Hti HT-9600

Contents

HTI HT-9600 PM2.5 Detector Particle Counter User Manual

Model: HT-9600 | Brand: Hti

1. Introduction

The HTI HT-9600 is a professional particle dust meter designed to detect PM2.5, PM10, and other particulate matter in the air. Equipped with a 2.8-inch LCD display, it also measures temperature and humidity, making it a versatile tool for monitoring air quality in various environments. This manual provides detailed instructions for the safe and effective use of your device.

2. Safety Information

WARNING: This device is NOT a personal safety device. It is intended for environmental monitoring and not for personal protection or medical use.

  • Always operate the device within the specified operating environment conditions.
  • Do not expose the device to extreme temperatures, humidity, or direct sunlight.
  • Keep the device away from strong electromagnetic fields.
  • Ensure the sampling air inlet is clear of obstructions during operation.
  • Handle the device with care to prevent damage to the semiconductor laser and other sensitive components.

3. Product Overview

3.1 Components and Controls

Familiarize yourself with the main parts and buttons of your HT-9600 particle counter:

Diagram of HTI HT-9600 particle counter showing buttons and ports

Key Components:

  • Power Key: Turns the device on or off.
  • RUN Button: Initiates or stops the sampling process.
  • ENTER Key: Confirms selections or enters menus.
  • SHIFT Key: Used for various functions, often in combination with other keys.
  • BACK Key: Returns to the previous screen or menu.
  • Page Up/Down Buttons (Triangles): Navigates through menus or data records.
  • USB Interface: For data transfer and charging.
  • Sampling Air Inlet: Where air is drawn in for particle analysis.
  • Temperature and Humidity Detection Port: Sensor for environmental temperature and humidity.

3.2 Display Interface

The 2.8-inch LCD screen displays real-time data and menu options. Key information includes:

  • Particulate Matter Readings: Displays concentrations for 0.3μm, 2.5μm (PM2.5), and 10.0μm (PM10) particles, typically in pieces per liter or μg/m³.
  • Temperature: Current ambient temperature.
  • Humidity: Current ambient relative humidity.
  • Battery Indicator: Shows remaining battery life.
  • Time and Date: Current system time and date.
  • Sampling Time: Duration of the current sampling cycle.
Close-up of the HTI HT-9600 LCD display showing particle matter, temperature, and humidity readings

4. Specifications

FeatureSpecification
Test Particle Size0.3μm / 2.5μm / 10.0μm
Light SourceSemiconductor laser
Temperature Range0℃ ~ 50℃ (32℉ ~ 122℉)
Temperature Accuracy±0.5℃
Humidity Range0 ~ 100%RH
Humidity Accuracy±3%RH
Display Screen320 * 240 2.8in LCD
Counting Efficiency100 ± 10% @ 0.5 μM
Sampling Time50 seconds
Sampling Frequency1-99 / continuous
Alarm Level1-100000 (FED209E standard), 2-9 (ISO14464-1 standard)
Data Storage2000 records
InterfaceUSB
Operating Environment-10℃ ~ 50℃, 10 ~ 90% RH
Storage Environment-20℃ ~ 60℃ < 90%RH
Battery IncludedYes (9V NEDA 1604 IEC 6F22 type)
CertificationsCE, RoHS

5. Setup

5.1 Battery Installation/Charging

The HT-9600 comes with a 9V battery. If not already installed, open the battery compartment on the back of the device and insert the 9V NEDA 1604 IEC 6F22 battery, ensuring correct polarity. The device can be charged via the USB interface using the provided cable and a compatible USB power adapter (not always included).

Front and back view of the HTI HT-9600 particle counter, showing the battery compartment

5.2 Initial Power On

  1. Press and hold the Power Key (red button) until the LCD screen illuminates.
  2. The device will perform a self-check and display the main measurement interface.

6. Operating Instructions

6.1 Taking a Measurement

  1. Ensure the device is powered on and the sampling air inlet is unobstructed.
  2. Place the device in the area where you wish to measure air quality.
  3. Press the RUN button to start the sampling process. The device will draw in air and analyze particle concentrations.
  4. The sampling time is typically 50 seconds. During this period, the display will update with real-time particle counts, temperature, and humidity.
  5. Once the sampling is complete, the results will be displayed and automatically stored in the device's memory.

Video Description: A demonstration of the HTI HT-9600 particle counter in operation, showing real-time particle counts for 0.3μm, 2.5μm, and 10μm particles, along with temperature and humidity readings. The video also illustrates how to navigate through stored records.

6.2 Understanding Readings

  • 0.3μm, 2.5μm, 10μm: These values represent the count of particles per liter (Piece/L) or mass concentration (μg/m³) for particles of these respective sizes. PM2.5 refers to particles 2.5 micrometers or smaller, and PM10 refers to particles 10 micrometers or smaller.
  • Temperature (°C/°F) & Humidity (%RH): These indicate the ambient environmental conditions.

6.3 Accessing Stored Records

  1. From the main screen, use the Page Up/Down buttons to navigate to the 'Records' menu or similar data viewing option.
  2. Press ENTER to view the stored measurement data.
  3. Use the Page Up/Down buttons to scroll through different records. The device can store up to 2000 data sets.
  4. Press BACK to exit the records view.

6.4 Adjusting Settings

The device may have settings for units, sampling frequency, alarm levels, and time/date. Refer to the on-screen menus and use the SHIFT, ENTER, and Page Up/Down buttons to navigate and adjust these settings as needed.

7. Maintenance

  • Cleaning: Use a soft, dry cloth to clean the exterior of the device. Do not use abrasive cleaners or solvents. Keep the display screen clean for optimal visibility.
  • Air Inlet: Periodically check the sampling air inlet for any dust or debris accumulation. Gently clean if necessary, ensuring no foreign objects enter the sensor.
  • Battery Care: If the device will not be used for an extended period, remove the battery to prevent leakage. Recharge the battery regularly to maintain its lifespan.
  • Storage: Store the device in its protective case in a cool, dry place, away from direct sunlight and extreme temperatures, within the specified storage environment.
HTI HT-9600 particle counter stored in its protective carrying case

8. Troubleshooting

  • Device not powering on: Check battery level. Ensure the battery is correctly installed and charged.
  • Inaccurate readings: Ensure the sampling air inlet is clear. Avoid taking measurements in areas with extreme drafts or very high humidity, which can affect sensor performance. Allow the device to stabilize in the environment for a few minutes before taking readings.
  • Display issues: If the screen is frozen or unresponsive, try restarting the device by holding the Power Key.
  • Data transfer issues: Ensure the USB cable is securely connected to both the device and the computer. Check computer's device manager for proper driver installation.

9. User Tips

  • Optimal Placement: For general air quality monitoring, place the device in the center of a room, away from windows, doors, or direct airflow from HVAC systems, to get a representative reading.
  • Understanding Particle Counts: The device measures particles in 'pieces per liter'. High numbers, especially for 0.3μm and 2.5μm, indicate poorer air quality. For example, a reading of 12,000 pieces/L for 0.3μm particles and 50 pieces/L for 2.5μm particles suggests a significant presence of very fine particulate matter.
  • Regular Monitoring: Take readings at different times of the day and under various conditions (e.g., after cooking, during cleaning, with windows open/closed) to understand the dynamics of your indoor air quality.
  • Data Logging: Utilize the 2000-record data storage capacity to track air quality trends over time. This can help identify sources of pollution or evaluate the effectiveness of air purification efforts.

10. Warranty and Support

For warranty information, technical support, or service inquiries, please contact the seller or manufacturer directly. Keep your purchase receipt as proof of purchase.


Documents and downloads

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.