Resolution14 bits
SineHarmonic Suppression≥45dBc(<1MHz); ≥40dBc(1MHz~60MHz)
Total Harmonic Distortion<0.8% (20Hz~20kHz, 0dBm)
SquareRise/Fall Time≤10ns (Vpp<5V)
Overshoot≤7.5%
Duty Cycle0.1%~99.9%
Sawtooth waveLinearity>98% (0.01Hz~10kHz)
Output CharacteristicsAmplitude (50Ω)<20MHz: 10mVpp~20Vpp; >20MHz: 10mVpp~5Vpp
Resolution1mV (Attenuate 20dB)
Amplitude Stability±0.5%/ 5 Hours
Amplitude Flatness±5%(<10MHz); ±10%(>10MHz)
Waveform OutputImpedance50Ω±10% (Typical)
ProtectionAll channels can work more than 60 seconds when the load is short-circuited.
DC OffsetOffset RangeFrequency≤20MHz: ±10V; Frequency>20MHz: ±2.5V
Offset Resolution1mV
TTL OutputElectrical Level Amplitude>3Vpp
Fan-out>8 TTL LOAD
Rise/Fall Time≤10ns
CMOS OutputLow Electric Level<0.3V
High Electric Level1V~10V
Rise/Fall Time≤20ns
Frequency MeterRange0.01Hz~100MHz
SensitivityGate Time 3 grades (1S, 10S, 100S) adjustable
CounterRange0-4294967295
CouplingDC, AC
Working ModeManual
Pulse Width MeasurementResolution 5nS. Max. 10S
Voltage Input Range0.7Vpp~20Vpp
Pulse Width MeasurementMeasurement Resolution 5ns, Max. Limit 10s
Period MeasurementResolution 5ns, Max. Limit 10s
Sweep (Only CH1 available)Sweep TypeLinear or Logarithm
Sweep ObjectsFrequency, Amplitude, Offset, Duty Cycle
Sweep Time0.01S~999.99S/Step
Sweep RangeStarting position and Finishing position can be set arbitrarily.
General SpecificationsDisplay2.4 inch, TFT Color Display.
Save & Load Amount20 (01 to 20, 01 for start default value)
InterfaceTypeUSB to Serial interface
ProtocolCommand line mode, providing communication protocols.
Communicating Speed9600bps
PowerVoltage RangeDC5V±0.5V
TechnicSMD, LSI, Reliable and durable
BuzzerCan be turned on/off by setting.
EnvironmentTemp.: 0~40℃, Humidity: <80%
Package Size140mm (Length) x 95mm (Width) x 54.2mm (Height)
Package WeightNet Weight: 450g, Gross Weight: 750g

8. Troubleshooting

  • No Power: Ensure the DC 5V power adapter is correctly connected and plugged into a working outlet. Check the power switch on the rear panel is in the 'ON' position.
  • No Signal Output: Verify that the signal output cables are securely connected to the CH1/CH2 BNC connectors and to your testing equipment. Check the waveform settings (frequency, amplitude, offset) on the generator's display.
  • Incorrect Frequency/Amplitude: Perform the frequency calibration procedure as described in Section 6. Check the amplitude and offset settings.
  • PC Software Connection Issues: Ensure the USB cable is properly connected. Verify that the correct drivers are installed (usually provided on the CD or available from the manufacturer's website).
  • Device Unresponsive: Try restarting the device by turning it off and on using the rear power switch. If the issue persists, contact technical support.

9. User Tips

  • Portability: The compact size and 5V DC power input make this generator highly portable. Consider using a compatible power bank for field use or when AC power is unavailable.
  • FeelTech FY6300 connected to a power bank via USB
    The FY6300 can be powered by a power bank for portable use.
  • Software Utilization: For complex or custom waveforms, leverage the PC software. It provides a graphical interface for drawing arbitrary waveforms and offers more detailed control.
  • Calibration Importance: Regularly check and calibrate the frequency output, especially if precise measurements are critical for your applications.
  • Cable Management: Use appropriate BNC cables and adapters for your specific testing needs to ensure signal integrity.

10. Maintenance

  • Cleaning: Use a soft, dry cloth to clean the exterior of the device. Do not use abrasive cleaners or solvents.
  • Storage: Store the device in a cool, dry place away from direct sunlight, extreme temperatures, and high humidity.
  • Handling: Handle the device with care to avoid physical damage. Do not drop or expose it to excessive vibration.
  • Ventilation: Ensure the ventilation slots on the side panels are not obstructed to prevent overheating.